J Thorac Cardiovasc Surg 2002;123:161-167
© 2002 The American Association for Thoracic Surgery
General Thoracic Surgery |
From the Department of Thoracic and Cardiovascular Surgery,a the Division of Surgical Pathology,b Kansai Medical University, Moriguchi, and the Division of Surgical Pathology,c International Center for Diagnostic Pathology, Kobe, Japan.
Received for publication April 3, 2001. Revisions requested May 14, 2001; revisions received June 4, 2001. Accepted for publication July 3, 2001. Address for reprints: Yukihito Saito, MD, Associate Professor, Department of Thoracic and Cardiovascular Surgery, Kansai Medical University, 10-15 Fumizonocho, Moriguchi, Osaka 570-8507, Japan (E-mail: saitoy@ takii.kmu.ac.jp).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The Dumon stent, a silicone tube, has been well established, especially for treatment of stenoses with intraluminal tumor growth or granulomatous tissue. A metallic stent, the nitinol self-expandable airway stent, is usually used in patients with airway stenosis because of extrinsic tumoral compression.
Metallic stents are sometimes used in patients with benign bronchial stenosis after lung transplantation caused by granulation of the bronchial anastomosis site or in pediatric patients with congenital tracheal stenosis and tracheobronchomalacia. Metallic stents have the following advantages compared with silicone stents: (1) simplicity of insertion and fixation; (2) better clearance secretions; (3) accommodation to varying tracheal dimensions; and (4) a low internal/external diameter ratio. However, metallic stents are not without problems. Once covered by epithelium, they are not easy to remove with a conventional bronchoscopic procedure alone, and an operation may be required.
Silicone stents can be removed, but they have several problems, such as disturbance of the physiologic mucociliary function of the tracheobronchial epithelium under the stent. Silicone stents are relatively thick and cause mucociliary function to be lost in the stented area. Secretions tend to accumulate in the lumen, which can cause obstruction.
In the pediatric patient who is growing, it may be necessary to exchange existing stents for larger ones. However, this is not easily done if the stent is metallic. Bioabsorbable airway stents, on the other hand, offer benefits: extraction of the device is unnecessary, and normal airway is preserved after stent resorption.
The aim of this study was to examine the biocompatibility and mechanical strength of a new tubular bioabsorbable stent made of poly-L-lactic acid (PLLA).
| Materials and methods |
|---|
|
|
|---|
|
The rabbits were anesthetized with subcutaneous injection of atropine (0.75 mg) and inhalation of 1.5% halothane. The animals maintained spontaneous breathing without intubation. In the midline the cervical trachea was prepared under direct visualization for implantation of the stent. Unnecessary dissection was avoided so as not to interfere with the blood circulation of the trachea. The trachea was opened transversely between cartilage rings for two thirds of its circumference. The stent was implanted intratracheally and fixed with the same continuous 5-0 polypropylene suture used to close the tracheostomy.
In group A the rabbits were killed at 8 weeks (5 rabbits) after stent implantation. In group B the animals were killed at 8 weeks (5 rabbits), 20 weeks (5 rabbits), and 40 weeks (4 rabbits) after stent implantation, at which time tracheal endoscopy was done. The cervical trachea was excised and divided longitudinally into two pieces, one fixed in formalin solution for histologic examination and the other in buffered glutaraldehyde solution for scanning electron microscopic study. All of the surgical and euthanasia procedures were performed in accordance with the "Guide for the Care and Use of Laboratory Animals" prepared by the Institute of Laboratory Animal Resources, National Research Council, and published by the National Academy Press, revised 1996.
In vitro evaluation of mechanical strength in 2 types of airway stents
We tested PLLA stents and silicone stents in vitro for their mechanical strength. Using a digital force gauge (DFG-0.2K; Shimpo Corp, Kyoto, Japan) and a noncontact profile projector (PJ 300; Mitsutoyo Corp, Kyoto, Japan), we subjected stents to area loads at room temperature (22°-24°C) and measured their deformation rates. Area load was applied over the full length of the stent with a 50-mm2 plate mounted on a digital force gauge with a measuring range of ±1.960 N and an accuracy of ±1 mN. The stent diameter before and after loading was measured 6 times each on the screen of a noncontact profile projector to an accuracy of 1/1000 mm (Figure 2). We calculated the deformation rate as follows:

|
| Results |
|---|
|
|
|---|
In group A, which received silicone stents, 3 rabbits died within 4 weeks of stent implantation as a result of airway obstruction by secretions inside the stent lumen. The 4-week survival of the rabbits in group A was 62.5%. In the remaining rabbits stridor developed within 8 weeks after stent implantation. Therefore, all rabbits surviving up to 8 weeks after stenting were killed in the eighth postoperative week. Obstructing material was visible inside the stents of all rabbits (Figure 3).
|
|
|
|
|
In vitro evaluation of mechanical strength in 2 types of airway stents
For diameters between 4 and 6 mm, the deformation rate of the silicone stent was lower than that of the PLLA stent. However, the mechanical strength of PLLA stents increased as a function of their diameter. At a diameter of 8 mm, the deformation rate of PLLA stents was similar to that of silicone stents. At diameters of 10 and 13 mm, the deformation rate of PLLA stents was significantly lower than that of silicone stents (Figure 7).
|
| Discussion |
|---|
|
|
|---|
The ideal stent should be simple to insert, fix, and remove; be biocompatible; not obstruct the airway; allow clearance of secretions; and accommodate to varying tracheal dimensions and shapes.
3,4
Metallic expandable wire stents, such as the Wallstent,
5 have a low internal/external diameter ratio. They cannot usually migrate after insertion, and the mucociliary clearance function is better maintained. However, once covered by epithelium, it is not easy to remove them with a bronchoscopic procedure. An operation may be required.
Silicone and metallic stents used in the tracheobronchial tree for the treatment of airway problems both have disadvantages, as mentioned above. Long-term results in tracheobronchial airway stenting after implantation has not fully been discussed
6 because airway stents are usually only used for short-term maintenance of quality of life in patients who have advanced malignant airway diseases, mainly caused by lung cancer.
In the pediatric patient who is growing, it may be necessary to exchange existing stents for larger ones. However, this is not easily done if the stent is metallic. Although silicone stent can be exchanged, it is unfavorable for stenting the small airway in a child because of its thick wall. Its most disadvantageous feature is its disturbance of the physiologic mucociliary function of the small airway.
The new tubular bioabsorbable knitted airway stent has advantages, such as a low internal/external diameter ratio and good clearance secretions. Bioabsorbable stents are optimal if the need for stenting the airway is only temporary. For benign and post-transplantation stenosis in particular, a bioabsorbable stent can be an alternative means of treatment.
The biocompatibility of bioabsorbable materials, such as PLLA and polyglycolic acid, has been good in other organs. Rods of bioabsorbable PLLA and pledgets of polyglycolic acid are in clinical use for the fixation of bone fractures and reinforcement of suture lines in lung parenchyma, with favorable results.
7-9 In our study PLLA was chosen for an absorbable bronchial stent because it has the longest biodegradation time of the basic molecules available. Its biodegradation time is about 6 months. The bioabsorbable PLLA stent examined in our study is a different type from the spiral stent previously reported.
10,11
The new type of PLLA stents are mechanically strong enough to retain their tubular shape against extrinsic mechanical force. The mechanical strength of the PLLA stent increased as a function of diameter. At diameters of 4 and 6 mm, the PLLA stents we tested were mechanically weaker than the silicone stents; however, knitted tubular stents made of PLLA wire thicker than 170 µm can be as strong as silicone stents. At diameters of 10 mm or more, the mechanical strength of PLLA stents was stronger than that of silicone stents. Therefore, for patients with various degrees of airway diseases, knitted tubular PLLA stents can be used like silicone stents.
In conclusion, the biocompatibility of airway stents made of PLLA is good enough to make it a promising material for airway stents in clinical use. This bioabsorbable airway stent can be useful in the treatment of tracheomalacia in a child who is growing.
This bioabsorbable knitted tubular stent and its construction require further investigation before clinical applications begin.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. H. Thornton, R. L. Gordon, R. K. Kerlan, J. M. LaBerge, M. W. Wilson, K. A. Wolanske, M. B. Gotway, G. S. Hastings, and J. A. Golden Outcomes of tracheobronchial stent placement for benign disease. Radiology, July 1, 2006; 240(1): 273 - 282. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Noppen, G. Stratakos, J. D'Haese, M. Meysman, and W. Vinken Removal of Covered Self-Expandable Metallic Airway Stents in Benign Disorders: Indications, Technique, and Outcomes Chest, February 1, 2005; 127(2): 482 - 487. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Saito, K.-i. Minami, H. Kaneda, T. Okada, T. Maniwa, Y. Araki, H. Imamura, H. Yamada, K. Igaki, and H. Tamai New Tubular Bioabsorbable Knitted Airway Stent: Feasibility Assessment for Delivery and Deployment in a Dog Model Ann. Thorac. Surg., October 1, 2004; 78(4): 1438 - 1440. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Ernst, D. Feller-Kopman, H. D. Becker, and A. C. Mehta Central Airway Obstruction Am. J. Respir. Crit. Care Med., June 15, 2004; 169(12): 1278 - 1297. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Dodge-Khatami, H. W. Niessen, L. H Koole, M. G Klein, T. M van Gulik, and B. A. de Mol Tracheal Replacement in Rabbits with a New Composite Silicone-Metallic Prosthesis Asian Cardiovasc Thorac Ann, September 1, 2003; 11(3): 245 - 249. [Abstract] [Full Text] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||