J Thorac Cardiovasc Surg 1995;109:74-80
© 1995 Mosby, Inc.
CARDIOPULMONARY BYPASS, |
Munich, Germany, Bronx, N.Y.
Address for reprints: Herbert O. Vetter, MD, Department of Cardiac Surgery, Grosshadern Medical Center, University of Munich, 81366 Munich, Germany.
Abstract
The three components of the Novacor left ventricular assist system, compact controller, battery, and back-up battery, have been miniaturized in the development of the wearable system. Therefore patients can be fully mobilized receiving mechanical circulatory support while awaiting heart transplantation. Between February 1992 and April 1994 a total of eight patients with decompensated heart failure (6 dilated cardiomyopathy, 1 acute myocarditis, 1 ischemic cardiomyopathy) were treated with the Novacor left ventricular assist system. In the most recent four cases the wearable system (N100P) was used. Patients' ages ranged from 17 to 49 years. In five patients severe failure of the right side of the heart was present at the time of implantation. Hemodynamic stabilization was achieved in all patients during the 2 to 122 days (mean 30.8 ± 42.5 days) of support. The following parameters were measured on average before and 24 hours after implantation of the left ventricular assist system: mean arterial pressure 70 ± 11 versus 87 ± 13 mm Hg (p < 0.05), cardiac index 1.71 ± 0.42 versus 3.23 ± 0.74 L/min/m2 (p < 0.05), pulmonary capillary wedge pressure 27.1 ± 4.4 versus 9.9 ± 5.2 mm Hg (p < 0.01), mean pulmonary pressure 41 ± 9 versus 27 ± 6 mm Hg (p < 0.05), and right ventricular ejection fraction 16.7% ± 10.3% versus 22.0% ± 11.6% (not significant). Patients who received the wearable system were capable of managing their own power supply during the bridging period and were able to walk to the hospital park and shopping area. One patient had a serious pulmonary infection, which was treated successfully, and two patients had a cerebrovascular accident, which resolved in one and resulted in a minor residual deficit in the other. All eight patients received a heart transplant. One patient died early after transplantation and seven patients are alive and well. In summary, the wearable Novacor left ventricular assist system provides major advantages regarding quality of life of patients during mechanical circulatory support. However, there is a remaining risk of thromboembolism despite anticoagulation therapy. (J THORAC CARDIOVASC SURG 1995;109:74-80)
Cardiac transplantation has become an accepted method in clinical therapy for end-stage heart disease. However, the shortage of donors is responsible for a high percentage of patients dying while awaiting heart transplantation. Within the past decade electrically powered ventricular assist devices
1 and pneumatically driven artificial hearts
2 have been used in clinical programs to rescue critically ill cardiac transplant candidates. Because the average waiting time for a donor heart increased during the past few years this situation has resulted in prolonged periods of mechanical circulatory support. However, the incidence of thromboembolism remains a major threat in patients who receive long-term support. Anticoagulation protocols and changes in device design are important factors for the outcome of such patients.
3 Since 1993 the large external console of the Novacor left ventricular assist system (LVAS) has been miniaturized: the result is a system that can be worn on a belt and that allows a higher degree of mobility and quality of life to patients during the waiting period.
MATERIAL AND METHODS
Assist device
The Novacor N100 LVAS (Novacor Division, Baxter Healthcare Corp., Oakland, Calif.), is an electromagnetically actuated implantable pump that consists of a seamless one-piece smooth polyurethane sac that is bonded to symmetrically opposed dual pusher-plates. The housing is a lightweight shell that incorporates bovine pericardial valve fittings and an energy converter. The implanted pump unit is connected to the extracorporeal control console or wearable controller via a percutaneous lead (Fig. 1). Inflow and outflow Dacron fabric grafts, which both perforate the diaphragm, connect the pump to the left ventricular apex and to the ascending aorta. The maximal stroke volume of 70 ml can provide a pump output up to 10 L/min. With operation in synchronous counterpulsation cardiac rates of up to 240 beats/min can be achieved.
4
|
Patients
Between February 1992 and April 1994 the Novacor LVAS was implanted in eight patients as a bridge to cardiac transplantation. The wearable system was used for the most recent four patients. Dilated cardiomyopathy in six patients, end-stage ischemic heart disease in one patient, and acute myocarditis in one patient were the underlying diseases when the patients were evaluated as heart transplant candidates. Before Novacor implantation two patients were supported by intraaortic balloon counterpulsation. The mean patient age was 31.6 ± 12.9 years (range 16 to 49 years). The average body surface area was 1.83 ± 0.10 m2 and ranged from 1.72 to 1.98 m2 .
The clinical status and hemodynamics of all patients were deteriorating despite inotropic or vasodilator support, or both therapies. Preoperative patient data are given in
Table I. Cardiac output and parameters of function of the right side of the heart were evaluated by means of a rapid-response thermodilution catheter. In five of the eight patients right ventricular ejection fraction was only 4% and in the others the value was 13%, 15%, and 16%, respectively. Renal function was reduced with oliguria or anuria and a mean serum creatinine level of 1.50 ± 0.54 mg/dl. Mean total bilirubin level was 2.11 ± 1.51 mg/dl and the mean value of aspartate aminotransferase was 198 ± 246 mg/dl (range 70 to 660 mg/dl) before Novacor LVAS implantation.
|
Anticoagulation protocol
The perioperative anticoagulation treatment consisted of 2 million units of aprotinin and regular heparinization (300 IU/kg body weight) for cardiopulmonary bypass. Systemic heparinization was fully antagonized with protamine sulfate at the end of cardiopulmonary bypass.
In the postoperative phase intravenous administration of heparin was started when the activated partial thromboplastin time (aPTT) was less than 50 seconds or the activated coagulation time was less than 140 seconds. Heparinization was considered sufficient if the aPTT ranged between 60 and 80 seconds. When patients were mobilized, the anticoagulation treatment was changed to oral phenprocoumon (Marcumar), maintaining the prothrombin time level at an international normalized ratio of 3.0 to 4.5.
7 Antithrombin III was substituted when levels fell below 80%. The protocol for postoperative anticoagulation was changed after a thromboembolic event in patient 7. It now includes heparin intravenously and then subcutaneously, acetylsalicylic acid 100 mg/d, and dipyridamole 75 mg/8 hourly.
After operation the following tests were frequently used to monitor anticoagulation therapy: aPTT, prothrombin time, platelet count, fibrinogen, activated coagulation time, antithrombin III,
-2-haptoglobin, and hemopexin-S For evaluation of platelet function a specific test, "bleeding time ex vivo," according to Kratzer and Born
8 was applied (Thrombostat 4000, Baxter, Unterschleissheim, Germany).
Statistics
Data are given as mean values ± standard deviation. The
2 test was used to calculate significant differences between groups; Fisher's exact test was used in small sample sizes. Yates' correction for continuity was done. A p value of 0.05 was taken as the level of significance.
RESULTS
Hemodynamic data
The hemodynamic situation was significantly improved after implantation of the LVAS (Figs. 2 and 3). Mean arterial pressure increased from 70 ± 11 mm Hg to 87 ± 13 mm Hg (p < 0.05). Mean cardiac index, 1.71 ± 0.42 L/min/m2 before Novacor LVAS implantation, rose to 3.23 ± 0.74 L/min/m2 24 hours after the start of mechanical circulatory support (p < 0.01). Average pulmonary capillary wedge pressure was 27.1 ± 4.4 mm Hg (range 22 to 34 mm Hg) before operation and decreased to 9.9 ± 5.2 mm Hg (range 4 to 13 mm Hg after operation (p < 0.05).
|
|
|
Complications
Only one patient acquired a documented serious pulmonary infection caused by Legionella bozemanii. He was treated successfully with erythromycin, ciprofloxacin, and fluconazole.
A tendency of prolonged bleeding was observed in most of the patients. Surgical intervention was necessary in five patients; three had a cardiac tamponade within the first postoperative week and a rethoracotomy because of cardiac tamponade was done in two patients who received long-term support. One patient underwent revision of the subcutaneous area of the abdominal pocket because of a hematoma.
Two patients who received long-term support had an embolic accident during left ventricular circulatory support (
Table II). These occurred on days 40 (severe impairment of visual function) and 31 (hemiparesis) of support, respectively. The possible source of this thromboembolism was found after explanation of the pump. Various amounts of thrombotic material were found on the proximal and distal sides of the inflow pericardial valve prostheses (Fig. 5). The neurologic lesions resolved completely in one patient and only a minor neurologic deficit remained at the time of transplant in the other patient.
|
|
50 days) support (DISCUSSION
This series of patients who underwent bridge to cardiac transplantation with the Novacor LVAS shows that mechanical circulatory support can be done with low mortality while providing significant hemodynamic improvement and physical rehabilitation. Hemodynamic data showed that parameters of function of the left side of the heart can be improved significantly within 24 hours of the institution of circulatory support. In the majority of patients pump operation in synchronous counterpulsation could be achieved immediately after Novacor LVAS implantation and provided a pump output ranging between 4.5 and 8.0 L/min. As a result impaired renal function improved rapidly (within hours or within the first 2 postoperative days). Continuous hemofiltration was necessary only in two patients in the early postoperative phase. Pennington and associates
9 reported in a series of 44 cases that used different methods of circulatory support that there is excellent functional recovery and survival of such patients.
Despite severe impairment, function of the right side of the heart improved significantly during isolated left ventricular support. Implantation of a right ventricular assist device was not indicated. In a clinical study Kormos and colleagues
10 demonstrated an improvement of the function of the right side of the heart in most patients after implantation of the Novacor system.
In our center the Novacor LVAS has proved to be a highly reliable system for assisting the failing heart. No technical faults have been detected during operation. The patients who received long-term support were trained to manage their own power supply: they were able to change and recharge batteries and prepare for out-of-hospital walks on their own. Pump noise was not a problem for the patients. Development of the wearable system was a major step forward in providing mobility for patients in need of long-term support and improved quality of life during the waiting period.
Thromboembolism remains the major threat during mechanical circulatory support. A reason for the thromboembolic event in one patient was the normalization of platelet function in the absence of acetylsalicylic acid or dipyridamole treatment during a period of low pump output (<3.5 L/min) caused by hypovolemia. Because a triple drug therapy consisting of heparin (intravenously or subcutaneously), acetylsalicylic acid, and dipyridamole has been used in the most recent two patients, no more thromboembolic complications have been experienced. Some fibrinous depositions on the prosthetic inflow valve have been frequently observed. Multiple factors including coagulation state, flow, and thrombotic deposition within the prosthetic device or native heart may contribute to thromboembolic complications.
The question as to whether a patient who has a residual deficit after a cerebrovascular event should receive transplantation before complete rehabilitation may be an issue. However, after 10 to 14 days, patients should be able to undergo cardiopulmonary bypass with minimal risk.
In summary, hemodynamic and metabolic recompensation of patients who receive the Novacor LVAS is possible within a relatively short period of support. Postoperative anticoagulation is important to minimize the risk of thromboembolic complications. Early mobilization and physical activities are easier to do in patients with use of the wearable system.
Appendix: DISCUSSION
Dr. Eric A. Rose (New York, N.Y.)
I congratulate you on these results and your illustration of the concept originally proposed by Bob Frazier and Victor Poirier that a vented system that is not totally implantable allows a reasonable quality of life in these patients.
I have two questions. One is the issue of anticoagulation. I think it still remains speculative that the addition of antiplatelet agents to the already vigorous anticoagulation regimen that you described will be effective in preventing thromboembolism in these patients.
Second, is the incidence of thromboembolism with the Novacor device low enough to consider the use of this type of device for long-term support?
Dr. Vetter
I did not want to give the impression that adding anti-platelet aggregation medication will prevent thromboembolism. In the two patients who had a cerebrovascular accident there were additional negative influences present. In one patient there was, as a result of hypovolemia, a rather low pump output situation that might have contributed to the appearance of thrombus formation at the inflow valve. The other patient was in atrial fibrillation at the time of the cerebrovascular accident.
Another patient was supported by the device for 122 days and received triple drug therapy (heparin intravenously, acetylsalicylic acid, and dipyridamole) He went through bridging without thromboembolism. Therefore I think that the Novacor LVAS in its present design is a suitable device for patients who need long-term support. Whether it is a device to send patients home with, I do not know yet. I think monitoring problems have to be solved in the future.
Dr. Mortimer J. Buckley (Boston, Mass.)
May I ask how you handled the right ventricular failure? In the Novacor LVAS, as I understand it, you cannot put in a biventricular system.
Dr. Vetter
As I showed, our first four patients used the system only for a short term. The reason for that was in three of these patients there was poor right ventricular function. Right ventricular ejection fraction was 4%. We always wanted to manage without institution of a right ventricular support system because at that time we did not have the Berlin Heart (Berlin Heart, Fehling Medical, Berlin, Germany), and we did not have heparinized circuits for the Bio-Medicus pump (Bio-Medicus, Inc., Eden Prairie, Minn.). We would rather use medications, such as prostacycline, to improve postoperatively function of the right side of the heart. Therefore in these patients we went on with transplantation. One can see, however, timing is important for such patients, and we showed that a short time of bridging does not mean the outcome must be bad.
Acknowledgments
We gratefully thank the medical students Mr. Willi Roll and Mr. Gernot Schindler for their support and Mr. Steeger and Mrs. von Yorck from the Department of Photography and Graphics at Grosshadern Hospital.
Footnotes
From the Department of Cardiac Surgerya (Head Prof. Dr. B. Reichard), the Institute of Anesthesiologyb (Head Prof. Dr. K. Peter), and the Institute of Clinical Chemistryc (Head Prof. Dr. D. Seidel), University of Munich, Grosshadern Hospital Munich, Germany. ![]()
Read at the Seventy-fourth Annual Meeting of The American Association for Thoracic Surgery, New York, N.Y., April 24-27, 1994. ![]()
*Gore-Tex Surgical Membrane. Gore-Tex is a registered trademark of W. L. Gore & Associates, Putzbrunn, Germany. ![]()
References
This article has been cited by other articles:
![]() |
M. Rinaldi, F. Pagani, F. Gazzoli, A. Alloni, D. Ricci, and M. Vigano Left ventricular assistance from bridge to transplantation to destination therapy. The Pavia experience Eur. Heart J. Suppl., November 1, 2004; 6(suppl_F): F81 - F86. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. G. Nabavi, J. Stockmann, C. Schmid, M. Schneider, D. Hammel, H. H. Scheld, and E. B. Ringelstein Doppler microembolic load predicts risk of thromboembolic complications in Novacor patients J. Thorac. Cardiovasc. Surg., July 1, 2003; 126(1): 160 - 167. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Rothenburger, M. J. Wilhelm, D. Hammel, C. Schmidt, T. D. T. Tjan, D. Bocker, H. H. Scheld, and C. Schmid Treatment of Thrombus Formation Associated With the MicroMed DeBakey VAD Using Recombinant Tissue Plasminogen Activator Circulation, September 24, 2002; 106(12_suppl_1): I-189 - I-192. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. A. Vilchez, M. C. McEllistrem, L. H. Harrison, K. R. McCurry, R. L. Kormos, and S. Kusne Relapsing bacteremia in patients with ventricular assist device: an emergent complication of extended circulatory support Ann. Thorac. Surg., July 1, 2001; 72(1): 96 - 101. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Koster, M. Loebe, R. Hansen, E. V. Potapov, G. P. Noon, H. Kuppe, and R. Hetzer Alterations in coagulation after implantation of a pulsatile Novacor LVAD and the axial flow micromed DeBakey LVAD Ann. Thorac. Surg., August 1, 2000; 70(2): 533 - 537. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Di Bella, F. Pagani, C. Banfi, E. Ardemagni, A. Capo, C. Klersy, and M. Vigano Results with the Novacor assist system and evaluation of long-term assistance Eur. J. Cardiothorac. Surg., July 1, 2000; 18(1): 112 - 116. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Schmitz, A. Welz, O. Dewald, R. Kozlik-Feldmann, H. Netz, and B. Reichart Switch from a BIVAD to a LVAD in a boy with Kawasaki disease Ann. Thorac. Surg., April 1, 2000; 69(4): 1270 - 1271. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. El-Banayosy, M. Deng, D.Y. Loisance, H. Vetter, E. Gronda, M. Loebe, and M. Vigano The European experience of Novacor left ventricular assist (LVAS) therapy as a bridge to transplant: a retrospective multi-centre study Eur. J. Cardiothorac. Surg., June 1, 1999; 15(6): 835 - 841. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Kasirajan, N. G. Smedira, J. Perl II, and P. M. McCarthy Cerebral embolism associated with left ventricular assist device support and successful therapy with intraarterial urokinase Ann. Thorac. Surg., April 1, 1999; 67(4): 1148 - 1150. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Schmid, M. Weyand, D. G. Nabavi, D. Hammel, M. C. Deng, E. B. Ringelstein, and H. H. Scheld Cerebral and Systemic Embolization During Left Ventricular Support With the Novacor N100 Device Ann. Thorac. Surg., June 1, 1998; 65(6): 1703 - 1710. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Vigano, S. Scuri, F. Cobelli, C. Opasich, F. M. Pagani, G. Minzioni, L. Martinelli, L. Tavazi, and M. Vigano Staged discharge out of hospital of the Novacor left ventricular assist system (LVAS) recipients Eur. J. Cardiothorac. Surg., April 1, 1997; 11(suppl): S45 - S50. [Abstract] [PDF] |
||||
![]() |
D.G. Nabavi, D. Georgiadis, T. Mumme, C. Schmid, T.G. Mackay, H.H. Scheld, and E.B. Ringelstein Clinical Relevance of Intracranial Microembolic Signals in Patients With Left Ventricular Assist Devices : A Prospective Study Stroke, May 1, 1996; 27(5): 891 - 896. [Abstract] [Full Text] |
||||
![]() |
Short-Term Experience with Portable LVADs for Transplant Candidates Journal Watch Cardiology, March 1, 1995; 1995(301): 8 - 8. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||