J Thorac Cardiovasc Surg 2001;122:946-954
© 2001 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease (ACD) |
From the Department of Biomedical Engineering,a Cleveland Clinic Foundation, Cleveland, Ohio; Division of Cardiothoracic Surgery,b University of Wisconsin, Madison, Wis; and Department of Mechanical Engineering,c University of Washington, Seattle, Wash.
Received for publication Oct 9, 2000. Revisions requested Jan 22, 2001; revisions received April 4, 2001. Accepted for publication April 6, 2001. Address for reprints: Karyn S. Kunzelman, PhD, University of Wisconsin, CSC H4/368, 600 Highland Ave, Madison, WI 53704 (E-mail: karynk{at}surgery.wisc.edu).
Abstract
Objectives: Progressive aortic root dilatation and an increased aortic root elastic modulus have been documented in persons with Marfan syndrome. To examine the effect of aortic root dilatation and increased elastic modulus on leaflet stress, strain, and coaptation, we used a finite-element model.
Methods: The normal model incorporated the geometry, tissue thickness, and anisotropic elastic moduli of normal human roots and valves. Four Marfan models were evaluated, in which the diameter of the aortic root was dilated by 5%, 15%, 30%, and 50%. Aortic root elastic modulus in the 4 Marfan models was doubled. Under diastolic pressure, regional stresses and strains were evaluated, and the percentage of leaflet coaptation was calculated.
Results: Root dilatation and stiffening significantly increased regional leaflet stress and strain compared with normal levels. Stress increases ranged from 80% to 360% and strain increases ranged from 60% to 200% in the 50% dilated Marfan model. Leaflet stresses and strains were disproportionately high at the attachment edge and coaptation area. Leaflet coaptation was decreased by approximately 20% in the 50% root dilatation model.
Conclusions: Increasing root dilatation and root elastic modulus to simulate Marfan syndrome significantly increases leaflet stress and strain and reduces coaptation in an otherwise normal aortic valve. These alterations may influence the decision to use valve-sparing aortic root replacement procedures in patients with Marfan syndrome.
The numerous abnormalities associated with the Marfan syndrome
1,2 have been linked to a genetic defect in fibrillin.
3 Fibrillin is a component of the microfibrils making up the elastic meshwork in the aorta and other large vessels.
4-6 The defect in fibrillin found in the Marfan syndrome results in disorganization and fragmentation of the elastic meshwork. Normally, elastic fibers are responsible for maintaining blood vessels at their normal dimensions by providing the ability to expand with the cyclic increase of blood pressure and then allowing complete recovery to the initial state on removal of the pressure load. In contrast, collagen fibers, which are much stiffer than elastic fibers, are responsible for preventing the aortic wall from stretching too far.
7 Elastic fragmentation therefore reduces the ability of the aortic wall to fully recover from the cyclic distending pressure, resulting in permanent stretching and dilatation. This dilatation shifts the pressure load from the fragmented elastic network to the stiffer collagen fibers, resulting in an increased elastic modulus in the aortic wall.
4,6,7
The altered aortic root geometry and elastic modulus may be linked to observed valvular changes, such as increased size, fibrotic thickening of the coaptation area, and marginal rolling sometimes found in aortic valves in the Marfan syndrome.
5,8,9 The normal function of the valve is dependent on the compliant tissue properties and rounded shape of the normal, undilated aortic root.
10,11 In the case of the Marfan syndrome, the combined effect of aortic root dilatation and increased aortic root elastic modulus could increase the stress on the aortic valve leaflets, potentially resulting in leaflet alterations. Although the tissue constituent makeup, geometry, and material properties of the aortic root in Marfan syndrome have been characterized, there has not yet been a systematic study of the relationship between these pathologic changes and the biomechanical function of the aortic valve. The objective of this investigation was to examine the influence of dilatation of a stiffened Marfan syndrome aortic root on aortic valve function (stress, strain, and coaptation) by using a 3-dimensional and anatomically realistic finite-element model of the aortic root and valve.
12
Methods
The detailed development of the original 3-dimensional model of the aortic root and valve has been described previously
12 and is summarized here.
Model geometry and element development
Finite-element analysis is a computational technique in which an object with a complicated structure is divided into smaller sections (ie, elements) that are interconnected by common points (ie, nodes). This discretization enables the use of algebraic equations to describe the individual structural state at each node. The solution of the system of equations yields the stress and strain at any point in the entire object. Our finite-element model was developed by means of ANSYS software (version 5.3; ANSYS Inc, Canonsburg, Pa) run on a DEC Alphastation 400 4/233 (Digital Equipment Corporation, Maynard, Mass). Normal human aortic valve and root specimens in a zero-pressure state were used to establish the geometry for the model by using magnetic resonance imaging to obtain the data coordinates and hence the nodes to represent the shape of the aortic root and valve. Elastic 6-node triangular shell elements were generated from these nodes to create 5000 elements in the aortic root and 1815 elements in the 3 valve leaflets.
Normal model boundary conditions
Geometric boundary constraints were assigned to all models. First, contact elements were incorporated into the leaflet surfaces to permit sliding and to prevent the leaflets from passing through each other or through the root wall. Second, the lowest layer of the aortic root base was restricted to permit only radial displacement. Finally, the physiologic longitudinal stretch found normally in arteries
13 was imposed by applying tension at the top of the ascending aorta and at the distal ends of the coronary ostia.
Marfan model boundary conditions
To simulate aortic root dilatations of 5%, 15%, 30%, or 50%, we applied radially directed forces to the root in the 4 Marfan models.
14 This action also served to pull the leaflets apart and create the slight central deficiency found clinically.
8,9,15 Only the root geometry was altered; all other conditions remained as in the normal model. Because these radial forces established preliminary stresses and strains that might not be present in an actual aortic root dilated as a result of Marfan syndrome, the force-induced root results were subtracted from the final results to obtain the root stresses and strains caused by the pressure load alone. However, because our objective was to determine the effect of the Marfan-associated root alterations on the leaflet stress and strain, the force-induced preliminary values were not subtracted from the leaflet results.
Pressure loading
To represent the early diastolic loading of the aortic valve and root (when peak pressure across the valve is generated), we applied simulated physiologic pressures to the valve and root structure in 2 phases (Figure 1). In the first phase, the aortic root alone was pressurized by a linearly increasing load from the zero-pressure state to the end-systolic aortic pressure level. In the second phase, physiologic pressure loads were applied to the valve, to the end-systolic root, and to the region of the root underneath the valve. These physiologic pressures were calculated as the pressure differences between the aortic root and left ventricle (load applied to aortic valve), between the aortic root and chest (root), and between the left ventricle and chest (subvalvular root). Loading started at end-systole, just after valve closure, and finished at the end of isovolumic relaxation, when peak pressure across the valve was reached.
|
|
Output analysis
For both the leaflets and root walls, the magnitude and location of the principal tensile stresses and strains were recorded at the end of physiologic loading, when maximum transvalvular pressure occurred. Therefore stresses and strains in the model at that time point are likely representative of the maximum values attained during the entire cardiac cycle. Stress (
) is defined as the force (F) per unit area (A):
= F/A
Strain (
) is defined as the percentage extension or stretch of the tissue:
=
L /Lo
where
L is change in length, and Lo is original length. Elastic modulus (E) is defined as the change in stress (
) divided by the change in strain (
):
E = 
/

Regional magnitudes were calculated by grouping sets of elements to define specific model components (belly, coaptation area, free margin, attachment edge, annular sinus wall, and sinotubular junction sinus wall). These components were analyzed to calculate the peak average, which was the average value of the 5% of elements with the highest values (peak standard deviation was calculated similarly). The coaptation area, defined as the percentage of leaflet area contacting the adjacent leaflet surfaces, was also examined throughout the preliminary root-loading and subsequent physiologic root/valve-loading phases. Leaflet coaptation creates a large, tight seal across the closed valve; reduced coaptation could indicate the potential for valvular regurgitation. The stress and strain results were compared with analysis of variance to determine significant differences with respect to root dilatation and aortic root elastic modulus.
Results
Valve stress, strain, and coaptation
The models simulating Marfan syndrome demonstrated progressive increases in peak average valve stress, corresponding to increasing amounts of root dilatation compared with that seen in the normal root model (Figures 2-4). In the 5% dilatation model the increase in leaflet stress was nominal, on the order of 0% to 32%. However, by the 50% dilatation model, the increase in valve stresses was much higher, ranging from 77% to 357%. The greatest increases in leaflet stress in all the Marfan models were evident at the attachment edge. This pattern was different from that found in the normal root and valve,
12 where leaflet stresses were lower at the attachment edge, and were somewhat lower at the coaptation area than in the leaflet belly and free margin caused by the support of the adjacent leaflet. In contrast, the Marfan syndrome models showed significant increases in stress at the coaptation area, as a result of reduced coaptation.
|
|
|
|
|
Root stress and strain
The stress values calculated for the aortic root walls (caused by pressure loading only) varied by less than 10% between the Marfan models and the normal root model (Figure 7). Strain magnitudes, however, did differ between models. In the 5% and 15% Marfan models the strain in the root was decreased by 36% and 15%, respectively, compared with that of the normal root. In contrast, for the 30% and 50% Marfan models, the strain in the root was increased by 30% and 45%, respectively, compared with that of the normal root.
|
These finite-element models demonstrated that the increased aortic root elastic modulus and progressive root dilatation found in Marfan syndrome results in an increase in stress and strain in aortic valve leaflets, as well as significantly reduced leaflet coaptation.
The initial root dilatation in the Marfan models (5% and 15%) was accompanied by progressively increasing leaflet stresses, particularly at the attachment edge and belly. The overall leaflet stress pattern in these 2 Marfan models, however, was similar to that of the normal valve leaflets in that the stresses at the attachment edge of the leaflet were lower in magnitude than the rest of the leaflet tissue (caused by stress sharing between the root and valve). In contrast, further dilatation of the Marfan root (30% and 50%) greatly increased the leaflet stresses, making them disproportionately high at the attachment edge and coaptation area. These regionally high magnitudes are due to reductions both in stress sharing at the attachment edge and in mutual compressive stress relief in the coaptation area. The increased leaflet stresses at the attachment edge were accompanied by a slight decrease in aortic root stress, which indicates a transfer of root stress to the valve leaflet, the opposite of the situation in the normal model.
In addition to increased leaflet stresses, the strains in the leaflets were altered, which directly affected leaflet coaptation. In the 5% dilated Marfan model strains were increased at the attachment edge and coaptation area but were actually decreased at the belly and free margin. These mixed results suggest that for small geometric changes to the root, a compensatory mechanism may occur through which valve deformation can be slightly adapted to respond to an altered environment. It may be through such a mechanism that the regional leaflet strains were altered to maintain coaptation, albeit at the expense of increased leaflet stress. However, with further increases in diameter of the Marfan root, the adaptation mechanism was not sufficient to counteract the dilatation, and the strain and coaptation patterns were significantly affected. With 15% and greater dilatation, strains were increased across the entire leaflet. By 30% and 50% dilatation, even though the leaflet was tightly stretched, not enough tissue was present to maintain normal coaptation, suggesting that the proposed compensatory mechanism for preservation of coaptation failed. Because the valves in these models could no longer maintain normal coaptation, these results indicate the possibility for regurgitation in these moderately dilated Marfan aortic roots.
The amount of strain in the aortic root sinuses also contributed to the results in this series of models. In the 5% and 15% dilated Marfan models aortic root strain was slightly decreased, which would suggest that the root was compensating to preserve coaptation. In contrast, increased root strains in the more dilated Marfan models (30% and 50%) again suggests that the compensatory mechanism has failed at these larger root diameters, and the valve cannot maintain normal coaptation.
The stress and strain results in the Marfan models were also compared with those of a series of idiopathic dilatation models that we previously reported.
14 Overall, the dilatation component of the Marfan models caused far greater increases in leaflet stresses than did the material change (30%-60% increases between subsequent dilatations vs 4%-12% differences between Marfan models and idiopathic dilatation models). However, the leaflet attachment edge stresses and strains were greater in the Marfan models than in corresponding idiopathic dilatation models. In particular, these stresses were significantly greater in the 5% and 15% Marfan models. The differences between the results of these 2 series of models emphasizes the contributing role of the elastic modulus of the aortic root in governing the stresses at the leaflet attachment edge. Alterations to the root modulus appear to be especially important in the early stages of aortic root dilatation and would undoubtedly influence leaflet stresses in the absence of root dilatation.
Limitations
As with any modeling study, there are some inherent limitations in our model, as described in our initial report.
12 First, we chose to simulate only the final closing phase of the valve because the transvalvular pressure and tensile stresses are highest during this period. Second, because of software restrictions, the material properties were assumed to be constant in the physiologic range of the stress-strain curve. As a result, the modeled stresses and strains were likely lower than those in the actual root and valve. However, these models analyzed diastolic function only, in which the valve tissue reportedly functions in the linear and higher elastic modulus region of its stress-strain curve.
17 Furthermore, the normal aortic root elastic modulus was calculated from reported in vivo measurements.
18 Third, the analysis assumed that the initial zero-pressure state in the root and valve was equivalent to a zero-stress state. Because of residual stresses caused by the differential makeup of the tissue layers
21,22 and normal dynamic root-valve motion, a zero-stress state may not exist in vivo. However, the magnitude of these initial stresses was assumed to be negligible in comparison with the tissue stresses at peak pressure. Fourth, a small deformation analysis was used on the basis of the rationale that the applied simulated pressures did not lead to significant changes in the model geometry. We therefore assumed that any incrementally increased deformations that could be gained from a large deformation analysis would not be large enough to justify the extracomputational burden. Finally, mild root dilatation in Marfan syndrome is believed to be limited to the sinus region and only later spread upward to the proximal ascending aorta.
6,9 However, the degree of root dilatation in all the Marfan models was assumed to be uniform from the anulus to above the sinotubular junction (the root base was not dilated). This action likely overestimated the stresses and strains above the upper root in the 5% and 15% dilated Marfan models.
Despite these limitations, the normal and Marfan models were designed by using identical conditions; only specific geometric and material properties (aortic root diameter and elastic modulus) were altered between models. Therefore, comparison between models provides a relative estimate of the differences that could be expected in vivo as a result of dilatation of a Marfan aortic root.
Clinical and surgical implications
These finite-element model results indicate that even with limited root dilatation associated with the Marfan syndrome, the simulated leaflets experience stress and strain changes before any loss of coaptation. These changes have been associated with leaflet calcification,
23 may lead to tissue remodeling to bear the higher stresses, and could explain the reports of aortic valve thickening
5,8,9 and regurgitation
1,24,25 found in persons with the Marfan syndrome. The greater attachment edge stresses and strains in the Marfan models compared with those in the idiopathic dilatation models suggest that there may be an earlier critical stage for the valve-sparing root-replacement procedure in persons with the Marfan syndrome. However, in patients with Marfan syndrome with advancing root dilatation, their aortic valve leaflets may no longer be normal and might not benefit as much as desired from a valve-sparing procedure. Of course, an additional factor in determining the optimal surgical timing is the inherent risk and complications of the surgical procedure. Further research regarding the timing of tissue remodeling in the aortic valve would be helpful to surgeons in determining the optimal time and procedure for replacement of the diseased aortic root in Marfan syndrome.
Acknowledgments
We thank the Whitaker Foundation and the National Partnership for Advanced Computational Infrastructure for their support of this research. This research was performed in the Department of Bioengineering and the Division of Cardiothoracic Surgery at the University of Washington.
References
This article has been cited by other articles:
![]() |
I. El-Hamamsy, K. Balachandran, M. H. Yacoub, L. M. Stevens, P. Sarathchandra, P. M. Taylor, A. P. Yoganathan, and A. H. Chester Endothelium-Dependent Regulation of the Mechanical Properties of Aortic Valve Cusps J. Am. Coll. Cardiol., April 21, 2009; 53(16): 1448 - 1455. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. Gundiah, K. Kam, P. B. Matthews, J. Guccione, H. A. Dwyer, D. Saloner, T. A.M. Chuter, T. S. Guy, M. B. Ratcliffe, and E. E. Tseng Asymmetric Mechanical Properties of Porcine Aortic Sinuses Ann. Thorac. Surg., May 1, 2008; 85(5): 1631 - 1638. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Hashim and D. Richens Finite element method in cardiac surgery Interactive CardioVascular and Thoracic Surgery, February 1, 2006; 5(1): 5 - 8. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Albes, U. A. Stock, and M. Hartrumpf Restitution of the Aortic Valve: What is New, What is Proven, and What is Obsolete? Ann. Thorac. Surg., October 1, 2005; 80(4): 1540 - 1549. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Tagusari, H. Ogino, J. Kobayashi, K. Bando, K. Minatoya, H. Sasaki, K. Niwaya, Y. Okita, M. Ando, T. Yagihara, et al. Should the transverse aortic arch be replaced simultaneously with aortic root replacement for annuloaortic ectasia in Marfan syndrome? J. Thorac. Cardiovasc. Surg., May 1, 2004; 127(5): 1373 - 1380. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Kirali, V. Erentug, M. B Rabus, A. Izgi, N. U Bozbuga, K. Erkanli, E. Akinci, and C. Yakut Extensive Aortic Surgery in Marfan Syndrome: 16-Year Experience Asian Cardiovasc Thorac Ann, December 1, 2003; 11(4): 337 - 341. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||