J Thorac Cardiovasc Surg 1998;116:236-240
© 1998 Mosby, Inc.
Surgery for Adult Cardiovascular Disease |
From the Divisions of Cardiovascular Surgery of The Toronto Hospital and the University of Toronto, Toronto, Ontario, Canada.
Received for publication Feb. 13, 1998. Revisions requested March 26, 1998; revisions received April 13, 1998. Accepted for publication April 13, 1998. Address for reprints: Tirone E. David, MD, 200 Elizabeth St.13EN219, Toronto, Ontario, Canada M5G 2C4.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
This study was carried out to compare the late outcomes of aortic valve replacement with stentless and stented porcine aortic valves.
| Patients and methods |
|---|
|
|
|---|
The Toronto SPV bioprosthesis was an investigational device during this study, and it was implanted in patients who fulfilled the following requirements: older than 34 years of age, no active aortic valve infection, no renal failure requiring dialysis, and an expected life span of at least 2 years. The Hancock II was implanted whenever appropriate. Because of these differences in patient selection, a case-control study was developed on the basis of preoperative variables predictive of death after AVR. These variables were identified by a stepwise logistic regression analysis for the entire patient population with porcine bioprostheses. The following preoperative variables were independent predictors of death after AVR: advanced age, New York Heart Association functional class IV, left ventricular ejection fraction below 30%, and coronary artery disease. With this information, a case-match study was developed on the basis of the patients' age (±5 years), New York Heart Association functional class (I, II, III together, or IV ), left ventricular ejection fraction (higher or lower than 30%), and coronary artery disease. Only survivors of the operation were entered into the case-match study. One hundred ninety-eight pairs of patients were matched. Table I shows the clinical profile of the matched patients. Both groups had the same incidence of associated diseases, such as peripheral vascular disease, diabetes, hypertension, and chronic obstructive lung disease. There were no differences in the intraoperative management of patients as far as conduction of anesthesia, cardiopulmonary bypass, or myocardial protection.
|
All statistical analyses were performed with the SAS (SAS Institute, Inc., Cary, N.C.) and the BMDP/DYN (BMDP Statistical Software, Berkeley, Calif.) programs. The clinical profile of the two groups was compared by
2 analysis. Longitudinal data, such as actuarial survival and freedom from morbid events, were estimated by the product-limit or by the Kaplan-Meier method. Values were expressed as means and the standard errors of the means. The longitudinal comparisons of the two groups of patients were evaluated by the Mantel-Cox statistic. The risk-adjusted benefit of the Toronto SPV bioprosthesis was evaluated multivariately in a Cox proportional hazard model.
| Results |
|---|
|
|
|---|
|
|
|
Infective endocarditis
Infective endocarditis developed in two patients with the Toronto SPV (one early and one late), and in six patients with the Hancock II (one early and five late). The freedom from infective endocarditis at 8 years was 98% ± 2% for the Toronto SPV and 94% ± 2% for the Hancock II (p = 0.05).
Reoperation
Redo AVR was necessary in two patients with the Toronto SPV bioprosthesis (both for endocarditis) and in eight patients with the Hancock II (six for endocarditis and two for primary tissue failure). The freedom from redo AVR at 8 years was 98% ± 2% for the Toronto SPV bioprosthesis and 83% ± 9% for the Hancock II (p = 0.03).
Valve-related morbidity
Fig. 3 shows the freedom from any valve-related complication for both groups of patients. The difference was statistically significant in favor of the Toronto SPV bioprosthesis (p = 0.05).
|
65 years), left ventricular dysfunction (ejection fraction, >30% and
30%), New York Heart Association (classes I through III and class IV), coronary artery disease, and valve type (Toronto SPV and Hancock II). Aortic stenosis was also tested because there were more patients with aortic stenoses in the Toronto SPV bioprosthesis than in the Hancock II group (Table I
|
| Discussion |
|---|
|
|
|---|
Assessing potential bias and confounding factors
Patients were matched by an epidemiologist who had no knowledge of the valves or of the clinical outcomes. Careful review of each patient's clinical profile failed to indicate any bias as far as the severity of the cardiac disease or of the associated diseases such as diabetes, hypertension, peripheral vascular disease, renal function impairment, or any malignancy that may have affected the surgeon's choice of the valve in the matched patients. However, there was a higher number of noncardiac deaths among patients with the Hancock II valve, and this finding could not be explained by examining the preoperative variables.
Table I shows a higher incidence of aortic stenosis among patients who received the Toronto SPV bioprosthesis when compared with patients who received the Hancock II bioprosthesis. This is probably due to the fact that the Toronto SPV bioprosthesis was not implanted in patients with a dilated aortic root. Nevertheless, aortic stenosis was not a predictor of clinical outcomes by univariate analysis (
2 and Fisher's exact test) or multivariate analysis (Cox regression analysis) in the entire population or in any of its subgroups (Hancock II sampled and not sampled or Toronto SPV).
To evaluate the potential bias of the Hancock II patients, the sampled group of (198 patients) was compared with the not sampled (468 patients). The actuarial survival at 8 years was 69% ± 5% for the sampled group and 70% ± 3% for the not sampled group of Hancock II patients (p = 0.34). The freedom from cardiac deaths and the freedom from valve-related deaths were also similar for the sampled and not sampled groups of Hancock II.
The clinical outcomes after AVR with the Toronto SPV bioprosthesis have been exceptionally good, particularly when compared with other reports on stented bioprostheses.
6,10-13 In our experience with 242 patients with a mean age of 62 years and coronary artery disease in one third of those patients, the actuarial survival including the operative mortality rate was 89% at 9 years.
6 Similarly, in the premarked clinical trial involving over 600 patients (mean age, 67 years; coronary artery disease, 40%) sponsored by the manufacturer of this valve to obtain its Food and Drug Administration approval, the actuarial survival at 6 years was approximately 90%.
10 The rates of valve-related complications have also been lower than those observed with other stented bioprostheses.
6,10-13 In most reports on late outcomes after AVR with stented biologic valves, the actuarial survival at 5 years ranged around 70% to 80% and at 10 years ranged around 50% to 60%.
11-13
Hemodynamic performance of the Toronto SPV bioprosthesis
Numerous studies have shown that the Toronto SPV bioprosthesis has excellent hemodynamic features that are comparable with those of the aortic valve homograft.
3-5,14 Serial echocardiographic studies in large numbers of patients who underwent an AVR with the Toronto SPV bioprosthesis demonstrated a progressive reduction in the mean transvalvular gradients and an increase in the effective valve orifices during the first year after implantation.
3-4,6 At the end of postoperative year 1, the mean transvalvular gradients have been consistently below 5 mm Hg in all patients, regardless of the size of valve implanted.
3-6 This favorable hemodynamic feature probably inhibits the stimulus responsible for left ventricular hypertrophy, and remodeling of the ventricle begins soon after operation.
3,5 Although regression of left ventricular hypertrophy also occurs after AVR with stented valves,
15-17 most patients still have residual hypertrophy after 2 to 3 years.
16 Residual hypertrophy is rare after AVR with the Toronto SPV bioprosthesis after postoperative year 2.
3 In addition, it has been shown that early postoperative left ventricular function after AVR is better with the Toronto SPV bioprosthesis than with stented porcine and mechanical valves.
5,15 This observation may explain the relatively low operative mortality rate for AVR with the Toronto SPV bioprosthesis in spite of longer aortic crossclamp times than with stented valves.
4-6 In our experience with porcine valves, the operative mortality rate for AVR with the Toronto SPV bioprosthesis was 0.8%, whereas for the Hancock II it was 4%.
4,11
The long-term effects of residual left ventricular hypertrophy after AVR are unknown, but the clinical experience with systemic hypertension indicates that left ventricular hypertrophy adversely affects long-term survival.
7,8 Thus it is conceivable that AVR with the Toronto SPV bioprosthesis enhances survival because of its hemodynamic features. This may well explain the differences in cardiac and valve-related deaths in our case-control study. Indeed, the major difference in clinical outcomes between our two groups of patients was in the number of late cardiac deaths. During a mean follow-up of 43 months, one third as many late deaths occurred in patients with the Toronto SPV valve as in patients with the Hancock II bioprosthesis. The frequency of valve-related complications were also significantly lower in patients who received stentless valves. This reduction in complications may also have indirectly affected survival.
Aortic valve homografts have excellent hemodynamic features and, when implanted in the subcoronary position, provide similar flows to the normal aortic valve.
14 Thus one would expect that AVR with an aortic valve homograft would enhance late survival, which is what we found with the SPV bioprosthesis. Although this may not have been observed in the past, recently published results on AVR with homograft indicate a survival of 85% to 95% at 8 to 10 years, including operative mortality rates.
18-20 Although the patient population in those series were different from ours, the results suggest that aortic valve homografts provide better clinical outcomes than those obtained with stented biologic valves. Probably the same could be said about AVR with pulmonary autograft because it is also a stentless biologic valve with superlative hemodynamic qualities. Finally, we have treated numerous patients with patient-prosthesis mismatch after AVR when mechanical and stented bioprostheses were implanted, but we have never documented a case of patient-prosthesis mismatch in more than 500 AVRs with stentless biologic valves of all types.
Durability of the Toronto SPV bioprosthesis
The Toronto SPV bioprosthesis has been used only since July 1991. There has been no documented case of primary tissue degeneration, but it has been used almost exclusively in older patients.
6,10 However, we implanted 29 custom-made stentless porcine aortic valves similar to the Toronto SPV bioprosthesis just over one decade ago,
1 and the experience with those valves suggests that their durability is no different from that of stented porcine valves.
6 If this is the case with the Toronto SPV bioprosthesis, and if this valve indeed enhances the life span after AVR, the probability of reoperation for valve failure will be greater with this valve than with stented ones because more patients will be at risk of outliving their bioprosthesis.
Limitations of the study
This was a retrospective study, and the two groups of patients may have been quite different in spite of our efforts to eliminate any potential bias and to exclude any confounding factor that may have influenced the clinical outcome. Other limitations are the relatively small number of patients in each group and the duration of follow-up.
| Conclusions |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. A. Yankah, M. Pasic, M. Musci, J. Stein, C. Detschades, H. Siniawski, and R. Hetzer Aortic valve replacement with the Mitroflow pericardial bioprosthesis: Durability results up to 21 years J. Thorac. Cardiovasc. Surg., September 1, 2008; 136(3): 688 - 696. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stelzer Stentless Aortic Valve Replacement: Porcine and Pericardial Card. Surg. Adult, January 1, 2008; 3(2008): 915 - 934. [Full Text] |
||||
![]() |
C. R. Bridges, S. M. O'Brien, J. C. Cleveland, E. B. Savage, J. S. Gammie, F. H. Edwards, E. D. Peterson, and F. L. Grover Association between indices of prosthesis internal orifice size and operative mortality after isolated aortic valve replacement J. Thorac. Cardiovasc. Surg., April 1, 2007; 133(4): 1012 - 1021. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bove, Y. Van Belleghem, K. Francois, F. Caes, H. Van Overbeke, and G. Van Nooten Stentless and stented aortic valve replacement in elderly patients: factors affecting midterm clinical and hemodynamical outcome Eur. J. Cardiothorac. Surg., November 1, 2006; 30(5): 706 - 713. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Borger, S. M. Carson, J. Ivanov, V. Rao, H. E. Scully, C. M. Feindel, and T. E. David Stentless Aortic Valves are Hemodynamically Superior to Stented Valves During Mid-Term Follow-Up: A Large Retrospective Study Ann. Thorac. Surg., December 1, 2005; 80(6): 2180 - 2185. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Totaro, N. Degno, A. Zaidi, A. Youhana, and V. Argano Carpentier-Edwards PERIMOUNT Magna bioprosthesis: A stented valve with stentless performance? J. Thorac. Cardiovasc. Surg., December 1, 2005; 130(6): 1668 - 1674. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Polvani, F. Barili, L. Dainese, M. Muratori, M. Porqueddu, A. Sala, and P. Biglioli Long-Term Results After Aortic Valve Replacement With the Bravo 400 Stentless Xenograft Ann. Thorac. Surg., August 1, 2005; 80(2): 495 - 501. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. A. Halees, M. A. Shahid, A. A. Sanei, A. Sallehuddin, and C. Duran Up to 16 years follow-up of aortic valve reconstruction with pericardium: a stentless readily available cheap valve? Eur. J. Cardiothorac. Surg., August 1, 2005; 28(2): 200 - 205. [Abstract] [Full Text] [PDF] |
||||
![]() |
I Florath, A Albert, U Rosendahl, T Alexander, I C Ennker, and J Ennker Mid term outcome and quality of life after aortic valve replacement in elderly people: mechanical versus stentless biological valves Heart, August 1, 2005; 91(8): 1023 - 1029. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tamim, T. Bove, Y. Van Belleghem, K. Francois, Y. Taeymans, and G. J Van Nooten Stentless vs. Stented Aortic Valve Replacement: Left Ventricular Mass Regression Asian Cardiovasc Thorac Ann, June 1, 2005; 13(2): 112 - 118. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tamim, T. Bove, Y. Van Belleghem, F. Caes, K. Francois, and G. J Van Nooten Aortic Valve Replacement with Toronto SPV in Elderly Patients: 10-Year Results Asian Cardiovasc Thorac Ann, June 1, 2005; 13(2): 143 - 148. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. F. O'Brien, M. A.H. Gardner, B. Garlick, H. Jalali, J. A. Gordon, S. L. Whitehouse, W. E. Strugnell, and R. Slaughter CryoLife-O'Brien Stentless Valve: 10-Year Results of 402 Implants Ann. Thorac. Surg., March 1, 2005; 79(3): 757 - 766. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. G. Gleason, T. E. David, J. S. Coselli, J. W. Hammon Jr, and J. E. Bavaria St. Jude Medical Toronto biologic aortic root prosthesis: Early FDA phase II IDE study results Ann. Thorac. Surg., September 1, 2004; 78(3): 786 - 793. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. John, C. Glauner, M. A. Manoutcheri, V. Ziaukas, G. M. Mahesh, and H. Warnecke Aortic Valve Replacement with the Medtronic Freestyle Stentless Bioprosthesis Asian Cardiovasc Thorac Ann, September 1, 2004; 12(3): 213 - 217. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Bach Choice of prosthetic heart valves:update for the next generation J. Am. Coll. Cardiol., November 19, 2003; 42(10): 1717 - 1719. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Rahimtoola The next generation of prosthetic heart valves needs a proven track record of patient outcomes at >=15 to 20 years J. Am. Coll. Cardiol., November 19, 2003; 42(10): 1720 - 1721. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ennker, U. Rosendahl, I. C. Ennker, S. Bauer, and I. Florath Risk in Elderly Patients After Stentless Versus Stented Aortic Valve Surgery Asian Cardiovasc Thorac Ann, March 1, 2003; 11(1): 37 - 41. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. A. Vricella, M. A. Coady, and M. D. Black Initial experience with a stentless porcine bioprosthesis for right ventricular outflow tract reconstruction in children J. Thorac. Cardiovasc. Surg., March 1, 2003; 125(3): 727 - 728. [Full Text] [PDF] |
||||
![]() |
T. Fukui, S. Suehiro, T. Shibata, K. Hattori, H. Hirai, and T. Aoyama Aortic root replacement with Freestyle stentless valve for complex aortic root infection J. Thorac. Cardiovasc. Surg., January 1, 2003; 125(1): 200 - 203. [Full Text] [PDF] |
||||
![]() |
F. J. Schoen and R. F. Padera Jr. Cardiac Surgical Pathology Card. Surg. Adult, January 1, 2003; 2(2003): 119 - 185. [Full Text] |
||||
![]() |
D. B. Doty and J. R. Doty Stentless Aortic Valve Replacement: Bioprostheses Card. Surg. Adult, January 1, 2003; 2(2003): 889 - 898. [Full Text] |
||||
![]() |
G. Dellgren, M.J. Eriksson, L.A. Brodin, and K. Radegran Eleven years' experience with the Biocor stentless aortic bioprosthesis: clinical and hemodynamic follow-up with long-term relative survival rate Eur. J. Cardiothorac. Surg., December 1, 2002; 22(6): 912 - 921. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Bevilacqua, J. Gianetti, A. Ripoli, U. Paradossi, A. Giuseppe Cerillo, M. Glauber, M. L. Sacha Matteucci, M. Senni, A. Gamba, E. Quaini, et al. Aortic valve disease with severe ventricular dysfunction: stentless valve for better recovery Ann. Thorac. Surg., December 1, 2002; 74(6): 2016 - 2021. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. B. Luciani, G. Casali, S. Auriemma, F. Santini, and A. Mazzucco Survival after stentless and stented xenograft aortic valve replacement: a concurrent, controlled trial Ann. Thorac. Surg., November 1, 2002; 74(5): 1443 - 1449. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. Bach, P. C. Cartier, N. D. Kon, K. G. Johnson, G. M. Deeb, and D. B. Doty Impact of implant technique following freestyle stentless aortic valve replacement Ann. Thorac. Surg., October 1, 2002; 74(4): 1107 - 1113. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. Castro, J. M. Arcidi Jr, A. L. Fisher, and V. A. Gaudiani Routine enlargement of the small aortic root: a preventive strategy to minimize mismatch Ann. Thorac. Surg., July 1, 2002; 74(1): 31 - 36. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. D. Kon, R. D. Riley, S. M. Adair, D. W. Kitzman, and A. R. Cordell Eight-year results of aortic root replacement with the freestyle stentless porcine aortic root bioprosthesis Ann. Thorac. Surg., June 1, 2002; 73(6): 1817 - 1821. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Bloomfield Choice of heart valve prosthesis Heart, June 1, 2002; 87(6): 583 - 589. [Full Text] [PDF] |
||||
![]() |
G. Dellgren, C. M. Feindel, J. Bos, J. Ivanov, and T. E. David Aortic valve replacement with the Toronto SPV: long-term clinical and hemodynamic results Eur. J. Cardiothorac. Surg., April 1, 2002; 21(4): 698 - 702. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David, J. Ivanov, M. J. Eriksson, J. Bos, C. M. Feindel, and H. Rakowski Dilation of the sinotubular junction causes aortic insufficiency after aortic valve replacement with the Toronto SPV bioprosthesis J. Thorac. Cardiovasc. Surg., November 1, 2001; 122(5): 929 - 934. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Silberman, J. Shaheen, O. Merin, D. Fink, N. Shapira, N. Liviatan-Strauss, and D. Bitran Exercise hemodynamics of aortic prostheses: comparison between stentless bioprostheses and mechanical valves Ann. Thorac. Surg., October 1, 2001; 72(4): 1217 - 1221. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Thomson, W.R. E. Jamieson, J. G. Dumesnil, J. J. Burgess, C. M. Peniston, J. Metras, J. A. Sullivan, J. C.W. Parrott, A. Maitland, and I. J. Cybulsky Medtronic mosaic porcine bioprosthesis: midterm investigational trial results Ann. Thorac. Surg., May 1, 2001; 71 (2007): S269 - S272. [Abstract] [Full Text] [PDF] |
||||
![]() |
W.R. E. Jamieson, M. T. Janusz, J. MacNab, and C. Henderson Hemodynamic comparison of second- and third-generation stented bioprostheses in aortic valve replacement Ann. Thorac. Surg., May 1, 2001; 71 (2007): S282 - S284. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. B. Luciani, F. Santini, S. Auriemma, L. Barozzi, P. Bertolini, and A. Mazzucco Long-term results after aortic valve replacement with the Biocor PSB stentless xenograft in the elderly Ann. Thorac. Surg., May 1, 2001; 71 (2007): S306 - S310. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. P. Kappetein, J. Braun, L. H.B. Baur, A. Prat, K. Peels, M. G. Hazekamp, P. H. Schoof, and H. A. Huysmans Outcome and follow-up of aortic valve replacement with the freestyle stentless bioprosthesis Ann. Thorac. Surg., February 1, 2001; 71(2): 601 - 608. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Westaby, M. Horton, X. Y. Jin, T. Katsumata, O. Ahmed, S. Saito, H.-H. Li, and G. L. Grunkemeier Survival advantage of stentless aortic bioprostheses Ann. Thorac. Surg., September 1, 2000; 70(3): 785 - 791. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Riley, J. W. Hammon Jr, S. M. Adair, A. R. Cordell, and N. D. Kon Stentless aortic valve replacement with freestyle or Toronto SPV: an early comparison Ann. Thorac. Surg., July 1, 2000; 70(1): 48 - 51. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Silberman, N. Shapira, and D. Bitran Comparison between stentless xenografts and homografts Ann. Thorac. Surg., June 1, 2000; 69(6): 1989 - 1989. [Full Text] [PDF] |
||||
![]() |
C. J. Knott-Craig, R. C. Elkins, K. Santangelo, C. McCue, and M. M. Lane Aortic valve replacement: comparison of late survival between autografts and homografts Ann. Thorac. Surg., May 1, 2000; 69(5): 1327 - 1331. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Walther, V. Falk, G. Langebartels, M. Kruger, U. Bernhardt, A. Diegeler, J. Gummert, R. Autschbach, and F. W. Mohr Prospectively Randomized Evaluation of Stentless Versus Conventional Biological Aortic Valves : Impact on Early Regression of Left Ventricular Hypertrophy Circulation, November 9, 1999; 100 (2009): II-6 - II-10. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. L. Yun, C. F. Sintek, A. D. Fletcher, T. A. Pfeffer, G. S. Kochamba, M. R. Hyde, J. O. Torpoco, and S. Khonsari Aortic Valve Replacement With the Freestyle Stentless Bioprosthesis : Five-Year Experience Circulation, November 9, 1999; 100 (2009): II-17 - II-23. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. D. Kon, A. R. Cordell, S. M. Adair, J. E. Dobbins, and D. W. Kitzman Aortic root replacement with the freestyle stentless porcine aortic root bioprosthesis Ann. Thorac. Surg., June 1, 1999; 67(6): 1609 - 1615. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Rao, G. T. Christakis, J. Sever, S. E. Fremes, G. Bhatnagar, G. Cohen, M. A. Borger, L. Abouzahr, and B. S. Goldman A NOVEL COMPARISON OF STENTLESS VERSUS STENTED VALVES IN THE SMALL AORTIC ROOT J. Thorac. Cardiovasc. Surg., March 1, 1999; 117(3): 431 - 438. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Hvass, G. M. Palatianos, R. Frassani, C. Puricelli, and M. O'Brien MULTICENTER STUDY OF STENTLESS VALVE REPLACEMENT IN THE SMALL AORTIC ROOT J. Thorac. Cardiovasc. Surg., February 1, 1999; 117(2): 267 - 272. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |