J Thorac Cardiovasc Surg 1996;112:273-280
© 1996 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
From the Departments of Cardiology and Cardiac Surgery, Galician General Hospital and University of Santiago Faculty of Medicine, Santiago de Compostela, Spain.
Received for publication March 16, 1995 Revisions requested June 20, 1995; revisions received August 9, 1995 Accepted for publication Jan. 25, 1996. Address for reprints: José Ramón González-Juanatey, MD, PhD, Departmento de Cardioloxía, Hospital Xeral de Galicia, Galeras s/n, 15705 Santiago de Compostela, Spain.
Abstract
Discussion of aortic valve replacement has primarily concerned the choice between tissue and mechanical prostheses. Less emphasis has been placed on prosthesis size. Despite technical advances increasing prosthesis orifice area, small valves implanted in the unenlarged aortic root may not be significantly less obstructive than the stenotic native valves they replace. Methods: In this work we studied 52 patients (31 women, 21 men; mean age 59.2 years) in whom valve prostheses sized 19, 21, 23, or 25 mm (30 bioprostheses and 22 tilting disc valves) had been implanted to replace stenotic aortic valves. Most patients with 19 or 21 mm prostheses were women. Doppler and conventional echocardiographic studies were performed in the 10 days preceding the operation and between 10 and 40 months (mean 18 months) after the operation. The patients receiving larger valve sizes had significantly larger body surface areas than those receiving smaller valve sizes (mainly women). Results: No significant differences were observed between preoperative and postoperative diameters or left ventricular systolic function parameters, but left ventricular mass and mass index decreased in all four groups (albeit nonsignificantly in the 19 mm group, and with less statistical significance in the 21 mm group than in the 23 and 25 mm groups). Postoperative peak and mean transvalvular pressure drops were significantly greater in the 19 mm group than in the other groups, and the 21 mm group had significantly greater transvalvular pressure drops than the 25 mm group. Postoperative effective valve area was significantly smaller in the 19 mm group than in the 21 mm group, and significantly smaller in the 21 mm group than in the 23 and 25 mm groups. Conclusion: We conclude that despite undeniable recent improvements in the design of artificial heart valves, 19 mm aortic prostheses continue to create significant obstruction of the left ventricular outflow tract and, possibly as a consequence of this, fail to bring about significant reduction in left ventricular hypertrophy. (J THORAC CARDIOVASC SURG 1996;112:273-80)
In general, replacing severely dysfunctional aortic valves improves performance and prognosis. The surgical management of patients with small aortic rings is nevertheless controversial because of the suboptimal hemodynamics of small sizes of even the most advanced types of prostheses; prognosis is worse for patients with small implanted valves than for those with larger prostheses.
1-5
For patients having a small aortic ring, the usual alternative to small valve implantation is to reduce valve prosthesispatient mismatch by implanting a larger valve after enlargement of the aortic ring. The drawback of this approach is that it increases the duration and, in principle, the risk of the operation.
4,6,7 Another alternative might be the use of valvular homografts, which have excellent hemodynamics but which are not readily available to most centers.
8
Aortic valve stenosis induces left ventricular hypertrophy (LVH) as an adaptive response to the chronic overload caused by the valve disease. LVH is the causal link between aortic valve stenosis and myocardiac ischemia, diastolic and systolic ventricular dysfunction, and ventricular arrhythmias associated with sudden death. A number of studies, mostly of patients with systemic arterial hypertension, have confirmed the adverse influence of LVH on life expectancy
9,10; among patients with aortic valve prostheses implanted to redress aortic stenosis, those with echocardiographically detected LVH have a significantly higher postoperative death rate than those without.
11 After aortic valve replacement to reduce the obstruction to ejection, left ventricular mass (LVM) tends to decrease; this decrease may reduce the risk of ischemia, dysfunction, and electrical instability and in any case is known to be associated with improved cardiac performance and prognosis.
Doppler echocardiography is currently the standard technique for noninvasive evaluation of the structure and function of heart valve prostheses. Measurements of mean and peak transvalvular pressure drops, effective valve area, and cardiac output obtained by Doppler echocardiography correlate satisfactorily with those obtained by invasive methods.
8,12-14
We report here the results of a retrospective study of the hemodynamics of various designs and sizes of aortic valve prosthesis and their relationship to postimplantation reduction in LVH.
Patients and methods
Protocol
Between January 1989 and December 1994, 68 patients in our center underwent replacement of the aortic valve alone on account of predominantly aortic valvular stenosis. None of these patients underwent, or had previously undergone, a cardiovascular operation of any other kind. Thirty-seven other patients who were operated on for severe aortic stenosis in the same period were excluded from the study because they also had other types of cardiac disease: 18 had moderate or severe aortic insufficiency, 11 mitral valve disease, 5 coronary artery disease, 2 aortic insufficiency and coronary artery disease, and 1 mitral valve disease and coronary artery disease. The follow-up protocol of all 68 patients with "pure" aortic stenosis included evaluation of valve function by Doppler echocardiography. We studied retrospectively 52 of these patients who fulfilled the following criteria: (1) before the operation they had sinus rhythm and greater stenosis of a native aortic valve than of any other valve, and they had no other cardiac disease; (2) the operation consisted only in the implantation of an aortic valve prosthesis, with no aortic ring enlargement procedure; (3) a Doppler echocardiogram good enough for evaluation of left ventricular structure and function and aortic valve disease had been taken within the 10 days preceding the operation; and (4) 12 months or more after valve implantation the patients were in New York Heart Association (NYHA) functional classes I or II.
Of the 52 patients studied, 12 received 19 mm valves: 2 MitroFlow pericardial bioprostheses (MitroFlow International Inc., Richmond, Canada), 6 Labcor-Santiago pericardial bioprostheses (Labcor Laboratories, Belo Horizonte, Brazil), and 4 CarboMedics bileaflet prostheses (CarboMedics Inc., Austin, Tex.); 15 received 21 mm valves (1 MitroFlow pericardial bioprosthesis, 7 Labcor-Santiago bioprostheses, and 7 CarboMedics valves); 16 received 23 mm valves (9 Labcor-Santiago bioprostheses and 7 CarboMedics valves); and 9 received 25 mm valves (5 Labcor-Santiago bioprostheses and 4 CarboMedics valves). All patients continued to have sinus rhythm during postoperative follow-up. Follow-up echocardiographic studies took place 10 to 40 months after the operations (mean 18 months).
Of the 16 patients with "pure" aortic stenosis not included in the study, only two had 19 mm valve prostheses. One of these, who was in NYHA class III, died suddenly 33 days after the operation and the other was excluded from the study because of atrial fibrillation before the operation. One year after the operation, no patient with a 19 mm prosthesis was in NYHA class III or IV.
Echocardiography
Echocardiograms were performed with a Hewlett-Packard Sonos 1000 apparatus (Hewlett-Packard Co., Palo Alto, Calif.) with 3.5 MHz transducers for images and 2.5 MHz transducers for Doppler effect measurements. All echocardiograms were recorded on SuperVHS videotape (Panasonic, Matsushita Electric Works, Osaka, Japan) for subsequent analysis.
M-mode measurements
M-mode echocardiograms were taken for measurement (in millimeters) of telediastolic interventricular septum thickness (IVST), left ventricular posterior wall thickness (PWT), and telediastolic (LVEDD) and telesystolic (LVESD) left ventricular diameters; simultaneous two-dimensional longitudinal or transverse parasternal views were used to ensure that the measurements were taken immediately above the papillary muscles. All measurements were repeated for four to six successive heartbeats, and the values accepted for each patient were the means of these four to six values.
LVM was calculated from Devereux and Reichek's formula
15: LVM (in grams) = 1.04 x ([LVEDD + IVST + PWT])3 - LVEDD3) - 13.6.
LVM index (LVMI, in grams per square meter) was defined as LVM/BSA, where BSA is body surface area in square meters. Left ventricular fractional shortening (percent) was calculated as 100 x (LVEDD - LVESD)/LVEDD. Left ventricular ejection fraction (percent) was calculated as 100 x (LVEDD
3 - LVESD
3)/LVEDD
3.
Doppler effect measurements
With the guidance of color flow images, systolic velocities in the outflow tract near the valve were measured in pulsed Doppler mode from the apical four-chamber view. Velocities through native and prosthetic aortic valves were measured in the continuous Doppler mode from apical, suprasternal, and right parasternal views; once the view giving the greatest velocities had been found, the velocities measured from this view during three cardiac cycles with durations close to the patient's mean beat-to-beat interval were averaged to yield a single value for that patient. Pulsed Doppler mode was used to search for increased velocities within the left ventricle by stepping the sample volume from the apex to the valve.
Peak and mean systolic transvalvular pressure drops calculated from peak and mean flow velocities by the modified Bernoulli equation are in this article denoted PSG and MSG, respectively. PSGc denotes peak pressure drops calculated by means of the Bernoulli equation with correction for prevalvular velocities: PSGc = 4(Vvalv
2 - Vlvol
2), where Vvalv is the peak velocity through the prosthesis and Vlvot the peak velocity in the left ventricular outflow tract just below the prosthesis.
16 The area (A) of the left ventricular outflow tract was calculated as p(D/2)
2, where D was the mean diameter of the tract in three successive images recorded at the beginning of systole from the longitudinal parasternal direction. Cardiac output (CO) was calculated as A x VTIlvot x HR, where HR is heart rate and VTIlvot is the time integral of the velocity in the left ventricular outflow tract.
17 Cardiac index was calculated as CO/BSA. Effective valve areas were calculated by means of both the standard continuity equation (VAstan = SV/VTIvalv, where SV is stroke volume and VTIvalv the time integral of the velocity through the valve) and the simplified equation (VAsimp = A x Vlvot/Vvalv).
18,19 The effective area index, a measure of how well the flow area of the valve matches body size, was calculated as VAstan/BSA.
Intralaboratory variability
To test intralaboratory variability in measurements of peak and mean gradients and effective valve areas, we repeated determinations of pulsed and continuous wave parameters and of left ventricular outflow tract diameter in 21 patients within 1 month of the postoperative echocardiographic examination. Variability was expressed as mean percent error, percent error being calculated as the absolute difference between the two observations divided by the mean of the two and expressed as a percentage (
Table I).
|
Results
Tables II and
III list preoperative characteristics of the four groups of patients defined by the size of the aortic prosthesis implanted. The patients in the larger valve size groups had significantly larger body surface areas than those in the smaller valve size groups, and they were approximately equally distributed between the sexes, whereas most patients receiving 19 or 21 mm prostheses were women. There were no statistically significant differences among the size groups as regards heart rate, blood pressure, degree of stenosis, or overall left ventricular systolic function. LVM and LVMI were significantly smaller in the 19 mm group than in any of the others, but the latter did not differ significantly one from another.
|
|
In 45 of the 52 prostheses (86.5%), the greatest velocities were recorded from apical views. The greatest subvalvular velocity measured after the operation was 1.65 m/sec. There was a significant negative correlation between subvalvular diameter and outflow tract velocity (r = -0.65, p < 0.001; n = 52). Postoperative peak and mean transvalvular pressure drops were significantly greater in the 19 mm group than in the other groups. The 21 and 23 mm groups did not differ significantly in this respect, and neither did the 23 and 25 mm groups, but transvalvular pressure drops in the 21 mm group were significantly greater than those in the 25 mm group (
Table IV and Fig. 1). There were significant negative correlations between valve size and peak and mean transvalvular pressure drops (r = -0.64 and r = -0.72, respectively; p < 0.001 in both cases).
|
|
|
Discussion
Doppler echocardiography is currently the most widely used technique for clinical evaluation of artificial heart valves; estimates of functional parameters obtained through Doppler echocardiographic measurements correlate well with those obtained by invasive methods,
8,12,13 which can no longer be justified for the evaluation of normally functioning valves. Measurements obtained by the two kinds of technique are nevertheless not directly comparable; transvalvular pressure drop values determined by Doppler echocardiography are greater than those determined by invasive procedures.
1,20 For example, Ihlen and coworkers
20 reported that echocardiographically measured mean pressure drops across CarboMedics prostheses in the aortic position exceeded those measured with catheters by 4.4 ± 4.8 mm Hg. The reason for this difference appears to be that in clinical studies catheter tips are usually located more than 2 cm from the plane where pressure recovery occurs: in an in vitro study of the St Jude Medical valve
21 (St. Jude Medical, Inc., St. Paul, Minn.), echocardiographically measured pressure drops agreed fully with catheter measurements taken at valve level but were 10.3 ± 2.5 mm Hg greater than catheter measurements taken 3 cm from the valve.
Although aortic transvalvular pressure drops are influenced by transvalvular flow rate, which should be taken into account in evaluating their clinical significance, they probably depend most on the diameter of the valve ring, increasing with decreasing ring size. In this study there was a significant negative correlation between valve size and peak and mean transvalvular pressure drops.
In a recent study, we
14 found no significant difference between the hemodynamic performance of 19 mm pericardial bioprostheses and that of 19 mm bileaflet mechanical prostheses in the aortic position; as in other studies of the hemodynamics of small prostheses,
1,22,22-24 in which effective valve areas of 0.9 to 1.2 cm2 and mean PSGs as high as 40 mm Hg have been observed, transvalvular pressure drops and effective valve areas were far from satisfactory. These findings, together with recent reports that patients with small (19 or 21 mm) prostheses have a poorer prognosis than those with larger prostheses,
25 must be taken into account when planning operations for patients for whom aortic valve replacement is indicated.
In patients with aortic stenosis, left ventricular pressure is chronically elevated. LVH occurs as an adaptive response and, initially, keeps left ventricular stress (the force generated by each unit of myocardium) within the normal range. Hypertrophy can be quantitated by echocardiography. In relatively symmetric ventricles, without left ventricular aneurysm or diastolic shape abnormality caused by marked dilation of the right side of the heart (conditions shared by all the patients in this study), this can be done by means of measurements obtained in the M mode
15: echocardiographic estimates of LVM calculated from Devereux and Reichek's formula correlate well with angiographic estimates of LVM, with postmortem left ventricular weight, and with estimates obtained by applying Simpson's rule or an area-length formula to two-dimensional echocardiographic measurements. New imaging techniques, such as nuclear magnetic resonance imaging and ultrafast computed tomography, provide images of the beating heart with high temporal and spatial resolution allowing more precise calculation of volumes and LVM, but echocardiography is still the most cost-efficient method.
The treatment of aortic stenosis by valve replacement is usually followed by the reduction of LVH by an amount variously reported as being between 16% and 44%.
26-29 Most regression occurs during the first 6 months after the operation. However, most studies of this phenomenon have involved patients with valves of size 21 mm or more; the available information on patients with 19 mm valves is scant. Sim and associates,
30 in a retrospective study of 39 of 714 patients in whom aortic valve prostheses of size 19, 21, or 23 mm had been implanted (accompanied by aortic ring enlargement in 168 cases), found that patients with a 19 mm aortic prosthesis and no other prostheses exhibited no significant reduction in LVM a mean of 26 months after the operation, whereas those with 21 or 23 mm prostheses did exhibit regression, and to similar extents. However, inasmuch as the 10 patients with 19 mm prostheses had preoperative LVMIs of 129 ± 47 gm/m2 (i.e., only slightly above normal, despite their having mean aortic pressure drops and valve areas suggestive of severe aortic stenosis), whereas the 21 and 23 mm groups had preoperative LVMIs of 150 gm/m2 or more, Sim and coworkers
30 suggested that the surgeons in their institution might have chosen to implant 19 mm valves only in patients with no more than slight LVH. Certainly, the scant ventricular hypertrophy of Sim's 19 mm group, and the small size of the study (39 patients) in relation to the 714 patients in whom aortic valves had been implanted, oblige one to treat their findings with some reserve. Nevertheless, in our study we also found significant postoperative reduction in LVM in patients with 21 to 25 mm aortic prostheses but not in patients with 19 mm prostheses (despite the postoperative effective valve areas of the latter group having been, as in the study by Sim and associates, almost double their preoperative areas); although the 19 mm group had less severe LVH than the other groups, their preoperative LVMIs were far above the upper limit of the normal range (possibly because almost all the patients in this group were women).
The lack of a significant reduction in LVM in our 19 mm group may have been due to a number of factors: (1) regression of LVH being slower in these patients than in the recipients of larger valves (although all the patients with 19 mm prostheses had their follow-up examination more than 10 months after their operations); (2) this group being older than the others (age is an independent determinant of LVM
31); and (3) the suboptimal hemodynamics of 19 mm prostheses at rest and during exercise. We are currently evaluating hemodynamics during exercise as part of a longer-term (3- to 5-year) study of patients with aortic valve prostheses implanted to replace stenotic native valves. The failure of small aortic prostheses to reduce LVM may be related to the results of several studies showing their recipients to have a poorer prognosis than recipients of larger valves.
We conclude that despite undeniable recent improvements in the design of artificial heart valves, 19 mm aortic prostheses continue to create a significant obstruction of the left ventricular outflow tract and, possibly as a consequence of this, fail to bring about significant reduction in LVH. Aortic prostheses of this size should probably not be implanted in young or physically active patients or in patients with body surface areas greater than 1.7 m2. Instead, aortic valve replacement in these groups of patients should probably be effected by means of a homograft or by implantation of a 21 mm or larger valve after aortic root enlargement.
References
This article has been cited by other articles:
![]() |
B. A. Youdelman, H. Hirose, H. Jain, J. Y. Kresh, J. W.C. Entwistle III, and A. S. Wechsler Comparison of eight prosthetic aortic valves in a cadaver model. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1526 - 1532. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Taniguchi, T. Takahashi, K. Toda, H. Matsue, Y. Shudo, H. Shintani, M. Mitsuno, and Y. Sawa Left ventricular mass: impact on left ventricular contractile function and its reversibility in patients undergoing aortic valve replacement Eur. J. Cardiothorac. Surg., October 1, 2007; 32(4): 588 - 595. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Dhareshwar, T. M. Sundt III, J. A. Dearani, H. V. Schaff, D. J. Cook, and T. A. Orszulak Aortic root enlargement: What are the operative risks? J. Thorac. Cardiovasc. Surg., October 1, 2007; 134(4): 916 - 924. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Borger, A. F. Nette, M. Maganti, and C. M. Feindel Carpentier-Edwards Perimount Magna Valve Versus Medtronic Hancock II: A Matched Hemodynamic Comparison Ann. Thorac. Surg., June 1, 2007; 83(6): 2054 - 2058. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. I. Duncan, J. Lin, C. G. Koch, A. M. Gillinov, M. Xu, and N. J. Starr The Impact of Gender on In-Hospital Mortality and Morbidity After Isolated Aortic Valve Replacement Anesth. Analg., October 1, 2006; 103(4): 800 - 808. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Mohty-Echahidi, J. F. Malouf, S. E. Girard, H. V. Schaff, D. E. Grill, M. E. Enriquez-Sarano, and F. A. Miller Jr Impact of Prosthesis-Patient Mismatch on Long-Term Survival in Patients With Small St Jude Medical Mechanical Prostheses in the Aortic Position Circulation, January 24, 2006; 113(3): 420 - 426. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. G. Koch, F. Khandwala, F. G. Estafanous, F. D. Loop, and E. H. Blackstone Impact of Prosthesis-Patient Size on Functional Recovery After Aortic Valve Replacement Circulation, June 21, 2005; 111(24): 3221 - 3229. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Penta de Peppo, J. Zeitani, P. Nardi, G. Iaci, P. Polisca, R. De Paulis, and L. Chiariello Small "Functional" Size after Mechanical Aortic Valve Replacement: No Risk in Young to Middle-Age Patients Ann. Thorac. Surg., June 1, 2005; 79(6): 1915 - 1920. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Gjertsson, K. Caidahl, M. Farasati, A. Oden, and O. Bech-Hanssen Preoperative moderate to severe diastolic dysfunction: A novel Doppler echocardiographic long-term prognostic factor in patients with severe aortic stenosis J. Thorac. Cardiovasc. Surg., April 1, 2005; 129(4): 890 - 896. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Otero, J. L. Pomar, J. M. Revuelta, and J. J. Rufilanchas Comparative Evaluation of Small-Size Sorin Slimline and St. Jude HP Heart Valve Prostheses Ann. Thorac. Surg., April 1, 2005; 79(4): 1284 - 1290. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Tasca, F. Brunelli, M. Cirillo, M. DallaTomba, Z. Mhagna, G. Troise, and E. Quaini Impact of Valve Prosthesis-Patient Mismatch on Left Ventricular Mass Regression Following Aortic Valve Replacement Ann. Thorac. Surg., February 1, 2005; 79(2): 505 - 510. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Seitelberger, J. Bialy, R. Gottardi, G. Seebacher, R. Moidl, M. Mittelbock, P. Simon, and E. Wolner Relation between size of prosthesis and valve gradient: comparison of two aortic bioprosthesis Eur. J. Cardiothorac. Surg., March 1, 2004; 25(3): 358 - 363. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Ruel, F. D. Rubens, R. G. Masters, A. L. Pipe, P. Bedard, P. J. Hendry, B. K. Lam, I. G. Burwash, W. G. Goldstein, M. P. Brais, et al. Late incidence and predictors of persistent or recurrent heart failure in patients with aortic prosthetic valves J. Thorac. Cardiovasc. Surg., January 1, 2004; 127(1): 149 - 159. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. De Carlo, A. D. Milano, C. Nardi, G. Mecozzi, and U. Bortolotti Serial Doppler echocardiographic evaluation of small-sized sorin bicarbon prostheses J. Thorac. Cardiovasc. Surg., August 1, 2003; 126(2): 337 - 343. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. B. Chambers, L. Oo, A. Narracott, P. M. Lawford, and C. I. Blauth Nominal size in six bileaflet mechanical aortic valves: A comparison of orifice size and biologic equivalence J. Thorac. Cardiovasc. Surg., June 1, 2003; 125(6): 1388 - 1393. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. H. Freed, J. W. Tam, M. C. Moon, G. E.J. Harding, E. Ahmad, and E. A. Pascoe Nineteen-millimeter prosthetic aortic valves allow normalization of left ventricular mass in elderly women Ann. Thorac. Surg., December 1, 2002; 74(6): 2022 - 2025. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. Castro, J. M. Arcidi Jr, A. L. Fisher, and V. A. Gaudiani Routine enlargement of the small aortic root: a preventive strategy to minimize mismatch Ann. Thorac. Surg., July 1, 2002; 74(1): 31 - 36. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Hanayama, G. T. Christakis, H. R. Mallidi, C. D. Joyner, S. E. Fremes, C. D. Morgan, P. R.R. Mitoff, and B. S. Goldman Patient prosthesis mismatch is rare after aortic valve replacement: valve size may be irrelevant Ann. Thorac. Surg., June 1, 2002; 73(6): 1822 - 1829. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. D. Milano, M. De Carlo, G. Mecozzi, A. D'Alfonso, G. Scioti, C. Nardi, and U. Bortolotti Clinical outcome in patients with 19-mm and 21-mm St. Jude aortic prostheses: comparison at long-term follow-up Ann. Thorac. Surg., January 1, 2002; 73(1): 37 - 43. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. K. Choudhary, A. Mathur, P. Venugopal, B. Airan, R. Sharma, A. Bhan, A. Saxena, A. Sampath Kumar, S. K. Choudhary, A. Mathur, et al. Prosthesis Size in Aortic Valve Replacement: Surgeon-Related Variable Asian Cardiovasc Thorac Ann, December 1, 2000; 8(4): 333 - 338. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Pibarot and J. G. Dumesnil Hemodynamic and clinical impact of prosthesis-patient mismatch in the aortic valve position and its prevention J. Am. Coll. Cardiol., October 1, 2000; 36(4): 1131 - 1141. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. S. Khan, R. J. Siegel, M. A. DeRobertis, C. E. Blanche, R. M. Kass, W. Cheng, G. P. Fontana, and A. Trento Regression of hypertrophy after Carpentier-Edwards pericardial aortic valve replacement Ann. Thorac. Surg., February 1, 2000; 69(2): 531 - 535. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. B. Izzat, I. Kadir, B. Reeves, P. Wilde, A. J. Bryan, and G. D. Angelini Patient-prosthesis mismatch is negligible with modern small-size aortic valve prostheses Ann. Thorac. Surg., November 1, 1999; 68(5): 1657 - 1660. [Abstract] [Full Text] [PDF] |
||||
![]() |
J R Gonzalez-Juanatey, M V Fernandez, F G Sampedro, J M Garcia-Acuna, J B Garcia-Bengoechea, A A Cendon, and M G de la Pena Haemodynamic performance of aortic pericardial bioprostheses and bileaflet prostheses at rest and during exercise: implications for the surgical management of patients with small aortic roots Heart, August 1, 1999; 82(2): 149 - 155. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Bech-Hanssen, K. Caidahl, B. Wall, P. Myken, S. Larsson, and I. Wallentin INFLUENCE OF AORTIC VALVE REPLACEMENT, PROSTHESIS TYPE, AND SIZEON FUNCTIONAL OUTCOME AND VENTRICULAR MASS IN PATIENTS WITH AORTIC STENOSIS J. Thorac. Cardiovasc. Surg., July 1, 1999; 118(1): 57 - 65. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Bernal, R. Martin-Duran, J. M. Rabasa, and J. M. Revuelta The CarboMedics ""Top-Hat"" supraannular prosthesis Ann. Thorac. Surg., May 1, 1999; 67(5): 1299 - 1303. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Hvass, G. M. Palatianos, R. Frassani, C. Puricelli, and M. O'Brien MULTICENTER STUDY OF STENTLESS VALVE REPLACEMENT IN THE SMALL AORTIC ROOT J. Thorac. Cardiovasc. Surg., February 1, 1999; 117(2): 267 - 272. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Vijayanagar, G. Chan, R. Musunuru, N. Sastry, I. Siegman, N. Rattehalli, and M. Cortelli Aortic Valve Replacement Without Annular Enlargement in Patients with Small Aortic Roots Asian Cardiovasc Thorac Ann, December 1, 1998; 6(4): 279 - 284. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Paulis, MD, L. Sommariva, MD, L. Colagrande, MD, G. M. De Matteis, MD, S. Fratini, MD, F. Tomai, MD, C. Bassano, MD, A. P. de Peppo, MD, and L. Chiariello, MD REGRESSION OF LEFT VENTRICULAR HYPERTROPHY AFTER AORTIC VALVE REPLACEMENT FOR AORTIC STENOSIS WITH DIFFERENT VALVE SUBSTITUTES J. Thorac. Cardiovasc. Surg., October 1, 1998; 116(4): 590 - 598. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Bartels, R. G. Leyh, J.F. M. Bechtel, E. Joubert-Hubner, and H.-H. Sievers Discrepancies Between Sizer and Valve Dimensions: Implications for Small Aortic Root Ann. Thorac. Surg., June 1, 1998; 65(6): 1631 - 1633. [Abstract] [Full Text] [PDF] |
||||