JTCS KCI
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Author home page(s):
Ruggero De Paulis
Christoph Schmitz
Raffaele Scaffa
Luigi Chiariello
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by De Paulis, R.
Right arrow Articles by Reul, H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by De Paulis, R.
Right arrow Articles by Reul, H.

J Thorac Cardiovasc Surg 2005;130:1016-1021
© 2005 The American Association for Thoracic Surgery


Surgery for Acquired Cardiovascular Disease

In vitro evaluation of aortic valve prosthesis in a novel valved conduit with pseudosinuses of Valsalva

Ruggero De Paulis, MD a , * , Christoph Schmitz, Dr-Ing b , Raffaele Scaffa, MD a , Paolo Nardi, MD a , Luigi Chiariello, MD a , Helmut Reul, Dr-Ing b

a Cardiac Surgery Department, University of Tor Vergata, Rome, Italy
b Helmholtz-Institute for Biomedical Engineering, RWTH Aachen, Aachen, Germany

Received for publication December 31, 2004; revisions received April 14, 2005; accepted for publication April 22, 2005.

* Address for reprints: Ruggero De Paulis, MD, Cattedra di Cardiochirurgia, Università di Roma Tor Vergata, European Hospital, via Portuense 700, 00149 Roma, Italy (Email: depauli{at}tin.it).

OBJECTIVE: This study was undertaken to determine whether the presence of vortices immediately above a prosthetic aortic valve could negatively influence the in vitro hydrodynamic performances of a biologic or mechanical valve implanted in a new Dacron polyester fabric conduit that incorporates sculpted sinuses of Valsalva.

METHODS: With a computer-controlled pulse duplicator, the in vitro performance (pressure differences, closure and leakage volumes, and energy losses) of a 25-mm mechanical or biologic prosthesis implanted in a standard Dacron straight conduit or in the new Dacron graft with a sculpted sinus were analyzed and compared.

RESULTS: The mechanical and biologic prostheses at 7 L/min cardiac output showed pressure drops across the valve of 8.72 mm Hg and 13.45 mm Hg, respectively, when inserted in the new Valsalva-style graft and of 7.97 mm Hg and 12.94 mm Hg, respectively, when inserted in the standard graft. The closure and leakage volumes for mechanical valves were higher than those for biologic valves; however, the presence or absence of sinuses did not result in significant differences in closure and leakage volumes. The maximal total energy losses were 5.89% and 9.49% for mechanical and biologic valves, respectively. No differences were evident between the two different Dacron grafts for each prosthetic heart valve.

CONCLUSION: The normal opening and closing behavior of a prosthetic aortic valve was not altered or modified by a different root shape above the heart valve. The presence of vortices inside the pseudosinuses of Valsalva did not influence the hydrodynamic properties of the biologic and mechanical valves tested.



17; 26; 35





This article has been cited by other articles:


Home page
J. Thorac. Cardiovasc. Surg.Home page
L. de Kerchove, M. Boodhwani, D. Glineur, M. Vandyck, J.-L. Vanoverschelde, P. Noirhomme, and G. El Khoury
Valve sparing-root replacement with the reimplantation technique to increase the durability of bicuspid aortic valve repair
J. Thorac. Cardiovasc. Surg., December 1, 2011; 142(6): 1430 - 1438.
[Abstract] [Full Text] [PDF]


Home page
Phil Trans R Soc AHome page
M. D. de Tullio, G. Pascazio, L. Weltert, R. De Paulis, and R. Verzicco
Evaluation of prosthetic-valved devices by means of numerical simulations
Phil Trans R Soc A, June 28, 2011; 369(1945): 2502 - 2509.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
A. N. Azadani, N. Jaussaud, P. B. Matthews, L. Ge, T. S. Guy, T. A.M. Chuter, and E. E. Tseng
Valve-in-Valve Implantation Using a Novel Supravalvular Transcatheter Aortic Valve: Proof of Concept
Ann. Thorac. Surg., December 1, 2009; 88(6): 1864 - 1869.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
M. Tabata, H. Takayama, M. E. Bowdish, C. R. Smith, and A. S. Stewart
Modified Bentall Operation With Bioprosthetic Valved Conduit: Columbia University Experience
Ann. Thorac. Surg., June 1, 2009; 87(6): 1969 - 1970.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS
Copyright © 2005 by The American Association for Thoracic Surgery.