J Thorac Cardiovasc Surg 1998;115:582-586
© 1998 Mosby, Inc.
SURGERY FOR ACQUIRED HEART DISEASE |
Supported by an unrestricted grant from the Shiley Heart Valve Research Center.
Received for publication June 5, 1997; revisions requested July 15, 1997; revisions received Nov. 24, 1997; accepted for publication Nov. 24, 1997. Address for reprints: Kenneth D. Hopper, MD, Department of Radiology Pennsylvania State University, 500 University Dr., P.O. Box 850, Hershey, PA 17033.
Abstract
Objective: Modified cineradiographic systems have been used clinically to detect partially broken outlet struts in normally functioning Björk-Shiley convexo-concave heart valves. Almost all such valves were explanted, presuming that full failure would likely follow. Inasmuch as the clinical setting only rarely permits examination of normally rated valves, the accuracy of radiographic detection cannot be clinically defined. This study uses the clinical radiographic technique in sheep implanted with known-status convexo-concave valves, comparing its accuracy and that of a newly developed, geometric image magnification radiography system.
Methods: Twenty-one sheep with mitral convexo-concave valves were studied on both systems. Five were used for extensive training. When operators were expert with both systems, images of four intact valves and 12 valves with outlet strut single leg separations, along with a seventeenth single leg separation valve used for calibration, were integrated into 112 image sets organized into a balanced incomplete block design for evaluation by eight trained, blinded reviewers.
Results: Cineradiography sensitivity was 24% versus 31% for direct image magnification. The odds ratio for detection of single leg separation by direct image magnification versus cineradiography was 2.0 (95% confidence interval, 0.76 to 5.9; p = 0.13). Cineradiography specificity was 93% versus 90% for direct image magnification. Sensitivity and specificity varied markedly by reviewer, with sensitivity ranging from 8% to 55% and specificity from 51% to 100% for the combined technologies.
Conclusions: The data support the need for more intensive training for convexo-concave valve imaging and further investigation of unconventional radiographic technologies. Clinical cineradiography of convexo-concave valves may detect as little as 25% of valves having a single leg separation, underestimating the prevalence of single leg separations and thereby implying more rapid progression to full fracture than is actually the case.
The Björk-Shiley convexo-concave (C/C) valve was introduced in 1979 for aortic and mitral valve replacement.
1,2 The valve has a pyrolytic carbon disc occluder that rotates about an inlet strut that is an integral part of the flange. Translation and rotation of the disc are restrained by an outlet strut that has two legs welded into the metal flange (Fig. 1). Between 1979 and 1986, approximately 86,000 C/C valves were believed to have been implanted worldwide. Shiley, Inc. (Irvine, Calif.) estimates that 41,000 C/C valves are currently implanted in living patients.
|
|
|
Twenty-one 29 mm C/C valves (18 with a 60-degree opening angle and three with a 70-degree opening angle) were implanted via a left thoracotomy as mitral replacements in anesthetized young adult Suffolk sheep. All animal procedures and care followed guidelines issued by the National Society for Medical Research and the National Institutes of Health.
4,5 Five of the implanted valves were for screening practice: two were intact and three had intentional, mechanically induced SLSs. The remaining 16 study sheep received four intact valves, seven valves with intentional SLS, and five clinical SLS valves. The mechanically induced SLSs were adjusted to place the break-surfaces in contact with less than 300 µm of offset between the two ends. The clinical SLS valves had been explanted from patients for reasons other than the result of radiographic screening. The status of all valves was verified by electron microscopy to be either intact or to have an SLS with limited offset (shift of strut alignment at the point of separation) and zero gap (strut sections remained in contact at the point of offset).
Extensive training of and practice by the two imaging teams were carried out on a modified Siemens Hicor-Coroskop system and the feinfocus system, respectively. Both systems were operated at peak energy (kVp: kilovoltage peak) so as to better detect C/C valve metallic defects.
6 Experience was gained with both systems using a mechanical heart phantom,
7 which approximated the normal attenuation of the human thorax and cardiac mitral valve motion.
8 The actual study was initiated only after the capabilities of both imaging teams were approved by the Shiley Heart Valve Research Center. The two systems and the training are fully described in Appendix A.
Separations of the outlet strut typically occur in or near the welds to the metal flange. In imaging these welds, the Shiley Heart Valve Research Center has developed five separate projections (views). The principal view profiles both welds away from the metal flange and has been termed the "tunnel view" (
Figs. 2 and
3). Internal and external oblique views of each outlet strut demonstrate each weld from two additional projections. A total of five views of each valve are possible. Because the tunnel view images both outlet strut welds, three views of each outlet strut were actually obtained.
The 16 study sheep, plus one sheep with a zero-gap, intentional SLS (used to generate a "calibration" set for the readers), were imaged in a randomized order on both systems. Each team (blinded to the valve status), the number of imaging sessions scheduled, and the findings of the other team optimized the visualization of both outlet strut welds, obtaining five views per valve. A navigational computer program assisted in initial C-arm positioning, saving scanning time.
9 Radiation exposure was directly measured on both systems by an air exposure product meter placed within the x-ray beam (Diamentor, model M2, PTW, Freiburg, Germany).
10
The acquired five views of each valve were prepared for reading according to a balanced incomplete block design
11 requiring 28 blocks of 16 image sets each and eight reviewers (Fig. 4). This design facilitated the large number of reviews needed to obtain adequate statistical power and a realistic SLS prevalence with only a limited number of study sheep. Each "image set" included a single tunnel view and two oblique views. Each block included image sets from the single calibration SLS valve, from one or two SLS valves, and from 13 to 14 intact valve image sets. The design alternatively paired each of the eight reviewers such that each block was read twice (
Fig. 4). Each reviewer read seven blocks of image sets per modality (112 sets) for a total of 224 image sets. Prevalence of SLSs among the 1792 image sets (896 per modality) ranged from 15% to 19%, as seen by any one reviewer. All aspects of the study design were blinded to the eight readers. No member of either imaging team served as a reader or had contact with any reader before or during the image interpretation sessions.
|
|
Considering all reviews except those of the calibration valve, sensitivity (fraction of correctly identified SLS valves) of the feinfocus technology (31%, 30 of 96) appeared higher than that of the modified Siemens system (24%, 23 of 96) (Table II). However, this difference was not statistically significant at conventional levels; the adjusted odds ratio of detecting an SLS by feinfocus compared with Siemens was 2.0 (95% confidence interval [CI], 0.76 to 5.9; p = 0.13). Specificity (fraction of correctly identified intact valves) of the Siemens system (93%, 695 of 744) was higher than that of the feinfocus system (90%, 668 of 744) (odds ratio = 1.96, 95% CI, 1.2 to 3.2; p = 0.003). The area under the receiver operating characteristic curve for feinfocus was slightly greater than for the Siemens system when the data were analyzed using all possible cut points along the image assessment grade scale (Fig. 5). Previously reported phantom and clinical studies
8,9,15 plus the results presented here have used a cut point of 4 to define an SLS.
|
|
Rates of sensitivity and specificity varied markedly by reviewer (see
Table II). For the feinfocus system, sensitivity ranged from 8% to 55% and specificity from 51% to 100%. For the Siemens system, the ranges were equally wide: 8% to 45% for sensitivity and 55% to 100% for specificity. The same variation appeared in the calibration SLS valve: sensitivities ranged from 0% to 86% for the feinfocus system and from 0% to 43% for the Siemens system. Two reviewers appeared to fare worse than the others on both imaging systems. Reviewer 3 exhibited much lower specificity (51% and 55%) than his peers; reviewer 8 exhibited much lower sensitivity (8%) with the feinfocus system, as well as a low sensitivity with the Siemens system (8%). A third reviewer (reviewer 7) also performed markedly worse than others in detecting SLS valves on the Siemens system. Eliminating reviewers 3 and 8 does not change the overall results because their aberrant performances offset each other.
Valve-specific results are displayed in Table III. On one extreme, two SLS valves, both clinical SLSs, were easily identified by means of both technologies (valves 11 and 12). On the other extreme, three were never seen on either technology (valves 3, 5, and 10). Between these two extremes, the differences in the two technologies appear in the imaging of specific valves, for example, valves 1 and 4. Taking the five clinical valves as a whole over both technologies, SLSs were detected in 40 of 80 cases (50%); by contrast, the seven intentional SLSs (excluding the calibration) were detected significantly less often (13 of 112 cases, 12%; p < 0.001).
|
Discussion
The first report of an outlet strut fracture appeared shortly after the C/C valve's introduction.
12 As of June 30, 1997, 603 outlet strut fractures have been reported to the manufacturer, approximately two thirds of which have been fatal. Failure to detect and report outlet strut fractures, especially among older patients with sudden death, is now compensated in Shiley's estimates of fracture incidence, using two known cohorts
13,14 for calibration.
15 The risk of outlet strut fracture varies from 0.01% to 3.29% per year depending on valve size, opening angle, implant position, weld date, and welder group. Weld dates in 1981 and early 1982, 29 mm flange size, 70-degree opening angle, and mitral implants in younger men indicated a greater likelihood of breakage.
15
The Shiley Heart Valve Research Center instituted a wide range of research efforts to find a noninvasive screening method to detect C/C valves in which one of the two outlet strut legs may have separated from the flangea single leg separation (SLS). Animal and in vitro studies have shown that the C/C valve apparently functions normally until both outlet strut legs separate from the flange and outlet strut fracture occurs. An SLS appears to be a precursor to complete outlet strut fracture and disc embolization.
9
Radiographic studies have correctly identified one aortic and 25 mitral SLS valves, incurring four false positives and one verified false negative in the process of studying 964 C/C valves with an annual risk of fracture of 0.1% or more in 842 patients.
15 Because almost 90% of the study valves received negative ratings and remain functioning in situ or buried with decedents, the accuracy of clinical radiographic SLS detection simply cannot be known. The radiographically detectable SLS prevalence was 3% in these higher epidemiologic-risk valves, implying a median SLS duration in the order of 2 years before fracture.
9,15 Inasmuch as the sensitivity is unknown, the actual prevalence may be much higher, implying longer persistence of the SLS condition and perhaps lesser need for valve replacement, particularly in older patients.
16
Over the past several years, the Shiley Heart Valve Research Center has aggressively pursued all avenues of research that held any promise of providing a means of detecting C/C valve SLSs in situ. The best of these, cineradiographic screening, has been studied in higher-risk C/C valve patients in selected sites in the United States and Europe. However, until the present investigation, no in vivo assessment of radiographic detection sensitivity and specificity has been undertaken and no systematic evaluation of unconventional radiographic technologies has been done. The sheep was used in this study because its thorax mimics the scatter and attenuation encountered in human imaging. Our expert imagers and reviewers were able to achieve only overall 24% sensitivity, 12% in mechanically induced SLSs and 50% in clinical SLS valves, suggesting that this may be the range of sensitivity under similar controlled circumstances in human subjects. The implication is that two or three of four radiographed SLS valves might be undetected in patients going about their normal activities. The need for valve replacement based on a radiographically identified SLS may not be as great as presumed, because a higher true SLS prevalence means that progression to complete fracture either is less likely or takes a longer time than the radiographic prevalence suggests.
16 In particular, older patients with substantial comorbidities might consider foregoing SLS valve replacement in view of the reported overall 15% operative mortality.
15 If the SLS condition can persist for years, this magnitude of operative risk might be unacceptable, save for younger patients with the highest risk-category valves.
Performance in this study varied by both reviewer and valve, suggesting that the abilities of Siemens and feinfocus technologies to detect SLSs in vivo depend heavily on the skills and experience of individual reviewers, as well as on the idiosyncrasies of individual SLSs. Individual reviewer sensitivities varied markedly, and overall reviewer performance varied widely by SLS valve. These results suggest that the overall performance not only is influenced by reviewer skills, but also is valve-specific, with some SLS valves identified correctly far more often than others.
Reviewer training and testing should improve the performance with these technologies, particularly with regard to specificity of the feinfocus system. For example, reviewer 3 overread nearly 50% of all image sets of normal valves as having an SLS. In contrast, several reviewers were nearly 100% correct when classifying image sets of normal valves, including reviewer 5, who was 100% correct for both technologies. One reviewer (reviewer 8) rarely identified an SLS with either technology.
Over all SLS valves, our results suggest a potential advantage for the feinfocus system, as summarized by a higher odds of detection with a small, but significant, loss of specificity, compared with the Siemens system. Improving sensitivity by seven percentage points means that out of 100 SLS valves, seven not detected by the Siemens system would potentially be identified by the feinfocus technology. The feinfocus system has been used on 12 patients with C/C valves at one center in Europe, resulting in one false positive and no true positive detections (R. W. Günther, personal communication, 1996).
In dealing with patients having C/C valves, especially in the higher risk groups, there are three potential courses of action: elective explantation, screening, and expectant management. de Mol and associates
17 have implied that prophylactic C/C valve replacement should be given more frequent consideration. Among 24 elective explants from 21 patients without valve dysfunction, they discovered seven unanticipated SLSs (29%). If the operative mortality was 0%, as it was among these 21 patients (mean age 51 years), explantation might be the most advisable course of action for high-risk valves.
The operative mortality from elective valve replacement, however, has ranged from 3% to 5% for mitral valve surgery in high-volume centers.
18-23 As with coronary bypass surgery, this mortality might increase with less experienced surgeons
24 and in older patients with serious cardiovascular or other medical diseases.
25-29 The operative mortality in 27 patients with radiographically suspected SLSs was 15%.
9,15
Elective screening of all patients with a C/C valve by a modified Siemens cineradiographic system is difficult to recommend on the basis of this study. With an overall prevalence of 11% (excluding the calibration valve data), the positive predictive value is 32% (23 of 72). Conversely, 68% of patients with positive test results might have normal valves. If the risk of operative mortality from explantation is about 5%, then routine explantation for positive tests would result in approximately three unnecessary deaths (operative mortalities in normal valves) per 100 patients with positive screens (68 x 0.05). The negative predictive value is 90% (695 of 768) in our sample, which means that 10% of those with negative screening would potentially have SLS valves.
The hazard of eventual fracture and sudden death among patients with a C/C valve SLS remains uncertain. Although one should extrapolate from our sheep data that the real incidence of clinical SLSs is higher than has been recognized radiographically, indicating slower progression to complete fracture, the SLSs in this study were intentionally among the more difficult to image with zero gaps and limited offsets.
The data from this study do not support universal cineradiographic screening of patients with C/C valves for SLS detection. They do indicate a need for more intensive training and selection of the physicians doing the scanning and image interpretation. With the SLS detection variability discovered in this study, perhaps pairing of the best readers would improve the sensitivity of radiographic screening. Last, the results of this study support further investigation of unconventional x-ray technologies such as direct image magnification in evaluating patients with C/C valves.
Appendix A: Expanded methods
Conventional angiographic systems are manufactured to operate best at 70 kVp (kilovoltage peak) to optimize the imaging of iodinated contrast material. However, the use of radiography for imaging metallic defects requires a higher energy (kilovoltage peak) so that the metal itself is partially penetrated. This higher energy also more completely penetrates background bone, making it less obscuring. These beneficial effects outweigh the deleterious effects of increased x-ray scatter. Abolfathi and coworkers
6 have verified this effect in in vitro imaging of C/C valves.
As a result, the Siemens Hicor-Coroskop system was modified to allow a peak voltage of 83 kVp and a peak power of 125 mA, using an exposure algorithm that favored peak voltage over peak power. The addition of a 135 mm focal length lens allowed larger valve images to be obtained on 35 mm cine film. Imaging on this system was performed at 30 frames per second with 4 to 6 seconds per view. An x-ray tube focal spot with a nominal size of 0.88 mm and a pulse width of 6.4 msec were used. Before study initiation and before each imaging session, this system passed a Siemens image quality assurance program. A minimum of 3.7 line-pairs per millimeter visualization on the film with the film density and grading indices within ±5% of specification was required before imaging began.
The feinfocus system uses nonpulsed electron gun technology, kilovoltage peak up to 180, direct magnification up to sixfold, and a high-resolution image (1024 x 1024 matrix) to achieve a spatial resolution of up to 18 line-pairs per millimeter. Electron gun technology eliminates the increasing peripheral blurring associated with direct magnification and allows a variable focal spot of from 40 µm to 300 µm. Imaging with this system was performed at 7 frames/sec with 3-second acquisitions per view. Digital data were reviewed on high-resolution monitors that maintained the original image quality. This system underwent a complete quality assurance check by the manufacturer just before study initiation and passed the manufacturer's line-pair test before and after each use.
A 12-month period of practice and testing ensured optimal imaging techniques by two separate imaging teams, one for each system. Each team was composed of an experienced physician plus one to three technologists. Initial training consisted of practice cine film review of 60 C/C valves imaged in a mechanical phantom.
7 For the feinfocus team, instruction and practice on the system itself included on-site assistance and training by the manufacturer. The team physicians passed a screening test, consisting of a series of 40 clinical studies with an undisclosed 20% SLS prevalence, as was required by the Shiley Heart Valve Research Center for its clinical screening sites.
The imaging teams practiced with a working heart valve phantom that simulated the attenuation and scatter of the human thorax and approximated normal cardiac anulus and valve motion.
8 This training sample included six intentional SLSs (zero gap) and three intact C/C valves imaged multiple times to create a randomized series of 120 sets of images with a 10% SLS prevalence.
During this same period, five sheep with C/C valves (three with intentional SLSs and two with intact SLSs) were imaged repeatedly on both systems. After this 12-month training period, the capabilities of both imaging teams were approved by senior staff of the Shiley Heart Valve Research Center.
Seven of the eight reviewers used for the study were selected by the sponsor (Shiley Heart Valve Research Center) from the three C/C valve clinical screening sites. The remaining reviewer was recruited from our institution from among four trained expert readers who had not participated on either imaging team. The readers included four cardiologists, three radiologists, and one cardiovascular surgeon. They assembled at a remote location, where they each independently reviewed their entire complement of 224 image sets over 3 days.
The Siemens image sets were read on Vanguard XR35 projectors (Vanguard Corporation, Melville, N.Y.). The feinfocus images were displayed on the Sun SPARC 20 workstation (Sun Microsystems, Inc., Mountain View, Calif.) with a high-resolution monitor equivalent to that on the actual feinfocus system itself.
Interpretation was by predetermined criteria (
Table I) as described in the statistical methods section. This scoring system was developed in prior phantom and sheep studies by Shiley and used in more than 890 clinical scans performed to date.
15 Grading with the feinfocus system differed from grading with the Siemens system only to compensate for the differing number of frames available for viewing. Reviewers practiced using these scoring systems with test image sets representing all five grades and could refer to these examples at any time. Reviewers were monitored and were prevented from discussing any material or cases among themselves.
The image assessment grade data from each reviewer were classified as either correct (1) or incorrect (0) for each image set read on each of the two technologies (Siemens, feinfocus) for each valve within the image set and block. Because of the repeated measures (reviewer, block, image set) for each valve, we used a stratified Mantel-Haenszel method,
30 adjusting for reviewer, block, and image set, to estimate the average odds ratio, as well as its 95% confidence interval, for the likelihood of the feinfocus system detecting an SLS relative to the Siemens technology. Estimates were produced for clinical SLS valves, intentional SLS valves, and for all SLS valves. Similarly, this same method compared specificity of the two technologies for the normal valves. These odds ratios were adjusted to control for reviewer, block, and image set. A paired t test was used to compare the radiation levels between the two imaging technologies. All calculations were performed with the use of the SAS statistical package (SAS Institute, Inc., Cary, N.C.), StatXact statistical package (Cytel Corporation, Boston, Mass.), and S-Plus (MathSoft, Seattle, Wash).
Footnotes
References
This article has been cited by other articles:
![]() |