J Thorac Cardiovasc Surg 1996;112:712-726
© 1996 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
Supported by grant HL36308 from the National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda Md.
Received for publication Feb. 9, 1996 Revisions requested March 18, 1996; revisions received April 9, 1996 Accepted for publication April 11, 1996. Address for reprints: L. Henry Edmunds Jr., MD, Department of Surgery, Four Silverstein, Hospital of the University of Pennsylvania, 3400 Spruce St., Philadelphia, PA 19104.
Abstract
Objectives. The first objective was to develop a quantitative method for tracking the three-dimensional geometry of the mitral valve. The second was to determine the complex interrelationships of various components of the mitral valve in vivo. Methods and results. Sixteen sonomicrometry transducers were placed around the mitral valve anulus, at the tips and bases of both papillary muscles, at the ventricular apex, across the ventricular epicardial short axis, and on the anterior chest wall before and during cardiopulmonary bypass in eight anesthetized sheep. Animals were studied later on 17 occasions. Reproducibility of derived chord lengths and three-dimensional coordinates from sonomicrometry array localization, longevity of transducer signals, and the dynamics of the mitral valve and left ventricle were studied. Reproducibility of distance measurements averages 1.6%; Procrustes analysis of three-dimensional arrays of coordinate locations predicts an average error of 2.2 mm. Duration of serial sonomicrometry array localization signals ranges between 60 and 151 days (mean 114 days). Sonomicrometry array localization demonstrates the saddle-shaped mitral anulus, its minimal orifice area immediately before end-diastole, and uneven, apical descent during systole. Papillary muscles shorten only 3.0 to 3.5 mm. Sonomicrometry array localization demonstrates nonuniform torsion of papillary muscle transducers around a longitudinal axis and shows rotation of papillary muscular bases toward each other during systole. Conclusion. Tagging of ventricular structures in experimental animals by sonomicrometry array localization images is highly reproducible and suitable for serial observations. In sheep the method provides unique, quantitative information regarding the interrelationship of mitral valvular and left ventricular structures throughout the cardiac cycle. (J THORAC CARDIOVASC SURG1996;112:712-26)
Sonomicrometry array localization (SAL) identifies an array of individual piezoelectric, sonomicrometry transducers affixed to the heart in three-dimensional space.
1 In SAL the three-dimensional coordinates of each transducer position are calculated from multiple intertransducer distance measurements by means of multidimensional scaling (MDS). The basic concepts of the method have been recently described and validated,
1 but the original technology required 15 to 20 minutes to obtain all of the intertransducer distance measurements during steady state conditions, did not include a fixed frame of reference, and did not produce comparably oriented two-dimensional views of transducer locations.
1
This report describes improved technology that includes an internal anatomic reference frame and permits the measurement of 120 intertransducer distances at 200 Hz. New customized software generates three-dimensional coordinates for each transducer and, when applied to the ovine mitral valve, produces two-dimensional "images" in three orthogonal views throughout the cardiac cycle. The reproducibility of distance measurements and two-dimensional views is assessed and the capability of the method for serial measurements over time is demonstrated. The complex three-dimensional geometric interrelationships of various components of the normal ovine mitral valve during the cardiac cycle is described with unexpected findings.
Methods
Hardware
The Sonometrics Series 5001 digital sonomicrometer (Sonometrics Corp., London, Ontario, Canada) measures distances in an aqueous medium by means of piezoelectric transducers, a high-frequency counter, and the time-of-flight principle. It is capable of measuring all 120 distances (n[n - 1]/2) between 16 transducers once every 5 msec. The sonomicrometer consists of four components: (1) a 486-class personal computer that controls sonomicrometry functions and runs applications software; (2) a customized timing board that controls the length of the energizing pulse and the sequence of transducer transmissions; (3) a customized distance-measuring circuit that translates the ultrasonic signal into a distance measurement; and (4) an eight-channel analog-to-digital board that is synchronized with the sonomicrometer distance measurements. The transducers are 3 mm hemispheric PZT-5A piezoelectric crystals (Crystal Biotech, Hopkinton, Mass.). When interfaced with the Sonometrics sonomicrometer, these transducers emit ultrasound at a frequency of 800 kHz.
Animal studies
So that the ovine mitral valve can be imaged, 16 transducers are placed as follows. Six transducers are sutured around the mitral valve anulus: one each at the anterior and posterior commissures, two along the anterior leaflet (dividing it into three approximately equal segments), and two along the posterior leaflet. Transducers are also sutured to the tips and bases of both papillary muscles. The short axis of the heart is defined by anterior and posterior epicardial transducers. A third epicardial transducer is placed at the ventricular apex. The final three transducers are sutured to the anterior chest wall and sternum, 3 to 5 cm apart, to form a triangular anatomic reference frame. Fig. 1 depicts the locations of the 13 cardiac transducers.
|
Through a sterile left lateral thoracotomy, the three chest wall transducers were sutured to the sternum and chest wall away from the lung. Three epicardial transducers were sutured to the short axis and apex of the ventricle. After heparin (120 mg intravenously), normothermic cardiopulmonary bypass was started and the heart was electrically fibrillated. The perfusion circuit consisted of a 28F wire-wrapped venous cannula (Medtronic Inc., Minneapolis, Minn.) placed into the right ventricle through the main pulmonary artery, a 14F wire-wrapped arterial cannula in the left carotid artery, a centrifugal pump with precalibrated electromagnetic flowmeter (Medtronic Inc.), and a bubble oxygenator/heat exchanger (model H-1700, Bard Cardiopulmonary Division, Tewksbury, Mass.). The large left atrial appendage was opened to expose the mitral valve. Under direct vision, 10 endocardial transducers were sutured to the mitral valve as described earlier (see Fig. 1). Polydioxanone hemoclips (Ethicon, Inc., Somerville, N.J.) were placed across the wire at tissue exit points to hold transducers in place. Wires to the papillary muscles were brought out through the left ventricular free wall below the plane of the valve. The annular transducer wires were brought out through the atriotomy. No wire crossed the plane of the mitral valve. After the atrium was closed, the animal was defibrillated and weaned from cardiopulmonary bypass. The 16 color-coded transducer wires were brought out through the chest wall and secured externally by a plastic patch sutured to the skin.
Animals were studied between 10 and 151 days after the instrumentation operation. A total of 17 studies were performed on eight sheep at various intervals. For SAL studies sheep were sedated with thiopental, intubated, and anesthetized with isoflurane. Animals were supine, their lungs mechanically ventilated, and their hearts atrially paced at 120 beats/min. Under sterile conditions a high-fidelity pressure transducer (SPC-350, Millar Instruments, Inc., Houston, Tex.) was passed through a cutdown into the left ventricle via the left carotid artery. Surface electrocardiogram and left ventricular pressure were monitored continuously (Hewlett-Packard 78534C monitor, Hewlett-Packard Company, Andover, Mass.). A Swan-Ganz catheter (131H-7F, Baxter Healthcare Corp., Irvine, Calif.) was introduced via the right internal jugular vein to measure pulmonary artery and capillary wedge pressures. Thermodilution cardiac output was measured in duplicate before sonomicrometry data were collected. Ventilation was suspended during measurements.
Sonomicrometer settings
Two sonomicrometer settings, pulse length and inhibit delay, are adjusted to maximize the number of measured chord lengths. The pulse length is the time during which a voltage is applied to a transducer to produce sound waves. The applied voltage also produces an electromagnetic pulse that prematurely activates receiving transducers; this artifact necessitates an inhibit delay. A longer pulse length improves signal quality but requires a longer inhibit delay, which increases the minimum distance that can be measured. Adjustments attempt to maximize pulse length and minimize inhibit delay. Data are collected in 5-second runs.
Sonomicrometry signal processing
Spikes caused by noise and intermittent level shifts caused by failure to track the leading wave front of the ultrasound signal are removed at a later time with software produced by Sonometrics Corporation. With the use of customized software (J. T. Streicher) any heartbeat during the data run can be chosen for analysis. Each chord length is examined within the chosen heartbeat. Some chord lengths are not successfully measured and other chord lengths are measured twice. Only one measurement of each chord length is selected for input into the MDS algorithm.
MDS is an iterative technique that calculates three-dimensional coordinates from a group of chord lengths.
2,3 MDS compensates for measurement error in in vivo studies.
1 MDS also accurately determines any intertransducer distances that are not directly measured (i.e., between two transducers) by establishing the coordinates of the two transducers and calculating the distance between them. Customized software was developed to implement the MDS algorithm (K. B. Gupta). The three-dimensional coordinates for all transducers at all time points in the cardiac cycle are determined.
Within the chosen heartbeat four time points of special interest in the cardiac cycle are identified by means of the electrocardiogram and pressures. (The aortic root pressure trace is first smoothed by applying a high-frequency filter once.) End-diastole is defined as the peak of the QRS complex; end-systole is defined as the maximum negative left ventricular dP/dt.
4 End isovolumic contraction is defined as the first time point at which the aortic root dP/dt is greater than zero. End isovolumic relaxation is defined as the time at which the left ventricular pressure is 10% of LVPmax and the left ventricular dP/dt is less than 0.
5
Imaging
The three chest wall transducers are used to determine a fixed reference frame and coordinate system in which the 13 transducers attached to the heart are located. Within this fixed reference frame three orthogonal viewing planes are established. These are related to the heart as follows: The first plane is determined by fitting a least-squares plane through the six annular transducers at end-diastole (Fig. 2, annular plane). The second plane is oriented orthogonal to the annular plane and passes through both the centroid of the mitral anulus and the transducer located at the anterior commissure (Fig. 2, sagittal plane). The third plane is oriented orthogonal to the other two planes and also passes through the centroid (Fig. 2, coronal plane). Once these planes are established at end-diastole, they do not change during the cardiac cycle. Individual transducers may move in and out of each viewing plane, but the planes remain fixed.
|
Ventricular torsion is calculated by means of the definitions depicted in Fig. 3. The midpoint of the line segment joining the anterior and posterior epicardial transducers is determined, and a longitudinal axis from that point to the apical transducer is defined. The orientation of a ventricular or annular transducer with respect to this axis is found by defining a unit vector along a line through the transducer perpendicular to the axis. The intersection of this line with the longitudinal axis provides the axial position (l) of the transducer. Arbitrarily, the vector to the posterior commissural transducer is assigned an orientation angle of zero and an axial length of zero (l = 0). The axial position of all transducers is defined at end-diastole. The change in orientation vector at any time point indicates rotation and is obtained by subtracting the vector angle at end-diastole from the angle at the time point of interest. Rotations that are in a counterclockwise direction as viewed from the base of the heart toward the apex are defined as positive; clockwise rotations are negative.
|
Estimation of reproducibility of intertransducer distance measurements and transducer coordinates
The protocol for determining the reproducibility of SAL distance measurements and transducer coordinates appears in Appendix 1. The effect of variations in sonomicrometry settings and the effects of within-analyst and between-analyst variability on reproducibility were studied, as was the reproducibility of sequential measurements in the same sheep.
Longevity of sonomicrometry signals
The longevity of a full array (n = 16) of sonomicrometry transducer signals was assessed in five sheep, which were monitored until loss of one or more signals or the animal's death.
Anatomy and histology
The attachment of the mitral leaflets to the heart was examined by cutting tissue blocks perpendicular to the plane of the mitral anulus through the valve leaflet, atrium, anulus, and ventricle at multiple locations around the anulus in a dog, baboon, two sheep, and a human cadaver. Photographs were taken of the cross sections. In addition, for the sheep, tissue blocks were fixed in formaldehyde, stained in hematoxylin and eosin, and microscopically sectioned perpendicular to the leaflet attachment to the heart.
The total area of the mitral leaflets was determined by planimetry in fresh hearts after euthanasia in three animals in which axial views were obtained during life by SAL. The ratio of leaflet area to maximal annular area was calculated for these sheep.
Results
Distance measurements
SAL produces distance measurements between any two transducers every 5 msec throughout the cardiac cycle. For the array of 16 transducers approximately 100 sets (depending on heart rate) of 120 distances between transducers are generated. These distance measurements calculated by MDS agree closely with direct measurements between each transducer pair.
1
Orthogonal two-dimensional views of the mitral valve
Three orthogonal views indicating the locations of all 16 transducers at end-diastole and end-systole are presented in Fig. 4. The images presented in this figure are from a single sheep but are representative of all eight. Although SAL measures chord lengths rather than arc lengths, the method estimates the circumference and area of the mitral anulus. Fig. 4 shows that the area of the mitral orifice in the axial view (Fig. 4, C) is smaller at end-diastole than at end-systole. The descent of the mitral anulus toward the ventricular apex during systole, the nearly constant position of the ventricular apex, and the "saddle-shaped" configuration
6 of the mitral anulus with depressed commissures are also apparent in these views.
|
|
|
|
|
|
|
SAL images of the ovine mitral anulus
Fig. 9 is an illustration of the three-dimensional shape of the mitral anulus at end-diastole and end-systole. An arbitrary surface passing through each of the annular transducers has been drawn to better demonstrate the "saddle shape" of the anulus, with apically depressed commisures, which are maintained throughout the cardiac cycle. All sheep studied had anuli of similar shapes.
|
|
|
Fig. 12, A shows the rotation of the tips and bases of both papillary muscles around a longitudinal ventricular axis. As viewed from the base of the heart, the base of the posterior papillary muscle rotates counterclockwise and the base and tip of the anterior papillary muscle rotate clockwise. The tip of the posterior papillary muscle undergoes little angular change. This rotational movement of the papillary muscles during the cardiac cycle indicates considerable changes in the direction of the forces tethering the mitral leaflets. Fig. 12, B shows that torsion along the posterior papillary muscle is greater in magnitude and opposite in direction to that along the anterior papillary muscle. During the majority of systole, the posterior papillary muscle is subject to a rapid increase in torsion. During late systole posterior papillary muscle torsion becomes constant for approximately 50 msec. With the onset of isovolumic relaxation the posterior papillary muscle sustains a rapid decrease in torsion that continues through the remainder of diastole at a slower rate.
|
SAL is a new method to accurately determine the three-dimensional locations of tagged cardiac components throughout the cardiac cycle in laboratory animals. The method produces reproducible distance measurements and two-dimensional views of transducer locations for periods ranging from several weeks to a few months. Our previous study
1 and this one show that signal processing and the MDS algorithm add minimal error to distance measurements between transducers and to three-dimensional transducer coordinates if all transducers continue to function. The addition of a fixed reference frame distinguishes translational movements from cardiac deformations and allows control of the orientation of imaging planes. Data can now be acquired at 200 Hz. These improvements overcome important limitations of our first-generation studies
7,8 and greatly expand the utility of the method. SAL can now be used to study cardiac inflow valves and other cardiac structures and to describe remodeling changes that occur over time after experimental myocardial infarction, injury, or disease.
SAL complements other methods of cardiac imaging and has both distinct advantages and limitations. SAL requires surgery to tag specific cardiac structures. Variations in placement of transducers between sheep preclude precise image comparisons between sheep. The ability to track the three-dimensional coordinates of tagged structures in animals serially for several months is an important advantage of the method. In contrast, quantitative echocardiography and magnetic resonance imaging depend on anatomic landmarks for location of cardiac structures.
9,10 Neither method can follow movements of islands of muscle within the heart during remodeling after myocardial infarction. On the other hand, both magnetic resonance imaging and quantitative echocardiography image intracardiac structures with increasing clarity, have good to excellent edge detection,
11 and can be used clinically. Each provides additional information, such as Doppler flow measurements or scanning photoacoustic microscopy grid lines
12 for calculations of regional strains and stresses.
13
SAL compares most closely with tantalum marker radiography and both complements and competes with that method. Both methods require surgery to place markers in or on anatomic structures within or on the heart, but once placed tantalum markers serve for the life of the animal, whereas sonomicrometry transducers eventually fail. More tantalum markers can be placed than sonomicrometry transducers, and tantalum markers can label more delicate, fast-moving structures because they do not have wires attached. Tantalum markers have been used in human transplanted hearts; current SAL technology cannot provide human data. Both methods have been validated for accuracy
14,15 and both provide reproducible data with small errors between measurements. Reproducibility of tantalum radiography has been assessed in vitro only.
14 Data presented in this article demonstrate in vivo reproducibility of SAL. The major limitations of tantalum marker radiography are high capital, engineering, and computer programming requirements; these requirements severely restrict its application. The relative simplicity and low costs of SAL allow other investigators to use this tracking technology in animal models.
Surgical placement of markers, whether sonomicrometry transducers or tantalum markers, reliably identifies the labeled anatomic structures and permits precise (see reproducibility studies herein) measurements between markers within that heart. Because different animals are not precisely identical and because no surgeon can place each marker in the same three-dimensional anatomic position in different hearts, both methods provide primarily qualitative comparisons between hearts and animals. Distances between markers are measured with great precision by both methods,
1,14,15 but the anatomic variations between hearts and differences in surgical placement significantly increase the standard deviation of measurements even if data are normalized for animal size. Nevertheless, the precision of serial comparisons within hearts over time provide important and otherwise unattainable quantitative information regarding ventricular deformations during the cardiac cycle and over days, weeks, and months (and years with tantalum).
Ventricular deformations during the cardiac cycle are complex and difficult to describe in quantitative terms. Together, the arrangement of ventricular muscle bundles and the time-dependent dynamics of the left ventricle preclude simplistic models of dynamic geometric changes in both healthy and diseased hearts. Ventricular torsion is particularly difficult to quantify by existing clinical imaging methods. Reliable data have been obtained with tantalum radiography.
16-18 SAL provides a relatively simple method for studying ventricular torsion and is particularly applicable to studies of the mitral valve. Varying placement of transducers allows endocardial, midwall, and epicardial ventricular torsion to be separately studied. With SAL, torsion can be measured relative to a ventricular longitudinal axis or relative to other axes that may have more relevance to a particular function or pathologic state.
The sheep is an almost ideal model of ischemic heart disease. Unlike other large animal models,
19-22 the sheep does not have preformed coronary arterial collateral vessels and has highly reproducible coronary anatomy.
23,24 Anteroapical left ventricular aneurysm,
23 postinfarction ventricular arrhythmias,
25,26 and acute and chronic ischemic mitral regurgitation
7,8,24,27 are reliably produced by occlusion of selected coronary arteries. These diseases occur in patients who have more varied coronary arterial anatomy,
28 but in patients correlation between coronary anatomy and development of ischemic mitral insufficiency or arrhythmias is much less predictable
29,30 than in sheep. The availability, cost, size, and docility of sheep are other important advantages.
The ovine mitral valve functions similarly to that of the human being, baboon, and dog but differs in anatomic details. In sheep the mitral leaflets attach to atrial muscle rather than to ventricular muscle. The valve also differs with respect to the length of free papillary muscle. The fact that these anatomic differences do not impair function raises the question of their importance in exercise and pathologic states, such as ischemic mitral regurgitation. The area of ovine mitral leaflets is approximately 1.6 times the area of the mitral orifice during isovolumic relaxation, when the valve is closed. This ratio has been estimated in cadavers
31 but not with in vivo measurements of mitral orifice area. The amount of actual papillary muscle shortening is approximately 2 to 3 mm in sheep
8 and dog.
5 However, the impact of this shortening on the closing geometry of the mitral valve may be relatively small compared with the descent of the mitral anulus toward the ventricular apex, rotational movement of the posterior papillary muscle around the longitudinal axis of the left ventricle, and concentric narrowing of the ventricle along the short axes. These spatial deformations of the normal ovine valve are probably relevant to the mechanisms that produce postinfarction mitral regurgitation.
7,8
This study illustrates an application of SAL for study of the dynamic, three-dimensional geometry ofthe normal ovine mitral valve. This work points out some unique anatomic characteristics of thesheep valve and provides new quantitative information regarding left ventricular deformations relativeto the mitral valve during the cardiac cycle. This new imaging technology combined with ovinemodels of ischemic heart disease is particularly promising for study of ventricular remodeling aftermyocardial infarction.
Appendix 1
Estimation of reproducibility of intertransducer distance measurements and transducer coordinates
Differences in sonomicrometry settings, selection of data for analysis, and processing may affect reproducibility of sequential distance measurements and coordinate locations. Therefore variations in sonomicrometry settings and the effects of within-analyst and between-analyst variability on reproducibility and the reproducibility of sequential measurements in the same sheep were studied.
Reproducibility between sonomicrometer settings
Within 2 minutes in a monitored, anesthetized, hemodynamically stable sheep five data sets, each lasting 5 seconds and each at a different sonomicrometer setting, were recorded. A single heartbeat was selected from each data set; a total of five heartbeats were processed. For each selected heartbeat, all intertransducer chord lengths were determined at end-diastole by means of MDS. The five measurements of each of the 120 end-diastolic chord lengths were averaged (n = 5) and a standard deviation for each of the 120 chord lengths was calculated. The normalized error for each individual chord length was estimated by dividing the standard deviation by the mean of the five measured chord lengths for each of the 120 distances. This normalized error for each of the 120 distances was averaged and tabulated as a mean, standard deviation, and range.
The five data sets were also compared by means of the least-squares Procrustes rotation algorithm.32-34 This technique compares three-dimensional coordinates from two separate data sets by superimposing the two data sets. The second data set Y is rotated and translated with respect to the first data set X so that the residual sum of the squares is minimized. The residual sum of the squares is defined:

where xr and yr are the individual points of X and Y, A' is the rotation matrix, and b the translation vector that operates on Y. Values of A' and b are obtained such that R2 is minimized.
The average absolute error E, which has units of length, is defined:

The minimum R2 and the corresponding E were calculated for each data set comparison. Calculations were performed on an 80486-based personal computer (Gateway 2000 Inc., North Sioux City, S.D.) with source code written in C (M. B. Ratcliffe).
Between any two heartbeats the Procrustes technique determines an average absolute error for the array of 16 transducers. Each of the five heartbeats processed was compared with each of the other four; therefore ten Procrustes comparisons were made. The mean of the ten average absolute errors was calculated. This procedure was followed twice, once by each of two analysts.
This protocol does not differentiate between within-analyst variability and the variability caused by changes in sonomicrometer settings. However, the variability associated with adjustment of the sonomicrometer can be estimated by the difference between these average normalized errors and the within-analyst normalized errors that were determined as described below.
Reproducibility within analysts
An analyst adjusted the sonomicrometer for pulse length and inhibit delay and recorded five data sets of five seconds each from an anesthetized, stable sheep. The same analyst selected and processed a single heartbeat from each of the five data sets obtained. An average normalized error was calculated as described earlier. The five heartbeats were also compared by means of the Procrustes algorithm. This complete protocol was repeated by a second analyst and again separately by both analysts on a second sheep.
Reproducibility between analysts
Five data sets, each at a different sonomicrometry setting, were recorded. Each of two analysts processed a single heartbeat (not necessarily the same heartbeat) from each of the five data sets, and all intertransducer distances were determined at end-diastole by means of MDS. This produced five pairs of heartbeats, one from each analyst at each sonomicrometer setting, which were compared as follows. The two measurements of the 120 end-diastolic chord lengths were averaged (n = 2) and the absolute difference for each of the 120 chord lengths was determined. The normalized error for each chord length was estimated by dividing the absolute difference by the mean of the two measurements, and the mean of these 120 normalized errors was found. Each pair of heartbeats was also compared by use of the Procrustes algorithm. The mean, standard deviation, and range of five normalized distance errors and the mean of five Procrustes absolute errors were calculated.
Combined reproducibility
An estimate of the effect of total potential variability on the reproducibility of the technique was performed. Five data sets, at a single sonomicrometer setting, were collected. The sonomicrometer settings were changed, and another five data sets were recorded. One heartbeat from each of the first five data sets was reduced by one analyst; the second analyst processed one heartbeat from each of the second five data sets. Each data set processed by analyst 1 was compared with each data set processed by analyst 2, for a total of 25 comparisons. For each comparison, a mean normalized chord length error was calculated and a Procrustes rotation was performed.
Reproducibility between study sessions
Data were collected from the same sheep during two study sessions 24 hours apart. The hemodynamic conditions on both days were similar. The sonomicrometer settings and analyst were the same for both sessions. The two data sets were compared by calculating the average difference and average percent difference of all 120 chord lengths, as well as applying the Procrustes algorithm to assess the differences in crystal coordinates between data sets.
Appendix 2
The Procrustes rotation is a least-squares method used to superimpose and compare two multidimensional configurations. Let X be an n x p matrix of the coordinates of n points obtained by one measurement method. Let Y be another n x p matrix of the coordinates of those same n points obtained by a different measurement method. The residual sum of squares of Y relative to X is:

where xr and yr are the individual points of X and Y respectively. The residual sum of squares can be minimized by rotating or translating Y relative to X as follows:
Ar 'yr b, r = 1 , . . . , n (4)
where A' is the transpose of a p x p orthogonal matrix. If equations 1 and 2 are combined in the following equation:

and values of A' and b are obtained such that R2 is minimized, then Y is superimposed on X. Values of A' and b are obtained in the following manner. Let
Z = Y'X (5)
where Y' is the transpose of Y. By singular value decomposition:
Z = VU' (6)
where V and U are orthogonal (p x p) matrices, U' is the transpose of U, and is a diagonal matrix of nonnegative elements. The minimizing solution for equation 3 is given by:
b = 0,A = VU' (7)
and further:
R2 = trXX' + trYY' 2tr (8)
where tr is the trace of a matrix, X' is the transpose of X, and Y' is the transpose of Y.
Acknowledgments
We thank Nicolas Gikakis, Michelle Money, and Dr. Yugi Hiramatsu for their invaluable help in these studies.
References
This article has been cited by other articles:
![]() |
W. Bouma, I. C.C. van der Horst, I. J. Wijdh-den Hamer, M. E. Erasmus, F. Zijlstra, M. A. Mariani, and T. Ebels Chronic ischaemic mitral regurgitation. Current treatment results and new mechanism-based surgical approaches Eur. J. Cardiothorac. Surg., January 1, 2010; 37(1): 170 - 185. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G. Yuen, D. T. Kettler, P. M. Novotny, R. D. Plowes, and R. D. Howe Robotic Motion Compensation for Beating Heart Intracardiac Surgery The International Journal of Robotics Research, October 1, 2009; 28(10): 1355 - 1372. [Abstract] [PDF] |
||||
![]() |
P. P. Caimmi, M. Diterlizzi, E. Grossini, E. I. Kapetanakis, M. Gavinelli, A. Carriero, and G. Vacca Impact of prosthetic mitral rings on aortomitral apparatus function: a cardiac magnetic resonance imaging study. Ann. Thorac. Surg., September 1, 2009; 88(3): 740 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Bothe, T. C. Nguyen, M. E. Roberts, T. A. Timek, A. Itoh, N. B. Ingels Jr., and D. C. Miller Presystolic mitral annular septal-lateral shortening is independent from left atrial and left ventricular contraction during acute volume depletion Eur. J. Cardiothorac. Surg., August 1, 2009; 36(2): 236 - 243. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Sonne, L. Sugeng, N. Watanabe, L. Weinert, K. Saito, M. Tsukiji, K. Yoshida, M. Takeuchi, V. Mor-Avi, and R. M. Lang Age and body surface area dependency of mitral valve and papillary apparatus parameters: assessment by real-time three-dimensional echocardiography Eur J Echocardiogr, March 1, 2009; 10(2): 287 - 294. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Nonaka, A. Marui, M. Fukuoka, T. Shimamoto, S. Masuyama, T. Ikeda, and M. Komeda Differences in mitral valve-left ventricle dimensions between a beating heart and during saline injection test Eur. J. Cardiothorac. Surg., October 1, 2008; 34(4): 755 - 759. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. O. Jensen, H. Jensen, M. Smerup, R. A. Levine, A. P. Yoganathan, H. Nygaard, J. M. Hasenkam, and S. L. Nielsen Saddle-Shaped Mitral Valve Annuloplasty Rings Experience Lower Forces Compared With Flat Rings Circulation, September 30, 2008; 118(14_suppl_1): S250 - S255. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. P. Ryan, B. M. Jackson, H. Hamamoto, T. J. Eperjesi, T. J. Plappert, M. St. John-Sutton, R. C. Gorman, and J. H. Gorman III The Influence of Annuloplasty Ring Geometry on Mitral Leaflet Curvature Ann. Thorac. Surg., September 1, 2008; 86(3): 749 - 760. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. C. Nguyen, A. Itoh, C. J. Carlhall, W. Bothe, T. A. Timek, D. B. Ennis, R. A. Oakes, D. Liang, G. T. Daughters, N. B. Ingels Jr., et al. The effect of pure mitral regurgitation on mitral annular geometry and three-dimensional saddle shape J. Thorac. Cardiovasc. Surg., September 1, 2008; 136(3): 557 - 565. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. P. Sengupta, A. J. Tajik, K. Chandrasekaran, and B. K. Khandheria Twist mechanics of the left ventricle principles and application. J. Am. Coll. Cardiol. Img., May 1, 2008; 1(3): 366 - 376. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Sakamoto, L. M. Parish, H. Hamamoto, L. P. Ryan, T. J. Eperjesi, T. J. Plappert, B. M. Jackson, M. G. St John-Sutton, J. H. Gorman III, and R. C. Gorman Effect of Reperfusion on Left Ventricular Regional Remodeling Strains After Myocardial Infarction Ann. Thorac. Surg., November 1, 2007; 84(5): 1528 - 1536. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. George, E. I. Kapetanakis, K. Dhadwal, S. Al-Ruzzeh, E. Rigopoulou, M. Bulanade, T. Athanasiou, and M. Amrani A three-dimensional echocardiographic comparison of a deep pericardial stitch versus an apical suction device for heart positioning during beating heart surgery Eur. J. Cardiothorac. Surg., October 1, 2007; 32(4): 604 - 610. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. P. Ryan, B. M. Jackson, Y. Enomoto, L. Parish, T. J. Plappert, M. G. St. John-Sutton, R. C. Gorman, and J. H. Gorman III Description of regional mitral annular nonplanarity in healthy human subjects: A novel methodology J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 644 - 648. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S Sacks and A. P Yoganathan Heart valve function: a biomechanical perspective Phil Trans R Soc B, August 29, 2007; 362(1484): 1369 - 1391. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. P. Ryan, B. M. Jackson, L. M. Parish, T. J. Plappert, M. G. St. John-Sutton, J. H. Gorman III, and R. C. Gorman Regional and Global Patterns of Annular Remodeling in Ischemic Mitral Regurgitation Ann. Thorac. Surg., August 1, 2007; 84(2): 553 - 559. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Joudinaud, C. L. Kegel, E. M. Flecher, P. A. Weber, E. Lansac, U. Hvass, and C. M.G. Duran The papillary muscles as shock absorbers of the mitral valve complex. An experimental study Eur. J. Cardiothorac. Surg., July 1, 2007; 32(1): 96 - 101. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Carlsson, M. Ugander, H. Mosen, T. Buhre, and H. Arheden Atrioventricular plane displacement is the major contributor to left ventricular pumping in healthy adults, athletes, and patients with dilated cardiomyopathy Am J Physiol Heart Circ Physiol, March 1, 2007; 292(3): H1452 - H1459. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Sakamoto, L. M. Parish, H. Hamamoto, Y. Enomoto, A. Zeeshan, T. Plappert, B. M. Jackson, M. G. St. John-Sutton, R. C. Gorman, and J. H. Gorman III Effects of hemodynamic alterations on anterior mitral leaflet curvature during systole J. Thorac. Cardiovasc. Surg., December 1, 2006; 132(6): 1414 - 1419. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Sacks, Y. Enomoto, J. R. Graybill, W. D. Merryman, A. Zeeshan, A. P. Yoganathan, R. J. Levy, R. C. Gorman, and J. H. Gorman III In-vivo dynamic deformation of the mitral valve anterior leaflet. Ann. Thorac. Surg., October 1, 2006; 82(4): 1369 - 1377. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. A Goetz, E. Lansac, H.-S. Lim, S. A Stevens, P. A Weber, and C. M. Duran Kinking of the Atrioventricular Plane During the Cardiac Cycle Asian Cardiovasc Thorac Ann, October 1, 2006; 14(5): 394 - 398. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. G. Leshnower, H. Sakamoto, A. Zeeshan, L. M. Parish, R. Hinmon, T. Plappert, B. M. Jackson, J. H. Gorman III, and R. C. Gorman Role of acetaminophen in acute myocardial infarction Am J Physiol Heart Circ Physiol, June 1, 2006; 290(6): H2424 - H2431. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek, J. R. Glasson, D. T. Lai, D. Liang, G. T. Daughters, N. B. Ingels Jr, and D. C. Miller Annular Height-to-Commissural Width Ratio of Annulolasty Rings In Vivo Circulation, August 30, 2005; 112(9_suppl): I-423 - I-428. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Watanabe, Y. Ogasawara, Y. Yamaura, N. Wada, T. Kawamoto, E. Toyota, T. Akasaka, and K. Yoshida Mitral Annulus Flattens in Ischemic Mitral Regurgitation: Geometric Differences Between Inferior and Anterior Myocardial Infarction: A Real-Time 3-Dimensional Echocardiographic Study Circulation, August 30, 2005; 112(9_suppl): I-458 - I-462. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Watanabe, Y. Ogasawara, Y. Yamaura, T. Kawamoto, E. Toyota, T. Akasaka, and K. Yoshida Quantitation of mitral valve tenting in ischemic mitral regurgitation by transthoracic real-time three-dimensional echocardiography J. Am. Coll. Cardiol., March 1, 2005; 45(5): 763 - 769. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. X. Qin, T. Shiota, H. Tsujino, G. Saracino, R. D. White, N. L. Greenberg, J. Kwan, Z. B. Popovic, D. A. Agler, W. J. Stewart, et al. Mitral annular motion as a surrogate for left ventricular ejection fraction: real-time three-dimensional echocardiography and magnetic resonance imaging studies Eur J Echocardiogr, December 1, 2004; 5(6): 407 - 415. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. M. Parish, B. M. Jackson, Y. Enomoto, R. C. Gorman, and J. H. Gorman III The Dynamic Anterior Mitral Annulus Ann. Thorac. Surg., October 1, 2004; 78(4): 1248 - 1255. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. Gorman III, B. M. Jackson, S. L. Moainie, Y. Enomoto, and R. C. Gorman Influence of inotropy and chronotropy on the mitral valve sphincter mechanism Ann. Thorac. Surg., March 1, 2004; 77(3): 852 - 857. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. Gorman III, B. M. Jackson, Y. Enomoto, and R. C. Gorman The effect of regional ischemia on mitral valve annular saddle shape Ann. Thorac. Surg., February 1, 2004; 77(2): 544 - 548. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Lansac, H.S. Lim, Y. Shomura, K.H. Lim, N.T. Rice, W. Goetz, C. Acar, and C.M.G. Duran A four-dimensional study of the aortic root dynamics Eur. J. Cardiothorac. Surg., October 1, 2002; 22(4): 497 - 503. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. M. Jackson, J. H. Gorman III, S. L. Moainie, T. S. Guy, N. Narula, J. Narula, M. G. St. John-Sutton, L. H. Edmunds Jr, and R. C. Gorman Extension of borderzone myocardium in postinfarction dilated cardiomyopathy J. Am. Coll. Cardiol., September 18, 2002; 40(6): 1160 - 1167. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. L. Moainie, T. S. Guy, J. H. Gorman III, T. Plappert, B. M. Jackson, M. G. St. John-Sutton, L. H. Edmunds Jr, and R. C. Gorman Infarct restraint attenuates remodeling and reduces chronic ischemic mitral regurgitation after postero-lateral infarction Ann. Thorac. Surg., August 1, 2002; 74(2): 444 - 449. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Leotta, G. Patejunas, G. Murphy, J. Szokol, L. McGregor, J. Carbray, A. Hamawy, D. Winchester, N. Hackett, R. Crystal, et al. Gene therapy with adenovirus-mediated myocardial transfer of vascular endothelial growth factor 121 improves cardiac performance in a pacing model of congestive heart failure J. Thorac. Cardiovasc. Surg., June 1, 2002; 123(6): 1101 - 1113. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Lansac, K. H. Lim, Y. Shomura, W. A. Goetz, H. S. Lim, N. T. Rice, H. Saber, and C. M. G. Duran Dynamic balance of the aortomitral junction J. Thorac. Cardiovasc. Surg., May 1, 2002; 123(5): 911 - 918. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Timek and D. C. Miller Experimental and clinical assessment of mitral annular area and dynamics: what are we actually measuring? Ann. Thorac. Surg., September 1, 2001; 72(3): 966 - 974. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. Mellors and C. J. Barclay The energetics of rat papillary muscles undergoing realistic strain patterns J. Exp. Biol., January 11, 2001; 204(21): 3765 - 3777. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. H. Gorman III, R. C. Gorman, T. Plappert, B. M. Jackson, Y. Hiramatsu, M. G. St. John-Sutton, and L. H. Edmunds Jr. Infarct Size And Location Determine Development Of Mitral Regurgitation In The Sheep Model J. Thorac. Cardiovasc. Surg., March 1, 1998; 115(3): 615 - 622. [Abstract] [Full Text] |
||||
![]() |
J. H. Gorman III, R. C. Gorman, B. M. Jackson, Y. Hiramatsu, N. Gikakis, S. T. Kelley, M. G. St. J. Sutton, T. Plappert, and L. H. Edmunds Jr Distortions of the Mitral Valve in Acute Ischemic Mitral Regurgitation Ann. Thorac. Surg., October 1, 1997; 64(4): 1026 - 1031. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||