J Thorac Cardiovasc Surg 2005;130:1016-1021
© 2005 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
a Cardiac Surgery Department, University of Tor Vergata, Rome, Italy
b Helmholtz-Institute for Biomedical Engineering, RWTH Aachen, Aachen, Germany
Received for publication December 31, 2004; revisions received April 14, 2005; accepted for publication April 22, 2005. * Address for reprints: Ruggero De Paulis, MD, Cattedra di Cardiochirurgia, Università di Roma Tor Vergata, European Hospital, via Portuense 700, 00149 Roma, Italy (Email: depauli{at}tin.it).
OBJECTIVE: This study was undertaken to determine whether the presence of vortices immediately above a prosthetic aortic valve could negatively influence the in vitro hydrodynamic performances of a biologic or mechanical valve implanted in a new Dacron polyester fabric conduit that incorporates sculpted sinuses of Valsalva.
METHODS: With a computer-controlled pulse duplicator, the in vitro performance (pressure differences, closure and leakage volumes, and energy losses) of a 25-mm mechanical or biologic prosthesis implanted in a standard Dacron straight conduit or in the new Dacron graft with a sculpted sinus were analyzed and compared.
RESULTS: The mechanical and biologic prostheses at 7 L/min cardiac output showed pressure drops across the valve of 8.72 mm Hg and 13.45 mm Hg, respectively, when inserted in the new Valsalva-style graft and of 7.97 mm Hg and 12.94 mm Hg, respectively, when inserted in the standard graft. The closure and leakage volumes for mechanical valves were higher than those for biologic valves; however, the presence or absence of sinuses did not result in significant differences in closure and leakage volumes. The maximal total energy losses were 5.89% and 9.49% for mechanical and biologic valves, respectively. No differences were evident between the two different Dacron grafts for each prosthetic heart valve.
CONCLUSION: The normal opening and closing behavior of a prosthetic aortic valve was not altered or modified by a different root shape above the heart valve. The presence of vortices inside the pseudosinuses of Valsalva did not influence the hydrodynamic properties of the biologic and mechanical valves tested.
This article has been cited by other articles:
![]() |
L. de Kerchove, M. Boodhwani, D. Glineur, M. Vandyck, J.-L. Vanoverschelde, P. Noirhomme, and G. El Khoury Valve sparing-root replacement with the reimplantation technique to increase the durability of bicuspid aortic valve repair J. Thorac. Cardiovasc. Surg., December 1, 2011; 142(6): 1430 - 1438. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. D. de Tullio, G. Pascazio, L. Weltert, R. De Paulis, and R. Verzicco Evaluation of prosthetic-valved devices by means of numerical simulations Phil Trans R Soc A, June 28, 2011; 369(1945): 2502 - 2509. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Azadani, N. Jaussaud, P. B. Matthews, L. Ge, T. S. Guy, T. A.M. Chuter, and E. E. Tseng Valve-in-Valve Implantation Using a Novel Supravalvular Transcatheter Aortic Valve: Proof of Concept Ann. Thorac. Surg., December 1, 2009; 88(6): 1864 - 1869. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tabata, H. Takayama, M. E. Bowdish, C. R. Smith, and A. S. Stewart Modified Bentall Operation With Bioprosthetic Valved Conduit: Columbia University Experience Ann. Thorac. Surg., June 1, 2009; 87(6): 1969 - 1970. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |