J Thorac Cardiovasc Surg 2000;120:473-477
© 2000 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
From Loma Linda University Medical Center, Loma Linda, Calif,a and CryoLife, Inc, Marietta, Ga.b
Address for reprints: Steven R. Gundry, MD, Professor and Head, Division of Cardiothoracic Surgery, Loma Linda University, 11175 Campus St, Loma Linda, CA 92354 (E-mail: sgundry{at}som.llu.edu ).
Objective: As heart surgery becomes increasingly focused on minimally invasive techniques, it has become apparent that conventional techniques of anastomosis will need to be severely altered or abandoned. Toward that end, we developed and tested in vitro and in vivo coronary artery bypass graft anastomoses using a biologic glue formulated from bovine albumin and glutaraldehyde. We used a double-balloon catheter as a temporary internal stent to create and seal the anastomosis during gluing.
Methods: Initially, anastomoses were made between cryopreserved human saphenous vein segments and coronary arteries in vitro on 12 intact bovine hearts. A total of 42 anastomoses were created with the catheter system introduced into the distal end of the graft, exiting the back wall, and entering the anterior wall of the coronary artery. Two balloons (one in the graft and one in the coronary artery) held the anastomosis stable while the biologic glue was applied externally and allowed to set for 2 minutes. The balloon catheter was then removed from the end of the graft simulating a side-to-side internal thoracic artery anastomosis. After the graft had been flushed to assure distal end patency, the open end of the graft was clipped, turning the anastomosis into an end-to-side graft. A pressure transducer was then attached to the graft and saline solution forcefully infused.
Results: All grafts easily held a pressure of 300 mm Hg; 10 grafts were tested up to 560 mm Hg without leaks. Distal and proximal coronary artery patency was checked by examining flow out of the coronary ostia and by cutting arteries distal to the grafts. All anastomoses were patent on being opened and no glue was seen intraluminally. Subsequently, 3 anastomoses of the left internal thoracic artery to the left anterior descending artery have been constructed in goats, with autopsies at 24 hours, 10 months, and 1 year revealing patent anastomoses.
Conclusion: A biologic glue and catheter system has been developed that allows a coronary anastomosis with a high bursting strength to be performed. When the system has been further developed and tested, truly minimally invasive heart surgery may be possible.
This article has been cited by other articles:
![]() |
D. M. Sidle and C. S. Maas Determination of Shear Strength of Periosteum Attached to Bone With BioGlue Surgical Adhesive Arch Facial Plast Surg, September 1, 2008; 10(5): 316 - 320. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Schiller, H. Rudorf, C. B. Welzel, M. J. Kiderlen, C. Probst, O. Dewald, and A. Welz Sutureless anastomoses of rabbit carotid arteries with BioGlue. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1513 - 1518. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Eto, S. Morita, M. Sugiura, T. Yoshimura, R. Tominaga, and T. Matsuda Elastomeric surgical sealant for hemostasis of cardiovascular anastomosis under full heparinization Eur. J. Cardiothorac. Surg., November 1, 2007; 32(5): 730 - 734. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Masuda, Y. Saiki, S. Kawatsu, I. Yoshioka, H. Fujiwara, S. Kawamoto, S. Sai, A. Iguchi, N. Sakamoto, T. Ohashi, et al. Trial of new vascular clips for aortic anastomosis in a canine model J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 723 - 730. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. J. Van Nooten, P. Somers, R. Forsyth, K. Narine, Y. Van Belleghem, S. Jacobs, and F. De Somer Autologous glue: Part of the sticky mystery unraveled J. Thorac. Cardiovasc. Surg., August 1, 2007; 134(2): 415 - 423. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wippermann, C. Konstas, T. Wahlers, and J. M. Albes Feasibility study of sutureless distal coronary anastomoses with degradable Y-shunt and tissue adhesives in a porcine off-pump model Interactive CardioVascular and Thoracic Surgery, December 1, 2006; 5(6): 676 - 679. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wippermann, C. Konstas, M. Breuer, H. Kosmehl, T. Wahlers, and J. M. Albes Long-term effects in distal coronary anastomoses using different adhesives in a porcine off-pump model. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 325 - 331. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Szafranek, S.R. Podila, W. Al-Khyatt, and E.N. Kulatilake Aseptic mediastinal cyst caused by BioGlue 7 months after cardiac surgery J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 1202 - 1203. [Full Text] [PDF] |
||||
![]() |
U. Klima, T. Kofidis, A. Boening, and A. Haverich Long term angiographic monitoring reveals safety and efficacy of the Converge end-to-side anastomotic coupler Interactive CardioVascular and Thoracic Surgery, December 1, 2005; 4(6): 505 - 508. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. M. Sidle, B. M. Loos, A. L. Ramirez, S. S. Kabaker, and C. S. Maas Use of BioGlue Surgical Adhesive for Brow Fixation in Endoscopic Browplasty Arch Facial Plast Surg, November 1, 2005; 7(6): 393 - 397. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Boening, F. Schoeneich, A. Lichtenberg, E. Bagaev, J. T. Cremer, and U. Klima First clinical results with a 30{degrees} end-to-side coronary anastomosis coupler Eur. J. Cardiothorac. Surg., May 1, 2005; 27(5): 876 - 881. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Furst and A. Banerjee Release of Glutaraldehyde From an Albumin-Glutaraldehyde Tissue Adhesive Causes Significant In Vitro and In Vivo Toxicity Ann. Thorac. Surg., May 1, 2005; 79(5): 1522 - 1528. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Klima, H. MacVaugh III, E. Bagaev, M. Maringka, S. Kirschner, J. Beilner, and A. Haverich Magnetic Vascular Port in Minimally Invasive Direct Coronary Artery Bypass Grafting Circulation, September 14, 2004; 110(11_suppl_1): II-55 - II-60. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Van Belleghem, R. G. Forsyth, K. Narine, A. Moerman, Y. Taeymans, and G. J. Van Nooten Bovine glue (BioGlue) is catabolized by enzymatic reaction in the vascular dog model Ann. Thorac. Surg., June 1, 2004; 77(6): 2177 - 2181. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Carrel, L. Englberger, D. Keller, S. Windecker, B. Meier, and F. Eckstein Clinical and angiographic results after mechanical connection for distal anastomosis in coronary surgery J. Thorac. Cardiovasc. Surg., June 1, 2004; 127(6): 1632 - 1640. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Carrel, F. S. Eckstein, L. Englberger, P. A. Berdat, and J. Schmidli Clinical experience with devices for facilitated anastomoses in coronary artery bypass surgery Ann. Thorac. Surg., March 1, 2004; 77(3): 1110 - 1120. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Klima, V. Falk, M. Maringka, S. Bargenda, S. Badack, A. Moritz, F. Mohr, A. Haverich, and G. Wimmer-Greinecker Magnetic vascular coupling for distal anastomosis in coronary artery bypass grafting: A multicenter trial J. Thorac. Cardiovasc. Surg., November 1, 2003; 126(5): 1568 - 1574. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. D'Ancona, T. A. Salerno, P. S. Aoukar, and H. Karamanoukian Use of nitinol-U clips and flow characteristics of LIMA-LAD anastomoses Interactive CardioVascular and Thoracic Surgery, September 1, 2003; 2(3): 237 - 240. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Mack, R. W. Emery, L. R. Ley, P. A. Cole, A. Leonard, J. R. Edgerton, T. M. Dewey, M. J. Magee, and T. S. Flavin Initial experience with proximal anastomoses performed with a mechanical connector Ann. Thorac. Surg., June 1, 2003; 75(6): 1866 - 1871. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P. Buijsrogge, C. W. J. Verlaan, P. F. Grundeman, and C. Borst Briefly occlusive coronary anastomosis with tissue adhesive J. Thorac. Cardiovasc. Surg., February 1, 2003; 125(2): 385 - 390. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Tozzi, A. F. Corno, and L. K. von Segesser Sutureless coronary anastomoses: revival of old concepts Eur. J. Cardiothorac. Surg., October 1, 2002; 22(4): 565 - 570. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Martens, M. Dietrich, M. Doss, A. Moritz, and G. Wimmer-Greinecker The Heartflo device for distal coronary anastomoses: clinical experiences in 60 patients Ann. Thorac. Surg., October 1, 2002; 74(4): 1139 - 1143. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. S. Eckstein, L. F. Bonilla, H. Schaff, L. Englberger, S. Windecker, P. Hindrichs, and T. P. Carrel Two generations of the St. Jude Medical ATG coronary connector systems for coronary artery anastomoses in coronary artery bypass grafting Ann. Thorac. Surg., October 1, 2002; 74(4): S1363 - 1367. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. S. Eckstein, L. F. Bonilla, L. Englberger, F. Eberli, S. Windecker, T. A. Berg, M. Romanens, F. F. Immer, and T. P. Carrel First Clinical Results With a New Mechanical Connector for Distal Coronary Artery Anastomoses in CABG Circulation, September 24, 2002; 106(12_suppl_1): I-1 - I-4. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Passage, H. Jalali, R. K.W. Tam, S. Harrocks, and M. F. O'Brien BioGlue surgical adhesive--an appraisal of its indications in cardiac surgery Ann. Thorac. Surg., August 1, 2002; 74(2): 432 - 437. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P. Buijsrogge, C. W.J. Verlaan, M. H.P. van Rijen, P. F. Grundeman, and C. Borst Coronary end-to-side sleeve anastomosis using adhesive in off-pump bypass grafting in the pig Ann. Thorac. Surg., May 1, 2002; 73(5): 1451 - 1456. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. A. LeMaire, Z. C. Schmittling, J. S. Coselli, A. Undar, B. A. Deady, F. J. Clubb Jr, and C. D. Fraser Jr BioGlue surgical adhesive impairs aortic growth and causes anastomotic strictures Ann. Thorac. Surg., May 1, 2002; 73(5): 1500 - 1506. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. S. Eckstein, L. F. Bonilla, L. Englberger, F. F. Immer, T. A. Berg, J. Schmidli, and T. P. Carrel The St Jude Medical symmetry aortic connector system for proximal vein graft anastomoses in coronary artery bypass grafting J. Thorac. Cardiovasc. Surg., April 1, 2002; 123(4): 777 - 782. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ono, R. K. Wolf, D. Angouras, and E. W. Schneeberger Early experience of coronary artery bypass grafting with a new self-closing clip device J. Thorac. Cardiovasc. Surg., April 1, 2002; 123(4): 783 - 787. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. S. Eckstein, L. F. Bonilla, L. Englberger, E. Stauffer, T. A. Berg, J. Schmidli, and T. P. Carrel Minimizing aortic manipulation during OPCAB using the symmetry aortic connector system for proximal vein graft anastomoses Ann. Thorac. Surg., September 1, 2001; 72(3): S995 - 998. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Raanani, D. A. Latter, L. E. Errett, D. B. Bonneau, Y. Leclerc, and G. C. Salasidis Use of ""BioGlue"" in aortic surgical repair Ann. Thorac. Surg., August 1, 2001; 72(2): 638 - 640. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |