J Thorac Cardiovasc Surg 1998;115:582-586
© 1998 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
Supported by an unrestricted grant from the Shiley Heart Valve Research Center.
Received for publication June 5, 1997; revisions requested July 15, 1997; revisions received Nov. 24, 1997; accepted for publication Nov. 24, 1997. Address for reprints: Kenneth D. Hopper, MD, Department of Radiology Pennsylvania State University, 500 University Dr., P.O. Box 850, Hershey, PA 17033.
Abstract
Objective: Modified cineradiographic systems have been used clinically to detect partially broken outlet struts in normally functioning Björk-Shiley convexo-concave heart valves. Almost all such valves were explanted, presuming that full failure would likely follow. Inasmuch as the clinical setting only rarely permits examination of normally rated valves, the accuracy of radiographic detection cannot be clinically defined. This study uses the clinical radiographic technique in sheep implanted with known-status convexo-concave valves, comparing its accuracy and that of a newly developed, geometric image magnification radiography system.
Methods: Twenty-one sheep with mitral convexo-concave valves were studied on both systems. Five were used for extensive training. When operators were expert with both systems, images of four intact valves and 12 valves with outlet strut single leg separations, along with a seventeenth single leg separation valve used for calibration, were integrated into 112 image sets organized into a balanced incomplete block design for evaluation by eight trained, blinded reviewers.
Results: Cineradiography sensitivity was 24% versus 31% for direct image magnification. The odds ratio for detection of single leg separation by direct image magnification versus cineradiography was 2.0 (95% confidence interval, 0.76 to 5.9; p = 0.13). Cineradiography specificity was 93% versus 90% for direct image magnification. Sensitivity and specificity varied markedly by reviewer, with sensitivity ranging from 8% to 55% and specificity from 51% to 100% for the combined technologies.
Conclusions: The data support the need for more intensive training for convexo-concave valve imaging and further investigation of unconventional radiographic technologies. Clinical cineradiography of convexo-concave valves may detect as little as 25% of valves having a single leg separation, underestimating the prevalence of single leg separations and thereby implying more rapid progression to full fracture than is actually the case.
This article has been cited by other articles:
![]() |
E. H. Blackstone Could It Happen Again?: The Bjork-Shiley Convexo-Concave Heart Valve Story Circulation, May 31, 2005; 111(21): 2717 - 2719. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |