|
|
||||||||
J Thorac Cardiovasc Surg 1994;108:57-62
© 1994 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
Rome, Italy
From the Department of Cardiac Surgery, University of Rome, Tor Vergata, Rome, Italy.
Received for publication June 28, 1993. Accepted for publication Oct. 17, 1993. Address for reprints: Ruggero De Paulis, MD, Cattedra di Cardiochirurgia, Universita di Roma, Tor Vergata, European Hospital, via Portuense 694, 00149 Roma, Italy.
Abstract
A Doppler echocardiographic study was performed to evaluate the hemodynamic performances of small diameter CarboMedics aortic valves (CarboMedics, Inc., Austin, Tex.) in patients with a mismatch between the prosthetic valve and body surface area. Fourteen patients receiving either a 19 mm (n= 7) or a 21 mm valve (n= 7) prosthesis were studied. Only patients with a body surface area greater than 1.65 m2were included in the study. Pulsed and continuous wave Doppler echocardiography was performed at rest and 2 minutes after treadmill exercise with the Bruce protocol. Peak and mean gradients across the valve prosthesis were estimated; effective orifice area, performance index, and discharge coefficient of the valve prosthesis were calculated. All patients achieved a significant increase in heart rate, systolic blood pressure, and cardiac output with exercise. Mean gradients ± standard deviation of the mean at rest and 2 minutes after exercise were 20.1 ± 7.1 mm Hg and 21.8 ± 9 mm Hg for the 19 mm prosthesis and 12.3 ± 3.4 mm Hg and 15.9 ± 3.9 mm Hg for the 21 mm prosthesis. The 19 mm valve prosthesis significantly increased the effective orifice area with exercise (1.02 ± 0.2 versus 1.20 ± 0.3; p< 0.05), whereas it was almost unmodified for the 21 mm valve (1.38 ± 0.2 versus 1.39 ± 0.3; p= not significant). Therefore, despite a similar increase in cardiac output with exercise, only the 21 mm valve prosthesis showed a significant increase in peak (25.4 ± 5.2 versus 34.9 ± 8.1 mm Hg) and mean gradient (p< 0.01). We conclude that small diameter CarboMedics valves have satisfactory hemodynamic performances even after strenuous exercise in patients with large body surface areas. The hemodynamic performances of the 19 mm valve prosthesis seem to be optimized with exercise. (J THORACCARDIOVASCSURG1994;108:57-62)
The CarboMedics aortic valve prosthesis (CarboMedics, Inc., Austin, Tex.) is a bileaflet valve incorporating several design modifications of existing bileaflet valves that should result in improved performances. Among these improved designs is the elimination of pivot guards, struts, and orifice projection, which is thought to reduce turbulence, and a reduced leaflet excursion that should ensure a more rapid and synchronous opening and closure. The plane of each leaflet forms angles of 25 and 78 degrees relative to the plane of the orifice ring in the closed and open position, respectively.
To evaluate the hemodynamic performances of small diameter CarboMedics prosthetic valves, we studied a selected patient population with a relatively great body surface area (BSA) and a small aortic anulus by means of Doppler echocardiography at rest and after isotonic exercise.
Patients and Methods
Patient population.
The study population comprised 14 consecutive patients (four men, ten women) ranging from 54 to 83 years of age (mean 67.2 ± 6.7 years) who underwent aortic valve replacement with a CarboMedics bileaflet valve prosthesis between January 1991 and January 1993. Only patients with a BSA greater than 1.65 m2 were included in the study. Seven of them received a 19 mm valve prosthesis, and seven had a 21 mm valve prosthesis. Five patients underwent valve replacement for severe aortic stenosis, two for valve regurgitation, and the remaining seven for mixed lesions. In two cases a mitral commissurotomy was associated, and in two cases a single coronary bypass was required. The interval between surgical intervention and follow-up Doppler evaluation ranged from 2 to 24 months (mean 11.6 ± 5.6). All patients were free of symptoms at follow-up examination and were in New York Heart Association class I or II. Medical ethics committee approval and informed patient consent for participation to the study was obtained in all cases. Patient characteristics were similar in both valve size groups and are reported in
Table I.
|
Doppler measurements and calculations.
Effective orifice area.
The effective orifice area (EOA) of the valve was calculated with the continuity equation by the simplified peak velocity method as EOA = CSA (Pk VLVOT/Pk Vjet), where CSA is the cross-sectional area of the LVOT, and Pk VLVOT and Pk Vjet are the maximum velocity in the LVOT and across the valve, respectively. This simplified method has shown an extremely good correlation with that of the original continuity equation.
3 Furthermore, cross-sectional area of the LVOT was determined assuming that the sewing ring of the valve is the cross-sectional area of the LVOT. This eliminates the variability of the various measurements and should be the method of choice when the valve size is known.
3
Effective area index.
The effective area index (EAI) that is a measure of how well the effective orifice area of the valve matches the BSA was calculated as EAI = EOA/BSA. This index was used as a confirmation of a good patient selection, that is, the presence of a mismatch between valve size and body surface area. According to Dumesnil and associates
4, it would appear that an indexed prosthetic valve area greater than 0.9 cm 2/m 2 would be a minimal requirement to minimize the postoperative gradient.
Performance index.
The performance index is a measure of how effectively the external dimension of the valve is used in providing forward flow, normalized to the valve size, and it is defined as the effective orifice area divided by the sewing ring area.
5 Most articles provide the performance index, making possible the comparison with other valve types.
Discharge coefficient.
The discharge coefficient is calculated from the effective orifice area divided by the calculated orifice area and is a measure of how effectively a valve uses its flow area.
Cardiac output.
The cardiac output (CO) was calculated using the formula CO = time velocity integral x effective orifice area x heart rate.
6
All parameters were calculated independently by two expert echocardiographers. A third investigator reviewed the echocardiograms in blinded manner to the results if the first two investigators were not in agreement. The magnitude of the difference and its variability (two standard deviations)
7 for maximum velocity values in patients with aortic valve prosthesis between interobserver and intraobserver agreement previously obtained in our laboratory were 0.028 ± 0.34 and 0.028 ± 0.36 m/s, respectively.
Specifications of the CarboMedics valve.
The CarboMedics valve is an all-pyrolytic carbon bileaflet valve. The valve size indicates the sewing ring diameter that is larger than the orifice diameter. The sewing ring area, the orifice diameter, the internal orifice area, and the calculated orifice area for the 19 and 21 mm valve prosthesis were obtained from the manufacturer and are reported in
Table II. Because of the presence of the two metallic leaflets the calculated orifice area is smaller than that derived from the orifice diameter.
|
Results
During exercise, all patients reached at least 85% of the maximal age predicted heart rate, as stated in the study protocol. The mean exercise duration was 5.9 ± 1.6 minutes ranging from 3.3 to 9.1 minutes. Heart rate and systolic blood pressure significantly increased in all patients. A significant increase in CO with exercise was observed in all cases, and no significant differences in exercise parameters were found between patients receiving different valve sizes (
Table I).
Doppler echocardiography findings at rest and after exercise are shown in
Table III. Only trivial valve regurgitation, compatible with that reported by the manufacturer, and without clinical significance, was recorded in all patients. As expected, patients with a 19 mm prosthesis had higher mean and instantaneous peak gradient than did those with a 21 mm prosthesis. However, despite a similar increase in CO with exercise, the proportional increase in flow velocity across the valve and therefore the gradient was significantly inferior for the 19 mm valve (
Table III). The effective area index of both valves confirmed the inadequacy of effective orifice area for each BSA (
Table III). The effective orifice areas in the resting condition were slightly inferior to the theoretical values reported from the manufacturer (1.02 versus 1.06 cm 2 for the 19 mm and 1.38 versus 1.41 cm 2 for the 21 mm prosthesis, respectively) (
Tables II and III); with exercise, however, a significant increase in effective orifice area was evident for the 19 mm, whereas it was almost unmodified for the 21 mm prosthesis. This finding was also reflected by the performance index which was similar after exercise for the two valve sizes.
|
Doppler ultrasonography has shown to be a dependable method to measure flow velocity and to estimate pressure gradients across valve prostheses. Sager and associates
8 have found a significant correlation between mean Doppler and catheterization gradients in patients with Hancock (Johnson & Johnson Cardiovascular, King of Prussia, Pa.) or Björk-Shiley (Shiley, Inc., Irvine, Calif.) aortic prostheses. More recently, in a study conducted in patients with different biologic and mechanical valves, Burstow and associates
9 confirmed the correlation (r = 0.94) between Doppler measurements and those obtained with simultaneous catheterization of the left ventricular and aortic chambers, thus concluding that Doppler echocardiography can be reliably used for the measurement of prosthetic gradients. Central-flow aortic prosthesis, like the St. Jude Medical valve (St. Jude Medical, Inc., St. Paul, Minn.) and all kinds of biologic valves, should theoretically allow accurate Doppler evaluation as do native aortic valves. However, although this has been demonstrated for normal and stenotic biologic valves,
10,11 it may not be true for bileaflet mechanical valves. Baumgartner and associates
12 demonstrated discrepancies between Doppler and catheter gradients in the St. Jude Medical prosthetic valves; Doppler gradients were always higher because of the presence of high localized pressure gradients within the valve and a significant pressure recovery after blood exits the valve. Because the CarboMedics valve is a bileaflet mechanical valve with characteristics similar to those of the St. Jude Medical valve, the same conclusion should apply. In fact, a recent clinical study comparing invasive and noninvasive techniques for the CarboMedics valve in aortic position confirmed that mean Doppler gradients exceeded mean catheter gradients by 4.3 ± 4.8 mm Hg.
13 We should therefore consider that the Doppler-defined gradients of our study could be a slight overestimation of catheter-derived gradients.
Another important point that needs to be taken into account is the problem of valve-patient mismatch. Jones and associates
14 evaluated the hemodynamic and clinical performances of standard Hancock bioprosthesis and recommended never to use the 19 mm prosthesis and to use the 21 mm only in very small patients. Schaff and associates
15 invasively evaluated the 19 mm Björk-Shiley valve concluding that it gives satisfactory hemodynamic performances provided that BSA is less than 1.7 m 2. Finally, Wortham and associates
16 evaluated the 19 and 21 mm St. Jude Medical valve prostheses, suggesting that the use of the 19 mm size should be restricted to patients with a BSA less than 1.8 m 2; however, the average BSA of their patients was only 1.49 m 2. In our selected series only two patients had a BSA less than 1.70 m 2, and the effective area indexes were extremely low (0.59 and 0.77 cm 2 for 19 and 21 mm prosthesis, respectively). These findings testify to the important mismatch between flow area and BSA in our patients. By comparison in a study by Ihlen and associates,
13 who used the same index, the effective area index for CarboMedics valve in aortic position was 0.72 and 0.89 cm 2 for the valve sizes 19 and 21 mm, respectively.
Therefore, considering the possibility of a gradient overestimation by the Doppler technique in this type of valve and the presence of valve prosthesisBSA mismatch, our results seem to indicate that the CarboMedics valve is an acceptable device with favorable hemodynamic performances even after strenuous exercise. The effective flow areas at rest were similar to the in vitro theoretic values, thus showing that the hemodynamic potential is fully realized in vivo for both valve sizes. It is noteworthy that the discharge coefficient in the 19 mm valve prosthesis has the tendency to be greater with exercise, implying a more efficient use of the orifice area. This tendency is in contrast with the study of Gray and associates,
17 which included a reduction of the discharge coefficient in St. Jude Medical valves after infusion of isoproterenol; this effect was ascribed to the turbulence occurring under conditions of higher flow. Because the discharge coefficient is an index of how effectively a valve uses its flow area, it is also an indirect measure of the efficiency of the moving parts of the valve as well as the optimization of the flow path across the prosthesis. Therefore, although the 19 mm valve has a low performance index as demonstrated by others
13 and confirmed by us, the design of the internal area and the leaflets is optimized. We can therefore speculate, according to our results, that the hemodynamic performances of the smallest CarboMedics valve is optimized after exercise probably through a more rapid and synchronous opening of the two leaflets, thus limiting turbulence. Furthermore, the design of leaflets and the pivot and hinge mechanism seem to minimize the areas of flow separation and turbulent shear stress that would cause a decrease in discharge coefficient. However, further investigations are necessary to define the reason for the apparent superior performance during exercise of the 19 mm CarboMedics valve when compared with the 21 mm valve.
Our data are consistent with the few studies available in the literature reporting a Doppler echocardiographic evaluation of patients with the CarboMedics valve. Ihlen and associates
13 reported maximal resting gradients of 30 and 21 mm Hg for patients with 19 and 21 mm valve prosthesis respectively, whereas Globitz and associates
18 found instantaneous peak gradients of 43 and 31 mm Hg, respectively. Our study also demonstrates that maximum and mean gradients increase only slightly with exercise, which shows the optimization of flow dymamics through the valve. Most studies evaluating the St. Jude Medical valve reported catheterization mean gradients varying from 7 to 14 mm Hg for the 19 mm valve and an average of 5 mm Hg for the 21 mm valve.
17,19 Similar values were reported on the catheterization-derived gradients for the CarboMedics valve.
13 However, although the hemodynamic performances of the two valves are comparable, the performance indexes found in our study, as well as those reported by Ihlen and associates,
13 are not higher than 0.40 cm 2; the St. Jude Medical valve, on the other hand, always shows a performance index varying from 0.56 to 0.71 cm 2 in different reports.
17,19 Furthermore, a performance index between 0.37 and 0.44 cm 2 has been reported for small Björk-Shiley valves,
10,20 and a performance index of 0.53 cm 2 has been reported for the 21 mm Medtronic Hall mechanical valve (Medtronic, Inc., Minneapolis, Minn.).
21 This finding led to the conclusion that, although the CarboMedics valve can be defined as a successful construction demonstrated by an in vivo discharge index equal to 1, when compared with most mechanical valves, it shows ineffective use of the total area of the prosthesis.
Conclusions and limitations of the study.
Several investigators have shown the value of Doppler echocardiography in the hemodynamic assessment of prosthetic biologic or mechanical valves. Although still controversial, a good correlation with catheterization study has been demonstrated in several studies.
8-13 The use of Doppler echocardiography has allowed us to obtain hemodynamic information after exercise more strenuous than would have been possible during catheterization.
A possible limitation in our measurements is that a reactive hypertrophy in the LVOT might develop in a certain number of elderly patients with aortic stenosis. Therefore, the gradient across the LVOT can be created not only by the prosthesis but also by the outflow tract hypertrophy. Although only five patients (three with the 19 mm and two with the 21 mm valve) had pure aortic valve stenosis, we cannot rule out the possibility of a less accurate echocardiographic assessment in these patients.
Our study involved only 14 patients, seven for each group, and this could account for less accurate statistical conclusions. However, the majority of reports are related to a small group of patients especially because they need to be categorized according to valve type and size. In our study, a further limitation was that only patients with a small aortic anulus and large BSA were selected, thus greatly narrowing the number of suitable patients. However, the evaluation of this small but particular group of patients allowed us to conclude that this type of prosthesis can be safely used in most clinical cases. A reduction in the ratio between sewing ring valve area and effective orifice area seems desirable.
References
This article has been cited by other articles:
![]() |
J. Chambers, J. Roxburgh, C. Blauth, J. O'Riordan, F. Hodson, and H. Rimington A randomized comparison of the MCRI On-X and CarboMedics Top Hat bileaflet mechanical replacement aortic valves: Early postoperative hemodynamic function and clinical events J. Thorac. Cardiovasc. Surg., September 1, 2005; 130(3): 759 - 764. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Niinami, S. Aomi, H. Tomioka, K. Nakano, and H. Koyanagi A comparison of the in vivo performance of the 19-mm St. Jude medical hemodynamic plus and 21-mm standard valve Ann. Thorac. Surg., October 1, 2002; 74(4): 1120 - 1124. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Vitale, I. Caldarera, C. Muneretto, R. Sinatra, A. Scafuri, E. Di Rosa, A. Contini, N. Tedesco, A. Pierangeli, M. Abbate, et al. Clinical evaluation of St Jude Medical Hemodynamic Plus versus standard aortic valve prostheses: The Italian multicenter, prospective, randomized study J. Thorac. Cardiovasc. Surg., October 1, 2001; 122(4): 691 - 698. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. E. Girard, F. A. Miller Jr., T. A. Orszulak, C. J. Mullany, S. Montgomery, W. D. Edwards, H. D. Tazelaar, J. F. Malouf, and A. J. Tajik Reoperation for prosthetic aortic valve obstruction in the era of echocardiography: trends in diagnostic testing and comparison with surgical findings J. Am. Coll. Cardiol., February 1, 2001; 37(2): 579 - 584. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Takakura, T. Sasaki, K. Hashimoto, T. Hachiya, K. Onoguchi, M. Oshiumi, and S. Takeuchi Hemodynamic evaluation of 19-mm Carpentier-Edwards pericardial bioprosthesis in aortic position Ann. Thorac. Surg., February 1, 2001; 71(2): 609 - 613. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Pibarot and J. G. Dumesnil Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention J. Am. Coll. Cardiol., October 1, 2000; 36(4): 1131 - 1141. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. D. Maslow, J. M. Haering, S. Heindel, J. Mashikian, R. Levine, and P. Douglas An Evaluation of Prosthetic Aortic Valves Using Transesophageal Echocardiography: The Double-Envelope Technique Anesth. Analg., September 1, 2000; 91(3): 509 - 516. [Abstract] [Full Text] [PDF] |
||||
![]() |
J R Gonzalez-Juanatey, M V Fernandez, F G Sampedro, J M Garcia-Acuna, J B Garcia-Bengoechea, A A Cendon, and M G de la Pena Haemodynamic performance of aortic pericardial bioprostheses and bileaflet prostheses at rest and during exercise: implications for the surgical management of patients with small aortic roots Heart, August 1, 1999; 82(2): 149 - 155. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Bernal, R. Martin-Duran, J. M. Rabasa, and J. M. Revuelta The CarboMedics "Top-Hat" supraannular prosthesis Ann. Thorac. Surg., May 1, 1999; 67(5): 1299 - 1303. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Vijayanagar, G. Chan, R. Musunuru, N. Sastry, I. Siegman, N. Rattehalli, and M. Cortelli Aortic Valve Replacement Without Annular Enlargement in Patients with Small Aortic Roots Asian Cardiovasc Thorac Ann, December 1, 1998; 6(4): 279 - 284. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Cam, H. Gercekoglu, S. Celik, M. Gursurer, G. Tayyareci, H. Karabulut, A. Narin, T. Tezel, and B. Yigiter Dobutamine Stress Test to Evaluate Different Sizes of Prosthetic Aortic Valves Asian Cardiovasc Thorac Ann, September 1, 1998; 6(3): 166 - 173. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Gonzalez-Juanatey Assessment of the hemodynamic performance of small-size aortic valve prostheses J. Thorac. Cardiovasc. Surg., June 1, 1997; 113(6): 1122 - 1123. [Full Text] |
||||
![]() |
R. De Paulis, L. Sommariva, G. M. De Matteis, E. Caprara, F. Tomai, A. P. de Peppo, P. Polisca, C. Bassano, and L. Chiariello EXTENT AND PATTERN OF REGRESSION OF LEFT VENTRICULAR HYPERTROPHY IN PATIENTS WITH SMALL SIZE CARBOMEDICS AORTIC VALVES J. Thorac. Cardiovasc. Surg., May 1, 1997; 113(5): 901 - 909. [Abstract] [Full Text] |
||||
![]() |
M. L. McDonald, R. C. Daly, H. V. Schaff, C. J. Mullany, F. A. Miller, J. J. Morris, and T. A. Orszulak Hemodynamic Performance of Small Aortic Valve Bioprostheses: Is There a Difference? Ann. Thorac. Surg., February 1, 1997; 63(2): 362 - 366. [Abstract] [Full Text] |
||||
![]() |
T. Carrel, U. Zingg, R. Jenni, B. Aeschbacher, and M. I. Turina Early In Vivo Experience With the Hemodynamic Plus St. Jude Medical Heart Valves in Patients With Narrowed Aortic Annulus Ann. Thorac. Surg., May 1, 1996; 61(5): 1418 - 1422. [Abstract] [Full Text] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini COMPARISON OF HEMODYNAMIC PERFORMANCES OF ST. JUDE MEDICAL AND CARBOMEDICS 21 MM AORTIC PROSTHESES BY MEANS OF DOBUTAMINE STRESS ECHOCARDIOGRAPHY J. Thorac. Cardiovasc. Surg., February 1, 1996; 111(2): 408 - 415. [Abstract] [Full Text] |
||||
![]() |
M. B. Izzat, M. Caputo, and G. D. Angelini Evaluation of the hemodynamic performance of stentless porcine aortic valves Ann. Thorac. Surg., November 1, 1995; 60(5): 1461 - 1461. [PDF] |
||||
![]() |
M. B. Izzat, I. Birdi, P. Wilde, A. J. Bryan, and G. D. Angelini Evaluation of the hemodynamic performance of small carbomedics aortic prosthesis using dobutamine-stress doppler echocardiography Ann. Thorac. Surg., October 1, 1995; 60(4): 1048 - 1052. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |