J Thorac Cardiovasc Surg 1995;110:504-510
© 1995 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
Seattle, Wash.
From The Hope Heart Institute, the Providence Medical Center, and the Department of Surgery, University of Washington School of Medicine, Seattle, Wash.
Received for publication Sept. 2, 1994. Accepted for publication Dec. 28, 1994. Address for reprints: Lester R. Sauvage, MD, 528 18th Ave., Seattle, WA 98122.
Abstract
The internal thoracic artery is preferable to the saphenous vein for use as a conduit for coronary artery bypass. More extensive use is possible if this artery is mobilized in a skeletonized form to provide greater length. Internal thoracic arteries are usually mobilized with cauterization. This study compared the effectiveness and effects on neighboring areas of division of the branches of the canine internal thoracic artery with bipolar cauterization and monopolar cauterization. Branch closure was significantly more secure in the bipolar cauterization group, with bleeding in 25 (9%) of 279 branches of 15 internal thoracic arteries treated with monopolar cauterization, in contrast to bleeding in 4 (1.3%) of 306 branches in the 15 internal thoracic arteries treated with bipolar cauterization, which were paired with the group treated with monopolar cauterization. The group of internal thoracic arteries treated with monopolar cauterization had a significantly higher prevalence of leakage when luminal pressure was increased from 120 to 160 mm Hg. Scanning electron microscopy demonstrated partial loss of endothelial cells on the flow surface of internal thoracic arteries treated with bipolar cauterization, compared with almost complete loss of endothelial cells around the orificial areas after monopolar cauterization. Secondary bipolar cauterization treatment caused only slightly more damage than primary treatment, but secondary monopolar cauterization was much more severe and extensive than primary treatment. These data suggest that bipolar cauterization is preferable to monopolar cauterization for skeletonized dissection of the internal thoracic artery. (J THORAC CARDIOVASC SURG 1995;110:504-10)
Studies have shown that the internal thoracic artery (ITA) used as a coronary artery bypass graft has a long-term result that is significantly superior to results obtained with the saphenous vein
1,2 We have developed techniques to completely revascularize the entire coronary artery system with only the ITAs.
3,4 This requires maximum lengths of the ITA, with skeletonized mobilization. Most surgeons dissect the ITA as a pedicle graft, or as a free graft with surrounding tissue, with the use of monopolar cauterization (MPC) for hemostasis, reinforced as needed by hemoclips.
5,6 Although this method provides a graft sufficiently long for a left ITA graft to the left anterior descending artery, it may not yield one that can reach the distal right coronary artery or its proximal posterior descending or left ventricular extension branches. Some reports have mentioned surgical methods of ITA graft mobilization, including skeletonized dissection
5-8 ; however, there has been little description of the type of cauterization used and the effect of its application. In 1987 Keeley
9 noted that bipolar cauterization (BPC) allows precise control of current and avoids random spraying of heat. For the past several years we have found that mobilization of the ITAs in a skeletonized manner is facilitated by use of BPC. This study was undertaken to compare the results of mobilization of the ITAs with MPC versus BPC, with special attention paid to the quality and security of branch closure and the extent and degree of tissue damage.
MATERIAL AND METHODS
Care of the dogs used in this study complied with the "Principles of Laboratory Animal Care" and the "Guide for the Care and Use of Laboratory Animals" (National Institutes of Health Publication No. 80-23, revised 1985). Seventeen randomly selected dogs weighing from 21.0 to 37.0 kg (average 23.6 ± 3.7 kg) were studied in a paired comparison form, in which one ITA was dissected by MPC and the contralateral ITA was dissected by BPC; the sides were alternated by cases. The surgical dissection was done with the use of proper anesthesia, which was induced with 5 to 10 ml 4% thiamylal intravenously and maintained with a combination of 0.5% to 1% halothane and a mixture of nitrous oxide and oxygen in a 2:1 ratio via an endotracheal tube connected to a closed-circuit respirator.
The chest was opened through a median sternotomy to allow access to both ITAs for skeletonized dissection. The cautery unit used in this study was a Bard System 5000 (Davol Electro Medical System, Englewood, Colo.), which can perform both MPC and BPC in either a cutting or a coagulant mode. The coagulant mode can be further selected as a pinpoint or spray type in MPC application. In this study the coagulant mode was used for both BPC and MPC dissection; during MPC dissection, the pinpoint application was selected.
For BPC dissection, the power was set at 5 W, and bipolar cautery forceps with a 0.7 mm fine tip (catalog No. NL 3785-112, V. Mueller/Baxter, McGaw Park, Ill.) were used. The branches were cauterized three times, 1 second per time, by gently squeezing and then reapplying the forceps along a 2 mm wide zone, which was 2 mm away from the stem of the ITA external wall. The branch was then cut with scissors in the middle of the cauterized zone (Fig. 1, A). If there was bleeding, the ends were held with the same forceps for recauterization until the bleeding stopped.
|
The ITA blood flow was measured in vivo before and after dissection by a Transonic T-208 flowmeter (Transonic System Inc, Ithaca, N.Y.) with a 3 mm probe. The ITA branches were counted and the ITA length was measured under the following three conditions: in vivo before dissection, in vitro after removal, and in vitro with saline solution inflation at 120 mm Hg.
The harvested ITAs were then put on a dry sponge and inflated at 120 mm Hg with saline solution for 1 minute so that any saline leakage from the branch was absorbed by the sponge and could be weighed for quantitative leakage evaluation. The ITAs were then submerged in a container of saline solution and inflated with air. The inflation pressure was gradually increased to 120 mm Hg and the number of branches that showed leakage of air bubbles was noted. This was repeated at 160 mm Hg and 200 mm Hg.
The ITAs were further dynamically inflated and fixed with 10% formalin at 120 mm Hg in a perfusion system that overcame branch leakage and maintained stable pressure for 24 hours; the ITAs were then kept in formalin for an additional week. Two tissue blocks containing the ITA stem and a branch perpendicular to it were taken from each ITA for histologic study. One tissue block was cut into continuous serial cross sections along the branch from the cauterized end toward the ITA stem (Fig. 2, A) to evaluate the extent of the effect of cauterization on the tissue. The other block was prepared with continuous serial cross sections along the ITA stem and across the orifice of the branch to evaluate the integrity of the ITA wall (Fig. 2, B). In both series, the histologic sections were 6 µm thick and were examined at 30 µm intervals (approximately 130 sections per specimen). All sections were stained with hematoxylin and eosin (H&E) stain. In addition, tissue blocks containing the branch that had secondary cauterization because of bleeding from initial cauterization failure were histologically evaluated in a similar manner.
|
2 or Wilcoxon tests were used for data analysis and statistical comparison.
RESULTS
The results, including the physical data, the security of branch closure, and pressure-leakage effects from cauterization for both the MPC and BPC groups, are presented in
Table I. There were no significant differences between physical aspects of the two groups; however, all the quantitative test results showed the branch closure was significantly more secure in the BPC group.
|
Light microscopy study of H&E-stained slides showed that there were no remarkable histologic changes in either the wall structure of the ITA stem area or the extreme proximal end of the branch near the ITA stem in either the BPC or MPC groups, except for a few MPC-harvested ITAs that showed intimal corrugation and detachment at the proximal ends of the branches (Fig 3, A through D).
|
On the flow surface, SEM demonstrated that the lumina of branch orifices treated with BPC were more precise and freer of cellular debris than those of the MPC-treated arteries. Although there was endothelial cell loss around the orificial areas of BPC-treated ITAs, some endothelial cells still remained on the flow surface. However, for the MPC-treated ITAs, there were accumulations of red blood cells and platelets, with almost a complete loss of endothelial cells (Fig. 4, A through D).
|
The mechanisms of MPC and BPC electric current conduction are quite different. For MPC, the current flows from the generator to the cautery pen (the active electrode) through the subject's body to the inactive dispersive electrode, which is in contact with the subject, and then back to the generator. In BPC, the current flows only from one forceps tip to the other, and a limited amount of tissue is cauterized in between. Because of the difference in the mechanism, the power setting required for BPC is usually much lower than that for MPC. In this study, 5 W power was used for BPC, compared with 15 W for MPC. Thus BPC allows control in delicate areas, and the damage is also confined to the tissue between the tips of the cauterization forceps.
10,11 Because of this, BPC can be used safely much closer to the phrenic nerve than is possible with MPC, which is a consideration of importance in freeing the ITAs high in the chest.
Our SEM study revealed that the ITA flow surface, especially around the branch orifice, had more damage with MPC than with BPC, and there was thrombus formation at the orifice of the branch treated by MPC (Fig. 4). The more flow surface damage with thrombus accumulation, the more the likelihood of luminal occlusion, especially in a low-flow situation. Those changes are probably a result of spreading of the cautery current, and this spreading is more extensive in the MPC mode than in the BPC mode. Lehtola, Verkkala, and Jarvinen
12 have also described similar surface damage caused by cautery current, either through direct contact with the wall or through conduction by metal hemoclips.
Sealing the ends of a divided branch with MPC is mainly achieved by the heat generated from contact of the cautery pen at one side of the branch wall If the cautery pen cuts through the wall before it is heat-sealed to the other side, bleeding occurs. With BPC, the walls are pressed together and then securely annealed by the current that passes between the opposed tips of the cautery pen (Fig. 1, Fig. 3, E and F). This explains why, in our study, the branch end closure was far more secure in the BPC group than in the MPC group, even with the use of less power, as shown in
Table I.
Table I also reflects a greater potential risk of branch leakage in the MPC group after a pressure increase from 120 to 200 mm Hg that is significantly higher (20 of 264, 7.6%) than that in the BPC group (4 of 300, 1.3%). This is of practical importance in regard to the danger of postoperative bleeding from both ends of divided branches.
After initial cauterization, 9% of the branches in the MPC group bled, compared with 13% in the BPC group (
Table I). Bleeding from a divided branch end results in subadventitial hematoma. In many instances, this hematoma will cause regional vascular spasm and decrease flow; it may also obscure vision needed for a precise secondary cauterization, resulting in unnecessary tissue damage and making an anastomosis more difficult, should it be required in that area. Histologically, severe and substantial tissue damage was observed in the ITA wall around the branch orifices after secondary MPC treatment, in contrast to the slight changes observed with BPC (Fig. 3, G and H). This indicates an additional risk of adverse effects on branches prepared by MPC.
In the BPC group, four branches, all more than 2 mm in caliber, required secondary cauterization (
Table I). This suggests that with BPC, when the branch caliber is greater than 2 mm, each cauterization time should be increased to slightly more than 1 second and the number of applications should also be increased; otherwise hemoclips will be necessary to achieve reliable branch division and closure.
Freeing the ITA as a pedicle must be done with the MPC technique. The BPC technique may be of value in further securing branch closure after pedicle development; however, the use of hemoclips is a reliable means to reinforce branch closure under these circumstances.
The security and safety of mobilizing the ITA in a skeletonized manner is greatly facilitated by use of BPC for secure bonding of the branch walls before they are divided.
This study demonstrates that the use of BPC for skeletonized dissection of the ITA will provide a better flow surface and more secure closure at the divided end of the branch than MPC.
References
This article has been cited by other articles:
![]() |
F. Onorati, A. Esposito, F. Pezzo, A. di Virgilio, P. Mastroroberto, and A. Renzulli Hospital Outcome Analysis After Different Techniques of Left Internal Mammary Grafts Harvesting Ann. Thorac. Surg., December 1, 2007; 84(6): 1912 - 1919. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Vassiliades Jr, N. Cosgriff, A. Denham, J. Olson, and D. H. Maul Superiority of Using Bipolar Radiofrequency Energy for Internal Mammary Artery Harvesting Ann. Thorac. Surg., April 1, 2007; 83(4): 1508 - 1512. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Athanasiou, M.-C. Crossman, G. Asimakopoulos, A. Cherian, A. Weerasinghe, B. Glenville, and R. Casula Should the internal thoracic artery be skeletonized? Ann. Thorac. Surg., June 1, 2004; 77(6): 2238 - 2246. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-M. Huang, N.-H. Chow, H.-L. Lee, and T.-J. Wu The Value of Color Flow Doppler Ultrasonography of the Superior Thyroid Artery in the Surgical Management of Graves Disease Arch Surg, February 1, 2003; 138(2): 146 - 151. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Jeanmart, L. P. Perrault, N. Desjardins, O. Chavanon, M. Carrier, and J. D. Fonger Arterial balloon catheter: a new atraumatic device for dilating arterial grafts Ann. Thorac. Surg., September 1, 2001; 72(3): 810 - 815. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Deja, S. Wos, K. S. Golba, P. Zurek, W. Domaradzki, R. Bachowski, and T. J. Spyt Intraoperative and laboratory evaluation of skeletonized versus pedicled internal thoracic artery Ann. Thorac. Surg., December 1, 1999; 68(6): 2164 - 2168. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Crouch, D. P. O’Hair, J. P. Keuler, T. P. Barragry, P. H. Werner, and L. H. Kleinman Open versus endoscopic saphenous vein harvesting: wound complications and vein quality Ann. Thorac. Surg., October 1, 1999; 68(4): 1513 - 1516. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Sasajima, M. H.-D. Wu, Q. Shi, N. Hayashida, and L. R. Sauvage Effect of Skeletonizing Dissection on the Internal Thoracic Artery Ann. Thorac. Surg., April 1, 1998; 65(4): 1009 - 1013. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Cunningham Skeletonization of internal thoracic artery grafts J. Thorac. Cardiovasc. Surg., February 1, 1996; 111(2): 489 - 490. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||