J Thorac Cardiovasc Surg 1994;108:604-608
© 1994 Mosby, Inc.
CARDIOPULMONARY BYPASS, |
St. Louis, Mo.
From the Division of Cardiothoracic Surgery, Department of Surgery, St. Louis University Health Sciences Center, St. Louis, Mo.
Received for publication Jan. 28, 1994. Accepted for publication May 31, 1994. Address for reprints: D. Glenn Pennington, MD, Department of Surgery, 3635 Vista Ave. at Grand Blvd., St. Louis, MO 63110.
Abstract
The Novacor left ventricular assist system (Baxter Healthcare Corp., Novacor Div., Oakland, Calif.) is currently undergoing clinical investigation as a bridge to cardiac transplantation. The original animal and clinical protocols recommended placement of the pump/energy converter within the peritoneal space. During preclinical cadaver fitting trials, we determined that it might be difficult to place this device in the peritoneal cavity of tall, slender patients, patients having a small muscular abdomen, and many women. We were also concerned that peritoneal placement would increase the risk and severity of infection and that abdominal organs might be injured by direct contact with the pump. In studies in cadavers, we therefore developed a technique using preperitoneal placement of the pump/energy converter. This technique has now been used successfully in more than 200 patients. A detailed description of this insertion technique is presented to facilitate its current use for the implantation of temporary devices and to plan for the implantation of permanent systems. (J THORACCARDIOVASCSURG1994;108:604-8)
This article has been cited by other articles:
![]() |
D. Loisance Mechanical Circulatory Support: a Clinical Reality Asian Cardiovasc Thorac Ann, October 1, 2008; 16(5): 419 - 431. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Zimpfer, P. Zrunek, S. Sandner, H. Schima, M. Grimm, A. Zuckermann, E. Wolner, and G. Wieselthaler Post-transplant survival after lowering fixed pulmonary hypertension using left ventricular assist devices Eur. J. Cardiothorac. Surg., April 1, 2007; 31(4): 698 - 702. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Zimpfer, P. Zrunek, W. Roethy, M. Czerny, H. Schima, L. Huber, M. Grimm, A. Rajek, E. Wolner, and G. Wieselthaler Left ventricular assist devices decrease fixed pulmonary hypertension in cardiac transplant candidates J. Thorac. Cardiovasc. Surg., March 1, 2007; 133(3): 689 - 695. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Pavie, N. Reiss, and S. Aubert Implantation technique of the Novacor left ventricular assist device MMCTS, February 19, 2007; 2007(0219): 1529. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Matsumiya, Y. Miyamoto, N. Fukushima, O. Monta, Y. Sawa, and H. Matsuda Intraabdominal placement of a Novacor ventricular assist system by using an expanded polytetrafluoroethylene pouch Ann. Thorac. Surg., May 1, 2004; 77(5): 1851 - 1853. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Rao, M. C. Oz, M. A. Flannery, K. A. Catanese, M. Argenziano, and Y. Naka Revised screening scale to predict survival after insertion of a left ventricular assist device J. Thorac. Cardiovasc. Surg., April 1, 2003; 125(4): 855 - 862. [Abstract] [Full Text] |
||||
![]() |
S. L. Tittle, D. Mandapati, G. S. Kopf, and J. A. Elefteriades Alternate technique for implantation of left ventricular assist system: left thoracotomy for reoperative cases Ann. Thorac. Surg., March 1, 2002; 73(3): 994 - 996. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. N. Pierson III, R. Howser, T. Donaldson, W. H. Merrill, R. J. Dignan, D. C. Drinkwater Jr, K. G. Christian, J. Butler, D. Chomsky, J. R. Wilson, et al. Left ventricular assist device implantation via left thoracotomy: alternative to repeat sternotomy Ann. Thorac. Surg., March 1, 2002; 73(3): 997 - 999. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. R. McBride, K. S. Naunheim, A. C. Fiore, R. G. Johnson, D. A. Moroney, J. A. Brannan, and M. T. Swartz Risk analysis in patients bridged to transplantation Ann. Thorac. Surg., June 1, 2001; 71(6): 1839 - 1844. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Reichenbach, D. J. Farrar, and J. D. Hill A versatile intracorporeal ventricular assist device based on the Thoratec VAD system Ann. Thorac. Surg., March 1, 2001; 71 (2007): S171 - S175. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Noirhomme, L. Jacquet, M. Underwood, G. El Khoury, M. Goenen, and R. Dion The effect of chronic mechanical circulatory support on neuroendocrine activation in patients with end-stage heart failure Eur. J. Cardiothorac. Surg., July 1, 1999; 16(1): 63 - 67. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. Hunt and O.H. Frazier Mechanical Circulatory Support and Cardiac Transplantation Circulation, May 26, 1998; 97(20): 2079 - 2090. [Full Text] [PDF] |
||||
![]() |
J. Muller, G. Wallukat, Y.-G. Weng, M. Dandel, S. Spiegelsberger, S. Semrau, K. Brandes, V. Theodoridis, M. Loebe, R. Meyer, et al. Weaning From Mechanical Cardiac Support in Patients With Idiopathic Dilated Cardiomyopathy Circulation, July 15, 1997; 96(2): 542 - 549. [Abstract] [Full Text] |
||||
![]() |
W. E. Richenbacher and W. S. Pierce Mechanical Circulatory Support Ann. Thorac. Surg., November 1, 1996; 62(5): 1558 - 1559. [Full Text] |
||||
![]() |
D. G. Pennington Extended Support With Permanent Systems: Percutaneous Versus Totally Implantable Ann. Thorac. Surg., January 1, 1996; 61(1): 403 - 406. [Abstract] [Full Text] |
||||
![]() |
J.-R. E. Monties, J. Trinkl, T. Mesana, P. J. Havlik, and J.-L. Y. Demunck Cora Valveless Pulsatile Rotary Pump: New Design and Control Ann. Thorac. Surg., January 1, 1996; 61(1): 463 - 468. [Abstract] [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |