J Thorac Cardiovasc Surg 1999;118:316-323
© 1999 Mosby, Inc.
SURGERY FOR ACQUIRED CARDIOVASCULAR DISEASE |
From the Heart Lung Institute, Utrecht University Hospital, Utrecht, The Netherlands.
Address for reprints: Paul F. Gründeman, MD, PhD, Experimental Cardiology Laboratory, Heart Lung Institute, Utrecht University Hospital (Rm G02.523), PO Box 85500, 3508 GA Utrecht, The Netherlands.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The objectives of study I were (1) to monitor changes in ventricular geometry, hemodynamics, and valvular function during vertical displacement of the beating porcine heart by the Octopus tissue stabilizer (Medtronic, Inc, Minneapolis, Minn)
5,7,8,10 and (2) to assess the modifying effect of the Trendelenburg maneuver (whole-body head-down positioning) on the changes in ventricular geometry, hemodynamics, and valve function.
In study II we investigated the effect on hemodynamics of mechanically supporting either the right or the left ventricle during vertical displacement of the beating porcine heart.
| Methods |
|---|
|
|
|---|
Anesthesia and instrumentation.
The animals were premedicated, anesthetized, and instrumented for hemodynamic measurements as described before.
8 To reduce the mechanical irritability of the heart, we gradually administered intravenous propranolol (range 15-25 mg) to obtain a heart rate between 50 and 70 beats/min. After median sternotomy, the heart was suspended in a pericardial cradle. A bipolar pacing lead was sutured on the right atrial appendix, whereafter the heart was paced at a fixed rate of 80 beats/min. An ultrasound transit time flow probe (Transonic Inc, Ithaca, NY, size 20 or 24 mm) was placed around the aorta for on-line measurement of the cardiac output. Stroke volume was calculated by dividing cardiac output by heart rate (fixed rate 80/min). For stepwise cardiac retraction,
8 two Octopus tissue stabilizers
5,10 were used, with an omniplane ultrasound probe (6.2. MHz, Sonos 2000, Hewlett-Packard, Amstelveen, The Netherlands) fixed in between (Fig 1). The ultrasound probe was remotely steered to the heart surface until it touched the immobilized epicardium for the on-line collection of 2-dimensional and M-mode echocardiographic images (Fig 2). During maneuvers, the ultrasound probe remained exactly positioned in between and in the same plane as the Octopus tentacles. Four 3F air-filled balloon-tipped catheters, which served as echographic beacons, were placed in the lumen of the right ventricle and marked precisely the RV long-axis plane (first pair: RV outflow tract and RV apical portion) or the RV short-axis plane (second pair: border of tricuspid valve and midportion of interventricular grove). Care was taken that the plane of the short-axis view of the left ventricle crossed the apices of both LV papillary muscles.
|
|
After completion of the experimental protocol of study I (see further), in 5 animals (study II) the right atrium and pulmonary artery trunk and the left atrium and left carotid artery were cannulated for right and left heart bypass (RHB and LHB, respectively), as described before.
11 A single centrifugal pump (BP 80, Medtronic Bio-Medicus, Eden Prairie, Minn) was connected to the withdrawal and return cannulas by 2 Y-connectors. The pump circuit was used as either RHB or LHB by instantaneously changing tubing clamps. A test run was performed at maximum flow to bypass the right ventricle and the left ventricle (bypass flow > 4 L/min) in the anatomic position. It was assured that the tip of the 38F LHB withdrawal cannula was suspended in the left atrial lumen by applying slight tension to the purse-string suture while the heart was retracted anteriorly. By palpation, the tip position was again verified.
Experimental protocol
Study I (n = 6).
Echocardiographic recordings were taken at each phase during at least 5 cardiac cycles and videotaped for analysis. In the same 2-dimensional view (probe at zero degrees), end-diastolic chamber dimensions (square millimeters) were obtained from the short-axis LV cross-sectional area and the long-axis RV cross-sectional area. In the M-mode view (position zero degrees), the distance between the LV posterior wall and RV free wall was measured in millimeters. After 90 degrees rotation of the ultrasound omniplane probe (probe in position 90 degrees), the RV short-axis cross-sectional area was measured. The LV long-axis view could not be consistantly obtained in the same 2-dimensional plane and was further not measured. In addition, Doppler echocardiographic images were made of the mitral and aortic valves after each maneuver. The function of the pulmonary and tricuspid valves was judged in 2 separate experiments with a hand-held ultrasound Doppler probe surfing over the epicardium using different echographic cross sections. Baseline cardiovascular values and echocardiographic measurements were recorded after fixation of the Octopus tentacles to the posterior wall after at least 15 minutes of pacing (phase 1, anatomic position). Subsequently, values were taken 3 minutes (phases 2-5) and 15 minutes (phases 3 and 5) after stabilization after each intervention.
The heart was vertically displaced by 90 degrees without delay with the Octopus tissue stabilizer (phase 2, first displacement) until the apex pointed upward. Circumflex branches on the back side of the heart became well exposed to the operative field
5,7,8 without the aid of additional supportive tools. Subsequently, after stabilization of hemodynamics, the operating table was tilted 20 degrees in the head-down position (Trendelenburg maneuver) without changing the position of the heart relative to the body (phase 3). After the Trendelenburg maneuver, the operating table was returned to the horizontal position while the heart remained retracted for 3 minutes (phase 4, second displacement). Subsequently, the heart was put back in the pericardial cradle (phase 5, anatomic position). The entire experimental protocol was repeated after a period of stabilization.
The first 3 consecutive phases in the experimental protocol are schematically depicted on the abcissa of Fig 3.
|
Statistical analysis.
First, data sets of 2 subsequently executed protocols from each animal were averaged and mean values were further used for analysis. Single video still frames at end-diastole and mid-systole were grabbed from original tapes and were off-line analyzed by an independent observer (C.M.M.). Data in Tables I and II are presented as mean ± standard deviation (SD) (absolute values). Hemodynamic variables and echocardiographic variables in the Results section and in Fig 3
are depicted as mean ± standard error of the mean (SEM) (percentage of basal values). Statistical analysis was performed by multivariate analysis of variance to assess the influence of displacement, and a paired Student t test was used to assess the modifying effect of the Trendelenburg maneuver or RHB/LHB compared with protocol control values.
|
|
| Results |
|---|
|
|
|---|
Study I. Echocardiography during displacement of the beating heart
(Table I
, Fig 2
, and Fig 3
, left panel ). Displacement of the beating heart (phase 2) caused the same degree of deterioration of the circulation as observed before.
7,8 RV end-diastolic pressure increased to 165% ± 19% of baseline values. In contrast, no significant change in LV end-diastolic pressure was observed (101% ± 9%). Echographically, a significant part of the RV free wall was pressed to the interventricular septum (Fig 2
, left panel, displaced ), whereas the RV outflow tract was somewhat narrowed but remained patent. Diastolic dimensions of the long-axis and short-axis RV cross-sectional areas decreased to 38% ± 6% and 40% ± 5%, respectively. Diastolic dimensions of the short-axis LV cross-sectional area decreased to 79% ± 3%, whereas the LV muscle was shaped elliptically. In M-mode, the distance between the LV posterior wall and RV free wall decreased by 17% ± 2%. Early diastolic LV septal motion was directed toward the RV free wall, thereby further decreasing RV dimension. In no case was incompetence of any valve observed during the entire cardiac cycle. Specifically, the posterior leaflet of the mitral valve did not show prolapse. The 2-dimensional geometry of the cardiac valves seemed grossly unaffected. The superior and inferior caval veins remained fully patent on displacement.
Tilting whole-body head-down 20-degree Trendelenburg position.
Three minutes into the Trendelenburg phase (phase 3), stroke volume had increased to 96% ± 8% and mean arterial pressure had normalized at the expense of increased ventricular preloads. Echocardiographically, a part of the RV free wall was released from the interventricular septum, and the RV outflow tract (in the short-axis view) increased its dimension. Diastolic dimensions of the long-axis and short-axis RV cross-sectional areas increased to 68% ± 8% and 57% ± 1%, respectively. Diastolic short-axis LV cross-sectional dimension increased to 96% ± 5%, whereas the left ventricle remained shaped elliptically. The distance between the LV posterior wall and RV free wall increased to 92% ± 2%. Again, in no case was dysfunction of any valve observed during the entire cardiac cycle after the preloads were increased.
Return of the operating table to the horizontal position.
After the operating table had been returned to the horizontal position (phase 4, the heart remained retracted 90 degrees), most values returned to pre-Trendelenburg levels.
Return of the beating heart to the anatomic position.
Shortly after return to the anatomic position (phase 5), the right ventricle was transiently dilated. Echocardiographically, RV diastolic dimensions were variably enlarged. Long-axis RV cross-sectional area increased from 1357 ± 262 mm2 to 1756 ± 273 mm2 (to 130% ± 6%, P = .001, vs baseline values). Central hemodynamic status improved rapidly and RV wall motion increased concomitantly. Diastolic RV dimension decreased to baseline values.
Study II. RHB or LHB in the retracted heart
(Table II
, Fig 3
, right panel ). The condition of the animal had somewhat deteriorated because of the previously executed protocol of study I, hemodilution, and subsequent cannulation. At 3 minutes displacement, RHB was started. Average RHB flow reached 100% of baseline cardiac output, which increased diastolic short-axis LV cross-sectional dimension, LV preload, and cardiac output to baseline values. In contrast, LHB increased mean arterial pressure to only 72% ± 11% (P = .035). Frequently, the LHB withdrawal cannula was pressed against the left atrial wall, which forced reduction of LHB flow. In 2 cases, air aspiration to the centrifugal pump occurred. Maximum LHB flow reached only 55% of baseline cardiac output.
| Discussion |
|---|
|
|
|---|
Second, the borderline circulatory status during displacement was markedly improved to almost baseline status by blood volume redistribution (Trendelenburg maneuver). At the same time, LV cross-sectional area, but not LV shape, returned to baseline.
Third, cardiac retraction did not cause valvular dysfunction.
Fourth, RHB effectively restored central hemodynamics in the tilted heart, whereas LHB only marginally improved circulatory status.
Epicardial 2-dimensional echocardiography of the beating heart.
The immobilized epicardial surface contained a distal circumflex branch, and its presentation by the Octopus tissue stabilizers simulated full exposure for coronary surgery without the use of additional supportive tools (Fig 1
).
In the 80-kg pig, 2-dimensional echocardiography of the beating heart proved technically feasible by fixing the ultrasound probe between two Octopus suction pods, which took hold of the posterior side of the heart (Fig 1
). The area immobilized by the retractor/ultrasound system comprized approximately a small matchbox area of cardiac tissue that does not affect cardiac function.
7 Interestingly, cardiac wall motion and systolic wall thickening between the tentacles was almost absent transmurally in the line of measurement of the echo probe. The latter phenomenon was observed immediately after attachment of the suction device. The function of the mitral valve and the aortic valve could be adequately assessed in all experiments by Doppler ultrasound in the LV long-axis/RV short-axis view.
Displacement of the beating heart.
The deterioration in circulation during displacement was characterized by severe reduction in stroke volume despite an elevated RV preload without a concomitant change of LV preload and with unimpaired valve function. Thus cardiac retraction causes biventricular pump failure. We hypothesized that the function of the thin-walled right ventricle was primarily affected. When the beating heart was retracted anteriorly, RV luminal dimension in two planes decreased by about 60%, and RV geometry was considerably distorted, evidenced by crumpling of the muscle. For a substantial portion, the thin-walled RV muscle was sqeezed between the left ventricle and the surrounding tissue, that is, the pericardial sac and right sternal border. A significant part of the RV free wall endocardium was pressed against the interventricular septum during the entire cardiac cycle. In diastole, the right ventricle did not expand sufficiently. A rim of RV tissue protruded over the interventricular groove, but that part of the ventricle in fact contributed well to RV output. The augmented RV preload corresponded to the absence of torque or compression of the caval veins and/or tricuspid valve dysfunction and/or pulmonary valve dysfunction. Apparently, the RV lumen available for diastolic expansion was reduced, which amounts to an acquired temporary RV hypoplasia .
In contrast to the severe changes in RV geometry, the changes in LV geometry were only moderate. The shape of the left ventricle changed from circular to elliptic, probably because of a thicker, less deformable muscle. LV function, however, also seemed impaired because stroke volume decreased despite unchanged LV preload and left atrial pressure. Load-independent indices of cardiac contractility were not assessed. Propranolol exerts negative inotropic effects on contractility and may worsen heart failure by displacement. Therefore a complete explanation of the changes of cardiac performance cannot be given.
RV support.
In study II, we tested the hypothesis that RV failure is predominant over left heart failure in a retracted heart. In all cases, selectively increasing the preload of the tilted left ventricle by deviating blood around the displaced right ventricle by means of a short-circuit RHB system fully restored LV stroke volume and, as a result, mean arterial pressure. RHB flow in this study reached baseline cardiac output. In contrast to healthy pigs, patients with poor LV function and dilated hearts may not tolerate augmentation of LV preload in deformed tilted hearts (mean LV end-diastolic pressure increased from 6 to 11 mm Hg). These patients may need a form of LV mechanical support as well.
LV support.
In contrast to RV support, LV support resulted in only marginal improvement of mean arterial pressure. Maximum LHB flow was comparable with cardiac output in the displaced heart without any form of circulatory support. Attemps to fully restore systemic circulation failed in all cases because blood supply originating from the right ventricle was apparently insufficient. The left atrial wall easily collapsed and yielded a risk of air aspiration (in 2 of 5 cases). We rule out technical problems related to cannulation because the correct position of the tip of the withdrawal cannula was verified by palpation. From these observations we infer that the severe drop in stroke volume during anterior displacement of the beating porcine heart is primarily caused by right heart pump failure.
Effect of 20-degree Trendelenburg positioning.
As reported earlier,
7,8 the borderline circulatory status during displacement markedly improved almost to baseline status by blood volume redistribution (Trendelenburg maneuver). When the subject was tilted head down, the compressed right ventricle was (in part) mechanically pushed open by further augmenting filling pressure, demonstrated by an improved but incomplete recovery of RV dimensions, while the position of the heart relative to the body remained unchanged. At the same time, LV filling pressure increased, LV dimensions returned to baseline, and stroke volume normalized.
In selected patients, the Trendelenburg maneuver effectively mitigates circulatory deterioration on exposure of the posterior wall.
5 In two thirds of the cases, however, inotropic support remained necessary.
3
Recent observations by us showed that right oblique positioning of the pig after sternotomy facilitated exposure of the posterior LV wall and also mitigated the sequelae of cardiac retraction on mean arterial pressure.
The beating heart in the anatomic position.
As before,
7,8 we observed transient firm distention of the right ventricle on quickly replacing the heart into its anatomic position. The dilatation disappeared within several minutes and the central hemodynamic status improved. Apparently, the right ventricle needs to adjust after more than 30 minutes of deformation and increased preload.
Limitations.
The pigs chest wall geometry differs from that of the human chest wall; that is, it is less barrel-shaped. Cardiac retraction was carried out in the normal healthy pig heart with its apex slightly pointing rightward. Patients with impaired LV function may not tolerate vertical displacement of the beating heart. Therefore inferences made from this experimental study may be applicable to human beings only after careful consideration.
| Conclusion |
|---|
|
|
|---|
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. Velissaris, A. T.M. Tang, P. J. Wood, D. A. Hett, and S. K. Ohri Thyroid function during coronary surgery with and without cardiopulmonary bypass Eur. J. Cardiothorac. Surg., July 1, 2009; 36(1): 148 - 154. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Caputo, R. R. Anis, C. A. Rogers, N. Ahmad, S. I.A. Rizvi, A. Baumbach, K. R. Karsch, G. D. Angelini, and M. Oberhoff Coronary Collateral Circulation: Effect on Early and Midterm Outcomes After Off-Pump Coronary Artery Bypass Surgery Ann. Thorac. Surg., January 1, 2008; 85(1): 71 - 79. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Dewey and M. J. Mack Myocardial Revascularization without Cardiopulmonary Bypass Card. Surg. Adult, January 1, 2008; 3(2008): 633 - 654. [Full Text] |
||||
![]() |
H. Ustunsoy, H. Kazaz, M. A Celkan, C. Kayiran, R. Hayta, and E. Bayar Deep Pericardial Suture vs Apical Suction for Off-Pump Bypass Grafting Asian Cardiovasc Thorac Ann, April 1, 2007; 15(2): 123 - 126. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Glauber, A. Farneti, S. Bevilacqua, and J. Karimov Pump-assisted beating heart surgery MMCTS, February 19, 2007; 2007(0219): 943. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Fiore, N. Brienza, P. Cicala, P. Tunzi, N. Marraudino, L. d. L. T. Schinosa, and T. Fiore Superior mesenteric artery blood flow modifications during off-pump coronary surgery. Ann. Thorac. Surg., July 1, 2006; 82(1): 62 - 67. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Michaux, M. Filipovic, K. Skarvan, S. Schneiter, R. Schumann, H.-R. Zerkowski, F. Bernet, and M. D. Seeberger Effects of on-pump versus off-pump coronary artery bypass graft surgery on right ventricular function J. Thorac. Cardiovasc. Surg., June 1, 2006; 131(6): 1281 - 1288. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G. Raja and B. S. Rayen Levosimendan in Cardiac Surgery: Current Best Available Evidence Ann. Thorac. Surg., April 1, 2006; 81(4): 1536 - 1546. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. P. Wirtz, C. Schmidt, H. Van Aken, G. Brodner, D. Hammel, H. H. Scheld, and E. Berendes Temporary Right Heart Support With Percutaneous Jugular Access Ann. Thorac. Surg., February 1, 2006; 81(2): 701 - 705. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P. Alston Pumphead--or not! Does avoiding cardiopulmonary bypass for coronary artery bypass surgery result in less brain damage? Br. J. Anaesth., June 1, 2005; 94(6): 699 - 701. [Full Text] [PDF] |
||||
![]() |
J. C. Diephuis, K. G. M. Moons, A. N. Nierich, M. Bruens, D. van Dijk, and C. J. Kalkman Jugular bulb desaturation during coronary artery surgery: a comparison of off-pump and on-pump procedures Br. J. Anaesth., June 1, 2005; 94(6): 715 - 720. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Fiore, M. E. Latrofa, P. Tunzi, M. Traversa, C. Fondacone, N. Marraudino, L. de Luca Tupputi Schinosa, and T. Fiore Hemodynamics in off-pump surgery: normal versus compromised preoperative left ventricular function Eur. J. Cardiothorac. Surg., March 1, 2005; 27(3): 488 - 493. [Abstract] [Full Text] [PDF] |
||||
![]() |
W.-I. Chang, K.-B. Kim, J. H. Kim, B. M. Ham, and Y. L. Kim Hemodynamic Changes During Posterior Vessel Off-Pump Coronary Artery Bypass: Comparison Between Deep Pericardial Sutures and Vacuum-Assisted Apical Suction Device Ann. Thorac. Surg., December 1, 2004; 78(6): 2057 - 2062. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Velissaris, A. T. M. Tang, M. Murray, R. L. Mehta, P. J. Wood, D. A. Hett, and S. K. Ohri A prospective randomized study to evaluate stress response during beating-heart and conventional coronary revascularization Ann. Thorac. Surg., August 1, 2004; 78(2): 506 - 512. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. F. Grundeman, C. W. J. Verlaan, W. J. van Boven, and C. Borst Ninety-degree anterior cardiac displacement in off-pump coronary artery bypass grafting: the Starfish cardiac positioner preserves stroke volume and arterial pressure Ann. Thorac. Surg., August 1, 2004; 78(2): 679 - 684. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. K. Shinn, Y. J. Oh, S. H. Kim, J. H. Lee, C. S. Lee, and Y. L. Kwak Evaluation of serial haemodynamic changes during coronary artery anastomoses in patients undergoing off-pump coronary artery bypass graft surgery: initial experiences using two deep pericardial stay sutures and octopus tissue stabilizer Eur. J. Cardiothorac. Surg., June 1, 2004; 25(6): 978 - 984. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Suzuki, M. Okabe, M. Handa, F. Yasuda, and Y. Miyake Usefulness of preoperative intraaortic balloon pump therapy during off-pump coronary artery bypass grafting in high-risk patients Ann. Thorac. Surg., June 1, 2004; 77(6): 2056 - 2059. [Abstract] [Full Text] [PDF] |
||||
![]() |
P.-G. Chassot, P. van der Linden, M. Zaugg, X. M. Mueller, and D. R. Spahn Off-pump coronary artery bypass surgery: physiology and anaesthetic management{dagger} Br. J. Anaesth., March 1, 2004; 92(3): 400 - 413. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Decoene, T. Modine, S. Al-Ruzzeh, T. Athanasiou, D. Fawzi, R. Azzaoui, A. Pol, and G. Fayad Analysis of thoracic aortic blood flow during off-pump coronary artery bypass surgery Eur. J. Cardiothorac. Surg., January 1, 2004; 25(1): 26 - 34. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Hughes, J.M. Hasenkam, I.K. Severinsen, and D.A. Steinbruchel Right heart assist for beating heart coronary artery bypass grafting Eur. J. Cardiothorac. Surg., November 1, 2003; 24(5): 762 - 769. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-H. Kim, B. M. Ham, Y. L. Kim, J.-H. Bahk, H.-G. Ryu, Y.-S. Jeon, and K.-B. Kim Prophylactic milrinone during OPCAB of posterior vessels: implication in angina patients taking {beta}-blockers Eur. J. Cardiothorac. Surg., November 1, 2003; 24(5): 770 - 776. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. F. Grundeman, R. Budde, H. M. Beck, W.-J. van Boven, and C. Borst Endoscopic Exposure and Stabilization of Posterior and Inferior Branches Using the Endo-Starfish Cardiac Positioner and the Endo-Octopus Stabilizer for Closed-Chest Beating Heart Multivessel CABG: Hemodynamic Changes in the Pig Circulation, September 9, 2003; 108(90101): II-34 - 38. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Perek, M. Jemielity, J. Tomczyk, E. Camacho, and W. Dyszkiewicz Deep Pericardial Stitch Enables Hemodynamically Stable Exposure of Beating Heart Asian Cardiovasc Thorac Ann, September 1, 2003; 11(3): 203 - 207. [Abstract] [Full Text] |
||||
![]() |
T. Velissaris, A. Tang, M. Murray, A. El-Minshawy, D. Hett, and S. Ohri A prospective randomized study to evaluate splanchnic hypoxia during beating-heart and conventional coronary revascularization Eur. J. Cardiothorac. Surg., June 1, 2003; 23(6): 917 - 924. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Velissaris, R. G Stuklis, D. A Hett, and S. K Ohri Right Pleuropericardial Release: a Useful Technique in Off-Pump Coronary Surgery Asian Cardiovasc Thorac Ann, June 1, 2003; 11(2): 174 - 176. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. A. Cooper, J. S. Corvera, V. H. Thourani, J. D. Puskas, J. M. Craver, O. M. Lattouf, and R. A. Guyton Perfusion-assisted direct coronary artery bypass provides early reperfusion of ischemic myocardium and facilitates complete revascularization Ann. Thorac. Surg., April 1, 2003; 75(4): 1132 - 1139. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Hart Maintaining hemodynamic stability and myocardial performance during off-pump coronary bypass surgery Ann. Thorac. Surg., February 1, 2003; 75(2): S740 - 744. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. M. Dewey and M. J. Mack Myocardial Revascularization Without Cardiopulmonary Bypass Card. Surg. Adult, January 1, 2003; 2(2003): 609 - 625. [Full Text] |
||||
![]() |
M. Caputo, M. Yeatman, P. Narayan, G. Marchetto, R. Ascione, B. C. Reeves, and G. D. Angelini Effect of off-pump coronary surgery with right ventricular assist device on organ function and inflammatory response: a randomized controlled trial Ann. Thorac. Surg., December 1, 2002; 74(6): 2088 - 2095. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Velissaris, A. Tang, M. Jonas, and S. Ohri Haemodynamic changes during off-pump surgery Eur. J. Cardiothorac. Surg., November 1, 2002; 22(5): 852 - 852. [Full Text] [PDF] |
||||
![]() |
L. Torracca, J. J. Schreuder, A. Quarti, G. Ismeno, V. Franze, and O. Alfieri Acute effects of beating heart coronary surgery on left ventricular performance Ann. Thorac. Surg., October 1, 2002; 74(4): S1348 - 1352. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. M. Mueller, P.-G. Chassot, J. Zhou, K. M. Eisa, C. Chappuis, H. T. Tevaearai, and L. K. von Segesser Hemodynamics optimization during off-pump coronary artery bypass: the 'no compression' technique Eur. J. Cardiothorac. Surg., August 1, 2002; 22(2): 249 - 254. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Meyns, R. Autschbach, A. Boning, W. Konertz, K. Matschke, F. Schondube, K. Wiebe, and E. Fischer Coronary artery bypass grafting supported with intracardiac microaxial pumps versus normothermic cardiopulmonary bypass: a prospective randomized trial Eur. J. Cardiothorac. Surg., July 1, 2002; 22(1): 112 - 117. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. M Mueller and L. K von Segesser Circulatory support for OPCAB procedures Perfusion, July 1, 2002; 17(4): 305 - 312. [Abstract] [PDF] |
||||
![]() |
G. Matsumiya, S. Ohtake, K. Kagisaki, T. Takahashi, Y. Sawa, and H. Matsuda Right Heart Assist During Beating Bypass for Severe Left Ventricular Dysfunction Asian Cardiovasc Thorac Ann, June 1, 2002; 10(2): 155 - 158. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Yeatman, M. Caputo, P. Narayan, A. Kumar Ghosh, R. Ascione, I. Ryder, and G. D. Angelini Intracoronary shunts reduce transient intraoperative myocardial dysfunction during off-pump coronary operations Ann. Thorac. Surg., May 1, 2002; 73(5): 1411 - 1417. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. George, S. Al-Ruzzeh, and M. Amrani Mitral annulus distortion during beating heart surgery: a potential cause for hemodynamic disturbance--a three-dimensional echocardiography reconstruction study Ann. Thorac. Surg., May 1, 2002; 73(5): 1424 - 1430. [Abstract] [Full Text] [PDF] |
||||
![]() |
Q.-B. Do, C. Goyer, O. Chavanon, P. Couture, A. Denault, and R. Cartier Hemodynamic changes during off-pump CABG surgery Eur. J. Cardiothorac. Surg., March 1, 2002; 21(3): 385 - 390. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. M Mueller, M. Augstburger, Y. Boone, and L. K von Segesser Hemocompatibility of a coaxial pump catheter for less invasive heart surgery Perfusion, January 1, 2002; 17(1): 3 - 7. [Abstract] [PDF] |
||||
![]() |
S. Muraki, C. D. Morris, J. M. Budde, R. N. Otto, Z.-Q. Zhao, J. D. Puskas, R. A. Guyton, and J. Vinten-Johansen Preserved myocardial blood flow and oxygen supply-demand balance with active coronary perfusion during simulated off-pump coronary artery bypass grafting J. Thorac. Cardiovasc. Surg., January 1, 2002; 123(1): 53 - 62. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. F. Grundeman, C. Borst, C. W.J. Verlaan, S. Damen, and S. Mertens Hemodynamic changes with right lateral decubitus body positioning in the tilted porcine heart Ann. Thorac. Surg., December 1, 2001; 72(6): 1991 - 1996. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. van Dijk, A. P. Nierich, E. W.L. Jansen, H. M. Nathoe, W. J.L. Suyker, J. C. Diephuis, W.-J. van Boven, C. Borst, E. Buskens, D. E. Grobbee, et al. Early Outcome After Off-Pump Versus On-Pump Coronary Bypass Surgery: Results From a Randomized Study Circulation, October 9, 2001; 104(15): 1761 - 1766. [Abstract] [Full Text] [PDF] |
||||
![]() |
K.-B. Kim, C. Lim, H. Ahn, and J.-K. Yang Intraaortic balloon pump therapy facilitates posterior vessel off-pump coronary artery bypass grafting in high-risk patients Ann. Thorac. Surg., June 1, 2001; 71(6): 1964 - 1968. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P.R. Watters, R. Ascione, I. G. Ryder, F. Ciulli, A. A. Pitsis, and G. D. Angelini Haemodynamic changes during beating heart coronary surgery with the 'Bristol Technique' Eur. J. Cardiothorac. Surg., January 1, 2001; 19(1): 34 - 40. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M Toomasian and W. Aboul-Hosn Coronary artery bypass grafting using a miniature right ventricular support system Perfusion, December 1, 2000; 15(6): 521 - 526. [Abstract] [PDF] |
||||
![]() |
M. Mathison, J. R. Edgerton, J. L. Horswell, J. J. Akin, and M. J. Mack Analysis of hemodynamic changes during beating heart surgical procedures Ann. Thorac. Surg., October 1, 2000; 70(4): 1355 - 1360. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Suematsu, T. Ohtsuka, K. Miyaji, A. Murakami, T. Miyairi, Z. Eyileten, Y. Kotsuka, and S. Takamoto Right heart mini-pump bypass for coronary artery bypass grafting: experimental study Eur. J. Cardiothorac. Surg., September 1, 2000; 18(3): 276 - 281. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Hart, T. H. Spooner, J. Pym, T. F. Flavin, J. R. Edgerton, M. J. Mack, and E. W.L. Jansen A review of 1,582 consecutive Octopus off-pump coronary bypass patients Ann. Thorac. Surg., September 1, 2000; 70(3): 1017 - 1020. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Mathison, E. Buffolo, A. D. Jatene, F. B. Jatene, H. Reichenspurner, R. G. Matheny, H. Shennib, J. J. Akin, and M. J. Mack Right heart circulatory support facilitates coronary artery bypass without cardiopulmonary bypass Ann. Thorac. Surg., September 1, 2000; 70(3): 1083 - 1085. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. P. Nierich, J. Diephuis, E. W.L. Jansen, C. Borst, and J. T.A. Knape Heart displacement during off-pump CABG: how well is it tolerated? Ann. Thorac. Surg., August 1, 2000; 70(2): 466 - 472. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. R. Kshettry, T. F. Flavin, R. W. Emery, D. M. Nicoloff, K. V. Arom, and R. J. Petersen Does multivessel, off-pump coronary artery bypass reduce postoperative morbidity? Ann. Thorac. Surg., June 1, 2000; 69(6): 1725 - 1730. [Abstract] [Full Text] [PDF] |
||||
![]() |