J Thorac Cardiovasc Surg 1997;114:568-577
© 1997 Mosby, Inc.
SURGERY FOR CONGENITAL HEART DISEASE |
This work was funded by the British Heart Foundation. F.J.K. is funded by the Wellcome Trust.
Received for publication Sept. 23, 1996 revisions requested Nov. 19, 1997; revisions received April 30, 1997 accepted for publication May 2, 1997. Address for reprints: Fenella J. Kirkham, FRCP, Neurosciences Unit, Institute of Child Health (UCL), Mecklenburgh Square, London WC1N 2AP, United Kingdom.
Abstract
Objectives: Neurologic impairment, at least partly ischemic in origin, has been reported in up to 25% of infants undergoing cardiopulmonary bypass, with or without circulatory arrest. Controversy continues about the effect of pump flow, pulsatile or nonpulsatile, on the brain and in particular on cerebral blood flow. This study examines the relationship between pump flow rate and cerebral hemodynamics during pulsatile and nonpulsatile cardiopulmonary bypass. Method: Near-infrared spectroscopy was used to determine cerebral blood flow and cerebral blood volume (measured as concentration change) in a randomized crossover study. Pulsatile and nonpulsatile flow were used for six 5-minute intervals at each of three different pump flow rates (0.6, 1.2, and 2.4 L · m2 · min1) in 40 patients, median age 2 months (range 2 weeks to 20 years 5 months). The relations between pulsatile flow, pump flow rate, cerebral blood flow, hemoglobin concentration change (cerebral blood volume), mean arterial pressure, arterial carbon dioxide tension, and hematocrit value were prospectively examined by means of multivariate analysis. Results: Cerebral blood flow decreased 36% per L · m2 · min1 decrease in pump flow rate and was associated with changes in mean arterial pressure but did not differ according to pulsatility. Change in hemoglobin concentration was unrelated to changes in pulsatility of pump flow. Conclusion: Cerebral blood flow is related to pump flow rate. Pulsatile flow delivered with a Stöckert pump does not increase cerebral blood flow or alter hemoglobin concentration during cardiopulmonary bypass in children.
In complex congenital heart disease, pump flow is often reduced to improve surgical exposure, but the safe lower limit is uncertain. Conventional cardiopulmonary bypass (CPB), delivered by a roller or centrifugal pump, is nonpulsatile, although experimental evidence for adverse cerebral effects has been detected.
1 Pulsatile pumps, which apparently improve systemic, myocardial, and pituitary perfusion,
2 are available, but whether pulsatile flow holds any advantages for the brain remains a subject of controversy.
3-5 No previous work has been done in children, partly because appropriate equipment has only recently become available.
If the critical opening pressure of the capillary bed is lower with pulsatile flow, then at low pump flow rates, the capillary bed or parts of it might be "open" during pulsatile flow but effectively "closed" during nonpulsatile flow. Noninvasive near-infrared spectroscopy (NIRS) offers a unique way of investigating this because an overall increase in capillary blood volume, in association with a lower critical opening pressure, would be reflected by an increase in total hemoglobin concentration (HbVol). In addition, cerebral blood flow (CBF) may be measured repeatedly by means of a modification of the Fick principle with indocyanine green dye as the tracer.
6 Using these methods, this study tests the hypothesis that pulsatility affects CBF and cerebral blood volume (CBV) at different pump flow rates.
Patients and methods
The study was approved by the Research Ethics Committee of the Great Ormond Street Hospital for Children, and written informed consent was obtained from the parents of the children studied.
Patients.
Forty patients, 26 cyanotic, median age 2 months (range 2 weeks to 20 years 5 months), undergoing surgical correction of congenital heart defects, were studied. Diagnostic and procedural data are shown in
Table I.
|
CPB management.
CPB was conducted by means of hypothermic hemodilution techniques with the use of a pulsatile roller pump (Stöckert, Munich, Germany) and a hollow-fiber membrane oxygenator (Minimax, Medtronic U.K. Ltd., Hertz, United Kingdom, or Safemicro, Polystan, Denmark), placed distal to the roller pump. The target hematocrit value was in the range of 22% to 25%. Nasopharyngeal temperature varied according to clinical need, and the pump flow rate was changed according to protocol. Details of the priming protocols are published elsewhere.
7 The alpha-stat method of arterial blood gas management was used. Measurements were made during steady-state hypothermic CPB (18° C to 25.0° C). Three patients (patients 3, 29, and 40) reached a stable hypothermic temperature of 25° C, whereas another three patients (patients 9, 10, and 15) reached a stable temperature of 18° C; the rest of the patients were cooled to 22° C. Pulsatile flow settings used were a heart rate of 100 beats/min. The baseline flow variable was from 0% to 100%; the setting used was 20% during the pulsatile phase, with a run time of 50%.
Experimental protocol.
The principle of NIRS is that biologic tissues are relatively transparent to light in the near-infrared spectrum (700 to 1000 nm).
8 CBF can be estimated by means of a modification of the Fick principle that uses indocyanine green dye as an intravascular and nondiffusible tracer.
6 The sum of the changes in deoxygenated hemoglobin and oxygenated hemoglobin concentration gives the change in HbVol from an arbitrary time. If the hematocrit value is constant during this period, then the change in HbVol will be proportional to the change in CBV in the tissue sampled.
A commercial spectrophotometer (NIRO 500, Hamamatsu, Hamamatsu City, Japan) was used. Light at four wavelengths (776, 819, 843, and 913 nm) was conveyed to the child's head via a flexible fiberoptic bundle, which terminated in an optode that was applied to the frontoparietal region of the child's head after induction of anesthesia. Transmitted light emerging was collected via another optode and fiberoptic bundle leading to the photon detector in the spectrophotometer. The distance between the position of the two optodes was measured with mechanical calipers and was always greater than 3 cm. To prevent interference from background light, we wrapped the head in a light-tight cloth. Measurements were made continuously and collected into 0.5-second time bins. They were displayed instantaneously and stored on computer disk for subsequent analysis. The differential path-length factor, reflecting photon scattering, was 4.39 in children younger than 2 and 5.93 in those older than 2.
9 The arterial concentration of tracer was optically quantified by using a fiberoptic probe (Pulsion Medizintechnik, Munich, Germany) placed in the arterial line of the CPB circuit at its junction with the aorta. Indocyanine green dye (0.1 mg · kg1) was injected upstream to this probe. Left-to-right shunts did not affect the study because CBF measurements were taken only when the aorta was crossclamped.
Patients were assigned a program of low, medium, and high flow rates in a random order to be applied during stable hypothermic bypass. No additional time was taken for the measurements to be performed. During each assigned pump flow rate, pulsatile flow was delivered for 5 minutes, followed by 5 minutes of nonpulsatile flow (or vice versa) in a prospective randomized way. The HbVol for each pump flow rate and pulsatility setting was calculated by taking an average of all the HbVol measurements for 3 minutes after the pump flow change, before making a CBF measurement. CBF was measured before the pulsatility change, after 4 minutes of the assigned pump flow rate. Each CBF measurement takes approximately 10 seconds. During each CBF measurement, nasopharyngeal temperature, pump flow, mean arterial pressure, arterial carbon dioxide tension (PaCO 2), and hematocrit values were measured. A schematic representation of the protocol is shown in Fig. 1. Thus one patient might have half, quarter, and full flow with nonpulsatile followed by pulsatile flow and the next patient might have full, quarter, and half flow with pulsatile followed by nonpulsatile flow.
|
Statistical analysis.
Multiple linear regression models were used to determine the extent to which HbVol and CBF were related to pulsatility and pump flow rate. CBF was logged before modeling. A series of dummy variables were used to model patient-specific variability in the outcomes, thus accounting for differences resulting from age, cyanosis, and other patient-specific factors. Mean arterial pressure, PaCO 2, arterial oxygen tension (PaO 2), hematocrit value, and temperature were then entered into the model to adjust for the potentially confounding effect of within-individual associations between these variables and the outcomes. The relationships between CBF and HbVol with pulsatility and pump flow are represented after adjusting for all of the potential confounders. Computations were made with the use of the SPSS statistical package (SPSS, Inc., Chicago, Ill.).
Results
Clinical outcome.
Patients 1, 8, 14, 20, and 29 died, patient 1 of pneumonia, patient 8 of low cardiac output, patient 14 of systemic ventricular failure, patient 20 of cardiac failure, and patient 29 of pulmonary hemorrhage. Patients 2 and 6 required permanent pacemakers for complete heart block. Patient 13 had a left pneumothorax, patient 21 had partial dehiscence of a ventricular septal defect patch dehiscence (which necessitated another operation), patient 25 had mediastinitis, and patient 35 had a chylothorax on the left side and required a permanent pacemaker for sinus bradycardia. The other patients made uneventful recoveries. No neurologic problems developed in the postoperative period.
Doppler measurements.
In the one patient studied by Doppler ultrasonography, while nonpulsatile pump flow did not produce a pulsatile waveform, pulsatile pump flow did, but this was of lower amplitude than that produced by a beating heart (Fig. 2).
|
|
|
|
|
|
|
Discussion
This study yielded two major findings. First, CBF fell with decreasing pump flow rate, PaCO 2, and mean arterial pressure. Mean arterial pressure covaried with pump flow rate, suggesting that autoregulation was abolished. Second, pulsatile flow did not influence CBF or HbVol at any pump flow. The methodologic issues will be addressed before discussion of the findings.
The CBF technique uses NIRS and a modification of the Fick principle.
6 The initial description involved the use of oxygen as a tracer and has been validated in neonates.
10 For applications, such as CPB, in which it may be difficult to change oxygen tension or in which the signal-to-noise ratio for measurement of the input function must be improved, a more recent modification for which indocyanine green is used as the tracer
11 is preferred. Recent data from our laboratory indicate that CBF measurements obtained with this technique are highly correlated with those obtained with the use of microspheres in a piglet model. Although no validation studies have been conducted yet in human beings, the technique appears to be able to measure change in CBF.
12 NIRS is a sensitive technique for observation of changes in HbVol and can detect changes of 0.5 mmol or less. It is uniquely suited for investigating the relatively small changes in CBV expected with capillary recruitment. CBV changes can be inferred from the change in HbVol, assuming the hematocrit value is constant during periods of measurement.
13 The mean change in hematocrit value between pulsatile and nonpulsatile pump flow in this study was 0.26%, standard deviation 1.74%, which satisfies this constraint.
Anesthesia was chosen so that there would be no confounding effect on CBF. The variations in premedication are unlikely to have affected CBF measurements. Ketamine and halothane were used only on induction, at least 40 minutes before measurements of CBF.
Pump flow rate.
Low pump flow can reduce noncoronary collateral circulation and thus improve myocardial preservation.
14 At times, adequate surgical exposure may require very low pump flow rates to reduce venous return to the open heart, particularly in children with congenital heart disease in whom the repairs may be complex. The important question is whether a safe minimum pump flow exists, below which brain damage is likely to occur.
There is evidence for an ischemic threshold from animal experiments conducted during CPB, at least at very low pump flow of 5 ml · kg1 · min1.
15 Although in clinical practice pump flow rates are not as low as these, CBF might fall below that required for the metabolic demand of the tissue. The relation between CBF and pump flow rate remains controversial, despite a number of investigations in human beings, with various forms of acid-base management.
16-19 Those who have claimed that changing pump flow has no effect have studied adults and have reduced pump flow to no lower than half of the calculated full requirement. Pediatric studies are clearly justified, particularly because pump flows tend to be lower. Our data confirm the findings of Kern and associates
19 in children, that a significant relation exists between CBF and pump flow rate when alpha-stat pH management is used, with covariance between pump flow rate and mean arterial pressure. Any beneficial effect at low pump flow of increasing blood pressure pharmacologically, changing pH strategy, or otherwise modifying CPB could be addressed by means of the NIRS technique.
Pulsatile flow.
In the animal literature, the controversy focuses on whether pulsatile flow yields any benefits in terms of cerebral hemodynamics.
3,4,20 It has been suggested that the critical opening pressure of the capillary bed is lower with pulsatile than with nonpulsatile flow,
21 which might be important in maintaining adequate tissue perfusion at the low pump flows used during CPB. This might explain the higher CBF during pulsatile pump flow at cerebral perfusion pressures below 50 mm Hg in dogs.
22 This higher CBF has not, however, been reproduced in rabbits.
4 Species differences do exist, which means that these questions should be addressed in human beings.
Our data show no difference in CBF between pulsatile and nonpulsatile flow, which is consistent with the findings of the only previous human studies and with the studies of Hindman and colleagues,
4 who argued that their lapine model might reflect the human situation more closely than the canine studies. The change in HbVol was similarly not significantly different for pulsatile and nonpulsatile pump flow at the three measured flow rates. There may be a difference between children and adults. Young patients have more compliant vessels and a lower vascular resistance; thus pulsatile flow, which decreases mean arterial pressure,
23 may hold no advantage.
24 However, preliminary studies from our group suggest that pulsatile flow does not increase CBF in adults either.
25
It has been argued that a CPB machine switched to "pulsatile flow" may not produce a truly pulsatile waveform and will almost certainly behave differently from the living heart. In this study we used the Stöckert pump, because it is the one most frequently used in adult and pediatric CPB and is considered to be capable of delivering pulsatile flow, although it generates only 12.4% of pulsatile power of a living heart.
26 In addition, the membrane oxygenators now used have compliant membrane components and tend to absorb much of the energy generated by the pulsatile pump. The aim of this study was to observe the effect of a pump marketed to deliver pulsatile flow during CBF and CBV and not to compare the waveforms obtained by different flow levels between patients. The precise waveform may be of little importance,
4 but Doppler recording confirmed a pulsatile waveform in the middle cerebral artery. It is possible that the arbitrary time of 5 minutes before the pulsatility of pump flow was changed, chosen to ensure that each of the six randomized conditions could be achieved during the relatively short CPB period, did not allow for any significant effects to become apparent. Further studies using alternative pumps or longer time periods, but using the same methods, could be planned.
In summary, we have investigated the relation of CBF to several important potential predictors. Mean arterial pressure covaries with changes in pump flow rate and is associated with CBF. Pump flow rate is significantly associated with CBF after taking into account any changes that could be attributed to differences in mean arterial pressure. CBF is also associated with PaCO 2 but not pulsatility delivered by the Stöckert pump. HbVol is related to mean arterial pressure and pump flow rate but not to pulsatility. These data do not support a change to use of pulsatile pump flow with a Stöckert pump in children.
Acknowledgments
We thank M. de Leval, M. Davis, R. Maynard, G. Walsh, A. Dale, nurses, and operating department assistants at the Great Ormond Street Hospital for children, for their assistance and cooperation.
Footnotes
From the Departments of Neurosciences,a Anaesthetics,b Epidemiology and Biostatistics,c and Cardiothoracic Surgery,d Institute of Child Health (UCL)/Great Ormond Street Hospital for Children, Great Ormond Street, and Department of Pediatrics and Neonatal Medicine,e Royal Postgraduate Medical School, Du Cane Rd., London, United Kingdom. ![]()
This work was performed at Great Ormond Street Hospital for Children, London, United Kingdom. ![]()
References
This article has been cited by other articles:
![]() |
J. C. Hirsch, J. R. Charpie, R. G. Ohye, and J. G. Gurney Near-infrared spectroscopy: What we know and what we need to know--A systematic review of the congenital heart disease literature J. Thorac. Cardiovasc. Surg., January 1, 2009; 137(1): 154 - 159. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. D. Doblar Intraoperative Transcranial Ultrasonic Monitoring for Cardiac and Vascular Surgery Seminars in Cardiothoracic and Vascular Anesthesia, June 1, 2004; 8(2): 127 - 145. [Abstract] [PDF] |
||||
![]() |
E. R. Stephenson Jr and J. L. Myers Pediatric cardiopulmonary bypass Ann. Thorac. Surg., December 1, 2001; 72(6): 2176 - 2177. [Full Text] [PDF] |
||||
![]() |
H. H. Hovels-Gurich, M.-C. Seghaye, M. Sigler, F. Kotlarek, A. Bartl, J. Neuser, R. Minkenberg, B. J. Messmer, and G. von Bernuth Neurodevelopmental outcome related to cerebral risk factors in children after neonatal arterial switch operation Ann. Thorac. Surg., March 1, 2001; 71(3): 881 - 888. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Undar, T. Masai, S.-Q. Yang, J. Goddard-Finegold, O.H. Frazier, and C. D. Fraser Jr Effects of perfusion mode on regional and global organ blood flow in a neonatal piglet model Ann. Thorac. Surg., October 1, 1999; 68(4): 1336 - 1342. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J Elliott Recent advances in paediatric cardiopulmonary bypass Perfusion, July 1, 1999; 14(4): 237 - 246. [PDF] |
||||
![]() |
A. Undar and C. D. Fraser Jr. The relation between pump flow rate and pulsatility on cerebral hemodynamics during pediatric cardiopulmonary bypass J. Thorac. Cardiovasc. Surg., September 1, 1998; 116(3): 530 - 531. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||