J Thorac Cardiovasc Surg 1999;117:1144-1149
© 1999 Mosby, Inc.
SURGERY FOR ADULT CARDIOVASCULAR DISEASE |
From the Department of General Surgery, Northside Medical Center, Forum Health,a Youngstown, Ohio, and Cardiothoracic Surgery Research, Cardiovascular and Pulmonary Research Institute, and Department of Thoracic Surgery, Allegheny General Hospital,b Pittsburgh, Pa.
Supported by the Allegheny Heart Institute, Pittsburgh, Pa.
Received for publication Dec 11, 1998. Revisions requested Jan 22, 1999. Revisions received Feb 10, 1999. Accepted for publication Feb 16, 1999. Address for reprints: James A. Magovern, MD, Cardiothoracic Surgery Research, CVPRI, 14th Floor, South Tower, Allegheny General Hospital, 320 East North Ave, Pittsburgh, PA 15212.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Median sternotomy wound complications range from prolonged incisional pain to dehiscence and mediastinitis. These problems are often preceded by sternal instability, which has been shown to compromise wound integrity and promote bacterial infection.
2,5,6 This study presents a comprehensive analysis of the stability of human sternal closure when subjected to mechanical distracting forces and to increases in intrathoracic pressure. The goal is to elucidate the mechanisms for sternal dehiscence so that improved methods for sternal closure can be developed.
| Methods |
|---|
|
|
|---|
Cadaver preparation
Median sternotomy was performed with an oscillating saw and closed with No. 5 stainless steel wire that was symmetrically tightened by twisting with a large needle-driver. A total of 7 wires were placed: 2 at the manubrium and 5 around the sternum in intercostal spaces 2 to 6. The chest wall was instrumented to apply distracting forces across the sternal closure. Brass plates were placed above and below the second, fourth, and sixth ribs bilaterally to prevent rib fracture resulting from concentration of forces at the 6 fixation points. A custom-designed rigid cage was constructed from steel piping and placed over the cadaver to provide anchor points for the application of traction forces (Fig. 1).Thin steel cables (1/16-inch diameter) were attached to metal hose clamps that were secured around the brass plates near the sternal insertion. Cables on one side of the sternum were fixed directly to the steel piping. Cables on the other side were first connected to a single o-ring, then to a turnbuckle, a load-cell, and finally to the rigid cage. The turnbuckle was used to increase tension in the cable which, in turn, applied force to the sternal closure. The load cell (model LCDB-200, Omega Engineering, Inc, Stamford, Conn) was used in conjunction with a DP25 series digital indicator (Omega Engineering, Inc) to measure traction forces.
|
|
Measurement of sternal distraction
Four pairs of piezoelectric crystal transducers were used to measure distraction of the closure site (sonomicrometer model 120, Triton Technology, Inc, San Diego, Calif). In the lateral and rostral-caudal distraction experiments, 4 pairs of crystals were placed on the anterior surface of the sternum, 1 on each side of the closure, at intervals equidistant from each other. One pair was placed on the manubrium, 2 pairs on the sternum, and 1 pair at the sternoxiphoid junction. Ultrasonic gel was placed between the crystals to provide a medium for sound wave conduction. Anterior-posterior motion was measured by placing 1 crystal of each pair face-up on one side of the sternum while the other was suspended (face down) directly above the first via a small Allen wrench anchored to the opposite side. Vertical traction was applied to one side of the sternum such that the lower crystals were drawn toward the upper crystals. In all traction experiments, forces were applied in discrete 10-N increments (starting at a resting tension of 40 N) until 2.0 mm of distraction was noted in at least 1 set of crystals. This end point was chosen as an amount of motion that would be clinically important and as a means to prevent the wires from fracturing the sternum and causing complete dehiscence.
Intrathoracic pressure experiments were conducted after completion of the distracting force experiments in each cadaver. Because distraction studies showed the xiphoid region of the sternum to be the least stable, all crystals were placed at the lower end of the sternum and arranged to measure motion in lateral, anterior-posterior, and rostral-caudal directions simultaneously. Intrathoracic pressure was steadily increased until 2.0-mm motion was detected in at least 1 set of crystals.
In all cases, lateral traction experiments were performed first, followed by anterior-posterior and rostral-caudal distraction trials (in that order). Intrathoracic pressure studies were performed last. Sternal wires were tightened to original tension levels after each experiment to reverse loosening produced by forced sternal motion against the twisted wire closure.
Data collection and statistical analysis
During sternal traction, crystal separations were recorded in discrete data sets, one for each load level applied. Sonomicrometry waveforms were recorded continuously during Valsalva studies. Data were digitized at a rate of 25 samples/s and stored in an IBM 300PL personal computer (data acquisition package: WinDAQ, Dataq Instruments, Akron, Ohio). Analysis of variance was performed to determine whether differences in separation existed among the 4 sternal sites monitored in the 4 cadavers (TRUE EPISTAT, Epistat Services, Richardson, Tex). Where differences were found, Bonferroni-corrected paired t tests were performed to determine which sites yielded significant differences in separation distances. All summary data are expressed as mean ± SE.
| Results |
|---|
|
|
|---|
As previously mentioned, sternal distraction was stopped when 2.0-mm motion was detected in 1 set of crystals to prevent complete dehiscence. This occurred for lateral, anterior-posterior, and rostral-caudal forces of 220 ± 40, 263 ± 74, and 325 ± 30 N, respectively. Sternal motion was maximum near the xiphoid and minimum near the manubrium in all 4 cadavers, regardless of traction orientation (Table I). During lateral distraction, xiphoid and manubrial displacement averaged 1.85 ± 0.14 and 0.35 ± 0.12 mm, respectively ( P = .0003). Anterior-posterior distraction caused 1.99 ± 0.04mm xiphoid displacement and 0.26 ± 0.12mm manubrial displacement (P = .0004). Representative force-distraction curves (acquired from the most robust model of sternal closure tested) are shown in Fig. 3.
|
|
|
| Discussion |
|---|
|
|
|---|
A number of papers have been published over the past 30 years that describe technical modifications intended to reduce sternal wound complications. Sternal and parasternal figure-of-eight wire arrangements
11,12 along with the basket weave formation popularized by Robicsek, Daugherty, and Cook
13 are examples that use steel wire. Mechanical reinforcement of these closures may be achieved by simply uniting the pectoralis muscles over the sternum as described by Robicsek and Hamilton,
14 such that contraction of these muscles serves to hold the sternal halves together rather than pull them apart. Others have used wider materials such as nylon bands,
15 Mersilene ribbon
16 (Ethicon, Inc, Somerville, NJ), and steel bands
17 to accomplish wider force distribution. Nylon bands have subsequently been associated with a higher rate of sternal wound infection and are not recommended.
18 Although clinical success with Mersilene ribbon and steel bands has been reported in prospective randomized studies,
19,20 neither of these closure techniques has been widely accepted by the surgical community.
Steel wire remains popular because it is simple, inexpensive, and familiar. Sternal approximation is obtained by twisting the wire until an appropriate level of tension is felt by the surgeon. Because twisting weakens steel wire, some have recommended crimped metal sleeves
21 or crimped metal plates
22 to tighten and secure steel wire. Bench studies comparing twisted and crimped wire have demonstrated superior tensile strength with crimping.
6 Our study, however, suggests that if crimped steel wire were tested in human sternums, the wire would tear through the bone before the crimping mechanism failed.
Only a few published studies have used an objective, biomechanical approach to analyze closure techniques. One such study, performed by Cheng and associates,
23 compared different types of sternal closure in a human cadaver model. After sternotomy closure, the investigators removed the sternum, placed it in a biomechanical testing system, and measured motion that occurred under a lateral distracting force. They found the use of steel cerclage wiring to be a more stable closure method than those techniques that employ Mersilene ribbon or steel bands. Steel wire was superior because symmetric twisting allowed tension to be maintained across the sternal halves, resulting in snug approximation. In contrast, Mersilene ribbon required knot tying, which was not able to sustain adequate tension. Moreover, steel bands were too rigid to conform to the shape of the sternum, which precluded a secure fit. This important study provided direct visualization of different closure techniques under stress and allowed investigators to understand the flaws of some closure methods. However, limited explanation was given for the motion that occurred with cerclage steel wire closure.
A recent study by Ozaki and associates
24 used methods similar to those of Cheng and colleagues to compare rigid plate and steel wire fixation in sternal closure. In the initial phase of the study, separation was seen to occur with both techniques when wires or plates pulled through the sternum. Subsequently, a new rigid plate technique was developed that effectively distributed force across the sternal closure and was more secure. By understanding the cause of closure instability (ie, closure material cutting through the sternum), the investigators were able to develop a better method for sternal closure. Surprisingly, few other studies have attempted to characterize instability after sternotomy closure.
This report presents an innovative approach to the analysis of sternal closure stability in that distracting forces are applied across the sternum with all anatomic support structures still intact. Because this experimental model subjects the intact human thorax to normal physiologic forces, it replicates real-life sternal stress distributions more accurately than is possible in sternums isolated from the rib cage and surrounding tissues. This whole-body experimental setup was used to test the adequacy of wire closure techniques, quantify the force levels required to disrupt the closure site, identify regions of poor sternal stability, and provide information concerning closure failure modes (eg, wire penetration into the bone, wire fracture, and twist-tie loosening) needed to design more reliable approximation methods.
Several findings from this study merit further discussion. First, significant amounts of sternal motion were detected with the application of physiologic force and pressure loads to the thoracic cage. It therefore seems likely that activities of daily life, such as pulling oneself up from a lying position or coughing, place significant strain on the sternal closure site, which can cause the sternal wires to cut into the sternum. This can lead to sternal pain in some cases and to sternal dehiscence in others. Second, traction forces were observed to cause more sternal distraction in the lateral direction than similar forces applied along other (orthogonal) lines of motion. In this study, forces required to distract the sternal halves in the lateral direction were smaller by 16% and 32%, respectively, than those needed to move the sternum in the anterior-posterior and rostral-caudal directions. Likewise, modest intrathoracic pressures were seen to cause significant lateral separation of the sternal halves while producing little motion in the anterior-posterior and rostral-caudal directions. This in itself is not surprising, because only lateral separation can be achieved without the need to overcome bone-on-bone frictional forces, but these data are nonetheless important in establishing performance criteria for improved sternal closure methods. Finally, the areas of the sternum were not at equal risk for disruption. In each of these experiments, the lower half of the sternum was subject to more distraction than the manubrium. One possible explanation is that the manubrium is thicker than the lower sternum and may be stabilized by the clavicle, the more horizontal position of the ribs, and the relatively small volume of the upper thorax.
These data suggest that standard sternal closure methods for which multiple twisted wires are used may not always provide sufficient mechanical stability to allow proper wound healing, especially near the xiphoid. Perhaps the most obvious means to improve sternal stability (using this same closure technique) would be to simply secure an additional wire tie near the xiphoid process to further stabilize this portion of the sternum. The use of alternate weaving patterns such as the figure-of-eight technique could potentially enhance rostral-caudal stability but would be unlikely to reduce lateral instability caused by wire transmigration through the sternum. The addition of metal sleeves or grommets, acting in a manner analogous to shoelace eyelets, might be used to improve sternal stability by preventing the wire from cutting into the bone. Such reinforcements could also be used in conjunction with other wire closure techniques and would be especially important for older patients, in whom the sternum is thinner and more porous. We plan to test such reinforced wire closures, together with other closure techniques that similarly provide wider force distribution against the sternum (eg, basket weave, Mersilene ribbon, and steel banding) to quantify the degree to which sternal stability may be improved via these methods.
Experimental limitations
The use of embalmed cadavers can be considered a potential drawback because the mechanical properties of fixed tissues may differ from fresh bone and muscle.
25,26 To address this concern, we performed identical tests on 2 female cadavers of similar size and ageone fixed with formaldehyde and the other fresh. Very similar results were seen in these 2 cadavers, suggesting that little practical difference exists between fixed and fresh sternums tested under quasistatic loading conditions. These results are consistent with findings by Currey and colleagues,
27 who recently reported that similar tests on bovine bone were practically unaffected by formaldehyde fixation (although a significant decrease in impact strength was found).
Another factor separating this experimental model from actual clinical conditions is the time course of force application. Real-life sternal stresses tend to be transient in nature, lasting no more than a few seconds, whereas traction forces in this experiment were applied over more prolonged periods (15-20 seconds per force setting). Sternal motion, however, was seen to occur in discrete steps and in synchrony with force increases, indicating that wound separation can occur rapidly in response to modest traction forces. Still, force levels quoted here should be considered minimum values, recognizing that sternal closures may tolerate larger forces of very brief duration because of inertial damping effects.
| Conclusions |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. Schimmer, S.-P. Sommer, M. Bensch, T. Bohrer, I. Aleksic, and R. Leyh Sternal closure techniques and postoperative sternal wound complications in elderly patients. Eur. J. Cardiothorac. Surg., July 1, 2008; 34(1): 132 - 138. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. C.H. John Modified closure technique for reducing sternal dehiscence; a clinical and in vitro assessment Eur. J. Cardiothorac. Surg., May 1, 2008; 33(5): 769 - 773. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Zeitani, A. Penta de Peppo, A. Bianco, F. Nanni, A. Scafuri, F. Bertoldo, A. Salvati, S. Nardella, and L. Chiariello Performance of a Novel Sternal Synthesis Device After Median and Faulty Sternotomy: Mechanical Test and Early Clinical Experience Ann. Thorac. Surg., January 1, 2008; 85(1): 287 - 293. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Losanoff, M. D. Basson, S. A. Gruber, H. Huff, and F.-h. Hsieh Single Wire Versus Double Wire Loops for Median Sternotomy Closure: Experimental Biomechanical Study Using a Human Cadaveric Model Ann. Thorac. Surg., October 1, 2007; 84(4): 1288 - 1293. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Oto, R. Venkatachalam, Y. S. Morsi, S. Marasco, A. Pick, M. Rabinov, and F. Rosenfeldt A reinforced sternal wiring technique for transverse thoracosternotomy closure in bilateral lung transplantation: From biomechanical test to clinical application J. Thorac. Cardiovasc. Surg., July 1, 2007; 134(1): 218 - 224. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. El-Ansary, G. Waddington, and R. Adams Measurement of Non-Physiological Movement in Sternal Instability by Ultrasound Ann. Thorac. Surg., April 1, 2007; 83(4): 1513 - 1516. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Bhattacharya, I. Sau, M. Mohan, K. Hazari, R. Basu, and A. Kaul Sternal Bands for Closure of Midline Sternotomy Leads to Better Wound Healing Asian Cardiovasc Thorac Ann, February 1, 2007; 15(1): 59 - 63. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Pai, N. J. Gunja, E. L. Dupak, N. L. McMahon, T. P. Roth, J. F. Lalikos, R. M. Dunn, N. Francalancia, G. D. Pins, and K. L. Billiar In Vitro Comparison of Wire and Plate Fixation for Midline Sternotomies Ann. Thorac. Surg., September 1, 2005; 80(3): 962 - 968. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Bruhin, U. A. Stock, J.-P. Drucker, T. Azhari, J. Wippermann, J. M. Albes, D. Hintze, S. Eckardt, C. Konke, and T. Wahlers Numerical Simulation Techniques to Study the Structural Response of the Human Chest Following Median Sternotomy Ann. Thorac. Surg., August 1, 2005; 80(2): 623 - 630. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Molina, R. S.-L. Lew, and K. J. Hyland Postoperative sternal dehiscence in obese patients: Incidence and prevention Ann. Thorac. Surg., September 1, 2004; 78(3): 912 - 917. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. K. Dasika and J. A. Magovern Lower sternal reinforcement to improve median sternotomy closure: Reply Ann. Thorac. Surg., June 1, 2004; 77(6): 2261 - 2262. [Full Text] [PDF] |
||||
![]() |
R. Joshi, S. Abraham, and A. Sampath Kumar Interlocking Sternotomy: Initial Experience Asian Cardiovasc Thorac Ann, March 1, 2004; 12(1): 16 - 18. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Saito, A. Iguchi, M. Sakurai, and K. Tabayashi Biomechanical study of a Poly-L-Lactide (PLLA) sternal pin in sternal closure after cardiothoracic surgery Ann. Thorac. Surg., February 1, 2004; 77(2): 684 - 687. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E. Losanoff, A. D. Collier, C. C. Wagner-Mann, B. W. Richman, H. Huff, F.-h. Hsieh, A. Diaz-Arias, and J. W. Jones Biomechanical comparison of median sternotomy closures Ann. Thorac. Surg., January 1, 2004; 77(1): 203 - 209. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. E Losanoff, B. W Richman, and J. W Jones Bilateral Longitudinal Reinforcement of Median Sternotomy Closure Asian Cardiovasc Thorac Ann, December 1, 2003; 11(4): 378 - 378. [Full Text] [PDF] |
||||
![]() |
W. E. McGregor, M. Payne, D. R. Trumble, K. M. Farkas, and J. A. Magovern Improvement of sternal closure stability with reinforced steel wires Ann. Thorac. Surg., November 1, 2003; 76(5): 1631 - 1634. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. R. Casha and M. Gauci Mechanical comparison of three sternotomy closure techniques using a polyurethane foam sternal model Ann. Thorac. Surg., June 1, 2003; 75(6): 2009 - 2010. [Full Text] [PDF] |
||||
![]() |
U. K. Dasika, D. R. Trumble, and J. A. Magovern Lower sternal reinforcement improves the stability of sternal closure Ann. Thorac. Surg., May 1, 2003; 75(5): 1618 - 1621. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Glennie, D. E.T. Shepherd, and R. S. Jutley Strength of wired sternotomy closures: effect of number of wire twists Interactive CardioVascular and Thoracic Surgery, March 1, 2003; 2(1): 3 - 5. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. R. Trumble, W. E. McGregor, and J. A. Magovern Validation of a bone analog model for studies of sternal closure Ann. Thorac. Surg., September 1, 2002; 74(3): 739 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. J. Cohen and L. V. Griffin A biomechanical comparison of three sternotomy closure techniques Ann. Thorac. Surg., February 1, 2002; 73(2): 563 - 568. [Abstract] [Full Text] [PDF] |
||||
![]() |
R.S. Jutley, D.E.T. Shepherd, D.W.L. Hukins, and R.R. Jeffrey Calculating stress magnitude between sternotomy closures and sternum Eur. J. Cardiothorac. Surg., November 1, 2001; 20(5): 1071 - 1071. [Full Text] [PDF] |
||||
![]() |
A.R. Casha, M. Gauci, L. Yang, M. Saleh, P.H. Kay, and G.J. Cooper Fatigue testing median sternotomy closures Eur. J. Cardiothorac. Surg., March 1, 2001; 19(3): 249 - 253. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. R. Casha, L. Yang, and G. J. Cooper Measurement of chest wall forces on coughing with the use of human cadavers J. Thorac. Cardiovasc. Surg., December 1, 1999; 118(6): 1157 - 1158. [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||