J Thorac Cardiovasc Surg 2002;123:512-516
© 2002 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease (ACD) |
From the Department of Cardiothorcic Surgery, Joan and Sanford I. Weill Medical College of Cornell University, New York Presbyterian Hospital, New York, NY.
Received for publication May 17, 2001; revisions requested July 13, 2001; revisions received July 18, 2001; accepted for publication July 24, 2001. Address for reprints: Richard Brodman, MD, Department of Cardiothoracic Surgery, Weill Medical College, 525 East 68th St, New York, NY 10021 (E-mail: rfb20002{at}mail.med cornell.edu).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
| Subjects and methods |
|---|
|
|
|---|
Evaluation procedures
Preoperative testing performed on the first consecutive 100 patients in this study who had coronary artery bypass grafting with an RA free graft included photoelectric plethysmography with a calculated perfusion index (PI; modified Biox model 3700 pulse oximeter; Datex-Ohmeda, Louisville, Colo) to determine whether adequate collateral circulation was present and color flow and Doppler studies to prospectively evaluate changes in the collateral vessels within the forearm before and after RA harvest. The remaining 289 patients underwent screening for adequate collateral circulation alone by using the PI studies. At the initiation of the study, RA harvest was from the nondominant arm. Therefore, only this arm was evaluated. After the 25th patient, if the study result was positive (indicating inadequate collateral flow), the dominant arm was evaluated, and if adequate collateral circulation was present, the dominant arm was used for RA harvest. After the 60th patient to undergo RA harvest, the PI was performed, preoperatively and postoperatively, on both upper extremities. Thus, 26 (13%) patients with unilateral harvest included in this study did not have a PI generated for their unoperated arms.
Timing of studies
Postoperative testing duplicated the patients' preoperative testing protocol. Color flow and pulsed Doppler scanning (Acuson 128 Cardiovascular System; Acuson, Inc, Mountainview, Calif) of the RAs and ulnar arteries (UAs) was also performed. Preoperative testing, regardless of the studies performed, took place at any time before the surgical procedure but in general was performed 1 to 3 days before the operation. Postoperatively, photoelectric plethysmography was performed approximately 4 to 5 weeks after the operation, whereas postoperative Doppler evaluations were scheduled 8 to 10 weeks after surgical intervention.
Description of studies performed
Photoelectric plethysmography yielding both an oxygen saturation (SpO2) and a calculated PI was performed with a modified pulse oximeter.
11 The modified oximeter's digital probe was applied to the thumb, ensuring that the photoelectric sensor was correctly positioned over the nail bed, and a blood pressure cuff was applied to the arm. A baseline reading of the SpO2 and the perfusion to the thumbnail bed was obtained. The RA was located by means of palpation, and the blood pressure cuff was inflated to 200 mm Hg until the PI indicated approximately zero flow. The RA was compressed manually at the wrist, blood pressure cuff inflation was released, and compression of the RA was maintained for a period of approximately 45 seconds. Flow to the hand with RA occlusion is through the UA. After 45 seconds, the PI and SpO2 levels were recorded. The compression of the RA was released, the UA was located at the wrist, the blood pressure cuff was reinflated to 200 mm Hg, and UA was compressed. After 45 seconds, the PI and SpO2 were recorded. The procedure was then repeated on the contralateral arm. This protocol was used in an effort to duplicate somewhat what happens during the modified Allen test, interrupting blood flow to the hand, but our protocol does not require patient cooperation and does not disturb the digital sensor. The ratio of the PI with the RA occluded and the PI with no occlusion (baseline) was used as an index of blood flow to the hand independent of RA flow; this is the perfusion index ratio (PIR). The PIR with the RA occluded is indicative of the flow primarily through the UA. The criterion we used for a negative study indicating that RA harvesting would be safe was a ratio of 40% or greater.
12,13 Preoperative studies evaluated the contribution of both the RA and UA to thumb perfusion and SpO2. In the operated arms postoperatively, only the UA contribution could be assessed, whereas in the unoperated arms, both the contribution of the RA and that of the UA could be assessed.
Color flow and pulsed Doppler scanning were used to determine both the lumen diameter and flow velocities, respectively. The Doppler studies were performed by a 7.5-MHz phased array transducer. The Doppler beam, ranging from 3 to 10 MHz, is emitted from a ceramic crystal in the transducer head. The distal RAs and UAs were bilaterally evaluated for internal diameter and flow velocities at a point just proximal to the wrist crease. Cardiac monitor dots (Red Dot, model 2239; 3M Company, Medical Device Division, St Paul, Minn) were applied to the patient, and the patient was connected to the electrocardiography monitor of the Doppler machine. All recordings were made in systole as noted by the onset of the QRS complex from the electrocardiography tracings. Aqueous gel (Aquasonic 100; Aquasonic Parker Laboratory, Orange, NJ) was applied on the transducer head, and the probe was positioned at a 60° angle to the volar aspect of the forearm. Once the arterial signal was detected, the image was frozen on screen, and the diameter was calculated with internal calipers. Flow was determined by relocating the artery and toggling the pulse-velocity switch. Internal calipers were used to measure the flow velocities obtained. Preoperative studies evaluated both the RAs and UAs bilaterally. Postoperatively, the UA diameter and flow velocity were calculated in the operated arm or arms, and both the RA and UA diameters and flows were obtained in the unoperated arms.
Postoperative clinical evaluation included examination for compromise in hand function, including strength, grip, ability to perform fine motor function, and hand ischemia.
Analysis of the data was performed with SPSS for Windows, version 6.1 (SPSS, Inc, Chicago, Ill). Means, SDs, and ranges were calculated. Data were compared with paired t tests. The unoperated arms from patients with unilateral RA harvest served as controls.
| Results |
|---|
|
|
|---|
There was a statistically significant increase in the preoperative to postoperative SpO2 in the unoperated extremities, from 95.09% to 96.13% (n = 98, P < .001), and a corresponding statistically significant change in the operated arms of 94.99% to 95.78% (n = 223, P < .001). There was no significant difference in the baseline preoperative SpO2 or in the postoperative values between unoperated and operated arms (n = 110). These statistical differences do not translate into clinical significance because all saturations, preoperatively and postoperatively, in both operated and unoperated arms, are within the normal physiologic range of 92% or greater.
The results for the PI are summarized in Table 1. There was no difference in the baseline preoperative PI of the unoperated and operated extremities (n = 87) and in the postoperative baseline flows (n = 115). A significant difference existed between the calculated preoperative PIRs during RA occlusion between the unoperated and operated extremities, measuring 0.65 and 0.77, respectively (P = .001), when all patients evaluated were included. However, when the value for the unoperated patients who had positive results are removed, there is no difference in the baseline ratios during RA occlusion (PIR of 0.75 for the unoperated arms and 0.76 for the operated arms, P = .658). There was no difference in the PIR during UA occlusion (P = 653) between the unoperated extremities preoperatively to postoperatively (P = .083). A significant decline in the PI was noted in both the unoperated and operated arms after the operation (P
.001 for both): a 30.4% decline in the unoperated arms (n = 80) and a 6% further decline in the operated arms (n = 258), which represents the loss of the RA flow to the hand. This further decline in RA perfusion was not significant (P = .252).
|
|
|
Qualitative assessment of forearm and digital perfusion revealed no evidence of ischemia. Two patients reported mild sensitivity to extremely cold temperatures (approximately 0°F-20°F) during the winter; the sensitivity diminished by wearing warmer gloves. After RA harvest, there was no loss of fine or gross motor function, nor did patients have functional limitations. Ten patients reported transient dysesthesias in operated extremities along the distribution of the superficial branch of the radial nerve or the lateral antebrachial cutaneous nerve, which are not vascular in cause; all resolved spontaneously and completely without further sequelae.
| Discussion |
|---|
|
|
|---|
There has been unsubstantiated trepidation in the harvesting of the RA for fear of ischemic sequelae in the hand, sensitivity to cold temperature, or development of hand claudication. In the current surgical era, the issue of arterial insufficiency caused by a loss of radial blood flow has been addressed in the context of RA cannulation and radial forearm flaps.
7-9,19 Analogous to physiologic compensatory changes seen in other parts of the vascular system when vascular occlusion occurs as a result of thrombi or stenosis, the growth of the UA's diameter and flow velocity can be postulated to occur to compensate for the loss of the RA's contribution to the hand's blood supply. A few studies evaluated the effect of RA harvest on digital and UA blood flow preoperatively and postoperatively.
1-3 These studies demonstrate that compensatory changes in flow occur to maintain adequate hand circulation in selected patients.
Our own study used photoelectric plethysmography, providing both SpO2 and a measure of digital blood flow. Pulse oximetry determines the amount of hemoglobin SpO2 by comparing the differences in light absorbed by oxygenated and unoxygenated hemoglobin.
20 The normal range for SpO2 is 92% or greater. Pulse oximetry with an unmodified pulse oximeter alone is not adequate to measure flow; that is, the pulse oximeter simply measures the amount of oxygenated hemoglobin passing through its sensor. Hemoglobin may remain saturated at flows consistent with ischemia that are inadequate to support circulation to the hand.
19,21-23 In our study all mean values for SpO2 were within physiologic parameters, and no clinical significance corresponds to saturations of greater than 92%. Therefore, the slight increase in SpO2 noted postoperatively was merely a statistical observation. In 31 patients who had a positive PIR test result preoperatively (demonstrating inadequate collateral circulation),
24 an SpO2 of 92% or greater during RA occlusion was maintained. Two patients had saturations of less than 92% at baseline that did not further drop during RA compression. Of interest, in these 31 patients 20 had normal Allen tests results, 7 patients had positive Allen test results, and the test could not be performed in 4 patients. The results of the Allen test had a weak negative correlation that was not significant compared with the PIR with the RA occluded (r = -0.2338, P = .241).
The modified pulse oximeter used in our study is capable of assessing digital flow by using the principle of photoelectric plethysmography. PI correlates with digital blood flow with venous occlusion plethysmography (r = 0.94).
11 The PI is obtained by using a modified pulse oximeter; a standard oximeter finger probe is applied over the nail bed. PI is determined by the change in the path of infrared light emitted from the sensor through the digit in both systole and diastole; by measuring the varying intensity, the change in the path length can be calculated. Greater blood flow into the digit being scanned corresponds to a greater PI.
11 In this study postoperative PI measurements performed at rest declined significantly in both the operated and unoperated extremities. There was a 33% decline in the unoperated extremities and an additional 6% decrease in operated arms, which we believe represents the loss of the RA to the thumb's circulation. It must be noted that the PI, like oximetry, is subject to variation because of changes associated with vasoconstriction induced by ambient temperature and sympathetic tone reflecting changes in the physiologic state.
20,23,24 Venous congestion, as in cardiac failure, carbon monoxide concentration (typical in cigarette smokers), tachyarrhythmias, and anemia have also been shown to influence these measurements.
20 Therefore, the drop in the absolute values noted by the baseline flows may represent factors other than decline in digital perfusion. The further drop in the operated to the unoperated extremities cannot be fully explained by these factors except for the surgical removal of the RA. The integrity of the PI as a determinant of the UA's contribution to digital blood flow is demonstrated by lack of difference between the ratio obtained in the unoperated arms during RA occlusion before (PIR = 0.66) and after surgical intervention (PIR = 0.61, P = .139).
Ultrasound Doppler studies can not only provide measures of vessel diameter and blood flow velocity but can also determine the patency of the RApalmar arteryUA circuit during contralateral vessel occlusion.
1-3,25-29 The Doppler studies performed as part of this study were used to quantify the short-term effects of the loss of the RA on the contralateral vessel diameter and in flow to the hand and not to determine the patency of the collateral circuit. In our study, color flow Doppler scanning revealed an 11% increase in UA diameter, preoperatively to postoperatively, in the arms from which the RA was harvested. UA blood flow velocity, as measured with pulsed Doppler scanning, increased by about 20% postoperatively. Our findings are consistent with those of Mulvany and associates
30 and Pola and colleagues,
1 who found that increasing flow appears to be a powerful stimulus to growth, resulting in an increased lumen.
This noninvasive study prospectively examined the forearm and digital circulation in 386 upper extremities in patients who underwent coronary artery bypass grafting with RA conduits. A significant increase in UA diameter and blood flow velocity 8 weeks after RA harvest was noted. Thumbnail SpO2 remained within physiologic parameters with a slight, nonstatistically significant decline in the PI of the thumb of the arms in which the RA was harvested compared with the arms in which the RA has not been harvested. Clinically, blood flow remains adequate to meet the physiologic demands of the upper extremity without causing hand ischemia or functional limitations. The infrequently noted (3.7% of the operated forearms) transient dysethesias are not vascular in origin because they occurred in the distribution of the lateral antebrachial cutaneous nerve or the superficial branch of the radial nerve, both of which are vulnerable during harvest. Thus, the compensatory mechanisms available to maintain blood flow and perfusion appear adequate in the short term after RA harvest in properly selected patients to sustain arterial inflow to the hand without clinical functional limitations in those extremities deemed to have forearm and hand collateral circulation by preoperative assessment with photoelectric plethysmography.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Brzezinski, T. Luisetti, and M. J. London Radial Artery Cannulation: A Comprehensive Review of Recent Anatomic and Physiologic Investigations Anesth. Analg., December 1, 2009; 109(6): 1763 - 1781. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Kohonen, O. Teerenhovi, T. Terho, J. Laurikka, and M. Tarkka Is the Allen test reliable enough? Eur. J. Cardiothorac. Surg., December 1, 2007; 32(6): 902 - 905. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Jaworski, P. Siondalski, K. Jarmoszewicz, and J. Rogowski Arm temperature distribution in thermographic pictures after radial artery harvesting for coronary bypass operation Interactive CardioVascular and Thoracic Surgery, October 1, 2007; 6(5): 598 - 602. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. R. Sajja, G. Mannam, N. R. Pantula, and S. Sompalli Role of Radial Artery Graft in Coronary Artery Bypass Grafting Ann. Thorac. Surg., June 1, 2005; 79(6): 2180 - 2188. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Hata, M. Shiono, A. Sezai, M. Iida, A. Saitoh, T. Hattori, S. Wakui, M. Soeda, N. Negishi, and Y. Sezai Determining the best procedure for radial artery harvest: Prospective randomized trial for early postharvest complications J. Thorac. Cardiovasc. Surg., April 1, 2005; 129(4): 885 - 889. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-S. Lee, Y. J. Heo, and B.-C. Chang Long-term digital blood flow after radial artery harvesting for coronary artery bypass grafting Eur. J. Cardiothorac. Surg., January 1, 2005; 27(1): 99 - 103. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. G. Royse, C. F. Royse, A. Maleskar, and A. Garg Harvest of the radial artery for coronary artery surgery preserves maximal blood flow of the forearm Ann. Thorac. Surg., August 1, 2004; 78(2): 539 - 542. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. G. Wallach Cannulation Injury of the Radial Artery: Diagnosis and Treatment Algorithm Am. J. Crit. Care., July 1, 2004; 13(4): 315 - 319. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Manabe, N. Tabuchi, M. Toyama, T. Yoshizaki, M. Kato, H. Wu, M. Kotani, and M. Sunamori Oxygen pressure measurement during grip exercise reveals exercise intolerance after radial harvest Ann. Thorac. Surg., June 1, 2004; 77(6): 2066 - 2070. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-S. Lee, B.-C. Chang, and Y. J. Heo Digital blood flow after radial artery harvest for coronary artery bypass grafting Ann. Thorac. Surg., June 1, 2004; 77(6): 2071 - 2074. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. C.F. Chong, P. J.L. Ong, C. S. Hayward, P. Collins, and N. E. Moat Effects of radial artery harvesting on forearm function and blood flow Ann. Thorac. Surg., April 1, 2003; 75(4): 1171 - 1174. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||