J Thorac Cardiovasc Surg 1994;107:289-292
© 1994 Mosby, Inc.
CARDIOPULMONARY BYPASS, MYOCARDIAL MANAGEMENT, AND SUPPORT TECHNIQUES |
Groningen, The Netherlands
Supported by a grant from Bentley/Baxter, Inc., Uden, The Netherlands.
Received for publication Feb. 2, 1993. Accepted for publication May 24, 1993. Address for reprints: W. van Oeveren, PhD, Thorax Centre, Department of Cardiopulmonary Surgery, University Hospital of Groningen, Oostersingel 59, 9713 EZ Groningen, The Netherlands.
Abstract
Heparin coating of an extracorporeal circuit for cardiopulmonary bypass improves the hemocompatibility of the circuit and reduces the inflammatory response of the body. It has not been established, however, that heparin coating also improves postoperative hemostasis. We therefore performed a study in 30 patients who underwent a routine coronary artery bypass graft operation subjected to cardiopulmonary bypass with an uncoated (control) or a heparin-coated extracorporeal circuit (Duraflo II). We found significantly higher plasma levels of heparin in the Duraflo II group. However, we found no significant differences between the two groups with regard to other parameters of activation of the fibrinolytic and coagulation systems and to activation of platelets. Postoperative blood loss and donor blood transfusions were reduced in the Duraflo II group but not to a statistically significant extent. We conclude that heparin coating of an extracorporeal circuit improves anticoagulation but does not significantly reduce platelet activation, fibrinolysis, postoperative blood loss, and donor blood transfusions in routine coronary bypass operations. (J THORAC CARDIOVASC SURG 1994;107:289-92)
Cardiopulmonary bypass (CPB) activates both formed blood elements and humoral systems and thereby causes a systemic inflammatory response.
1-3 This inflammatory response can be reduced by heparin coating of the surface of the extracorporeal circuit (ECC).
4, 5 Coating of the surface of an ECC with heparin improves the hemocompatibility by thrombin binding and inactivation.
6-9 However, for hemostasis to be preserved, platelet activation and fibrinolytic activity also have to be reduced. In the present study we tested the hypothesis that heparin coating can also reduce platelet activation and fibrinolytic activity and as a consequence can lead to an improved postoperative hemostasis in routine coronary artery bypass graft (CABG) operations. Therefore we investigated the activation and consumption of blood components that are indicative for alteration in the hemostatic capacity in a study on 30 patients randomly allocated to heparin-coated or uncoated (control) devices. The clinical impact of the effects of heparin coating was assessed by measuring postoperative bleeding time, total amount of blood loss, and donor blood transfusions.
METHODS
Patients
The study was conducted in 30 patients who underwent an elective CABG operation with saphenous vein or internal mammary arteries (or both). These patients were less than 75 years of age, had no evidence of severe heart failure or renal or liver dysfunction, had no bleeding diathesis, did not have diabetes mellitus, had not used platelet-inhibiting drugs within 10 days before the operation, and had given informed consent. The study was approved by the medical ethical committee of our institution and was conducted in the Department of Thoracic Surgery of the University Hospital of Groningen.
The patients were randomly assigned for CPB with or without a heparin-coated ECC. After premedication of the patient with diazepam (10 to 15 mg), anesthesia was induced with sufentanyl (3 to 5 µg/kg) and muscle relaxation was achieved with pancuronium (100 to 140 µg/kg). Ventilation was controlled with air in oxygen. Analgesia was provided with sufentanyl and midazolam infusion. The antibiotic cefamandole was administered before the operation in a dosage of 2 gm intravenously 4 times daily up to 48 hours after the start of the operation. Dexamethasone (1 mg/kg) was administered 30 minutes before the operation.
10 After a median sternotomy, bovine lung heparin (300 IU/kg) was administered. Heparinization was checked by the activated clotting time, which had to be kept between 400 and 800 seconds during CPB.
CPB
Except for the cardiotomy reservoir, all components of the ECC were coated with heparin (Duraflo II, Bentley/Baxter Inc., Uden, The Netherlands). The ECC consisted of roller pumps and a membrane oxygenator (CM-50, Bentley/Baxter), a Bentley cardiotomy reservoir, and polyvinylchloride tubing, except for the pump tubing, which was silicone rubber. The ECC was primed with a gelatin-containing crystalloid solution (Gelifundol, Biotest Pharma GmbH, Dreieck, Germany) and contained 1500 IU of bovine heparin. CPB was performed with cardioplegic arrest and hypothermia (27° C nasopharyngeal temperature) with pump flows of 2.4 L/m2 per minute and maintenance of mean arterial blood pressures of 50 to 60 mm Hg. After CPB heparin was neutralized by means of protamine chloride (3 mg/kg, HoffmannlaRoche bv, Mijdrecht, The Netherlands).
Measurements
Bleeding time was measured according to the method described by Ivy, Nelson, and Bucher
11 on the day before the operation and 3 hours after arrival of the patient in the intensive care unit. During the operation blood samples were collected six times; after induction of anesthesia, 5 minutes before the start of CPB, after 5 minutes of CPB, just before release of the aortic crossclamp, at the end of CPB, and 20 minutes after protamine chloride administration. Platelet count and hematocrit value from citrated blood were measured by a cell counter (Cell-Dyn 610; Sequoia-Turner, Mountain View, Calif.). For biochemical assays, plasma was obtained by centrifuging whole blood in a chilled centrifuge at 1000 g for 10 minutes. Plasma heparin concentrations were measured by factor Xa inhibition by chromogenic assay (Kabi Diagnostica, Stockholm, Sweden) from the priming solution of the ECC and from plasma anticoagulated with citrate. Plasma with aprotinin-hirudin-citrate medium was prepared to determine fibrin degradation products (enzyme-linked immunosorbent assay, Organon Teknika, Turnhout, Belgium) and fibrinopeptide A concentrations (radioimmunoassay, Mallinckrodt, St. Louis, Mo.). Blood (2.5 ml) was collected into ß-thromboglobulin collecting tubes to determine plasma ß-thromboglobulin concentrations with radioimmunoassay (Amersham, Amersham, United Kingdom), and plasma with indomethacin-citrate medium was used for determination of thromboxane B2 concentrations (enzyme immunoassay, Cayman Chemical Co., Ann Arbor, Mich.). Postoperative blood loss through chest tube drainage and postoperative donor blood transfusions were measured up to 18 hours after the end of CPB.
Statistics
Values for biochemical tests from blood samples obtained during CPB were corrected for hemodilution by the prebypass hematocrit levels, except for heparin concentrations. Statistical analysis was performed by Student's t test. Results were expressed as mean ± standard error of the mean and a p value of less than 0.05 was considered to be significant.
RESULTS
Patients in the two groups did not differ significantly in regard to age, body weight, number of anastomoses, bypass time, and aortic crossclamp time (
Table I). Plasma heparin concentrations decreased immediately after the start of CPB because of dilution with the pump prime and decreased further in the control group but not in the Duraflo II group. Thus heparin concentrations were significantly higher in the Duraflo II group at the end of bypass (Fig. 1). Notably, we found 0.1 ± 0.05 IU/ml of heparin in the priming solution of the Duraflo II circuit before heparin was introduced.
|
|
Fibrinopeptide A concentrations remained low during the operation in both groups; in particular, during the heparinization period almost no fibrinopeptide A was generated (highest levels were 6.3 ± 1.4 ng/ml in the Duraflo II group and 6.8 ± 2.0 ng/ml in the control group). No significant differences were observed between the two groups.
Platelet count increased slightly after the start of CPB in both groups (highest values were 147% ± 10% of baseline in the Duraflo II group and 163% ± 45% of baseline in the control group) and remained stable thereafter. After administration of protamine chloride the platelet count decreased in both groups (129% ± 10% in the Duraflo II group and 138% ± 35% in the control group). We found no significant differences between the two groups.
After the operation the bleeding time appeared to be increased similarly in the two groupsfrom 6.6 to 10.5 minutes in the control group and from 7.5 to 11.0 minutes in the Duraflo II group. Differences between the two groups were not significant.
ß-Thromboglobulin increased significantly and progressively during CPB in both groups, the highest values being reached at the end of CPB (1910 ± 310 ng/ml in the Duraflo II group and 1810 ± 280 ng/ml in the control group). We found no significant differences between the two groups.
After the start of CPB thromboxane B2 increased slightly. After aortic crossclamping, however, a steep increase was found in both groups (highest values were 114 ± 10 ng/ml in the Duraflo II group and 104 ± 16 ng/ml in the control group). Differences between the two groups were not significant.
No statistically significant differences in blood loss and donor blood transfusions were observed between the two groups, although blood loss was somewhat lower in the Duraflo group (709 ± 101 ml in the Duraflo group versus 931 ± 136 ml in the control group).
In both groups one patient was excluded from the evaluation of blood loss and donor blood transfusions because of bleeding necessitating a rethoracotomy. One additional patient from the Duraflo II group was excluded because of the possibility of protamine chloride hypersensitivity (
Table II).
|
Heparin coating of the ECC appeared to prevent the reduction of plasma heparin levels. This could be explained by reduced activation of the clotting system and reduced consumption of heparin or by leaching of surface-bound heparin during CPB. Furthermore, we found that reduction of thrombin activity by heparin coating, which has been reported previously,
6, 7 was insufficient to preserve hemostasis during CPB, as indicated by a small but not significant reduction in postoperative blood loss and donor blood transfusions.
Hemostasis is dependent on the function of the clotting system, the platelets, and the fibrinolytic system. Evidence for the pivotal role of platelet activation and fibrinolysis in hemostasis during CPB has been obtained by studies with platelet inhibitors such as prostacyclin
12 and plasmin inhibitors such as aprotinin and tranexemic acid.
13, 14 Of the three main components of hemostasis, fibrinolysis might be of the greatest importance, because inhibitors of fibrinolysis have been most efficacious in reducing blood loss after CPB.
15, 16
The failure to preserve hemostasis by heparin coating can be explained in various ways. In this study we demonstrated that fibrinolysis and platelet activation were not inhibited at all by the heparin coating, as shown by the generation of fibrin degradation products and by the platelet release products ß-thromboglobulin and thromboxane B2. More specifically, hemostasis seemed in particular to be affected in the second period of CPB, after aortic crossclamping. This enhanced effect is indicated by the observation in the present study that activation products from clotting and fibrinolysis were markedly generated after aortic crossclamping. That period during CPB seems to initiate more detrimental effects to the blood than the first period of CPBduring aortic crossclamping. After aortic crossclamping we found that the hemostatic mechanism was affected despite heparin coating, as indicated by similar postoperative bleeding times, blood loss, and donor blood transfusions in the two groups. These effects are thought to be due to reperfusion of the heart and lungs, to rewarming, and to the introduction into the systemic circulation of cardiotomy suction blood, known to incorporate many blood-activating products such as thrombin, plasmin, and fibrin degradation products.
17 The result is activation of fibrinolysis and the clotting system and impairment of platelet function. Moreover, cardiotomy suction blood might have been activated additionally by the cardiotomy reservoir, the only uncoated component of the ECC. Past studies have shown that cardiotomy suction can destroy blood elements like platelets and can even obscure the improved hemocompatibility of a membrane oxygenator in comparison with that of a bubble oxygenator.
18 On the other hand, nowadays the total amount of cardiotomy blood suction in routine CABG operations is small.
We conclude that Duraflo II heparin coating of a CPB circuit with our technique does partly prevent activation of the clotting system, but it does not reduce fibrinolysis and platelet activation. Consequently, it does not significantly reduce postoperative bleeding time, blood loss, and donor blood transfusions in routine CABG operations. Additional measures are required to protect the blood during CPB, in particular in the period after aortic crossclamping, when pulmonary circulation and reperfusion of the ischemic heart are reestablished and cardiotomy suction blood is introduced into the systemic circulation.
References
This article has been cited by other articles:
![]() |
D. Reser, B. Seifert, M. Klein, T. Dreizler, P. Hasenclever, V. Falk, and C. Starck Retrospective analysis of outcome data with regards to the use of Phisio(R)-, Bioline(R)- or Softline(R)-coated cardiopulmonary bypass circuits in cardiac surgery Perfusion, November 1, 2012; 27(6): 530 - 534. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Raivio, R. Lassila, and J. Petaja Thrombin in Myocardial Ischemia-Reperfusion During Cardiac Surgery Ann. Thorac. Surg., July 1, 2009; 88(1): 318 - 325. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Murphy, E. A. Hessel II, and R. C. Groom Optimal Perfusion During Cardiopulmonary Bypass: An Evidence-Based Approach Anesth. Analg., May 1, 2009; 108(5): 1394 - 1417. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ranucci, A. Balduini, A. Ditta, A. Boncilli, and S. Brozzi A Systematic Review of Biocompatible Cardiopulmonary Bypass Circuits and Clinical Outcome Ann. Thorac. Surg., April 1, 2009; 87(4): 1311 - 1319. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Mangoush, S. Purkayastha, S. Haj-Yahia, J. Kinross, M. Hayward, F. Bartolozzi, A. Darzi, and T. Athanasiou Heparin-bonded circuits versus nonheparin-bonded circuits: an evaluation of their effect on clinical outcomes Eur J Cardiothorac Surg, June 1, 2007; 31(6): 1058 - 1069. [Abstract] [Full Text] [PDF] |
||||
![]() |
The Society of Thoracic Surgeons Blood Conservatio, V. A. Ferraris, S. P. Ferraris, S. P. Saha, E. A. Hessel II, C. K. Haan, B. D. Royston, C. R. Bridges, R. S.D. Higgins, G. Despotis, et al. Perioperative Blood Transfusion and Blood Conservation in Cardiac Surgery: The Society of Thoracic Surgeons and The Society of Cardiovascular Anesthesiologists Clinical Practice Guideline Ann. Thorac. Surg., May 1, 2007; 83(5_Supplement): S27 - S86. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. Jessen Pro: Heparin-Coated Circuits Should be Used for Cardiopulmonary Bypass Anesth. Analg., December 1, 2006; 103(6): 1365 - 1369. [Full Text] [PDF] |
||||
![]() |
S. Hamanaka, K. Tanemoto, E. Inagaki, T. Yamasawa, K. Yoshida, S. Mochizuki, M. Goto, T. Miyasaka, F. Kajiya, and N. Tanaka Direct Measurement of Nitric Oxide during Experimental Cardiopulmonary Bypass Journal of International Medical Research, May 1, 2005; 33(3): 295 - 300. [Abstract] [PDF] |
||||
![]() |
L. Lindholm, M. Westerberg, A. Bengtsson, R. Ekroth, E. Jensen, and A. Jeppsson A Closed Perfusion System With Heparin Coating and Centrifugal Pump Improves Cardiopulmonary Bypass Biocompatibility in Elderly Patients Ann. Thorac. Surg., December 1, 2004; 78(6): 2131 - 2138. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Noora, A. Lamy, K. M Smith, R. Kent, D. Batt, J. Fedoryshyn, and X. Wang The effect of oxygenator membranes on blood: a comparison of two oxygenators in open-heart surgery Perfusion, September 1, 2003; 18(5): 313 - 320. [Abstract] [PDF] |
||||
![]() |
G. M. Palatianos, C. N. Foroulis, M. I. Vassili, G. Astras, K. Triantafillou, E. Papadakis, A. A. Lidoriki, E. Iliopoulou, and E. N. Melissari A prospective, double-blind study on the efficacy of the bioline surface-heparinized extracorporeal perfusion circuit Ann. Thorac. Surg., July 1, 2003; 76(1): 129 - 135. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Borgdorff, D. Fekkes, and G. J. Tangelder Hypotension Caused by Extracorporeal Circulation: Serotonin From Pump-Activated Platelets Triggers Nitric Oxide Release Circulation, November 12, 2002; 106(20): 2588 - 2593. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S. Aldea, L. O. Soltow, W. L. Chandler, C. M. Triggs, C. R. Vocelka, G. I. Crockett, Y. T. Shin, W. E. Curtis, and E. D. Verrier Limitation of thrombin generation, platelet activation, and inflammation by elimination of cardiotomy suction in patients undergoing coronary artery bypass grafting treated with heparin-bonded circuits J. Thorac. Cardiovasc. Surg., April 1, 2002; 123(4): 742 - 755. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Borgdorff, R. H. van den Berg, M. A. Vis, G. C. van den Bos, and G. J. Tangelder PUMP-INDUCED PLATELET AGGREGATION IN ALBUMIN-COATED EXTRACORPOREAL SYSTEMS J. Thorac. Cardiovasc. Surg., November 1, 1999; 118(5): 946 - 952. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Wan, J.-L. LeClerc, M. Antoine, J.-M. DeSmet, A. P.C. Yim, and J.-L. Vincent Heparin-coated circuits reduce myocardial injury in heart or heart-lung transplantation: a prospective, randomized study Ann. Thorac. Surg., October 1, 1999; 68(4): 1230 - 1235. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Niimi, F. Ichinose, Y. Ishiguro, K. Terui, S. Uezono, S. Morita, and S. Yamane The Effects of Heparin Coating of Oxygenator Fibers on Platelet Adhesion and Protein Adsorption Anesth. Analg., September 1, 1999; 89(3): 573 - 573. [Abstract] [Full Text] [PDF] |
||||
![]() |
C B. Mahoney Heparin-bonded circuits: clinical outcomes and costs Perfusion, May 1, 1998; 13(3): 192 - 204. [Abstract] [PDF] |
||||
![]() |
Y. J. Gu, P. W. Boonstra, A. A. Rijnsburger, J. Haan, and W. van Oeveren Cardiopulmonary Bypass Circuit Treated With Surface-Modifying Additives: A Clinical Evaluation of Blood Compatibility Ann. Thorac. Surg., May 1, 1998; 65(5): 1342 - 1347. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. J. Gu, M. A. Mariani, W. van Oeveren, J. G. Grandjean, and P. W. Boonstra Reduction of the Inflammatory Response in Patients Undergoing Minimally Invasive Coronary Artery Bypass Grafting Ann. Thorac. Surg., February 1, 1998; 65(2): 420 - 424. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Shimono, Y. Shomura, I. Hioki, A. Shimamoto, H. Tenpaku, Y. Maze, K. Onoda, M. Takao, H. Shimpo, and I. Yada Silicone-Coated Polypropylene Hollow-Fiber Oxygenator: Experimental Evaluation and Preliminary Clinical Use Ann. Thorac. Surg., June 1, 1997; 63(6): 1730 - 1736. [Abstract] [Full Text] |
||||
![]() |
A. H. Kuitunen, L. J. Heikkila, and M. T. Salmenpera Cardiopulmonary Bypass With Heparin-Coated Circuits and Reduced Systemic Anticoagulation Ann. Thorac. Surg., February 1, 1997; 63(2): 438 - 444. [Abstract] [Full Text] |
||||
![]() |
J. M Courtney, B. M Matata, H. Q Yin, A. Esposito, A. Mahiout, D. P Taggart, and G. Lowe The influence of biomaterials on inflammatory responses to cardiopulmonary bypass Perfusion, May 1, 1996; 11(3): 220 - 228. [PDF] |
||||
![]() |
S. Daniel Review on the multifactorial aspects of bioincompatibility in CPB Perfusion, May 1, 1996; 11(3): 246 - 255. [Abstract] [PDF] |
||||
![]() |
B. P. Bidstrup Monitoring of Systemic Anticoagulation During Cardiopulmonary Bypass Ann. Thorac. Surg., March 1, 1996; 61(3): 781 - 782. [Full Text] |
||||
![]() |
R. C. Gorman, N. P. Ziats, A. K. Rao, N. Gikakis, L. Sun, M. M. H. Khan, N. Stenach, S. Sapatnekar, V. Chouhan, J. H. Gorman III, et al. SURFACE-BOUND HEPARIN FAILS TO REDUCE THROMBIN FORMATION DURING CLINICAL CARDIOPULMONARY BYPASS J. Thorac. Cardiovasc. Surg., January 1, 1996; 111(1): 1 - 12. [Abstract] [Full Text] |
||||
![]() |
P. W. Weerwind, J. G. Maessen, L. J. H. van Tits, R. K. Stad, E. J. Fransen, D. S. de Jong, and O. C. K. M. Penn INFLUENCE OF DURAFLO II HEPARIN-TREATED EXTRACORPOREAL CIRCUITS ON THE SYSTEMIC INFLAMMATORY RESPONSE IN PATIENTS HAVING CORONARY BYPASS J. Thorac. Cardiovasc. Surg., December 1, 1995; 110(6): 1633 - 1641. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||