J Thorac Cardiovasc Surg 2001;121:1101-1116
© 2001 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
From the Department of Neurologya and the Department of Cardiothoracic Surgery,b Martin-Luther-University Halle-Wittenberg, Halle/Saale, Germany, and the Department of Cardiac Surgery, Royal Infirmary, University of Glasgow,c Glasgow, United Kingdom.
Received for publication May 26, 2000. Revisions requested June 26, 2000;.revisions received Aug 24, 2000. Accepted for publication Nov 28, 2000. Address for reprints: D. Georgiadis, MD, Department of Neurology, University of Heidelberg, Im Neuenheimer Feld 110, 691210 Heidelberg, Germany (E-mail: dimitrios.georgiadis{at}medizin.uni-halle.de).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The purpose of this study was the comparison of MES prevalence and counts between patients having St Jude Medical valves (SJM; St Jude Medical, Inc, St Paul, Minn) and patients having the ATS valve (ATS Medical, Inc, Minneapolis, Minn).
| Patients and methods |
|---|
|
|
|---|
All patients had undergone examination of the extracranial arteries supplying the brain with continuous wave Doppler andin selected casesduplex ultrasound as part of the routine preoperative assessment. Presence of hemodynamically significant carotid disease (>70% luminal stenosis on continuous wave Doppler ultrasound), as well as perioperative stroke, served as exclusion criteria. A total of 239 patients were identified and received a written invitation to participate in this study. Of these invitees, 191 patients attended, 12 (6.3%) of whom were excluded because of insufficient temporal bone window. Thus, 179 patients, 98 men and 81 women, aged 61 ± 11 years (mean ± SD), were enrolled in this study after giving informed consent. One hundred ten patients had undergone surgery for aortic valve replacement (AVR), 39 for mitral valve replacement (MVR), and the remaining 30 for both aortic and mitral valve replacement (DVR). The interval between transcranial Doppler monitoring and valve insertion was 15 months (13-17 months; median and 95% confidence intervals). All patients were receiving warfarin at the time of the transcranial Doppler study (target international normalized ratio 3-3.5). The actual international normalized ratio at the time of monitoring was not evaluated.
Methods
A detailed neurologic examination was performed in all cases. Subsequently, the middle cerebral arteries were identified at a depth of 52 to 58 mm and continuously insonated for 60 minutes with the 2-MHz transducers of a pulsed ultrasound machine (Multi-Dop X-4; DWL Elektronische Systeme GmbH, Sipplingen, Germany). Settings of the Doppler machine were as follows: power 110 mW, sample volume 5 to 8 mm, low-pass filter 80 kHz, high-pass filter 100 Hz, and sweep 5 seconds. These settings were used throughout the study. Bilateral monitoring was not feasible in 17 patients because of insufficient temporal bone window on one side. All monitoring sessions were evaluated online and recorded on digital audiotapes using an 8-channel digital audiotape recorder (TASCAM DA 88; TASCAM America, Montebello, Calif) for later re-evaluation. This was performed by assigning 15 random tapes (30 hours of monitoring data) to a second examiner, blinded both to the clinical details of the involved patients and to the results of the first observer. Both observers were asked to note the exact position of each detected MES, based on the display of the digital audiotape recorder. MES were identified according to criteria established in a recent consensus.
7 Their count was expressed as mean between the right and left middle cerebral arteries in bilaterally monitored patients. Otherwise, the MES count detected in the monitored middle cerebral artery was used.
Statistical analysis
Normally distributed data were expressed as mean ± standard deviation (SD) and compared by means of the 2-sample t test. Non-normally distributed data were expressed as median and 95% confidence intervals and compared by means of the Mann-Whitney U test. The hypothesis that MES counts significantly differed depending on valve position was tested with the Kruskal-Wallis test for each valve type. The
2 test was used for comparison of frequency distributions; when the sample size was low, the Fisher exact test was used instead. The Spearman rank correlation was used to examine the influence of valve duration and valve size on MES counts.
A binary logistic regression model (using Minitab version 13) was applied to evaluate the influence of valve type, valve position, duration of valve implantation, cardiac rhythm, prevalence of neurologic complications, and presence of diabetes mellitus and arterial hypertension on the presence or absence of MES. All predictors were included as categorical factors, except for duration of valve implantation (continuous variable). For this model, MES counts were expressed as 0 (for MES-negative patients) or 1 (for MES-positive patients).
Interobserver variability was evaluated by means of the Cohen
statistic.
8 Values derived from this evaluation range between 1 (complete disagreement) and 1 (complete agreement). A value of 0 reflects lack of a relation between the evaluations of the 2 observers. Acceptable to good agreement is indicated by values between 0.4 and 0.75, and values greater than 0.75 indicate an excellent agreement.
8
| Results |
|---|
|
|
|---|
2 test). Heart rhythm was sinus in 60.4% and 55.7% and atrial fibrillation in 34.1% and 36.4% of patients with ATS and SJM valves, respectively; the remaining patients (5.5% and 7.9%, respectively) had a pacemaker. These differences were also insignificant (P = .72).
|
2 test, P = .009; AVR vs MVR, P = .1; AVR vs DVR, P = .004; MVR vs DVR, P = .2). In the group of patients with ATS valves, no significant differences in MES prevalence were evident depending on the valve position (
2 test among all patients, P = .1;
2 test, AVR vs MVR, P = .1; AVR vs DVR, P = .09; MVR vs DVR, P = .8). The contrary was true in patients with SJM valves (
2 test among all patients, P = .03;
2 test, AVR vs MVR, P = .4; Fisher exact test, AVR vs DVR, P = .007; Fisher exact test, MVR vs DVR, P = .1). The Kruskal-Wallis test revealed no significant differences in MES counts depending on valve position in patients with ATS or patients with SJM valves (P = .6 and P = .3, respectively). The exact MES counts for each position are listed inTable I
The Spearman rank correlation revealed a significant relationship between duration of valve implantation and MES counts when data from all patients were analyzed (
= 0.25, P = .001). This was also true for patients with SJM valves (
= 0.3, P = .005), but not for patients with ATS valves (
= 0.09, P = .41).
MES counts did not correlate with valve size, except in patients with an SJM valve in the aortic position (AVR overall,
= 0.19, P = .05; ATS valves,
= 0.001, P = .994; SJM valves,
= 0.29, P = .03; MVR overall,
= 0.07, P = .66; ATS valves,
= 0.086, P = .71; SJM valves,
= 0.09, P = .73; Spearman rank correlation).
Results of logistic regression analysis are displayed inTable II. Valve type was a significant predictor of MES counts, with SJM valves being a predictor of higher MES counts than ATS valves; presence of diabetes mellitus was a further significant predictor for lower MES counts.
|
statistic revealed an excellent agreement between the 2 observers (
= 0.82).
A total of 15 patients had had neurologic symptoms since valve insertion. These consisted of amaurosis fugax (n = 3), transient ischemic attack (n = 4), and ischemic stroke (n = 8). All transient ischemic attacks involved the anterior circulation. Ischemic strokes involved the anterior circulation in 6 patients (partial anterior circulation syndrome in all cases) and the posterior circulation in 2 patients (isolated brainstem infarction, n = 1; brainstem and cerebellar infarction, n = 1). Six of the 15 patients had SJM valves and 9 had ATS valves (
2 test, P = .5). The aortic valve was replaced in 6 patients (6/110, 5.4%), the mitral in a further 6 patients (6/39, 15.3%), and both valves in 4 patients (4/26, 15.3%); these differences did not reach statistical significance (all patients,
2 test, P = .1, AVR vs DVR, P = .22; AVR vs MVR, P = .08; MVR vs DVR, P = 1; all Fisher exact test).
History of cerebral embolic complications after valve insertion was not a predictor of MES counts(Table II
). Additionally, no statistically significant differences in MES counts were evident on direct comparison between patients with a history of cerebral embolic complications and the remaining patients (17 [8-31] vs 10 [5-18], P = .1; Mann-Whitney U test).
| Discussion |
|---|
|
|
|---|
The differences in MES counts between patients with ATS and patients with SJM valves are difficult to explain assuming formed underlying embolic material, particularly because the 2 patient groups were comparable in their clinical characteristics, and the same anticoagulant regimen was used for all patients. Recent reports from several research groups add weight to the hypothesis that the underlying material of MES in patients with prosthetic valves consists of nitrogen bubbles. The observed significant differences in MES counts between 2 valve types that are quite similar in design and material composition suggest that even slight differences in the local fluid mechanical development in the vicinity of the bileaflet valve can have a profound influence on the flow field distant to the valve. The hydrodynamic characteristics of SJM and ATS valves were compared in 2 previous studies: similar pressure drops across both valves were observed in the first study, under both steady and pulsatile flow. Nevertheless, pressure drop was lower in the ATS valve as compared with the SJM valve under incorporation of an enlargement instead of a straight downstream conduit.
9 Significant differences in mean increasing pressure drops and effective orifice areas were noted in the second study (both ATS > SJM).
10 Still, both reports examined only hydrodynamic performance with the valve in the mitral position. The causative role of the aforementioned hydrodynamic findings in the discrepancies in MES counts observed in this study remains to be evaluated.
The incidence of embolic complications in patients with SJM valves was examined in several studies
11-13 and reported between 1.09%
11 and 2.45%
13 per patient-year. Two clinical reports concerning the ATS valve have been published to date. One thromboembolic complication was noted in the first, describing 200 patients followed up for 1 to 3 years.
14 A single cerebral embolic complication is also described in the second report, in which results from 100 patients, followed up for 5 to 27 months, are presented.
15 Obviously, the small size of the aforementioned reports prohibits clinical comparisons between the 2 groups. The prevalence of cerebral embolic complications noted in our study is markedly higher than previously reported. It must be taken into account, though, that most symptoms were transient. In fact, most patients with transient ischemic attacks and amaurosis fugax reported these incidents only after detailed questioning, having denied neurologic complications beforehand. Considering the small sample size of the present study, definitive statements on the prevalence of cerebral embolic events are not warranted. Additionally, our patients were asked to attend the ultrasound examination on an outpatient basis; this fact may have biased our results, because only healthy or at least not heavily disabled patients were bound to attend. We presume that the high prevalence of amaurosis fugax in our study is due to the fact that the ophthalmic artery, with its small diameter, missing collaterals, and dramatic neurologic signs, represents a predilection site for symptomatic embolism.
MES counts were not associated with a history of cerebrovascular events in our study, which corresponds to the results of most previous studies. Obviously, patients were not examined immediately after the cerebrovascular events occurred, but in some cases as long as 2 years later, so that a potential relationship in the acute phase cannot be excluded. On the other hand, MES counts in patients with prosthetic heart valves remain relatively constant over time.
16
Duration of valve implantation was not identified as an independent predictor of MES counts in logistic regression analysis. Still, results from nonparametric correlation demonstrated a significant influence of valve duration on MES counts in patients with SJM valves but not in those with ATS valves. This finding is difficult to explain, because it suggests a significant difference in the behavior of the 2 valves. It must be emphasized that only 1 previous study examining a homogeneous group, namely patients with Carbomedics valves (Sulzer Carbomedics, Inc, Austin, Tex), described such an influence.
2 It has been observed that the incidence of MES increases with time of implantation for porcine valves implanted in an animal model, due to increasing pannus overgrowth on the valve.
17 It is possible that a small difference in tissue encroachment between the ATS and SJM valves may produce a similar effect with time. Although it appears possible that the influence of valve duration on MES counts differs depending on the valve type, we are unable to provide definitive information on this matter.
No previous study has examined the influence of diabetes mellitus on MES counts. The underlying mechanism of this potential effect is unclear. It may be that blood osmolality influences formation, as well as half-life, of cavitation bubbles, thus affecting MES counts. Obviously, several additional parameters would have to be included in future analyses, particularly adequacy of diabetes control and actual blood glucose level at the time of transcranial Doppler study.
Influence of MES on neuropsychologic deficit in patients with prosthetic valves was suggested in a recent study.
6 However, the sample size of this study was quite small, thus prohibiting definitive statements on this matter. Provided that large-scale studies confirm the influence of MES on cognitive function, this method is bound to become a standard test for these patients, at least in the evaluation of new valve types.
In conclusion, our study provided evidence of significant differences in MES counts between patients with ATS and SJM valves. A relation with the prevalence of neurologic complications could not be established. Influence of MES on cognitive function remains to be evaluated.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Skjelland, A. Michelsen, F. Brosstad, J. L. Svennevig, R. Brucher, and D. Russell Solid Cerebral Microemboli and Cerebrovascular Symptoms in Patients With Prosthetic Heart Valves Stroke, April 1, 2008; 39(4): 1159 - 1164. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Hiratsuka, S. Fukunaga, E. Tayama, K. Takagi, K. Arinaga, T. Shojima, H. Teshima, and S. Aoyagi High-intensity transient signals due to prosthetic valve obstruction: diagnostic and therapeutic implications Ann. Thorac. Surg., May 1, 2004; 77(5): 1615 - 1621. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Dalinin, P. S. Lingaas, K. Hatteland, and J. L Svennevig Carotid Doppler microembolic signals in patients one year after heart valve surgery Perfusion, December 1, 2003; 18(6): 333 - 337. [Abstract] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||