J Thorac Cardiovasc Surg 1998;115:913-924
© 1998 Mosby, Inc.
CARDIOPULMONARY SUPPORT AND PHYSIOLOGY |
This work was supported in part by the Medical Research Council of Canada (grant 96090P-3408-BME-CDAA-26321).
Received for publication Feb. 27, 1997. Revisions requested May 7, 1997; revisions received July 9, 1997. Accepted for publication Sept. 30, 1997. Address for reprints: Ganghong Tian, MD, PhD, Institute for Biodiagnostics, National Research Council Canada, 435 Ellice Ave., Winnipeg, Manitoba, Canada R3B 1Y6.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The concept of SARC might have originated from the technique of intermittent occlusion of the coronary sinus (IOCS). The latter has been used for short periods as a strategy to salvage ischemic myocardium in the beating heart and was shown to be able to reduce infarct size and improve regional myocardial function. However, it has not been directly demonstrated that IOCS indeed improves myocardial perfusion. In addition, because the conditions of the heart (beating vs arrested) and the techniques used (intermittent occlusion vs continuous perfusion of the coronary sinus) are different with these two methods, the beneficial effects observed with IOCS may not necessarily occur in SARC.
Magnetic resonance (MR) imaging shows that SARC significantly improves myocardial perfusion in the jeopardized area of the myocardium. Interestingly, almost no cardioplegic solution was delivered to the heart through the coronary sinus line during SARC under our experimental conditions. We speculate that elevated pressure at the coronary sinus during SARC redistributes the solution drained from adjacent normal tissue into the jeopardized myocardium.
| Material and methods |
|---|
|
|
|---|
Domestic pigs weighing 45 to 55 kg were sedated with an intramuscular injection of diazepam (5 to 10 mg) and ketamine (25 mg). Anesthesia was maintained with 1% to 2% isoflurane in a mixture of oxygen and nitrous oxide. The respiratory rate and volume were adjusted to keep the arterial blood gas values within the normal physiologic range. The brachiocephalic artery was cannulated at the level of the common carotid artery for arterial pressure monitoring, blood sampling, and infusion of cardioplegic solution. A sternotomy was performed. The brachiocephalic and subclavian arteries were dissected. The pericardium was opened longitudinally along the midline. The ascending aorta and the main pulmonary artery were isolated by threading umbilical tape around the origin of the descending aorta. Anticoagulation was provided by injection of heparin (3000 IU) into the superior vena cava. A cannula was inserted centrally in the brachiocephalic artery. The brachiocephalic artery, subclavian artery, descending aorta, and superior and inferior venae cavae were then clamped in succession. Heparinized cold (approximately 4° C) cardioplegic solution was infused into the aortic root (10 ml/kg body weight). The right and left atria were cut to allow drainage of the cardioplegic solution and to prevent warm blood in the lungs from returning to the heart. The heart was excised and then mounted in a Langendorff perfusion apparatus. A hydraulic occluder (IN VIVO METRIC, Healdsberg, Calif.) was placed at the origin of the left anterior descending artery (LAD). A 15F retrograde cannula (DLP Inc., Grand Rapids, Mich.) with a manually inflated balloon at the tip was positioned approximately 1 cm into the coronary sinus and then secured with a purse-string suture. We did not advance this cannula as far as possible because it might have impeded effluence of antegradely perfused cardioplegic solution. In this study, all of the hearts showed extension of the posterior interventricular vein. The flow of retrograde perfusion was constantly monitored with use of an in-line transonic flowmeter.
Perfusion solutions
Complete washout of the contrast agent before any subsequent injection of the agent is a prerequisite for accurate measurement of tissue perfusion. The amount of blood obtained from one pig was not sufficient to perform an imaging study. Thus in the present study we used an oxygenated crystalloid cardioplegic solution as the perfusion medium. The composition of this solution was as follows (in millimoles per liter): NaCl 100, KCl 16, MgCl2 16, ethylenediaminetetraacetic acid 0.5, glucose 11, NaHCO3 25, KH2PO4 1.2, CaCl2 1.75, and bovine serum albumin 0.5%. The concentration of free calcium in this solution was 1.1 to 1.2 mmol/L. The pH of the solution was adjusted to 7.4 to 7.5. Tepid cardioplegic solution has been shown to improve myocardial protection during cardiac operations. We believe that this temperature may combine the advantages of both hypothermic and normothermic cardioplegia. Therefore the temperature of the heart was maintained at 30° to 31° C throughout the protocol.
MR perfusion imaging
Myocardial perfusion was assessed with MR imaging with an MR contrast agent (gadolinium-diethylenetriamine pentaacetic acid). This technique is noninvasive and nondestructive, such that a heart can serve as its own control.
24,25 This technique also offers a very high spatial resolution and is excellent for detecting small regions of perfusion deficits.
T1-weighted MR imaging was used to assess myocardial perfusion in this study because this imaging method provides a higher magnitude of contrast agentinduced enhancement of signal intensity and has lower sensitivity to distortion caused by the interface between the myocardium and the air compared with the T2*-weighted imaging method. T1-weighted imaging was performed with a 7 tesla 40 cm horizontal bore magnet and a Helmholtz probe surrounding the whole heart. The pulse sequence to acquire T1-weighted images was an inversion-recovery gradient-echo with a short repetition time (10 msec) and short echo time (4 msec). The inversion time and flip angle of the pulse were 250 msec and 10°, respectively. Each image covered a 15 cm2 field of view with a 128 x 128 matrix size, leading to a pixel size of 1.17 mm2. One image was acquired every 1.25 seconds. The images were acquired along the short axis of the heart. This orientation of the images provides the best view of myocardial perfusion throughout the heart with a single slice. Fifty images were taken in a block. Three blocks of images were acquired for each injection of contrast agent to follow the dynamic changes in contrast agent level and its distribution in the myocardium. Gadolinium-diethylenetriamine pentaacetic acid (0.05 ml/kg body weight; Berlex Canada, Montreal, Canada) was used as the MR contrast agent.
Time courses of the signal intensity of the perfusion images in the LAD region and the non-LAD region were generated by averaging the signal intensities measured in each area (Fig. 1).
|
|
During the interval between the first and second periods and the interval between the second and third periods, the LAD occluder was released for 5 minutes to wash out the contrast agent delivered during the previous injection as well as to minimize any possible injury to the coronary vessels in the region. MR images were continuously acquired during these intervals to ensure the removal of all contrast agent from the heart before the subsequent injection.
Statistical analyses
Statistical analyses were performed with STATISTICA computer software (STATSOFT Inc., Tulsa, Okla.). All results are expressed as the mean plus or minus the standard error of the mean. Comparison of the time course of signal intensity from the LAD and non-LAD regions was performed by one-way analysis of variance with repeated measurements. One-way analysis of variance was also used to compare the maximum signal intensity obtained from the LAD region during various stages of the protocol. The distribution of cardioplegic solution in the LAD region during SARC was determined using EvIdent two- and three-dimensional image analysis software (Institute for Biodiagnostics, NRC, Winnipeg, Canada).
| Results |
|---|
|
|
|---|
|
|
Fig. 5 summarizes myocardial perfusion during antegrade delivery of cardioplegic solution (second period, top panel) and SARC (third period, middle panel) in the LAD and the non-LAD regions in group 1 hearts. During antegrade perfusion (second period), signal intensity in the LAD region remained almost unchanged whereas signal intensity in the non-LAD region showed a rapid and significant increase (p < 0.001), indicating that the myocardium in the LAD region was virtually not perfused. During SARC (Fig. 5,
middle panel), injection of contrast agent resulted in an increase of signal intensity in both regions, although the magnitude of signal intensity increase was significantly smaller in the LAD region than in the non-LAD region (p < 0.001). SARC resulted in a significant increase of signal intensity in the LAD region compared with that during antegrade cardioplegia (p < 0.001) (bottom panel). Although contrast agent was administered only into the aortic line in group 2 hearts, similar results were obtained and are shown in Fig. 6.
|
|
|
Representative patterns of cardioplegic distribution in the LAD region obtained during SARC are shown in Fig. 8. All the hearts in both groups showed heterogeneous perfusion to the occluded myocardium during SARC.
|
| Discussion |
|---|
|
|
|---|
In the hearts in group 1 SARC produced an increase in the maximum signal intensity in the LAD region (47.3% ± 8.6%) relative to that observed in the non-LAD region (100%) (Fig. 7
). Because it may not depend solely on the coronary flow, the maximum signal intensity cannot be used as a precise measurement of myocardial flow rate. On the other hand, comparison of the maximum signal intensity between the two regions was made in the same hearts. Other factors (which may have an impact on the maximum signal intensity, such as blood vessel volume and permeability) are expected to be comparable in these two regions. Accordingly, the ratio of the maximum signal intensity in two regions could be considered as an approximation of the ratio of myocardial flow. Based on the maximum signal intensity, myocardial flow resulting from SARC to the jeopardized myocardium is approximately 40% to 50% of that to the non-LAD region. The results demonstrate that SARC significantly improves myocardial perfusion in the jeopardized areas. On the other hand, the data also show that SARC provides heterogeneous perfusion within the occluded region (Fig. 8
).
In addition, we found that pressure values in the aorta (46.8 ± 3.2 mm Hg) and coronary sinus (46.7 ± 2.9 mm Hg) were similar during SARC in group 1 hearts, if there was no solution leakage at the coronary sinus. We also found that the pressure values in the coronary sinus and retrograde flow during SARC were highly dependent on balloon inflation. If the balloon was fully inflated and the orifice of the coronary sinus was completely sealed, there was no solution leakage in the coronary sinus. Under these conditions, coronary sinus pressure reached a maximum and the retrograde flow was almost zero (five hearts showed no retrograde flow and two showed a flow rate of 1 to 10 ml/min). This suggests that during SARC the myocardium in the LAD region was not perfused by the solution delivered through the retrograde line. It is our hypothesis that during SARC the myocardium at risk is mainly perfused by the solution delivered antegradely and drained from the adjacent tissue as a result of elevated pressure at the coronary sinus.
To test our hypothesis and to determine the effect of pressure at the coronary sinus during SARC on perfusion of the myocardium at risk, we studied the second group of hearts. In this group, contrast agent was introduced only into the aortic line during SARC. Retrograde perfusion of SARC was performed with use of a saline solution bag filled with cardioplegic solution and joined to the retrograde cannula to set the coronary sinus pressure to the designed levels. In this group, coronary sinus pressure was controlled at 38 mm Hg and the aortic pressure was maintained at 55 mm Hg. Under these conditions, we found that no solution was delivered to the heart through the retrograde cannula during SARC. However, MR images still show invasion of MR contrast agent into the LAD region during SARC (Figs. 3
, 4
, 6
, and 7
) although no contrast agent was administered into the retrograde line. This further demonstrates that the myocardium in occluded areas is mainly perfused by the solution drained from the adjacent normal tissue, not by solution delivered retrogradely.
Furthermore, MR images show that myocardial perfusion in the LAD region was significantly less in group 2 hearts than in group 1 hearts (23.46% ± 5.29% vs 47.3% ± 8.6%, p = 0.002), suggesting that pressure at the coronary sinus relative to the aortic pressure may play an important role in determining the flow of SARC solution to the risk areas of the myocardium. We are currently testing whether a higher pressure in the coronary sinus relative to that in the aorta will further improve myocardial perfusion. In addition, maximum signal intensity in the non-LAD region from group 1 hearts was significantly higher (p = 0.024) during SARC than during antegrade delivery of cardioplegic solution, even though total flow and amount of contrast agent used were identical during these two periods. At present, it is still uncertain whether higher maximum signal intensity in the non-LAD region during SARC relative to that during antegrade perfusion was caused by improved perfusion or other unknown factors.
In addition, with use of our two- and three-dimensional image analysis software (EvIdent), we have demonstrated in this study that SARC does not provide homogeneous perfusion to the jeopardized myocardium. In the clinical situation, however, the size of the area at risk caused by coronary stenosis or occlusion is usually much smaller than that used in this study. Heterogeneous perfusion of SARC, therefore, may not be so pronounced in the human heart.
It must be pointed out that the present study was performed on isolated pig hearts. The subjects were healthy and relatively young. It is possible that the development of coronary anastomoses and thebesian vessels is much less in the pig heart than in the human patient heart. Consequently, the capacity of SARC to perfuse the area at risk may be higher in the human being than it is in the pig. In addition, this study did not compare SARC with other cardioprotective techniques (such as antegrade cardioplegia, retrograde cardioplegia, and alternate antegrade/retrograde cardioplegia) in terms of their effects on myocardial energy metabolism and functional recovery. Moreover, the potential detrimental effect of SARC on the coronary vasculature was not evaluated in this study. Therefore, at the present, no judgment can be made regarding which technique is superior for heart protection. As mentioned previously, this study was done with the use of crystalloid solution because of the insufficient blood volume that was obtainable from the animal. Because of intrinsic physiologic characteristics of blood, we expect that our study would be more clinically relevant if it had been performed with blood cardioplegic solution.
Clinically, some surgeons use antegrade delivery of cardioplegic solution through a newly established vein bypass graft while cardioplegic solution is still being delivered retrogradely through the coronary sinus. This combination has been used to test vascular anastomoses and to improve myocardial perfusion in the territory supported by the grafted artery. This is a practical and interesting combination of both antegrade and retrograde cardioplegia. However, because the conditions of this combination and those of SARC used in our study are very different, it is expected that these two techniques would result in different myocardial perfusion.
In summary, this study demonstrates that SARC improves myocardial perfusion to the areas of the myocardium at risk. However, under our experimental conditions, almost no solution is delivered to the heart through the coronary sinus. The myocardium at risk is mainly perfused by the solution drained from adjacent normal tissue during SARC. Elevated pressure in the coronary sinus seems to be the principal factor responsible for improved perfusion of the myocardium at risk.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Thielmann, M. Neuhauser, A. Marr, U. Herold, M. Kamler, P. Massoudy, and H. Jakob Predictors and Outcomes of Coronary Artery Bypass Grafting in ST Elevation Myocardial Infarction Ann. Thorac. Surg., July 1, 2007; 84(1): 17 - 24. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Wang, H. Liu, T. A. Salerno, B. Xiang, G. Li, M. Gruwel, M. Jackson, D. Manley, B. Tomanek, R. Deslauriers, et al. Does Normothermic Normokalemic Simultaneous Antegrade/Retrograde Perfusion Improve Myocardial Oxygenation and Energy Metabolism for Hypertrophied Hearts? Ann. Thorac. Surg., May 1, 2007; 83(5): 1751 - 1758. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Tian, G. Dai, B. Xiang, J. Sun, W. G. Lindsay, and R. Deslauriers Effect on myocardial perfusion of simultaneous delivery of cardioplegic solution through a single coronary artery and the coronary sinus J. Thorac. Cardiovasc. Surg., November 1, 2001; 122(5): 1004 - 1010. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Ruengsakulrach and B. F. Buxton Anatomic and hemodynamic considerations influencing the efficiency of retrograde cardioplegia Ann. Thorac. Surg., April 1, 2001; 71(4): 1389 - 1395. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Borger, K. S. Wei, R. D. Weisel, J. S. Ikonomidis, V. Rao, G. Cohen, T. Shirai, A. S. Omran, S. C. Siu, and H. Rakowski Myocardial perfusion during warm antegrade and retrograde cardioplegia: a contrast echo study Ann. Thorac. Surg., September 1, 1999; 68(3): 955 - 961. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||