J Thorac Cardiovasc Surg 2000;120:1053-1063
© 2000 The American Association for Thoracic Surgery
Surgery for Congenital Heart Disease |
From the Departments of Physiologya and Engineering Science,c University of Auckland, Auckland, New Zealand; Department of Surgery,b University of Alabama at Birmingham, Birmingham, Ala; and the Department of Cardiology,d The Prince Charles Hospital, Brisbane, Australia.
Supported by a grant from the National Heart Foundation of Australia (grant No. G89B2741).
Address for reprints: David McGiffin, MD, Department of Surgery, Division of Cardiothoracic Surgery, The University of Alabama at Birmingham, 780 Lyons-Harrison Research Building, 1919 Seventh Ave South, Birmingham, AL 35294-0007.
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Reasons advanced for the development of aneurysms include the disparate elastic properties of the aorta and synthetic patch
2 and resection of the posterior intimal shelf.
9 On the other hand, it has been argued that fragmentation or loss of elastin layers in the peripatch region may reflect underlying tissue abnormality associated with the initial coarctation of the aorta rather than the repair.
4
A study of synthetic patch repair of coarctation by McGiffin and colleagues
18 used finite element analysis to examine aortic wall stress at the site of the repair. Aortic wall stress at the site of repair was shown to increase out of proportion to the increase in diameter of the aorta because of placement of the synthetic patch. Although this study was not designed to relate increases in aortic wall stress to subsequent aneurysm formation, the abnormally high aortic wall stress predicted opposite the patch suggests that the aorta may be operating close to its strain limit in this region. This would predispose the aorta to structural change that would likely result in elastin fiber fracturing and degeneration.
Important limitations of this analysis are that nonlinear effects were not considered in a full three-dimensional (3-D) analysis and that it did not yield information on the regional distribution of strain in the aorta. There is an evident requirement for a carefully controlled experimental study of the biomechanical effects of synthetic patch repair of aortic coarctation by using human tissue. It is important to note that whereas strain and deformation can be estimated in soft tissue, stress cannot be reliably measured.
This study was undertaken to characterize the distribution of strain in the aortic wall adjacent to a synthetic patch to test the hypothesis that patch repair of coarctation leads to substantial increases on aortic strain (and stress).
| Materials and methods |
|---|
|
|
|---|
Tissue procurement
Measurements were made in segments of human descending thoracic aorta. The tissue was obtained at the time of autopsy of trauma victims at the Queensland Government Institute of Forensic Pathology, Brisbane, Australia. Permission was obtained from relatives of the victims for removal of tissue as part of a general consent for use of allograft valves and tissue for scientific investigation. Ethical approval for the project was granted by the Ethics Committee of the Prince Charles Hospital, Brisbane, Australia. After procurement, the segments of descending thoracic aorta were trimmed of areolar tissue, packaged in nutrient medium (Medium 199; Commonwealth Serum Laboratories, Parkville, Victoria, Australia) and frozen at 0°C. It has previously been demonstrated that storage of aortic tissue by freezing in saline solution does not alter the material properties of the tissue.
19 Aorta was procured from 7 victims of trauma (4 male and 3 female), ages ranging from 9 to 25 years.
Tissue preparation
Frozen segments of aorta were removed from the packaging and thawed. An inflatable internal sleeve was formed by drawing a thin latex tube through the aorta and securing the aorta and lining at each end to a cannula(Fig 1). Circumferential columns of fine India ink dots (2-3 mm apart) were then marked on the adventitia of the aorta. A 7-0 polypropylene suture was then sutured to the adventitia so that it lay longitudinally on the posterior aspects of the aorta. This was used as a landmark during marker coordinate reconstruction.
|
|
Experimental protocol
The aorta was prestressed slightly to minimize buckling during inflation. A pressure of 60 mm Hg (8.16 kPa) was applied and the axial position of the upper cannula was adjusted until the aorta was cylindrical. The aorta was then subjected to 10 cycles of volume infusion and withdrawal at a fixed rate over a pressure range 10 to 120 mm Hg (1.36-16.32 kPa). The cycle duration (1-2 minutes) was fixed for each experiment. Pressure, camera positions, and biplane video images were acquired continuously.
When this control sequence was complete, the aorta and cannulas were removed from the recording system. A longitudinal incision was made in the anterior wall of the aorta, and an oval patch of Dacron graft was sutured to the aortotomy. Synthetic patches were fashioned with different dimensions to produce degrees of patch ballooning of the extent typically seen in surgical practice. Synthetic patch length ranged from 11.2 mm to 27.4 mm and patch width ranged from 8.5 mm to 20.4 mm. All grafts were sutured with 4-0 polypropylene. The aorta was then remounted in the recording system and the full sequence used for control testing was repeated.
On completion of the experiment, the aorta was removed and filled with silicone rubber molding material (Dow Corning RTV; Dow Corning, Midland, Mich) for subsequent determination of geometry. Dimensions were obtained for each specimen and key measures are presented inTable I.
|
Marker positions were reconstructed for the infusion phase of the ninth or tenth loading cycle. Between 27 and 51 points were tracked for the samples at pressures that were closest to 10, 20, 40, 60, 80, 100, and 120 mm Hg (1.36, 2.72, 5.44, 8.16, 10.88, 13.6, and 16.32 kPa).
Lateral boundary coordinates at these pressures were estimated at 2 positions below and/or above the patch and at 6 to 8 vertical positions for the control sequence. The coordinates of points along the edge of the patch were also reconstructed at the reference pressure (close to 10 mm Hg; 1.36 kPa).
The relationship of the 3-D coordinates of any marker point to its screen coordinates in biplane camera views depends on the distortion introduced by the lenses, perspective effects due to displacement of the point from the cameras, the angle of intersection of camera axes, and the distance of the cameras from the intersection point. Distortion introduced by the macro lenses in our system occurs at the periphery of the image
20 and can be neglected for the fields used in this study. Because biplane angle and camera position are known, it is necessary only to identify the linear perspective transformations. To do this, a calibration grid (200 x 300 mm with 1-mm ruling and cumulative accuracy better than 0.02 mm) was placed perpendicular to camera 1 at the center of the recording system and recorded for both cameras over their full range of positions. In each view, selected points were tracked for camera positions spanning 250 mm and parameters were fitted by the least squares method.
The accuracy of coordinate reconstruction was checked by comparing actual and estimated coordinates of selected points in records of the calibration grid. This procedure was carried out on several occasions for up to 25 uniformly distributed points in the range x = ± 10 mm, y = 0 mm, z = ± 10 mm. Mean reconstruction accuracy for each coordinate was x = 0.002 ± 0.074 mm, y = 0.008 ± 0.096 mm, z = 0.039 ± 0.055 mm. Maximum deviations were x = 0.2 mm, y = 0.3 mm, z = 0.2 mm and in all cases these occurred at the periphery of the field.
Reconstruction of aortic geometry
The geometry of the aorta was reconstructed by fitting a finite element surface to the marker points(Fig 3). The aorta was modeled as an initially cylindrical finite element mesh consisting of 4 surface elements. Marker points were converted from Cartesian to cylindrical polar coordinates (r,
, z) and the element boundaries were defined so that data points were contained in a single element (element 1), with the remaining elements evenly subdivided circumferentially. The interpolating basis functions for each of the 3 surface coordinates were bicubic Hermite polynomial. The error function minimized in the fitting process consisted of 2 terms. These were, first, the sum of squared residuals between data points and corresponding points on the fitted surface and, second, a quadratic smoothing term that penalized first and second derivatives of the displacement from initial mesh locationsfor a more detailed account, see Young and associates.
21 The use of this smoothing penalty, together with cylindrical topology, constrained the fitted surface at the boundaries of element 1.
|
|
Estimation of regional strain
The Green strain tensor associated with the deformation of the finite element mesh was estimated at the 9 quadrature points in element 1. The surface strain components E
, Ezz, E
z (circumferential, longitudinal and longitudinal-circumferential strain, respectively) were also decomposed into the principal*
strains E1, E2, referred to as major and minor strains, respectively. In this article, E
, Ezz, E1, and E2 are converted to extension ratios by using the relationship E = 0.5 (
2 1), where
is the extension ratio.
Strain was determined with respect to separate reference states for control and patched cases, because the sets of marker points imaged in the control study were not identical to those observed after patch placement. Under these circumstances, the strain analysis for the patched state does not reflect the residual deformation due to patch placement. To address this problem, we identified a subset of common material points in both control and patched states, and the initial strain in the patched reference state was estimated. In each experiment, deformation gradient tensors estimated for this state were used to correct the strain determined for the patched state.
Data analysis
All data are expressed as mean ± SD. Comparison of grouped data was made by multiway repeated measures analysis of variance (ANOVA).
| Results |
|---|
|
|
|---|
The precision with which the aortic surface was reconstructed in both control and patched states was wholly acceptable. Root mean squared (rms) errors for the finite element surface fits to 3-D marker coordinates ranged from 0.02 mm to 0.44 mm with a mean value of 0.13 mm.Fig 3
presents the fitted reference surface for a typical control experiment (rms error 0.138 mm). InFig 4
, fitted reference surfaces are compared with deformed surfaces for (a) control and (b) patched cases.Fig 4
, a, demonstrates that, before patch placement, marker points in reference and inflated states are well fit by a cylindrical surface patch. This was a consistent result for all control experiments. Patch angioplasty distorted the aorta adjacent to patch (compare reference states forFigs 4
, a and b), and the surfaces fitted at all levels of inflation were more complex than for corresponding control results. Root mean squared (rms) errors were greater in the patched than in the control case. Errors typically increased adjacent to the patch(Fig 4
, b, is the worst-case result).
Curvature
Principal (maximum and minimum) curvatures coincided with circumferential and longitudinal directions for both control and patched states.Fig 5, a, presents data for a typical experiment. In the control state there was a progressive reduction in mean circumferential curvature with increased pressure, whereas longitudinal curvature was unchanged. There was little variation in circumferential and longitudinal curvature over the 9 gauss points in element 1 for the control case. Overall, these results are entirely consistent with the inflation of a cylinder. Circumferential curvature was reduced for all pressures in the patched state, indicating that patch angioplasty flattens the adjacent aorta. On the other hand, there appears to be a small increase in longitudinal curvature in the patched case consistent with ballooning of the aorta immediately adjacent to the patch. The widened standard deviations indicate that patch angioplasty produces substantial regional variation of curvature in the region adjacent to the patch.
|
Regional strain
InFig 6, we present data for first and second principal (major and minor) extensions in element 1 for a typical experiment (a) and for all experiments (b). Control results inFig 6
, a indicate that major extensions increased progressively with inflation before patch implantation, whereas minor extensions remained relatively constant. There was little variation in the principal strains over the 9 gauss points in element 1.
|
Patch angioplasty altered the geometry of the aorta adjacent to the patch. The deformation associated with patch placement was characterized by estimating strain in the patched reference state relative to the control reference state(Table II). There was a modest but significant circumferential extension, and the variation of longitudinal extension and circumferential-longitudinal shear between experiments was relatively large.
|
Major extensions for all experiments at a pressure of 103 mm Hg (14.0 kPa) are presented inTable III for each of the 9 gauss points in element 1. Control principal extensions were relatively homogeneous across all gauss points. However, there was a consistent circumferential gradient in major extension in the aorta adjacent to the patch (P = .001). The greatest change in major extension occurred at gauss point 6 adjacent to the center of the patch. Mean values of the major extension at this site were 1.40 ± 0.08 and 1.55 ± 0.07 for control and patched states, respectively, with the greatest individual difference from 1.25 to 1.53. There was no systematic regional variation of minor extension, following patch implantation.
|
| Discussion |
|---|
|
|
|---|
To address these issues, regional strain was estimated in human aortic wall before and after simulated coarctation repair by using a synthetic patch. Arterial geometry and deformation have been studied in vivo by using various techniques including angiography, ultrasonography, intravascular devices, and ultrasonic crystals. Although an in vivo model would have been desirable in the present study, none of these techniques could have provided regional strain data and an in vitro preparation was selected instead. The preparation used could not possibly simulate the exact geometry of a coarctation or the subsequent repair, but it was designed to determine strain fields around a patch in an abnormally large aorta, which is the inevitable consequence of repairing a coarctation by patch angioplasty.
The intrinsic resolution of 3-D coordinate reconstruction by using our biplane imaging system was on the order of 0.1 mm. However, a number of factors combined to reduce the overall accuracy of coordinate reconstruction for marker points on the surface of the aorta. Surface marks were not always punctate and, in such cases, it was necessary to identify a specific feature of the mark and follow this through successive inflation steps. This sometimes became difficult as the mark itself deformed. It was also hard to identify the exact location of points in camera views where they appeared close to the boundary of the aorta.
The complexity of the finite element surface fitted to the reconstructed marker points was limited by the fact that the data points were bounded by a single element and by the smoothing constraints imposed in the fitting process. As a result, the fitted surfaces could not fully reproduce the variations in aortic geometry immediately adjacent to a synthetic patch. However, it is noteworthy that the average rms error for surface fits was 0.13 mm, which compares well with the expected precision of 3-D coordinate reconstruction.
The estimates of surface curvature and regional strain obtained for the unpatched aorta confirm that it deforms as a cylinder with pressurization. Moreover, we have demonstrated that simulated coarctation repair with a synthetic patch substantially reduces circumferential curvature and slightly increases longitudinal curvature in the reference state. These changes in geometry were associated with complex patterns of deformation in the aorta immediately adjacent to the patch in the reference state.
The distribution of maximum extensions in the aorta adjacent to the patch is of considerable interest. Maximum extensions were greatest near the patch and progressively approached control values lateral to this. This result is not dependent on orientation of the patch with respect to the biplane measurement system because the side of the aorta on which the patch was viewed was randomly alternated. This result is also not due to the correction for initial strain after patch angioplasty, because regional extension gradients were, in general, greater for the uncorrected data.
The changes in major extensions adjacent to the patch appear to be relatively modest: an average 21% increase overall from 1.39 to 1.47 at a pressure of 103 mm Hg (14.0 kPa), with a 35% increase in average extension from 1.40 to 1.54 at the gaussian quadrature point closest to the center of the patch. The largest individual principal extension (1.70 at 126 mm Hg; 17.14 kPa) was also observed in the aorta immediately adjacent to the patch.
These data are placed in context if we consider the material properties of the aorta. Circumferential wall tension
for a cylinder is given by
= pt r, where pt is transmural pressure and r is the radius of the aorta. For the control case, therefore, circumferential tension can be estimated from the data for curvature and transmural pressure to obtain the relationship between tension and extension presented inFig 7. After simulated coarctation repair, the deformation of the aorta adjacent to the patch is initially noncylindrical and it is not possible to estimate circumferential tension by using the simple membrane solution above. However, it is reasonable to make inferences about the circumferential tensions associated with measured circumferential extensions from the material behavior of the aorta in the control state. The overall relationship between circumferential tension and extension ratio inFig 7
is nonlinear with a progressive reduction in compliance for extensions greater than 30%. These control data have been fitted with a third-order polynomial, which has been used to associate circumferential tensions with the regional circumferential extensions measured after patch angioplasty. There are a number of reasons why such an approach must be seen as qualitative rather than quantitative. First, longitudinal and shear deformations that occur in the patched state may well influence the circumferential material behavior of the aorta. Second, in the patched case, we are extrapolating beyond the range of extension ratios observed in the control state. However, the fact that the range of maximum extensions observed in the patched case was limited to 70% is consistent with the material relationship presented inFig 7
. It follows that the increases in circumferential extension observed in the patched case must be associated with disproportionate increases in circumferential tension and even greater increases in stress if the effects of wall thinning are taken into account.
|
The latex balloon used to line the aorta before and after simulated patch repair did not appreciably influence the measured material properties of the aorta in this preparation. This method has previously been validated for the measurement of left ventricular intracavity volumes in small animals
23 and negligible additional pressure was required to inflate the balloon.
The findings of this study demonstrate that regional strain in the aortic wall is significantly increased after simulated coarctation repair with a synthetic patch. These observations cannot be attributed to measurement error because the technique used is robust and well established.
20 The increases in regional strain are associated with changes in local aortic wall geometry that are a direct result of patch implantation. The resultant flattening of the aorta in the circumferential direction must increase local stress and strain at equilibrium and it might be expected that the magnitude of these changes would be simply related to the relative circumferential dimensions of the patch and the aorta. In the present study, we were unable to establish any relationship between regional strain and patch size. This may be due to the complex geometry of the aorta in the region adjacent to the patch, but the observation that even a small patch may produce large increases in regional strain is noteworthy.
It is necessary to be circumspect in extrapolating from the findings of this acute in vitro study to the situation of a chronic patch aortoplasty. For instance, there will be ongoing tissue remodeling at the interface between patch and aorta after patch repair of coarctation. Scar tissue formation on the patch and the aorta immediately adjacent to it would be expected to relieve stress at the suture line and give rise to a more uniform strain distribution in the surrounding aortic wall than has been presented here. However, we have shown that strain is elevated throughout the aortic wall adjacent to the patch and that this is associated with circumferential flattening of the aorta due to patch implantation. It is unlikely that these geometric changes will be reversed by scar formation at the boundary of the patch. Turbulent flow, caused by changes in geometry at the site of repair, may also occur after patch aortoplasty. The rapid perturbations in mechanical loading due to turbulence are known to cause fragmentation and loss of elastin in blood vessels and this could contribute to aneurysm formation in the peripatch region. Finally, the current studies were carried out with patches of Dacron graft, but a variety of other materials including polytetrafluoroethylene
and glutaraldehyde-fixed bovine pericardium are also used for coarctation repair. We believe that our experimental model is representative within this context, because Dacron, polytetrafluoroethylene, and glutaraldehyde-fixed bovine pericardium are all highly inextensible at systemic pressure. However, a realistic finite element computer model is seen as the appropriate platform for systematic analysis of the effects of patch geometry, dimensions and material properties on the distribution of stress and strain in the surrounding aortic wall. The present study provides the data necessary to implement and validate such a model and we will use this approach in the future to extend this work.
In conclusion, the findings of this study suggest that synthetic patch repair of coarctation of the aorta may produce increases in wall strain (and disproportionate increases in stress) adjacent to the patch. The unpredictable appearance of true aneurysms in patients after coarctation repair by this technique may reflect the variability of regional strain in the aortic wall, which, even with a small synthetic patch, may be substantial. Because of this unfavorable biomechanical feature, synthetic patch repair of coarctation should not be preferentially used.
| Acknowledgments |
|---|
| Footnotes |
|---|
Gore-Tex patch, registered trademark of W. L. Gore & Associates, Inc, Flagstaff, Ariz. ![]()
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
Y.-Q. Lai, Q.-W. Zhou, H. Wei, C. Zhang, and Z.-G. Zhang Intrapulmonary Channel for One-Stage Correction of Aortic Arch Obstruction Asian Cardiovasc Thorac Ann, October 1, 2006; 14(5): 402 - 406. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Yang, R. Sodian, P. Fu, C. Luders, T. Lemke, J. Du, M. Hubler, Y. Weng, R. Meyer, and R. Hetzer In Vitro Fabrication of a Tissue Engineered Human Cardiovascular Patch for Future Use in Cardiovascular Surgery Ann. Thorac. Surg., January 1, 2006; 81(1): 57 - 63. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Kang, A. J. B. Clarke, I. A. Nicholson, and R. B. Chard Circulatory arrest for repair of postcoarctation site aneurysm Ann. Thorac. Surg., June 1, 2004; 77(6): 2029 - 2033. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Walhout, J. C. Lekkerkerker, G. H. Oron, F. J. Hitchcock, E. J. Meijboom, and G. B.W.E. Bennink Comparison of polytetrafluoroethylene patch aortoplasty and end-to-end anastomosis for coarctation of the aorta J. Thorac. Cardiovasc. Surg., August 1, 2003; 126(2): 521 - 528. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Mekkaoui, P. H. Rolland, A. Friggi, M. Rasigni, and T. G. Mesana Pressure-flow loops and instantaneous input impedance in the thoracic aorta: Another way to assess the effect of aortic bypass graft implantation on myocardial, brain, and subdiaphragmatic perfusion J. Thorac. Cardiovasc. Surg., March 1, 2003; 125(3): 699 - 710. [Abstract] [Full Text] [PDF] |
||||
![]() |