The Journal of Thoracic and Cardiovascular Surgery, Vol 81, 464-469, Copyright © 1981 by The American Association for Thoracic Surgery and The Western Thoracic Surgical Association
KV Arom, C Ellestad, FL Grover and JK Trinkle
Six venous cannulas (USCI No. 32, USCI No. 40, USCI No. 44, Sarns No. 40,
Sarns two-stage cavoatrial, and Ferguson Argyle No. 40) were tested for
efficiency of venous flow during cardiopulmonary bypass, with and without
aortic cross-clamping. Each cannula was tested six times in dog models
(twice in each of three dogs) and the data were averaged. The tip of the
cavoatrial Sarns catheter was positioned as recommended. Two No. 32 USCI
caval cannulas were placed either with or without caval snaring. The other
cannulas were placed in the right atrium. Arterial flow was constant at 80
ml/kg/min, and aortic pressure averaged 84 mm Hg. Central venous pressure
and the right ventricle excluded from the pulmonary artery so that blood
which was not drained via the venous cannula was measured. Each single
atrial cannula decompressed the right atrium (right atrial pressure 0 to 1
mm Hg) better than two caval cannulas (right atrial pressure 0 to 2 mm Hg).
Caval snaring did not empty the right atrium (right atrial pressure 2 to 5
mm Hg) as well as any single cannula. The cavoatrial Sarns catheter did not
empty the right atrium as well (right atrial pressure 2 to 4 mm Hg), but
drained blood from the inferior vena cava (central venous pressure 0 mm Hg)
better than other atrial cannulas (central venous pressure 1 to 5 mm Hg).
All of the atrial cannulas performed equally well. The right ventricular
vent flow varied between 63 and 70 ml/min with each of the atrial cannulas
but rose to 85 ml/min with the cavoatrial Sarns catheter and 190 and 74
ml/min during caval cannulation with and without snaring, respectively.
Aortic cross-clamping eliminated coronary sinus flow and decreased right
ventricular vent flow. Therefore, a single atrial cannula is more efficient
in draining blood from the right side of the heart than are two caval or a
cavoatrial cannula. This advantage is negated by aortic cross-clamping.
ARTICLES
Objective evaluation of the efficacy of various venous cannulas
This article has been cited by other articles:
![]() |
J. W. Hammon Extracorporeal Circulation: Perfusion System Card. Surg. Adult, January 1, 2008; 3(2008): 350 - 370. [Full Text] |
||||
![]() |
D. Jegger, S. Sundaram, K. Shah, I. Mallabiabarrena, G. Mucciolo, and L.K. von Segesser Using computational fluid dynamics to evaluate a novel venous cannula (Smart canula (R)) for use in cardiopulmonary bypass operating procedures Perfusion, July 1, 2007; 22(4): 257 - 265. [Abstract] [PDF] |
||||
![]() |
D. Jegger, A. F Corno, A. Mucciolo, G. Mucciolo, Y. Boone, J. Horisberger, I. Seigneul, M. Jachertz, and L. K von Segesser A prototype paediatric venous cannula with shape change in situ Perfusion, January 1, 2003; 18(1): 61 - 65. [Abstract] [PDF] |
||||
![]() |
E. A. Hessel II and L. H. Edmunds Jr. Extracorporeal Circulation: Perfusion Systems Card. Surg. Adult, January 1, 2003; 2(2003): 317 - 338. [Full Text] |
||||
![]() |
M. Kurusz, D. J Deyo, A. D Sholar, W. Tao, and J. B Zwischenberger Laboratory testing of femoral venous cannulae: effect of size, position and negative pressure on flow Perfusion, September 1, 1999; 14(5): 379 - 387. [Abstract] [PDF] |
||||
![]() |
J. B Riley, S. B Hardin, B. A Winn, and M. B Hurdle In vitro comparison of cavoatrial (dual stage) cannulae for use during cardiopulmonary bypass Perfusion, July 1, 1986; 1(3): 197 - 204. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |