J Thorac Cardiovasc Surg 1997;114:224-233
© 1997 Mosby, Inc.
SURGERY FOR CONGENITAL HEART DISEASE |
Received for publication Dec. 31, 1996. Accepted for publication April 16, 1997. Address for reprints: Yukio Ichikawa, MD, PhD, First Department of Surgery, Yokohama City University School of Medicine, 3-9 Fukuura, Kanazawa-ku, Yokohama 236, Japan.
Abstract
Objective: We evaluated a bovine jugular vein graft with a natural trileaflet valve for right ventricular outflow tract reconstruction in a canine study for an entire year. Methods: Heparinized bovine jugular vein grafts with a natural valve cross-linked with a hydrophilic polyepoxy compound of 18 to 20 mm in internal diameter were implanted in the right ventricularpulmonary arterial position in eight dogs, and the main natural pulmonary artery was ligated. Anticoagulants were not used after implantation. Five grafts were retrieved on day 182 after implantation and the other three grafts on days 196, 375, and 385, respectively, and were inspected by macroscopic and microscopic observation. Cardiac catheterization, followed by angiography and echocardiography, was done just before graft retrieval. Results: No graft kinking or regurgitation of the valve was observed. Echocardiography showed natural valve motion without thickening of the leaflets. Blood pressure in the conduits ranged from 18/9 to 31/4 mm Hg, in the right ventricle from 18/4 to 40/0 mm Hg, and the gradient varied from 0 to 14 mm Hg. The explanted conduits maintained their original shape, softness, and pliability with good coaptation of valves, without calcification or degenerative changes, except for one leaflet with slight deformation. In microscopic observation, endothelial cells lined the luminal surface of the conduit except for the areas adjacent to the valve. Conclusion: The graft worked perfectly as a right ventricularpulmonary arterial valved conduit without anticoagulant therapy for a long time. J Thorac Cardiovasc Surg 1997;114:224-33
Many kinds of extracardiac conduits have been developed, but the valves of these conduits do not maintain satisfactory hemodynamic function in the low-pressure pulmonary arterial system for very long after implantation. The major problems are the inadequacy of synthetic valves, stenosis caused by degenerative changes of the biologic conduits, or peeling of the pseudointima on the grafts.
1-13 Other limitations are size, availability, and antithrombogenicity of such bioprostheses for children. Many other problems are present as well, such as the quality of the raw materials and cross-linking of the bioprostheses in clinics. The ideal graft has to have ideal structure and hemodynamic properties. The cross-linking must provide biocompatibility and hydrophilic properties without cytotoxicity. We must also be able to obtain various sizes of conduits readily at any time. To solve these problems, we developed a conduit using bovine jugular veins, which contain a natural trileaflet valve. These veins with natural valves have the structure peculiar to venous tissue, which makes them adequate to function as a valved conduit in the low-pressure system. To preserve the mechanical properties of the natural tissues and to reduce the immunogenicity of the animal tissue, we treated the graft with a hydrophilic polyepoxy compound (PC) instead of the usual glutaraldehyde. Studies have been conducted comparing the mechanical properties of biologic prostheses cross-linked with PC or glutaraldehyde by measuring their compliance and pliability. An arterial vascular prosthesis cross-linked with PC showed more pliability than one cross-linked with glutaraldehyde.
14 The porcine valvular leaflets cross-linked with PC were more pliable than their glutaraldehyde counterparts in two-dimensional echocardiographic inspection. Epoxy-treated valves appeared to open more widely than glutaraldehyde-treated valves. Also, less calcium deposition and pannus ingrowth were observed in the epoxy compound valve than in its glutaraldehyde counterpart.
15
For antithrombogenicity, heparin was immobilized with protamine sulfate on the luminal surface of the graft. In this communication, we report the 1-year results with such a graft in a canine study. The results show that the new right ventricularpulmonary arterial conduit with a valve naturally functioning in a low-pressure system provides satisfactory hemodynamics and excellent preservation of the venous characteristics.
Materials and methods
Preparation of the conduit
A fresh bovine jugular vein with a natural valve of 18 to 20 mm in internal diameter was obtained from a slaughterhouse. It was soaked in distilled water for 1 hour and sonicated at 28 kc for 20 seconds to cause cell destruction. Cell debris and water-soluble proteins were removed by squeezing the vein in distilled water. Thus a natural tissue tube composed of collagen and elastic laminae was obtained (Fig. 1). A 0.7% protamine sulfate solution, pH 5.9, was impregnated into the luminal surface of the tube by air inflation at a pressure of 30 mm Hg. During the impregnation, the tube was crimped and cross-linked in a 4% solution of glycerol polyglycidyl ether polyepoxy compound (PC, Denacol EX-313; Nagase Chemical, Osaka, Japan), pH 9.5 ± 0.2 (0.2 mol/L carbonate-bicarbonate buffer), for 45 ± 3 hours at 37° C. After being rinsed with distilled water for more than 24 hours, the tube was soaked in a 1% heparin solution at pH 4.5 for 5 hours at 45° C and then washed again with distilled water for 24 hours. Then it was sterilized in a 4% solution of PC (Denacol Ex-810; Nagase) for 6 days at 38° C and preserved in 70% ethanol. Before implantation, the graft was squeezed with a dry sponge and immersed in physiologic saline solution three times to remove the ethanol and then soaked in physiologic saline solution for more than 20 minutes.
|
All animal care was in compliance with the "Principles of Laboratory Animal Care" formulated by the National Society for Medical Research and the "Guide for the Care and Use of Laboratory Animals" prepared by the National Academy of Sciences and published by the National Institutes of Health (NIH publication No. 80-23, revised 1985).
Cardiac catheterization study
A 7F thermodilution catheter was inserted through the left femoral vein under the same general anesthesia as described earlier. The right ventricularpulmonary arterial pressure gradient across the conduit was measured at 182 days in five dogs and at 196, 376, and 386 days after the operation in the other three dogs. Cardiac catheterization, followed by angiography and echocardiography, was done just before the graft retrieval.
Angiographic study
An angiographic catheter was inserted through the same vessel to allow for catheterization. Conventional right ventricular angiography with cut film was recorded in three dogs at 182 days while about 30 ml of contrast media was being manually injected into the right ventricle in the left anterior oblique position. Conventional pulmonary artery angiography with motion picture film was also performed in two dogs at 182 and 375 days, respectively, with the same procedure.
Echocardiography
A left thoracotomy was performed through the fourth intercostal space with the animal under general anesthesia. Echocardiography was conducted directly on the graft in two dogs at 182 days and in two dogs at 196 and 385 days, respectively.
Graft retrieval
All grafts were harvested from 182 days to 385 days after the operation with the aid of general anesthesia. Five grafts were explanted at 182 days and three grafts at 196, 375, and 385 days after the operation, respectively. Before harvesting, sodium heparin (100 IU/kg) was administered intravenously to prevent clotting. All retrieved grafts were rinsed with saline solution to remove excess intraluminal blood and were inspected macroscopically.
Observations
For light microscopic observation, 3 mm wide sections of the grafts were cut longitudinally from the right ventricular anastomosis to the pulmonary arterial anastomosis. Each sample was cut into four pieces, and identity marks were placed from the proximal to the distal area to observe the entire length of the graft in sequence. These specimens were fixed with 1% glutaraldehyde in phosphate buffered solution, 0.2 mol/L, pH 7.4, and embedded in a hydrophilic resin (Technovite 7100; Kulzer & Co. GmbH, Germany). Sections were stained with hematoxylin and eosin, von Kossa stain for the detection of calcification, Papanicolaou's method for the detection of factor VIII of endothelial cells, Victoria blue for collagen, and van Gieson elastin for the detection of elastic fiber.
Results
Prepared grafts
The bovine jugular vein grafts cross-linked with PC were soft and pliable like native venous tissue, that is, the grafts maintained their original mechanical properties. We could choose a suitable graft from various ones with an internal diameter of from 10 to 30 mm with two or three natural leaflets. The leaflets were also extremely thin and soft. In saline solution, the leaflets could open and close with a minimum flow of saline solution (Fig. 2). In the present study, grafts of 18 to 20 mm internal diameter with three leaflets were used. The outer surface of the graft was white and rough with connective tissue velour. The luminal surface was shiny and white (Fig. 3). The graft wall was tough enough to retain sutures.
|
|
|
|
|
|
|
|
|
|
|
The grafts made of bovine jugular vein with natural venous valves maintained natural valve function in a low-pressure area of the pulmonary arterial system for an entire 1-year observation period. No significant dilated area was present in any of the retrieved conduits. This finding was consistent with a short-term right ventricular angiographic observation in a preliminary study.
16 The pressure gradient across the conduit valve ranged from 0 to 14 mm Hg, which was a satisfactory value when compared with the mean gradient across the conduit (22.8 mm Hg) at completion of the operation with an extracardiac conduit in clinics.
11 In one dog (dog 1), scar formation was noted in the right ventricular muscle at the anastomotic site. This finding was consistent with the results of the angiographic examination. The high pressure gradient in this case was considered to be caused by this scar formation.
Although we did not use any anticoagulants after the graft implantation, thrombus adhesion at the leaflets was minimal, and these thrombi were extremely fresh without organization. Therefore the thrombi could have been created during the catheterization just before the graft retrieval. Endothelialization of the graft surface was also observed without intimal hyperplasia or degenerative changes such as calcification. All this evidence indicates that the graft worked perfectly as a right ventricularpulmonary arterial valved conduit for a long period of time and should be considered for clinical use.
In the current experiment, no control conduit was adopted because no suitable standard conduit for clinical use was available. Therefore we will discuss our results with the results reported in various references.
1-13 Total surgical repair of congenital heart lesions with pulmonary arterialright ventricular discontinuity was first accomplished with an aortic homograft in a patient with tetralogy of Fallot and pulmonary atresia.
17 Many kinds of conduits have been reported for right ventricular outflow tract reconstruction, but no conduits have maintained functional valves without degenerative changes or calcification without anticoagulant therapy.
1-13
Recently, success has been reported with cryopreserved allografts.
18,19 These grafts functioned for more than 10 years, which was a tremendous result compared with our 1-year experiment. Unfortunately, almost all showed valvular regurgitation. Selection of the size of the grafts was also discussed, and immunogenic problems as the pathogenesis of valve allograft failure were also pointed out. There are limitations to homografts, such as ready availability and size of the graft. On the other hand, we could choose from grafts with an internal diameter of from 10 to 30 mm. There were no limitations of availability and size of the graft in our bioprostheses.
Biologic valves made of porcine aortic valves have been used and are expected to maintain their function continuously. Their leaflets are, however, too rigid to work in the low-pressure system, because these grafts originate from aortic valves, not from pulmonary valves. It has been reported that the ratio of thickness of the media of the pulmonary trunk to that of the aorta was approximately 0.6,
20 that the amount of pulmonary elastic tissue per unit of medial surface area in a histologic section was only 50% to 60% of that in the aorta,
20 and that the total calcium content per gram of tissue in the pulmonary media was on an average less than half of that in the aortic counterpart.
21 Early calcification and stenosis of glutaraldehyde-treated porcine valves in children have been observed.
7,13
Another problem is the cross-linking method. The grafts previously reported have been cross-linked with glutaraldehyde, which makes biologic materials stiff and hydrophobic.
22,23 Cytotoxicity of residual glutaraldehyde released from the graft long after implantation can also induce another problem, foreign body reaction.
24 Host cells cannot migrate onto the cytotoxic area.
25 Therefore no endothelial cell lining can be hoped for in the glutaraldehyde-treated grafts.
Previously, cardiac wall substitute composed of collagen-coated ultrafine polyester mesh cross-linked by a hydrophilic polyepoxy compound (CUFP) or glutaraldehyde-treated equine pericardium was evaluated for right ventricular outflow tract implantation in dogs. There was endothelialization in the CUFP with fibroblasts and smooth musclelike cells, but no endothelialization was seen in the glutaraldehyde-treated graft without fibroblasts.
26 The glutaraldehyde treatment completely alters the original properties of the grafts.
To overcome these problems, we used three special technologies for the new valved conduit. First, as the raw material for the graft with valves, we used the bovine jugular vein, which has natural valves with adequate shape, structure, function, and mechanical properties to work in the low-pressure system without thrombus formation. These remarkable characteristics cannot be created by any one or combination of the various synthetic materials by means of currently available technology. Today's technology has produced various successful medical devices, but as raw materials for artificial organs, natural tissues and organs have still attractive properties both morphologically and functionally. The natural vein with venous valves is a typical case. If we can preserve these remarkable properties of natural tissues by some kind of chemical modification, we can produce ideal artificial organs in various fields.
Second, we used PC instead of glutaraldehyde for chemical cross-linking. PC cross-linking has been used for small-diameter vascular prostheses of biologic origin to preserve natural tissue elasticity and pliability.
27,30 PC contains ether bounds in its molecule. The ether bound is a free joint, and it makes the cross-linked graft soft and flexible. PC also contains hydroxyl groups in its molecule, which give the materials hydrophilicity. On the other hand, glutaraldehyde contains a continuous five-carbon chain, which is rigid and hydrophobic, and makes the cross-linked materials stiff and hydrophobic. Another advantage of PC is its low cytotoxicity compared with that of glutaraldehyde,
25 and materials cross-linked with PC do not show foreign body reaction after implantation.
29,31 Host cells can proliferate on the graft surface and migrate into the graft framework with the hydrophilicity and low cytotoxicity offered by PC. Consequently, PC cross-linking can provide a suitable environment for host cell migration, resulting in endothelial cell lining on the graft surface. Microscopic observation in the current study indicated that the endothelial cells migrated 1 to 2 cm away from the suture lines to near the leaflets, but no endothelialization was noted on the leaflets. Because endothelial cell division is limited by cell aging, endothelial lining to cover the entire surface of the graft is difficult.
32,33 In the present experiment, the valves maintained their original mechanical properties after the PC cross-linking.
In general, biologic materials that contain collagen are thrombogenic, because collagen molecules have many amino groups that are positively charged and can accumulate negatively charged platelets.
34,35 These amino groups, however, are used for the cross-linking reaction, and carboxyl groups in the collagen molecules become relatively dominant. Therefore the materials become negatively charged and become less thrombogenic,
36 since platelets, which have a negative charge, show repulsion toward the same charge. Moreover, biologic materials cross-linked with PC become hydrophilic, because PC contains hydroxyl groups. In general, hydrophilic substances in wet conditions are surrounded with water, and water does not initiate either platelet aggregation or fibrin precipitation. Therefore, after cross-linking with PC, the biologic materials acquire nonthrombogenic properties.
Third, we used a special heparinization method for powerful antithrombogenicity. Heparin was bound in the luminal surface of the graft in which protamine sulfate was impregnated in advance and bound covalently during the PC cross-linking. Protamine can bind heparin ionically and release it slowly after implantation. During the heparin release, fibrin cannot precipitate on the graft surface, but endothelial cells can migrate and proliferate on it.
27 Therefore, after heparin is gone, endothelial cells can prevent thrombus formation on the graft surface. Even in small-diameter grafts, the graft maintained its patency with the heparinization technology in the aorta-coronary bypass for the entire 1-year observation period.
29,31,37 The antithrombogenicity induced by the slow release of heparin proved to be effective in the pulmonary arterial system, and the graft worked perfectly without anticoagulant therapy. This is a great advantage when the graft is used in infants and young children in whom anticoagulant treatments cannot be used after implantation.
Another concern related to biologic materials for infants and young children is degenerative changes such as calcification. Collagen materials cross-linked with PC have proved to induce less calcification than those cross-linked with glutaraldehyde. In growing rats, the calcification rate of biologic valves cross-linked with PC was less than 1% compared with those cross-linked with glutaraldehyde.
38 Previous articles on PCcross-linked aortic valves implanted in both the systemic and the pulmonary arterial systems did not report any calcification.
38 In the present experiments, we did not use growing animals, but young dogs less than 1 year old, and the graft showed no calcification during the observation period. We can expect a similar result in human beings as well.
Peeling of the neointima has been reported as a complication of right ventricularpulmonary arterial conduits.
9,12 Our hypothesis is that the peeling is caused structurally by intimal hyperplasia associated with poor anchoring of the neointima on the graft surface and biologically poor cell affinity of the neointima to the graft surface. Highly porous Dacron fabric grafts do not show peeling.
39,40 In high cell-affinity grafts such as PCcross-linked biologic materials, neointimae on the graft surfaces are thin and biologically stable and do not become detached from the luminal surfaces. Our grafts had a thin neointima, lined with endothelial cells without intimal hyperplasia, which adhered stably to the graft. We believe that the probability of peeling in our grafts is very low.
In summary, we believe that the success of our graft is due to the combination of three techniques: (1) use of natural tissue as raw material, (2) improved cross-linking, and (3) heparinization for antithrombogenesis. The combination of these techniques is applicable not only in conduits, but also in various other implantable artificial organs in general. We are now continuing our experiments to make results with these grafts comparable with the allograft results obtained with 10-year observation.
References
This article has been cited by other articles:
![]() |
N. Sekarski, H. van Meir, M. E.B. Rijlaarsdam, P. H. Schoof, D. R. Koolbergen, J. Hruda, L. K. von Segesser, E. J. Meijboom, and M. G. Hazekamp Right Ventricular Outflow Tract Reconstruction With the Bovine Jugular Vein Graft: 5 Years' Experience With 133 Patients Ann. Thorac. Surg., August 1, 2007; 84(2): 599 - 605. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Sierra, J. T. Christenson, N. H. Lahlaidi, M. Beghetti, and A. Kalangos Right Ventricular Outflow Tract Reconstruction: What Conduit to Use? Homograft or Contegra? Ann. Thorac. Surg., August 1, 2007; 84(2): 606 - 611. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Chiappini, C. Barrea, and J. Rubay Right Ventricular Outflow Tract Reconstruction With Contegra Monocuspid Transannular Patch in Tetralogy of Fallot Ann. Thorac. Surg., January 1, 2007; 83(1): 185 - 187. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Rastan, T. Walther, I. Daehnert, J. Hambsch, F. W. Mohr, J. Janousek, and M. Kostelka Bovine jugular vein conduit for right ventricular outflow tract reconstruction: evaluation of risk factors for mid-term outcome. Ann. Thorac. Surg., October 1, 2006; 82(4): 1308 - 1315. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. W. Brown, M. Ruzmetov, M. D. Rodefeld, P. Vijay, and R. K. Darragh Valved bovine jugular vein conduits for right ventricular outflow tract reconstruction in children: an attractive alternative to pulmonary homograft. Ann. Thorac. Surg., September 1, 2006; 82(3): 909 - 916. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. L.S. Morales, B. E. Braud, K. S. Gunter, K. E. Carberry, K. A. Arrington, J. S. Heinle, E. D. McKenzie, and C. D. Fraser Jr Encouraging results for the Contegra conduit in the problematic right ventricle-to-pulmonary artery connection. J. Thorac. Cardiovasc. Surg., September 1, 2006; 132(3): 665 - 671. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. O. Shebani, S. McGuirk, M. Baghai, J. Stickley, J. V. De Giovanni, F. A. Bu'Lock, D. J. Barron, and W. J. Brawn Right ventricular outflow tract reconstruction using Contegra(R) valved conduit: natural history and conduit performance under pressure Eur. J. Cardiothorac. Surg., March 1, 2006; 29(3): 397 - 405. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. A. Quintessenza, J. P. Jacobs, V. O. Morell, J. M. Giroud, and R. J. Boucek Initial Experience With a Bicuspid Polytetrafluoroethylene Pulmonary Valve in 41 Children and Adults: A New Option For Right Ventricular Outflow Tract Reconstruction Ann. Thorac. Surg., March 1, 2005; 79(3): 924 - 931. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Gober, P. Berdat, M. Pavlovic, J.-P. Pfammatter, and T. P. Carrel Adverse Mid-Term Outcome Following RVOT Reconstruction Using the Contegra Valved Bovine Jugular Vein Ann. Thorac. Surg., February 1, 2005; 79(2): 625 - 631. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Meyns, L. Van Garsse, D. Boshoff, B. Eyskens, L. Mertens, M. Gewillig, S. Fieuws, E. Verbeken, and W. Daenen The Contegra conduit in the right ventricular outflow tract induces supravalvular stenosis J. Thorac. Cardiovasc. Surg., December 1, 2004; 128(6): 834 - 840. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. F. Corno, S. D. Qanadli, N. Sekarski, S. Artemisia, M. Hurni, P. Tozzi, and L. K. von Segesser Bovine Valved Xenograft in Pulmonary Position: Medium-Term Follow-Up With Excellent Hemodynamics and Freedom From Calcification Ann. Thorac. Surg., October 1, 2004; 78(4): 1382 - 1388. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Purohit, D. Kitchiner, and M. Pozzi Contegra bovine jugular vein right ventricle to pulmonary artery conduit in Ross procedure Ann. Thorac. Surg., May 1, 2004; 77(5): 1707 - 1710. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bottio, G. Thiene, V. Vida, M. D. Barbera, A. Angelini, G. Stellin, and G. Gerosa The bovine jugular vein conduit for right ventricular outflow tract reconstruction: A feasible alternative to homograft conduits? J. Thorac. Cardiovasc. Surg., April 1, 2004; 127(4): 1204 - 1207. [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline, D. Bonnet, T. A. Massih, G. Agnoletti, F. Iserin, F. Jaubert, D. Sidi, and P. Vouhe Use of bovine jugular vein to reconstruct the right ventricular outflow tract: early results J. Thorac. Cardiovasc. Surg., August 1, 2003; 126(2): 490 - 497. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Herijgers, S. Ozaki, E. Verbeken, A. Van Lommel, B. Meuris, E. Lesaffre, W. Daenen, and W. Flameng Valved jugular vein segments for right ventricular outflow tract reconstruction in young sheep J. Thorac. Cardiovasc. Surg., October 1, 2002; 124(4): 798 - 805. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bove, H. Demanet, P. Wauthy, J. P. Goldstein, H. Dessy, P. Viart, A. Deville, and F. E. Deuvaert Early results of valved bovine jugular vein conduit versus bicuspid homograft for right ventricular outflow tract reconstruction Ann. Thorac. Surg., August 1, 2002; 74(2): 536 - 541. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Chang, C.-C. Tsai, H.-C. Liang, and H.-W. Sung Reconstruction of the right ventricular outflow tract with a bovine jugular vein graft fixed with a naturally occurring crosslinking agent (genipin) in a canine model J. Thorac. Cardiovasc. Surg., December 1, 2001; 122(6): 1208 - 1218. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. F. Corno, M. Hurni, H. Griffin, X. Jeanrenaud, and L. K. von Segesser Glutaraldehyde-fixed bovine jugular vein as a substitute for the pulmonary valve in the Ross operation J. Thorac. Cardiovasc. Surg., September 1, 2001; 122(3): 493 - 494. [Full Text] [PDF] |
||||
![]() |
Y. Goto Should the anticalcification effect of epoxy compound in vascular bioprostheses be judged after implantation in mongrel dogs? J. Thorac. Cardiovasc. Surg., April 1, 1998; 115(4): 960 - 961. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||