The Journal of Thoracic and Cardiovascular Surgery, Vol 82, 391-397, Copyright © 1981 by The American Association for Thoracic Surgery and The Western Thoracic Surgical Association
EH Williams, VR Gonti, A Nishimura, LC Stout and V Ferrans
The clinical course of a patient with bioprosthetic failure of aortic and
mitral Ionescu-Shiley valves (which contain bovine pericardial leaflets) is
presented. Failure occurred in less than 4 months and was due to infection
that resulted in calcification and severe stenosis. On examination of
tissue specimens under light and electron microscopes, the pathological
process was seen to be similar o that occurring with infected porcine
prostheses.