J Thorac Cardiovasc Surg 1999;118:946-952
© 1999 Mosby, Inc.
CARDIOPULMONARY SUPPORT AND PHYSIOLOGY |
From the Laboratory for Physiology, Institute for Cardiovascular Research Vrije Universiteit (ICaR-VU), Amsterdam, The Netherlands.
Address for reprints: Piet Borgdorff, PhD, Laboratory for Physiology, Vrije Universiteit, Van der Boechorststraat 7, 1081 BT Amsterdam, The Netherlands (E-mail:p.borgdorff.physiol{at}med.vu.nl).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
One of these causes could be the use of roller or centrifugal pumps, indispensable in many extracorporeal circuits. Pumps may indirectly activate platelets by causing lysis of red blood cells, with subsequent release of adenosine diphosphate (ADP). Pumps have therefore continuously been improved to reduce hemolysis. Nevertheless, the problem of platelet activation in extracorporeal systems has not been resolved.
Little attention has been paid to the possibility that intermittent increases in shear stress caused by the pumping can activate platelets more directly. Effects of shear stress on platelet behavior have mainly been studied in vitro with viscometers
18-21 but are demonstrable in vivo as well. In a previous study on rats with an albumin-coated extracorporeal tube running from the carotid artery to the distal femoral artery we continuously measured aggregate formation and found that platelets easily aggregate in areas of increased shear stress caused by partial occlusion of the tube.
22
In this study we therefore tested the hypothesis that in a coated extracorporeal system the roller pump also evokes platelet aggregation. Albumin was employed as the coating to avoid possible direct effects of surface-bound heparin on platelets. Albumin coating has been reported to be as effective as heparin coating in preventing platelet aggregation and activation as a result of blood-material contact.
2,9,10,23 To elucidate the role of shear stress we studied whether pump-induced platelet aggregation could be prevented with aurintricarboxylic acid (ATA), a synthetic triphenylmethane dye that specifically prevents shear-induced platelet aggregation without affecting aggregation induced by ADP, collagen, epinephrine, thrombin, or arachidonic acid.
24,25 Experiments were done with human blood in vitro and with rat blood in vitro and in vivo.
| Methods |
|---|
|
|
|---|
Rat blood.
Male Wistar rats (325-400 g) were anesthetized with an intramuscular bolus of ketamine (10%, 60 mg/kg, Kombivet) and subsequent intraperitoneal administration of pentobarbital (Nembutal, 25 mg/kg). After cannulation of a common carotid artery, the animals received heparin (800 IU/kg) and 10 to 12 mL blood was slowly collected in a syringe. Values (mean ± SEM) for pH, PCO2, and PO2 were 7.37 ± 0.01, 42.5 ± 1.5 mm Hg, and 74.4 ± 3.4 mm Hg, respectively.
Setup for in vitro experiments.
To study the influence of the pump on platelet behavior we used a device in which blood flow was generated either by gravity or by a pump(Fig 1). This setup consisted of 2 syringes positioned at different heights and connected by medical-grade polyvinyl chloride tubing (1.5 mm inner diameter, 100 cm long). Halfway along this tube a piece of pump tube (1.65 mm inner diameter, 7 cm long; Gilson, Inc, Middleton, Wis) was inserted and loosely placed in a small noncommercial roller pump with rotator and roller diameters of 35 and 6 mm, respectively. At 10 cm from the pump the polyvinyl tube was serially connected to a glass capillary tube (0.6 mm inner diameter) on the stage of a microscope for continuous measurement of platelet aggregation, as has been described previously.
22 The system was filled with 7 to 8 mL blood. The difference in height between the 2 syringes (30 cm) caused a flow of approximately 6 mL/min that lasted about 1 minute; at that point the nearly empty upper syringe was automatically lowered with an electromotor and the full syringe was elevated to continue flow. In this way there was no need to decant the blood and contact of blood with air was minimized. To replace gravitational flow by pump flow a calibrated screw on the pump was tightened to set the pump so that it was just occlusive for blood. Then a tube end was detached and placed into the other reservoir and the pump was started. The whole system was either coated with albumin or not coated and was placed in a polymethyl methacrylate (Perspex; ICI Acrylics, Inc, Nederland, Tex) box with the temperature regulated at 32°C ± 1°C. The method of coating has been described previously.
26 Tests with radioactively labeled albumin showed that the coating consisted of 1 to 2 monolayers and was stable during at least 4 hours of blood perfusion.
|
After cannulation of the central part of a common carotid artery (20 gauge) and the distal part of the left femoral artery (22 gauge) the cannulas were connected with medical-grade polyvinyl chloride tubing (1.5 mm inner diameter, 30 cm long) filled with heparinized saline solution. During the connection procedure flow to the leg was interrupted for 2 to 3 minutes only. Part of the tube was loosely positioned in the roller pump. If the tube in the pump was not compressed, blood flowed spontaneously (2-4 mL/min) as a result of the difference between central carotid arterial and distal femoral arterial pressures. For transition to pump flow the tube was gradually compressed until flow stopped, and the pump was then started. Platelet aggregation in the tube between the pump and the femoral artery was continuously measured with the same photometric device used in the in vitro experiments. The whole system was coated with albumin. Aortic pressure was measured through a stainless steel T-piece inserted in the tube about 3 cm from the carotid cannula. Flow in the tubing was measured with a perivascular flow probe. A warming lamp above the extracorporeal circuit prevented the blood from cooling more than about 1°C.
All animal handling was in compliance with the "Guide for the Care and Use of Laboratory Animals" prepared by the Institute of Laboratory Animal Resources, National Research Council, and published by the National Academy Press, revised 1996.
Measurement of platelet behavior.
The device for continuous detection of platelet aggregates has been described previously.
22 Passage of aggregates through the glass capillary tube was indicated by short-duration voltage peaks from photocells. For quantification the signals were converted to uniform spikes with a spike processor and subsequently counted during fixed periods. In the in vivo experiments aggregate formation was related to the value found after injection of 2 ng ADP into the tube a few centimeters from the carotid cannula. The 2 ng ADP was dissolved in 20 µL saline solution and injected within half a second. This dose elicited strong platelet aggregation that just barely did not impede flow. Periodic inspection of blood samples under a microscope confirmed the results obtained with the continuous aggregometer; it also showed that aggregates seldom contained any leukocytes and that aggregates of fewer than 4 platelets were not detected.
For platelet counting and visual inspection of aggregates 0.1 mL blood was withdrawn from a syringe in the in vitro experiments and from the T-piece in the extracorporeal shunt in vivo. The blood was added to 150 µL of a solution of ethylenediaminetetraacetic acid (3.6 mg/mL saline solution), resulting in a final concentration of 6.6 mmol/L, to stabilize the number of platelets and aggregates. From the ethylenediaminetetraacetic aciddiluted sample, 20 µL was added to 2 mL Thromboplus (Sarstedt, Nürembrecht, Germany). Aggregates and platelets were counted in a Bürker counting chamber under phase contrast microscopy (x400). Platelets were also counted with a Coulter Counter (model ZF; Coulter Corporation, Hialeah, Fla). Hemolysis was quantified by colorimetric determination of free hemoglobin in plasma at 600 nm (procedure No. 527; Sigma Chemical Co, St Louis, Mo).
Aurintricarboxylic acid.
A trisodium salt of ATA (Aldrich Chemical Company, Bornem, Belgium) was dissolved (30 mg/mL) in phosphate buffer (pH 7.4). In the in vivo experiments 35 mg/kg was infused intravenously within 10 minutes in rats that had not been earlier exposed to pump perfusion. In a previous study
22 this dose had been found to effectively inhibit platelet aggregation in rats in which a mean shear stress of 231 dyne/cm2 was generated by partial occlusion of a tube between the carotid artery and the femoral artery. In the in vitro experiments final concentrations of 1 mg/mL in rat blood and 100 µg/mL in human blood
24,25 were used.
Statistics.
All values are expressed as mean ± SEM. Data were analyzed by 2-way analysis of variance. If a significant P value was present posttests for comparison between conditions (pump or no pump) at different times were performed with the Bonferroni multiple comparison procedure.
| Results |
|---|
|
|
|---|
|
|
|
Platelet counts in vitro.
During gravitational flow in coated tubes platelet count showed a small and gradual decrease during a period of 60 minutes(Fig 5) and no aggregation was observed. By contrast, when the pump was started after 10 minutes of gravitational flow(Fig 5
), platelet count significantly (P = .0001) decreased within 5 minutes in both human and rat blood. A further gradual decrease, similar to that noticed with gravitational flow, was seen thereafter. With ATA (n = 3) platelet counts after 60 minutes of pumping had diminished less than without ATA (78% ± 3% in human blood and 72% ± 4% in rat blood).
|
As appeared from platelet counts in blood that was kept 1 hour in Eppendorff vials, human platelets (loss to 78% ± 4% of control value, n = 11) seemed to be more vulnerable than rat platelets (loss to 99% ± 9% of control value, n = 22). Initial platelet counts for rat blood were more than 5 times the counts in human blood (1039 ± 65 x 109 cells/L vs 187 ± 17 x 109 cells/L).
Conditions for pump-induced platelet aggregation.
The pump-induced aggregation in vitro could only be observed when blood was handled with great care during the preparation procedure. It was often absent after contact of the blood with air or uncoated material, use of too much heparin, blood loss during cannulation, or vacuum sucking during blood collection. All these conditions caused platelet aggregation and platelet loss, as judged by microscopic inspection of the blood immediately after collection. Moreover, experiments with human blood (n = 5) anticoagulated with 3.8% citrate, which does not inhibit agonist-induced platelet aggregation, showed that pump-induced platelet aggregation was absent after even partial calcium chelation.
The importance of the occlusion setting of the pump was tested in vitro with rat blood (n = 3) by intentionally increasing its compression onto the tube. This was done at 20 minutes after the start of pump perfusion, when the initial pump-induced aggregation had leveled off. Elevated compression caused a gradual and continuous increase of aggregation that was clearly distinct from the immediate and temporary aggregation induced with the pump set to be just barely occlusive as in the other experiments. This effect was probably caused by hemolysis because in the experiments with increased compression the mean increase in plasma free hemoglobin levels after 60 minutes of pumping was 121 ± 32 mg/dL, whereas this value was 26 ± 13 mg/dL in the in vitro experiments in which the pump was set to be just barely occlusive (n = 9). In the in vivo experiments the increase in plasma free hemoglobin level after 2 hours of pump perfusion was less than 1.32 ± 2.5 mg/dL (n = 5).
| Discussion |
|---|
|
|
|---|
Contact of blood with our uncoated tubes caused immediate strong platelet aggregation and platelet loss, confirming that this occurrence is a "first pass" effect.
17 It may even be stronger in the uncoated capillary tubes of clinically used filters, oxygenators, and artificial kidneys. The last, for example, have a surface to blood volume ratio of 163 cm2/mL, whereas in our system corresponding values were 26 cm2/mL for the in vitro experiments and 12 cm2/mL for the in vivo experiments.
Our finding that the pump-induced aggregation could be prevented with ATA suggests that elevated shear stress plays a prominent role because ATA specifically inhibits shear-induced platelet aggregation by binding to the von Willebrand factor but does not suppress the reactions of platelets to other agonists.
24,25 The ATA dose used in this study has been shown to inhibit platelet aggregation caused by a shear stress of 231 dyne/cm2, which resulted from partial tube occlusion. In addition to shear stress pumping may have caused the release of platelet-activating substances from red blood cells as a result of hemolysis. Although this is a real possibility in clinical cardiopulmonary bypass, in which the increase in plasma free hemoglobin levels usually varies from 26 to 82 mg/dL,
27,28 it seems a less likely cause of platelet aggregation in our experiments. The increase in plasma free hemoglobin in our in vivo experiments was less than 1.32 mg/dL, indicating that hemolysis was limited. Even in our in vitro setup, in which erythrocytes probably were more vulnerable, free plasma hemoglobin concentration only increased to values in the lower range of those seen in clinical practice. Moreover, the time course of platelet aggregation caused by hemolysis could be distinguished from that related to elevated shear stress in our experiments. When we intentionally augmented hemolysis by tightening the rollers of the pump platelet aggregation increased slowly and steadily, whereas aggregation caused by pumping at the proper setting was a short lasting phenomenon.
Pump-induced aggregation could be prevented with ATA, which also attenuated the loss of platelets. Whether ATA in its commercially available form can be employed in human beings is not known, although several authors suggest its potential usefulness
21 and no toxic effects have been found in dogs.
25 Nevertheless, it might be safer to develop substances that are more specifically suited for human use. Like ATA, these substances should reversibly inhibit the binding of von Willebrand factor to platelet glycoprotein Ib receptors. One could think of recombinant fragments of von Willebrand factorlike VCL (leucine504-lysine728), a peptide which binds to glycoprotein Ib receptors in human blood.
29 In an experiment on a rat in vivo we used VCL (gift of Professor J. J. Sixma, University Hospital, Utrecht, The Netherlands). In the final concentration used (about 2 µmol/L) VCL largely prevented platelet aggregation during pumping in vivo.
Our results show that the use of a roller pump in extracorporeal circuits causes platelet aggregation. This seems to be the result of elevated shear stress in the blood within the pump tube because the platelet aggregation can be prevented with ATA. Our results also indicate that the effects of pumping on platelets are masked in uncoated circuits because of the dominant influence of blood-material contact.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P. Borgdorff and G. J. Tangelder Pump-induced platelet aggregation with subsequent hypotension: Its mechanism and prevention with clopidogrel J. Thorac. Cardiovasc. Surg., April 1, 2006; 131(4): 813 - 821. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Kanemitsu, M. Nishikawa, K. Onoda, T. Shimono, H. Shimpo, A. Yazaki, K. Tanaka, H. Shiku, and I. Yada Pharmacologic platelet anesthesia by glycoprotein IIb/IIIa complex antagonist and argatroban during in vitro extracorporeal circulation J. Thorac. Cardiovasc. Surg., August 1, 2003; 126(2): 428 - 435. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Borgdorff, D. Fekkes, and G. J. Tangelder Hypotension Caused by Extracorporeal Circulation: Serotonin From Pump-Activated Platelets Triggers Nitric Oxide Release Circulation, November 12, 2002; 106(20): 2588 - 2593. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Borgdorff, G. van den Bos, and G. J. Tangelder Extracorporeal circulation can induce hypotension by both blood-material contact and pump-induced platelet aggregation J. Thorac. Cardiovasc. Surg., July 1, 2000; 120(1): 12 - 19. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||