J Thorac Cardiovasc Surg 1994;108:221-230
© 1994 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
Belfast, Northern Ireland
From the Department of Cardiac Surgery, Royal Victoria Hospital, and the Department of Epidemiology and PublicHealth, Queen's University, Belfast, Northern Ireland.
Received for publication Aug. 16, 1993. Accepted for publication Mar. 16, 1994. Address for reprints: Hugh O'Kane, FRCS, Department of Cardiac Surgery, Royal Victoria Hospital, Belfast, BT12 6BA, Northern Ireland.
Abstract
From May 1979 until July 1992, 1184 patients, 2 to 86 years of age (median = 59 years), received 1367 St. Jude Medical prostheses (578 aortic, 440 mitral, 3 tricuspid, and 170 multiple). Overall early mortality was 4% with 2.4%, 4.3%, and 8.2% after aortic, mitral and multiple valve replacement, respectively. Follow-up was 100% complete (4936 patient-years). Actuarial survival at 10 years (including both early and late deaths) was 71% ± 4.1% after aortic valve replacement, 58.8% ± 5.1% after mitral valve replacement, and 58.9% ± 9.6% after multiple valve replacement. Multivariate analysis identified age, previous operations, diabetes mellitus, extent of coronary artery disease, preoperative New York Heart Association class, and additional procedures as independent prognostic factors for overall survival. All patients including 20 children (2 to 18 years of age) received sodium warfarin. The linearized risk per 100 patient-years for all embolic events (major and minor) was 2.4, 4.4, and 5.0 after aortic, mitral, and multiple valve replacement, respectively; when only major events were considered, the linearized risks were 0.33, 1.17, 1.54, respectively. Freedom from major systemic embolism at 10 years was 96% ± 2% after aortic valve replacement, 88% ± 4% after mitral valve replacement, and 89% ± 3% after multiple valve replacement. Five patients had valve thrombosis, four of whom definitely received inadequate anticoagulation therapy. Ninety-seven patients had 107 episodes of anticoagulant-related hemorrhage (2.2/100 patient-years) accounting for 17 late deaths; 14 of the deaths involved an international normalized ratio in excess of 3.5. To reduce the rate of thromboembolism without increasing the rate of anticoagulant-related hemorrhage, we propose that the international normalized ratio should be kept between 2.5 and 3.0. No recorded structural failure or significant hemolysis was found in the absence of periprosthetic leak. This experience encourages us to continue using the St. Jude Medical prosthesis. (J THORACCARDIOVASCSURG1994;108:221-30)
With the introduction of each new prosthetic valve, it behooves surgeons and cardiologists to monitor its performance carefully. It is important that any observation should be standardized, especially with regard to thromboembolic events, as suggested by McGoon,
1 so that a reasonable attempt can be made to compare prostheses with each other. We introduced the St. Jude Medical prosthesis into our unit in 1979 and herein record a 13-year follow-up presented strictly within the guidelines proposed by the Society of Thoracic Surgeons.
2
PATIENTS AND METHODS
Follow-up data were obtained during a 4-month closing interval (August 1992 until the end of November 1992). Between May 1979 and July 1992, 1184 patients received 1367 St. Jude Medical prostheses (St. Jude Medical, Inc., St. Paul, Minn.); 515 were male (43.5%) and 669 were female (56.5%) patients, ranging from 2 to 86 years of age (median = 59 years), with 11% being 70 years of age or older. Four hundred ten patients (35%) had previous cardiac operations. In the preoperative period, 56% of patients were in New York Heart Association (NYHA) functional class III, 26% were in class IV, and the remainder were in class I or II. Of the valves that were implanted one at a time, 578 were aortic, 440 were mitral, and three were tricuspid. One hundred fifty-eight patients underwent aortic and mitral valve replacement (AVR + MVR), five underwent mitral and tricuspid valve replacement (MVR + TVR), one underwent aortic and tricuspid valve replacement (AVR + TVR), and six underwent triple valve replacement (AVR + MVR + TVR). In 284 cases (24%), additional procedures were required. Coronary artery bypass grafting (CAGB) was performed in 13% (97 AVR, 45 MVR, 11 double valve), 83 patients had other valves repaired, 11 had correction of a congenital defect, 10 had replacement of the ascending aorta, 9 had resection of subaortic stenosis, and the remaining 18 had miscellaneous procedures.
Operative technique included cardiopulmonary bypass with moderate hypothermia (28° to 32° C), topical cooling, and either crystalloid or blood cardioplegic solution. A continuous suture technique with polypropylene was used to insert 1339 valves, an interrupted suture technique was used for 20 valve replacements (11 mitral and 9 aortic), and a mixed technique was used for eight MVRs. Each valve was soaked in a 1% neomycin solution before insertion.
Sodium warfarin therapy was started either 48 hours after the operation or as soon as the chest drains were removed; the aim was to achieve an international normalized ratio of 2.0 to 2.5 (prothrombin time ratio of 1.7 to 1.9.) One surgeon administered subcutaneous heparin (5000 IU thrice daily) simultaneously until this international normalized ratio range was achieved. Two surgeons used dipyridamole (100 mg daily) in addition.
All patients who were still alive during the closing interval were contacted by telephone or through their family doctors. In addition, hospital records were reviewed, and the cardiologists in charge of the patients were contacted as necessary. Clinical details on deceased patients were obtained from hospital records, family doctors, the Office of the Registrar General, and relatives. In 26% of the deceased patients, postmortem examination was carried out. Echocardiographic studies to exclude periprosthetic leaks were performed as necessary based on clinical indications. Follow-up was 100% complete with a total of 4936 patient-years (aortic 2441 patient-years, mitral 1782 patient-years, and multiple 713 patient-years). The mean follow-up was 4.2 years.
This report strictly follows the guidelines of the Society of Thoracic Surgeons
2 for reporting morbidity and mortality after cardiac valvular operations. It is worth emphasizing that the term periprosthetic leak is used to indicate leaks in the absence of infection. In addition, McGoon's
1 strictures concerning the reporting of embolic events were adhered to. Patients were specifically asked if they had experienced blurred or double vision, dizziness, numbness, weakness, inability to move a part or to speak, blackout, or a stroke. Dizziness was considered relevant if it occurred repeatedly and suddenly, in the absence of sudden movements or change of position, and if the patient had no ear problems. Dizzy spells occurring both before and after the operation were not included under the category of embolic events.
Statistical methods
Early mortality rates were compared in subgroups of the patients defined by potential prognostic variables. Univariate comparisons were performed with the
2 test (or Fisher's exact probability test). Logistic regression analysis was then used to define a subset of variables that were independently predictive of early mortality. Only variables that gave significant likelihood ratio statistics were retained in the model. Results were summarized as relative odds on early death with 95% confidence limits.
The actuarial method was used to summarize late mortality and overall mortality. Survival estimates are reported ± standard error of the estimate. Univariate comparisons were performed with Breslow's test. Cox's proportional hazards model was then used to identify independent prognostic variables. Partial-likelihood ratio tests were used, and only significant variables were retained in the model. Results were summarized as relative hazards with 95% confidence limits. Tests were conducted at the 5% level of significance. All analyses were performed with the SPSS for Windows software package (SPSS Inc., Chicago, Ill.).
RESULTS
Early mortality
Forty seven patients (4%) died within 30 days or during the same hospital admission even if the hospital stay was longer than 30 days. Sixty percent of these cases had a postmortem examination carried out. Mortality according to the location of the prostheses was 2.4% after AVR, 4.3% after MVR, and 8.2% after multiple valve replacement (
Table I). Thirteen patients died on the operating table, including three who had a ruptured left ventricle during MVR. The other causes of early mortality were multiple organ failure (n = 13), congestive heart failure (n = 5), arrhythmia (n = 4), tamponade (n = 2), intraoperative cerebrovascular accidents (n = 3), postoperative cerebral infarcts (n = 3), further aortic dissection (n = 1), prosthetic endocarditis (n = 1), and septicemia (n = 2).
|
|
Late mortality
One hundred sixty patients (14%) had died by the time of the closing interval, 26% of whom had postmortem examination. The causes of late death are listed (
Tables III and
IV). Diagnosis was deferred in two patients because we were unable to ascertain the exact cause of death. The sudden unexplained death in the AVR case without an autopsy was regarded as valve related (
Table IV). No recorded deaths were due to structural failure.
|
|
The Cox's model analysis indicated that age group, previous operation, diabetes mellitus, extent of coronary artery disease, and additional procedures were independently prognostic of late mortality (
Table V).
|
Overall survival
The overall survival (including both early and late mortality) is shown in Fig. 1. Of the 977 patients who were alive by the time of the study, 88% were in NYHA functional class I or II, and 12% were in class III or IV.
|
|
Systemic embolism
Ninety-eight patients (8.3%) had 174 embolic events (40 major and 134 minor).* Nine of the events were fatal, seven cerebral and two mesenteric. Sixty-eight patients had only one embolic event, eleven had two events, eleven had three events, and eight had more than three events. The great majority (89%) of embolic events were cerebral. In 41 patients, the international normalized ratio at the time of the event was within our accepted therapeutic range (2.0 to 2.5), whereas in 62 events it was below the therapeutic range (<2.0). The international normalized ratio at the time of the event was not available in 71 episodes, the majority of which (66 of 71) were minor and the patient therefore was not admitted to the hospital.
The freedom from all embolic events is shown in Fig. 2 and was significantly better after AVR than after either MVR or multiple valve replacement. No significant difference was found between the MVR group and the multiple valve replacement group.
|
|
Anticoagulant-related hemorrhage
Ninety-seven patients (8.2%) had 107 episodes of hemorrhage necessitating hospital admission or blood transfusion or resulting in death. Ten patients each had two hemorrhagic episodes, and two of these died of intracranial hemorrhage. In 101 episodes (94%), the international normalized ratio at the time of the episode was known. In 82 cases, it was above our therapeutic range, whereas in 19 it was within our range. Seventeen episodes were fatal. In 14 of these cases, the international normalized ratio was known at the time of the hemorrhage and it was found to be in excess of 3.5. The vast majority of these fatalities (15 of 17) were due to intracranial hemorrhage, with the remaining two deaths resulting from severe gastrointestinal bleeding. An additional eight patients had nine episodes of subdural hematoma, seven requiring emergency burr hole drainage. The linearized risk of anticoagulant-related hemorrhage was 2.2/100 patient-years for all patients, 1.7/100 patient-years for AVR, 2.8/100 patient-years for MVR, and 2.1/100 patient-years for multiple valve replacement.
Estimates for freedom from anticoagulant-related hemorrhage are shown in Fig. 4. The risk of anticoagulant-related hemorrhage was significantly greater with MVR than with AVR (p = 0.004). However, no significant difference was found between the AVR and multiple valve replacement groups.
|
Periprosthetic leak
Forty-three patients had 44 periprosthetic leaks and necessitated 14 reoperations in 13 patients, 13 in the mitral and one in the aortic area (p < 0.001). The linearized risk per 100 patient-years was 0.87 for all patients, 0.3 after AVR, 1.5 after MVR, and 1.26 after multiple valve replacement. The linearized risk per 100 patient-years for periprosthetic leak necessitating reoperation was 0.28 for all patients. The actuarial freedom from periprosthetic leak is shown in Fig. 5.
|
|
Freedom from all valve-related events
Embolic complications, valve thrombosis, anticoagulant-related hemorrhage, prosthetic valve endocarditis, periprosthetic leak, reoperation, and valve-related death were reviewed overall for AVR, MVR, and multiple valve replacement. The actuarial estimate of the percentage free from a valve-related event is shown in Fig. 7. A significant difference was found in the occurrence of valve-related events between the AVR group and each of the other groups.
|
Our results once again draw attention to the fact that thromboembolism and anticoagulant-related hemorrhage remain the most numerous and life-threatening complications of mechanical prosthetic valves. They were the major causes of our valve-related late mortality, accounting for 24 deaths (71%). In the 14 hemorrhage-related deaths in which the international normalized ratio was known, the ratio was above 3.5 at the time of the event. This finding would lead us to hypothesize that if the international normalized ratio was not allowed to exceed an upper limit of 3.0 the prevalence of anticoagulant-related hemorrhage would fall. At same time we must be aware of what could happen to the embolic events if we so adjusted the international normalized ratio. From our own results we know that the international normalized ratio was under 2.5 in all 103 embolic events in which it was known, which included 35 of the major events (88%). Combining these two observations, we postulate that an international normalized ratio in the range of 2.5 to 3.0 would be the optimum for reducing the amount of both embolic events and anticoagulant-related hemorrhage.
Our conclusion is in accordance with that of Horskotte and associates
3 who state that an international normalized ratio of 2.5 to 3.0 would reduce thromboembolic and bleeding events in patients with St. Jude Medical prostheses. These authors recommend a higher international normalized ratio (2.7 to 3.6) for patients with mitral St. Jude Medical prostheses who are in atrial fibrillation, but we would be reluctant to recommend an international normalized ratio above 3.0 because of our fear of fatal hemorrhage. In addition, although we found the rate of systemic embolism was higher in patients in atrial fibrillation (10.3%) compared with those in sinus rhythm (7.1%), the difference was not significant.
When we tried to do a comparative analysis of our prevalence of anticoagulant-related hemorrhage and thromboembolic complications with those in the literature, we were frustrated by a lack of conformity in the reported results. In none of the reports we looked at
4-14 could we find a record of the state of anticoagulation at the time of a hemorrhagic episode. Similarly for thromboembolism, no standard form of reporting events was found. Some reported only major emboli,
7 others reported events only after the first 6 months,
15 and with others
4,5 it was not clear how many patients had recurrent episodes or whether each event was included in their calculation of the linearized risks. Where we could find reasonably comparable definitions of thromboembolism
12,13, our overall embolic rate was similar. With those authors
7 who only recorded their major events, we found that we had a much lower prevalence of major events (0.3 versus 2.1 for AVR, 1.2 versus 4.3 for MVR, and 1.5 versus 4.6 for multiple valve replacement). This experience reinforces the difficulty in comparing the results from different groups that do not adhere to a standard definition. The excellent paper written by McGoon
1 in 1984 gives guidelines for reporting thromboembolic complications which could be followed before papers are submitted and accepted for publication. The adherence to these guidelines and the guidelines approved by the Council of the Society of Thoracic Surgeons
2 would remove some of the difficulty in comparing the results of different valve prostheses.
Prosthetic valve endocarditis was the next major cause of morbidity and mortality (after thromboembolism and anticoagulant-related hemorrhage). Our linearized prevalence of prosthetic valve endocarditis is similar to that of other authors
4,5,7,11 reporting experience with different prostheses but is lower than that reported by Lund and associates
14 with the Starr-Edwards prosthesis (Baxter Healthcare Corp., Edwards Division, Santa Ana, Calif.). Six of the seven patients who died had their prosthetic valve endocarditis late (i.e., after 3 months). Therefore, the reduction of prosthetic valve endocarditis depends not only on a rigid aseptic technique at the time of operation but just as importantly on the careful advice to patients concerning possible sources of bacteremia and their prompt treatment.
The analysis of the effect that CABG has on early and late mortality gave us interesting results. For early mortality, CABG was associated with a slight but not statistically significant increase in risk (
Table II), whereas for late mortality, CABG was identified as a factor that positively affected survival (
Table VI) when the extent of the patient's coronary artery disease is taken into account. The beneficial effect on late mortality is most likely due to a decrease in ischemic events and an improvement in left ventricular function over the years. Our overall survival results would encourage us to perform CABG in addition to valve replacement for its protective long-term effect, even though we found it to be associated with a slight increase in the risk of early mortality.
Surprisingly, we found that older age was neither a univariate nor a multivariate predictor of early mortality in our series. This removal of age as a risk factor is probably related to the energetic and detailed attention we pay to the immediate postoperative care of the elderly, especially regarding renal and respiratory problems. This experience, however, is not shared by others
16 who have identified age as an important risk factor both for MVR and AVR. Age still remained a negative prognostic factor for late mortality (
Table V) and for overall survival (
Table VI).
The increased early mortality that we found after MRV and multiple valve replacement could not be accounted for by either the NYHA preoperative class or the higher rate of previous operation in these cases, and the results of our analysis led us to conclude that this increased mortality was due to the greater technical difficulties that MVR and multiple valve replacement entail. The significantly higher early morality with aortic regurgitation as opposed to aortic stenosis in AVR was not totally unexpected and was probably related to the fact that left ventricular function is better preserved in aortic stenosis. Like other authors,
16 we found that the nature of the mitral leision (i.e., regurgitation, stenosis, or mixed) had no prognostic significance for early death.
Because a continuous suture technique
17 was used to implant the vast majority (98%) of all the valves in this series, we wondered whether this technique could account for the significantly higher rate of periprosthetic leak that we found in the MVR group compared with the AVR group. However, when we looked at the experience of others with various valves, we found that they too had a higher prevalence of periprosthetic leak with mitral compared with aortic valve replacement. For instance, Lindblom with the Björk-Shiley prosthesis (Shiley, Inc., Irvine, Calif.)
8,9 found a much greater (3.5 times) prevalence of periprosthetic leak in the mitral compared with the aortic position despite having used interrupted sutures in both locations. In addition, Keenan and associates
17 who used an interrupted technique with mitral Medtronic prostheses (Medtronic, Inc., Minneapolis, Minn.) reported a prevalence of periprosthetic leak similar to our own. We concluded, therefore, that no one suture technique could be implicated in causing periprosthetic leak but that there is an impression that the mitral location has a higher prevalence of periprosthetic leak than does the aortic.
Our overall early mortality of 4% compares favorably with that reported previously for the St. Jude Medical valve
3-6,19-24 and other mechanical prostheses.
7-12,18,25-27 Of the 47 early deaths, because of the high rate of postmortem examination (60%) and because of the close clinical monitoring especially with echocardiography and fluoroscopic valve screening, we are confident that mechanical failure was not a cause of death. It is worth pointing out that the comparison of mortality between different reports has little value in the absence of a comprehensive clinical profile of the patients.
The results of our analysis encourage us to continue using the St. Jude Medical prosthesis with its proven excellent hemodynamics and absence of structural failure over 13 years. One of the major advantages of this valve is its low profile, which makes it especially useful in children and adults with small ventricles, particularly women. We believe that, by refining our anticoagulation regimen as described previously in this article and by possible further modification such as the introduction of low-dose aspirin therapy
28 into the regimen, we may reduce further the prevalence of systemic embolism and anticoagulant-related hemorrhage.
Footnotes
*Present address: Division of Cardiothoracic Surgery, University of Ottawa Heart Institute, Ottawa Civic Hospital, 1053 Carling Ave., Ottawa, Ontario, Canada K1Y 4E9. ![]()
*"Minor representing a transient episode and major indicating permanent residual effect or the requirement for operation or other significant therapeutic modality...."
1 ![]()
References
year experience with the St. Jude Medical cardiac valve prosthesis. J THORAC CARDIOVASC SURG 1985;90:137-44.[Abstract]This article has been cited by other articles:
![]() |
M. Maganti, V. Rao, S. Armstrong, C. M. Feindel, H. E. Scully, and T. E. David Redo valvular surgery in elderly patients. Ann. Thorac. Surg., February 1, 2009; 87(2): 521 - 525. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. D. Desai and G. T. Christakis Bioprosthetic Aortic Valve Replacement: Stented Pericardial and Porcine Valves Card. Surg. Adult, January 1, 2008; 3(2008): 857 - 894. [Full Text] |
||||
![]() |
H. V. Schaff and R. M. Suri Multiple Valve Disease Card. Surg. Adult, January 1, 2008; 3(2008): 1129 - 1158. [Full Text] |
||||
![]() |
G. Gerosa, R. Carta, M. Montisci, L. Leoni, S. Iliceto, G. Rizzoli, and F. di Marco How to deal with recipients of valves prone to structural failure in the 2000s: Padua experience with the TRI Technologies valve. Ann. Thorac. Surg., September 1, 2006; 82(3): 858 - 864. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Jones, D. Lovell, G. W. Cran, and S. W. MacGowan Impact of coronary artery bypass grafting on survival after aortic valve replacement Interactive CardioVascular and Thoracic Surgery, June 1, 2006; 5(3): 327 - 330. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. S. De Santo, G. Romano, A. D. Corte, F. Tizzano, A. Petraio, C. Amarelli, M. De Feo, G. Dialetto, M. Scardone, and M. Cotrufo Mitral mechanical replacement in young rheumatic women: Analysis of long-term survival, valve-related complications, and pregnancy outcomes over a 3707-patient-year follow-up J. Thorac. Cardiovasc. Surg., July 1, 2005; 130(1): 13 - 19. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Tominaga, K. Kurisu, Y. Ochiai, Y. Tomita, M. Masuda, S. Morita, and H. Yasui A 10-Year Experience With the Carbomedics Cardiac Prosthesis Ann. Thorac. Surg., March 1, 2005; 79(3): 784 - 789. [Abstract] [Full Text] [PDF] |
||||
![]() |
J P A Puvimanasinghe, J J M Takkenberg, M B Edwards, M J C Eijkemans, E W Steyerberg, L A van Herwerden, K M Taylor, G L Grunkemeier, J D F Habbema, and A J J C Bogers Comparison of outcomes after aortic valve replacement with a mechanical valve or a bioprosthesis using microsimulation Heart, October 1, 2004; 90(10): 1172 - 1178. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Vink, R. A. Kraaijenhagen, and M. Levi Reply J. Am. Coll. Cardiol., September 1, 2004; 44(5): 1144 - 1145. [Full Text] [PDF] |
||||
![]() |
H. O. O'Kane The optimal intensity of vitamin K antagonists in patients with mechanical heart valves: A meta-analysis J. Am. Coll. Cardiol., September 1, 2004; 44(5): 1143 - 1143. [Full Text] [PDF] |
||||
![]() |
A. J. Murday, A. Hochstitzky, J. Mansfield, J. Miles, B. Taylor, E. Whitley, and T. Treasure A prospective controlled trial of st. jude versus starr edwards aortic and mitral valve prostheses Ann. Thorac. Surg., July 1, 2003; 76(1): 66 - 73. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. L. Grunkemeier and Y. Wu "Our complication rates are lower than theirs": Statistical critique of heart valve comparisons J. Thorac. Cardiovasc. Surg., February 1, 2003; 125(2): 290 - 300. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. D. Desai and G. T. Christakis Stented Mechanical/Bioprosthetic Aortic Valve Replacement Card. Surg. Adult, January 1, 2003; 2(2003): 825 - 856. [Full Text] |
||||
![]() |
T. Gudbjartsson, S. Aranki, and L. H. Cohn Mechanical/Bioprosthetic Mitral Valve Replacement Card. Surg. Adult, January 1, 2003; 2(2003): 951 - 986. [Full Text] |
||||
![]() |
H. V. Schaff and D. H. Marsh Multiple Valve Disease Card. Surg. Adult, January 1, 2003; 2(2003): 1017 - 1045. [Full Text] |
||||
![]() |
M. S. Edwards, G. B. Russell, A. F. Edwards, J. W. Hammon Jr, A. R. Cordell, and N. D. Kon Results of valve replacement with omniscience mechanical prostheses Ann. Thorac. Surg., September 1, 2002; 74(3): 665 - 670. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. H. H. Lim, M. Caputo, R. Ascione, J. Wild, R. West, G. D. Angelini, A. J. Bryan, and S. A. M. Calafiore Prospective randomized comparison of CarboMedics and St Jude Medical bileaflet mechanical heart valve prostheses: An interim report J. Thorac. Cardiovasc. Surg., January 1, 2002; 123(1): 21 - 32. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Jones, H. O'Kane, D. J. Gladstone, M. A.I. Sarsam, G. Campalani, S. W. MacGowan, J. Cleland, and G. W. Cran Repeat heart valve surgery: Risk factors for operative mortality J. Thorac. Cardiovasc. Surg., November 1, 2001; 122(5): 913 - 918. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Sidhu and H. O. O'Kane Self-managed anticoagulation: results from a two-year prospective randomized trial with heart valve patients Ann. Thorac. Surg., November 1, 2001; 72(5): 1523 - 1527. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y.-Y. Cen, D. D. Glower, K. Landolfo, J. E. Lowe, R. D. Davis, W. G. Wolfe, C. Pieper, and B. Peterson Comparison of survival after mitral valve replacement with biologic and mechanical valves in 1139 patients J. Thorac. Cardiovasc. Surg., September 1, 2001; 122(3): 569 - 577. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Khan, A. Trento, M. DeRobertis, R. M. Kass, M. Sandhu, L. S. C. Czer, C. Blanche, S. Raissi, G. P. Fontana, W. Cheng, et al. Twenty-year comparison of tissue and mechanical valve replacement J. Thorac. Cardiovasc. Surg., August 1, 2001; 122(2): 257 - 269. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Sidhu, H. O'Kane, N. Ali, D. J. Gladstone, M. A.I. Sarsam, G. Campalani, and S. W. MacGowan Mechanical or bioprosthetic valves in the elderly: a 20-year comparison Ann. Thorac. Surg., May 1, 2001; 71 (2007): S257 - S260. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-H. Li, J. Hahn, P. Urbanski, M. Torka, G. L. Grunkemeier, and R. W. Hacker Intermediate-term results with 1,019 Carbomedics aortic valves Ann. Thorac. Surg., April 1, 2001; 71(4): 1181 - 1187. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. P. Remadi, O. Baron, C. Roussel, P. Bizouarn, A. Habasch, P. Despins, J. L. Michaud, and D. Duveau Isolated Mitral Valve Replacement With St Jude Medical Prosthesis : Long-Term Results: A Follow-Up of 19 Years Circulation, March 20, 2001; 103(11): 1542 - 1545. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Ishii, T. Nitta, M. Fujii, H. Ogasawara, H. Iwaki, N. Ohkubo, and S. Tanaka Serial change in the atrial transport function after the radial incision approach Ann. Thorac. Surg., February 1, 2001; 71(2): 572 - 576. [Abstract] [Full Text] [PDF] |
||||
![]() |
D Vinereanu, A A Ionescu, and A G Fraser Assessment of left ventricular long axis contraction can detect early myocardial dysfunction in asymptomatic patients with severe aortic regurgitation Heart, January 1, 2001; 85(1): 30 - 36. [Abstract] [Full Text] |
||||
![]() |
O. Lund, S. L. Nielsen, H. Arildsen, L. B. Ilkjaer, and H. K. Pilegaard Standard aortic St. Jude valve at 18 years: performance profile and determinants of outcome Ann. Thorac. Surg., May 1, 2000; 69(5): 1459 - 1465. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Yau, Y. A. F. El-Ghoneimi, S. Armstrong, J. Ivanov, and T. E. David MITRAL VALVE REPAIR AND REPLACEMENT FOR RHEUMATIC DISEASE J. Thorac. Cardiovasc. Surg., January 1, 2000; 119(1): 53 - 61. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. L. Zellner, J. M. Kratz, A. J. Crumbley III, M. R. Stroud, S. M. Bradley, R. M. Sade, and F. A. Crawford Jr Long-term experience with the St. Jude medical valve prosthesis Ann. Thorac. Surg., October 1, 1999; 68(4): 1210 - 1218. [Abstract] [Full Text] [PDF] |
||||
![]() |
W.R. E. Jamieson, R. T. Miyagishima, G. L. Grunkemeier, E. Germann, C. Henderson, G. J. Fradet, L. H. Burr, and S. V. Lichtenstein Bileaflet mechanical prostheses performance in mitral position Eur. J. Cardiothorac. Surg., June 1, 1999; 15(6): 786 - 794. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Craver CarboMedics(R) Prosthetic Heart Valve Eur. J. Cardiothorac. Surg., January 1, 1999; 15(suppl_1): 3 - 11. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. C. Fiore, H. B. Barner, M. T. Swartz, L. R. McBride, A. J. Labovitz, K. J. Vaca, J. St. Vrain, G. L. Grunkemeier, and G. C. Kaiser Mitral valve replacement: randomized trial of St. Jude and Medtronic Hall prostheses Ann. Thorac. Surg., September 1, 1998; 66(3): 707 - 713. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-P. Remadi, P. Bizouarn, O. Baron, O. Al Habash, P. Despins, J.-L. Michaud, and D. Duveau Mitral valve replacement with the St. Jude medical prosthesis: a 15-year follow-up Ann. Thorac. Surg., September 1, 1998; 66(3): 762 - 767. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. M. Mueller, H. T. Tevaearai, F. Stumpe, A. P. Fischer, M. Hurni, P. Ruchat, and L. v. Segesser Long-term results of mitral-aortic valve operations J. Thorac. Cardiovasc. Surg., June 1, 1998; 115(6): 1298 - 1302. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. F. Katircioglu, B. Yamak, T. Ulus, L. Birincioglu, A. Saritas, O. Tasdemir, and K. Bayazit Mitral Valve Replacement with Bioprosthesis in Age Group Fifty to Sixty-Five Years Asian Cardiovasc Thorac Ann, March 1, 1998; 6(1): 23 - 27. [Abstract] [Full Text] [PDF] |
||||
![]() |
W.R. E. Jamieson, A. I. Munro, R. T. Miyagishima, G. L. Grunkemeier, L. H. Burr, S. V. Lichtenstein, and G. F. O. Tyers Multiple mechanical valve replacement surgery comparison of St. Jude Medical and CarboMedics prostheses Eur. J. Cardiothorac. Surg., February 1, 1998; 13(2): 151 - 159. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ninet, F. Tronc, J. Robin, A. Curtil, I. Aleksic, and G. Champsaur Mechanical versus biological isolated aortic valvular replacement after the age of 70: equivalent long-term results Eur. J. Cardiothorac. Surg., January 1, 1998; 13(1): 84 - 89. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Bernal, J. M. Rabasa, F. Gutierrez-Garcia, C. Morales, J. F. Nistal, and J. M. Revuelta The CarboMedics Valve: Experience With 1,049 Implants Ann. Thorac. Surg., January 1, 1998; 65(1): 137 - 143. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Hasegawa, S. Kitamura, S. Taniguchi, T. Kawata, K. Niwaya, K. Mizuguchi, H. Nishioka, and Y. Kameda Comparative Rest and Exercise Hemodynamics of Allograft and Prosthetic Valves in the Aortic Position Ann. Thorac. Surg., December 1, 1997; 64(6): 1753 - 1756. [Abstract] [Full Text] |
||||
![]() |
J.-F. Obadia, Y. A. Martelloni, O. H. Bastien, G. M. Durand de Gevigney, and J.-F. Chassignolle Long-Term Follow-up of Small (Size 20 and 21) Medtronic-Hall Aortic Valve Prostheses Ann. Thorac. Surg., August 1, 1997; 64(2): 421 - 425. [Abstract] [Full Text] |
||||
![]() |
T. E. David, V. L. Gott, L. A. Harker, G. E. Miller Jr, D. C. Naftel, A. G. G. Turpie, and C. W. Akins Mechanical Valves Ann. Thorac. Surg., November 1, 1996; 62(5): 1567 - 1570. [Full Text] |
||||
![]() |
W. Vongpatanasin, L. D. Hillis, and R. A. Lange Prosthetic Heart Valves N. Engl. J. Med., August 8, 1996; 335(6): 407 - 416. [Full Text] [PDF] |
||||
![]() |
C. W. Akins Results With Mechanical Cardiac Valvular Prostheses Ann. Thorac. Surg., December 1, 1995; 60(6): 1836 - 1844. [Abstract] [Full Text] |
||||
![]() |
R. G. Masters, A. L. Pipe, V. M. Walley, and W. J. Keon Comparative results with the St. Jude Medical and Medtronic Hall mechanical valves J. Thorac. Cardiovasc. Surg., September 1, 1995; 110(3): 663 - 671. [Abstract] [Full Text] |
||||
![]() |
G.-W. He, G. L. Grunkemeier, H. L. Gately, A. P. Furnary, and A. Starr Up to Thirty-Year Survival After Aortic Valve Replacement in the Small Aortic Root Ann. Thorac. Surg., May 1, 1995; 59(5): 1056 - 1062. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||