J Thorac Cardiovasc Surg 2002;123:768-776
© 2002 The American Association for Thoracic Surgery
Evolving Technology (ET) |
From the Divisions of Cardiovascular Surgery,a Cardiology,b and Radiology,c School of Medicine, Albert-Ludwigs-University Freiburg, Freiburg, Germany.
Dr. Lutter is supported by the Clinical Cardiovascular Research Center II and III at the Medical Center (grant B3 and 11-00-III), Albert-Ludwigs-University of Freiburg, Germany.
Received for publication May 15, 2001. Revisions requested Sept 14, 2001; revisions received Oct 1, 2001. Accepted for publication Oct 8, 2001. Address for reprints: Georg Lutter, MD, Division of Cardiovascular Surgery, University of Freiburg, Medical Center, Hugstetter Str 55, D-79106 Freiburg, Germany (E-mail: lutter{at}ch11.ukl.uni-freiburg.de).
| Abstract |
|---|
|
|
|---|

max of 5.4 ± 3.3 mm Hg for the descending aorta group, 5.4 ± 1.2 mm Hg for the supracoronary group, and 5.4 ± 1.1 mm Hg for the subcoronary group), which did not differ from their in vitro gradients. Two-dimensional echocardiography demonstrated complete valvular closure and opening in 5 of 5 cases. Angiography indicated only a physiologic jet of regurgitation (0°) in 8 animals and mild (I°) regurgitation in 3 animals. Color Doppler ultrasonography indicated no regurgitation in 5 of 5 cases and minor paravalvular leakage in 1 case. | Introduction |
|---|
|
|
|---|
|
A number of minimally invasive techniques for replacing heart valves have been experimentally developed over the last 10 years. Such techniques have been reported by Andersen and coworkers,
1 who indicated in 1992 that artificial aortic valves can be implanted in closed-chest animals by means of a transluminal catheter technique; by Pavcnik and colleagues,
2 who demonstrated in 1992 that a percutaneous transcatheter placement of an artificial caged-ball valve in animals is feasible; by Maozami and coworkers,
3 who showed in 1996 that a hemodynamically acceptable prosthetic aortic valve for transluminal placement is feasible, whereas an open chest model for implantation was used; and by Bonhoeffer and coworkers
4 in 2000, who successfully performed a percutaneous pulmonary valve implantation in 5 sheep. Although the instruments, devices, and techniques reported are clearly experimental, it appears that each of them still has one or more problems.
It would be especially advantageous if a defective heart valve could be removed by means of an endovascular procedure (ie, through the femoral artery). The procedure is then carried out percutaneously and transluminally by using the vascular system to convey appropriate devices to the aortic anulus to carry out the percutaneous aortic valve ablation procedure.
We set out to develop a technique for placement of a biologic valve in the aortic position through a transluminal approach.
5 This report presents our initial experience with transluminal aortic valve implantation in the circulatory system of anesthetized pigs after this experimental set up had been developed in more than 30 pilot studies.
| Materials and methods |
|---|
|
|
|---|
Pericardial valves were used in the other 3 animals. Pericardium of the same animals was selected for the creation of valves. Pericardium was cleaned, prepared, preserved with glutaraldehyde (0.6%), and finally constructed with an ultra-thin polytetrafluroethylene layer to the exterior for stabilization (by D.K.).
The above-described valves were sutured to the interior of a stent with 7-0 polypropylene sutures. This radially self-expanding vascular memory stent (21-28 mm in length and 15-23 mm in diameter) includes several cylindrical fine-wire sections that are interconnected into a single tubular structure. The size of the valved stent correlated with the width of the aorta (as determined with intra-arterial digital subtraction angiography [IA-DSA]) and body weight. The wires are made of a soft and highly malleable alloy consisting of nitinol (Nitinol Devices & Components Inc, Fremont, Calif). The tubular portion of the trimmed ascending aorta of the stentless aortic valve spanned the partial axial length of the nitinol stent (Figure 1). For the anchoring fixation, barbs are attached to the outside of the stent.
|
Catheter delivery system
The valved stent is inserted by means of a self-constructed, flexible, catheter-based (external diameter, 22F) delivery technique (overall length, 90 cm) through the iliac artery or the infrarenal aorta (Figure 2).
|
|
In vivo studies
Animals received humane care, as approved by the Center for Experimental Animal Research at Freiburg University and in compliance with the "Guide for the Care and Use of Laboratory Animals" prepared by the Institute of Laboratory Animal Resources, National Research Council, and published by the National Academy Press, revised 1996.
Pigs of the "German Landrace" weighing 64 to 76 kg were premedicated and anesthetized, as reported elsewhere.
6 Only in vitro-tested valved stents with a pressure gradient of less than 7 mm Hg and a regurgitation of I° or less were used for transluminal aortic valve implantation in vivo. First, a paracolic incision and an extraperitoneal approach toward the infrarenal aorta, aortic bifurcation, or the left iliac artery was prepared. This extraperitoneal access was chosen because femoral arteries of pigs weighing 65 to 75 kg are only 2.5 to 4.0 mm in diameter.
Second, through the iliac communis artery (n = 3), the infrarenal aorta (n = 10), or the iliac external artery (n = 1), a guidewire was retrogradely positioned into the proximal descending aorta or the left ventricle under continuous fluoroscopy. The valved stent (21-28 mm in length, Figure 1
) was then hand crimped onto the proximal portion (35 mm in length) of a catheter delivery system (Figure 2
). The diameter of the memory stent was dependent on body weight, and the diameter of the ascending or descending aorta was measured by using IA-DSA. The appropriate size of the valved stent was chosen from different available stent diameters, ranging from 15 to 23 mm. Because of the expandable character of the valved stent, minimal oversizing (1-2 mm) did not affect optimal valve function. The assembly was front loaded in a 22F long sheath to enable a precise positioning (Figure 2
). In case of deployment into the ascending aorta, the ostia of the coronary arteries were identified by means of angiography for correct placement of the valved stents. The valved stent was placed at a secure distance of several millimeters from the coronary ostia into the ascending aorta (supra-annular placement) and from the native aortic valve into the left ventricular outflow tract (subannular placement).
Third, after loading onto the previously positioned guidewire, the whole system was advanced and inserted into the iliac artery or infrarenal aorta through a 24F port. The position of the valved stent was easily tracked fluoroscopically (highly radiopaque stent material) and supported by means of transesophageal echocardiography (SSH-140A, Toshiba GmbH). The valved stent was slowly released (memory stent) by pulling the securing sheath back and deployed in the ascending (subcoronary or supracoronary position) or proximal descending (distal to the subclavian artery) aorta. The duration of deployment was 1.7 ± 1.2 minutes. Through the self-expanded valved stent, the catheter delivery system was subsequently pulled back and removed, leaving the valve assembly in the desired position. After deployment of the valved stent, the observation period of all experimental groups lasted 164 ± 48 minutes (range, 90-240 minutes). Data were obtained for all parameters at baseline and at the end of the observation period.
Experimental groups
Over a 2-year period, the in vitro and in vivo experimental model gradually evolved. After the first successful implantations of a valved stent in the pilot studies (n = 30, for details see "Discussion" section), the present preliminary study was started. The 14 animals fulfilled the following inclusion criteria: First, these valved stents demonstrated adequate performance in the in vitro experiments, and second, before implantation of these valved stents, their catheter delivery system has been successfully tested.
In the descending aorta group the in vitro-controlled homologous valved stents (21-28 mm in length, n = 6) were implanted into the proximal descending aorta (distal to the subclavian artery).
In the supracoronary group valved stents (n = 6, 28 mm of length) were implanted into the proximal ascending aorta (distal to the coronary ostia) of the anesthetized pigs.
In the subcoronary group 2 pigs received valved stents into the left ventricular outflow tract (shortly beyond the native valve), with a shorter stent of 21 mm in length.
Measurements
Assessment of transvalvular gradient
In vitro, pressure tranducers that were placed on the ventricular and the aortic side of the valved stent were used to analyze the transvalvular gradient shortly after implantation and after 4 hours (by D.K.). In vivo, blood pressure was measured online with two 5F Millar micromanometers placed in the proximal and distal area of the deployed valved stent.
Evaluation of valvular opening and closure
Two-dimensional ultrasonography in vitro (by D.K.) and 2-dimensional transesophageal echocardiography in vivo (by M.H.) were used to semiquantitatively grade the valvular opening and closure process (0, total valvular opening and closure; I, partial opening or incomplete closure of one or more leaflets; II, lack of movement of one leaflet; and III, lack of movement of 2 or more leaflets; Figure 4).
|
Regurgitation and paravalvular leakage
Color Doppler ultrasonography in vitro (by D.K.) and in vivo (by M.H.) and additional angiography in vivo were used to assess any valve insufficiency (0, no; I, mild; II, moderate; III, severe; and IV, fulminant regurgitation) and paravalvular leakage (0, none; I, mild; II, moderate; III, severe; and IV, fulminant leakage).
Mortality and complications
All complications (mechanical and others) and deaths were identified.
Duration of deployment
The time from pushing the valved stent out of the cranial catheter capsule (Figure 2
) and the total deployment into the left ventricular outflow tract or aortic lumen was recorded.
Macroscopic analysis
First, the desired positioning of the valved stent in situ was controlled (0, optimal positioning; I, <5 mm; II, >5 mm; III, >10 mm from the desired position). Second, the anchoring of the valved stents was determined in situ (0, sufficient anchoring; I, minimal shift; II, moderate shift; and III, twisting of the valved stent). Third, all grafts were explanted. Fourth, the valved stent was rinsed with distilled water to remove the blood and was macroscopically inspected. Gross leaflet structure and integrity were analyzed, referring to the grading of morphologic damage (0, no; I, mild; II, moderate; and III, severe morphologic damage) and thrombus formation (0, none; I, small; II, moderate; and III, fulminant thrombus formation).
Statistical analysis
Paired und unpaired t tests with the Bonferroni correction were used as appropriate. Values are reported as means ± SD. Data were used from all animals who survived the entire observation period (n = 11). Only data of valved stents that demonstrated sufficient anchoring (0°-I°) were analyzed.
| Results |
|---|
|
|
|---|

max, 4.5 ± 1.7 mm Hg) and pericardial valved stents (mean end-systolic 
max, 5.9 ± 1.4 mm Hg) were revealed. In addition, different pressure values compared with baseline were not observed.
Evaluation of valvular opening and closure
Semiquantitative assessment of the valvular opening and closure process demonstrated total valvular opening and closure in all cases.
Blood-flow characteristics
Color Doppler ultrasonography demonstrated systolic laminary blood flow in all cases (supravalvular, intravalvular, and subvalvular).
Regurgitation and paravalvular leakage
Minimal regurgitation was observed in 3 cases (0°-I°), whereas in 11 cases valve insufficiency was excluded (0°). Paravalvular leakage occurred in 3 of the cases because of a short distance between the valved stent and silicone tube caused by the barbs.
In vivo studies
Mortality and complications
One animal died of intractable ventricular fibrillation after repositioning of a guidewire with its catheter tip in the left ventricle. In addition, technical failure occurred in 2 pigs (supracoronary group) as a result of stent twisting (III°) in the ascending aorta. Therefore the length of the stent was increased from 21 to 28 mm, except in the subcoronary group. Afterward, no twisting occurred in subsequent animals. Eleven animals were analyzed after successful implantation. Surgical preparation, advancement of the catheter delivery system, and deployment of the valved stent were well tolerated in all but 2 animals, one with ventricular tachycardia and one with reversible ventricular fibrillation associated with placement into the ascending aorta. Bleeding caused by insertion of the catheter delivery system into the iliac artery or infrarenal aorta was controlled easily.
Duration of deployment
The duration of deployment of the valved stents was 1.7 ± 1.2 minutes on average. Differences between groups were not revealed.
Assessment of transvalvular gradient
At the end of the observation period, the valved stents did not reveal significantly higher transvalvular gradients (mean end-systolic 
max: 5.4 ± 3.3 mm Hg for the descending aorta group, 5.4 ± 1.3 mm Hg for the supracoronary group, and 5.5 ± 1.1 mm Hg for the subcoronary group) compared with their in vitro gradients.
Evaluation of valvular opening and closure
Two-dimensional echocardiography demonstrated complete valvular closure and opening (0°) in all 5 analyzed cases.
Blood-flow characteristics
Laminar blood flow was observed in 5 of 5 cases by using color Doppler ultrasonography to examine the transvalvular blood flow.
Regurgitation and paravalvular leakage
Angiography indicated no jet of regurgitation (0°) in 8 and mild (I°) regurgitation in 3 animals (2 in the supracoronary group and 1 in the subcoronary group). Color Doppler ultrasonography revealed no regurgitation in 5 of 5 cases. Paravalvular leakage (observed by means of IA-DSA) occurred in none of the descending aorta group and in one animal of the subcoronary and one of the supracoronary group. In contrast, color Doppler ultrasonography showed minor paravalvular leakage in 1 of 5 cases (descending group).
Macroscopic analysis
Good positioning (0°-I°) of the valved stent was reached in 8 of 11 cases. Two valved stents had 5 to 10 mm of distance (II°) and 2 valved stents had more than 10 mm of distance (III°) from the desired position in the descending artery (close to the subclavian artery). In one animal the supracoronary valved stent was implanted 6 to 7 mm (II°) too cranial (disturbing the outflow of the truncus brachiocephalicus), and in another case it shifted 2 to 3 mm (I°) toward the aortic arch. In the subcoronary group in one animal the valved stent was implanted 2 to 3 mm too proximal into the left ventricular outflow tract.
Macroscopic findings at autopsy demonstrated that the anchoring barbs of all stents were sufficiently embedded into the aortic wall in the descending aorta group (0°). In the ascending aorta group, in 2 cases the valved stent was sufficiently anchored in the aortic wall or the left ventricular outflow tract (0°), whereas 2 of the supracoronary group were not fully embedded (I°) because of minimal shift in the lumen. In the subcoronary group, in one animal only the distal barbs were anchored, whereas the proximal barbs did not reach the wall of the left ventricular outflow tract because of the positioning of the stent.
Morphologic damage did not occur in any of the placed stent assemblies (0°). The valves inside the stents were completely free of thrombus, except in one case (I°) in the subcoronary group, which presented a thrombus (2 x 3 mm) between the stent and the outflow tract. No aortic dissection, hematoma, or bleeding into the aortic wall was macroscopically seen in any of the pigs.
| Discussion |
|---|
|
|
|---|
Mortality and complications
First, 30 pilot studies served for the development of a suitable catheter delivery system, an optimal handling of the material, an anchoring mechanism, and a suitable visualization method. These pilot animals had a high complication and mortality rate (83%) caused by stent twisting, dislodgement, difficulties of advancing the catheter delivery system (around the aortic arch), vessel perforation (because of rigidity of the delivery system), and myocardial ischemia (caused by placing the stent too close to the coronary ostia).
However, after improvement of material and insertion techniques among the 14 animals of this study group, all but one survived the procedure and the following observation period. More secure and easier placement of the aortic valve assembly was accomplished after increasing the height of the stent (except for in the subcoronary group) in the ascending aorta. Furthermore, the anchoring barbs were necessary for stable deployment into the aorta.
Transvalvular gradient and regurgitation
Many authors have highlighted the necessity for caution in using in vitro results to assess a valve's performance in vivo.
8,9 Therefore the valved stents, which revealed a pressure gradient of less than 7 mm Hg and a regurgitation of I° or less in the circulatory system, were used and retested in vivo. As the gold standard, the natural aortic valve provides minimal obstruction to the ventricular outflow during systole. In this sense it is considered to provide maximum forward flow efficiency. By comparison, the prosthetic or bioprosthetic valve is more obstructive and hence less efficient than the natural valve. In all our experiments, transvalvular pressure measurements indicated very low gradients several hours after placement of the valved stents. Relevant stenotic transvalvular gradients were also not observed in other experimental studies
1,3,4 in which smaller series of valved stents were analyzed.
Although the size of the valved stent correlated with the width of the aorta (determined with IA-DSA) and body weight, mild regurgitation was observed in 3 of 11 cases and mild paravalvular leakage in 2 of 11 cases, mainly caused by a slightly tilted stent that could not be perfectly expanded in the ascending arched aorta. These findings demonstrate that the valved stent effectively prevents regurgitation when the valve has an optimal size and is anchored properly in the descending aorta group.
In general, a suitable positioning (which might require a slightly different stent design with the same valve design as proposed below) and size of the valved stent leads to unobstructed forward blood flow and prompt opening and closure. Therefore more experience is required with the angiographic sizing used in this study and also with echocardiographic sizing for subsequent insertion of valved stents. Furthermore, one can postulate that the dynamic interaction between the valved stent and aorta (which changes its size during the cardiac cycle) could be another reason for the mild regurgitation observed in these 3 animals, as opposed to the in vitro data of these same 3 valves, which were collected in a rigid system.
9
Stent design
In 1992, Dake and coworkers
10 started implantations of stent graft prostheses in patients with thoracic aneurysms. Since then, improvements in interventional catheterization and stent technology have made percutaneous stent placement a routine procedure, and more than 20 noncovered and covered stent designs have been developed to fit the multiple cardiovascular indications.
11 Therefore experience with this kind of introducer system and different stent assemblies has been already acquired in other applications.
Limitations of the study
Continuous electrocardiographic, hemodynamic, and pulse oximetric monitoring was used throughout the procedure. Nevertheless, hemodynamic parameters and global cardiac function should be analyzed in detail in further studies (eg, to describe the changes during implantation and deployment of valved stents). Changes of the coronary blood flow after supracoronary valved stent implantation should be included in subsequent studies. Unfortunately, this has also not been performed in other studies.
1-4 Because of the echo-dense structure of the stent, echocardiographic assessment could not adequately be performed in all cases.
Because of the unfeasibility of the aortic valved stent implantation into the aortic anulus, the valve had been evaluated in the subcoronary and supracoronary position. This experimental positioning of the valved stent was only used because native porcine leaflets were not removed in this study and will therefore occlude coronary ostia. As Andersen and colleagues
1 already showed in their porcine model, the implantation of a valved stent into the annular position is impossible because of the restriction of coronary blood flow.
In contrast, Pavcnik and colleagues
2 experimentally showed that transannular placement of a prosthetic caged-ball aortic valve in mongrel dogs is feasible. Their ring assembly was successfully placed below the ostia of coronary arteries and was securely anchored against the anulus of the aorta by the barbs, and no damage to the surrounding structures was macroscopically noted.
The degeneration of bioprosthetic valves has been widely studied in recent years. Various causative factors are involved that influence the durability of the valve to different degrees.
12 Because only acute studies were performed, the long-term durability of this kind of aortic valved stent is unknown. Questions regarding thrombogenicity, neointimalization, calcification, stability of position, histomorphologic examination, and function during long-term follow-up should be addressed. Percutaneous implantation is one of several steps toward an ideal percutaneous aortic valve replacement (ie, in human subjects with a calcified stenotic aortic valve). Development of various techniques for implementing this replacement should be also considered. Percutaneous aortic valve ablation techniques for aortic stenosis, a stable scaffold enabling those intra-aortic procedures, filters avoiding systemic embolization, and circulatory support during percutaneous aortic valve ablation and implantation with a femoro-femoral bypass and left ventricular venting should be developed. In our short-term study the duration of deployment was about 2 minutes. Nevertheless, a longer operation time (>3 minutes) in the ascending aorta will be necessary for percutaneous aortic valve ablation and implantation, and therefore hemodynamic instability can be expected. More complex procedures at the native aortic valve using ablation techniques should only be performed with a circulatory support.
Future techniques will allow transluminal aortic valve ablation, which will be the biggest challenge of this approach. Different ablation devices have to be tested to achieve this. Preliminary in vitro studies of our laboratory demonstrated the possibility of ablating human calcified aortic valves with 3 different types of lasers (CO2, Hol:YAG, and Erb:YAG lasers). Complete ablation of each aortic cusp took 2.5 hours with the CO2 laser and 1 hour with the Hol:YAG laser. The erbium laser was checked for the excision of a cusp. This also took 1 hour (unpublished data). In subsequent studies ablation time will be reduced by optimizing the laser parameters. Examination of the remaining debris solution of the lasered aortic cusps indicated debris particles with a size between 4 µm and 1 mm. Therefore these particles have to be eliminated by using a filter system, which has not yet been developed.
Although a great deal of work remains to be done, these preliminary results indicate that the development of a prosthetic aortic valved stent for transluminal placement is feasible.
1-5 Probably this valved stent would be applicable to aortic insufficiency caused by valves that are pliable enough to be displaced against the aortic wall in the subcoronary region but do not adequately conform to it. In patients with fulminant endocarditis and aortic insufficiency, the valved stent might be implanted in the descending aorta for hemodynamic stabilization. Another possibility will be the implantation of a valved stent into the pulmonary position, which has already been shown by Bonhoeffer and colleagues.
4
An ideal aortic valved stent may be achieved for long-term use by individually lining the entire aortic root (including the sinus), ascending aorta, and aortic arch with stent material (including a bioprosthesis) to avoid any shift caused by shear stress, with individually designed marked wholes for the coronary and supra-aortic arteries. An additional anchoring of stents into the supra-aortic vessels and the ascending valved stent should be considered.
| Conclusion |
|---|
|
|
|---|
| Discussion |
|---|
|
|
|---|
A second aspect is this: We also know from clinical practice, that if you leave even a small peripheral leak, it has some very often important deleterious effect, even a rather small leak, either hemolysis or insufficiency. Knowing that you will never be able to remove all the calcium formation, then you have a high risk of having peripheral leak. How can you solve that problem?
Dr Lutter. Thank you very much for your kind remarks, Dr Carpentier.
With respect to your first remark, to ablate the calcified, stenotic aortic valve will be a challenge, and this has been a big problem up to now. We have performed several pilot ablation studies and can say that different lasers can microscopically ablate the aortic valve, but we have not had techniques up to now that can, in the ascending aorta, ablate the calcified valve.
So you are right also with your second remark regarding the small leaks and the possible upcoming insufficiency if you perform a valve stent implantation in a calcified anulus. And you have to be sure that every particle, every embolus that might come up to the supra-aortic vessels, you have to filter, and this is truly a lot of work to be done. Therefore, we think that, as surgeons, we should start with this work; otherwise the cardiologists will start with this problem. Therefore, we think we have to have a scaffold in the ascending aorta, with which we can properly work.
Dr Hani Shennib (Montreal, Quebec, Canada). Dr Lutter, is this technology good or bad for surgeons? Could you see that this will ultimately be an interventional cardiology type of technique?
And also, when you propose that a femoro-femoral bypass would be a way to support the heart, is there a possibility of looking at coupling this with drugs, such as adenosine, to arrest the heart or to control hypertension?
And the other part of the question is that when you solve the problem, you create another problem. Is embolization control an issue here, and would you couple this with other technologies, such as protection catheters?
Dr Lutter. Regarding your first remark, is it bad or not for surgeons, this is a question for the future, and I cannot answer it now. So whether you can do it on a beating heart or not, that will be seen in the future.
And with respect to your second remark, regarding embolization control, we do not have it right now because if you have a total and very good filter in your ascending aorta, up to now, you have such resistance in the ascending aorta, which you cannot have on a beating heart.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A.-P. Kappetein, N. Piazza, J.-C. Laborde, P. P. de Jaegere, and P. W. Serruys Transapical implantation of a self-expanding aortic valve bioprosthesis -- animal feasibility study Eur. J. Cardiothorac. Surg., November 1, 2009; 36(5): 813 - 817. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Vergnat, R. Henaine, M. Kalejs, S. Bommeli, E. Ferrari, J.-F. Obadia, and L. K. Von Segesser A new self-expanding aortic stent valve with annular fixation: in vitro haemodynamic assessment Eur. J. Cardiothorac. Surg., June 1, 2009; 35(6): 970 - 976. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E. McRae, M. Rodger, and B. A. Bailey Transcatheter and Transapical Aortic Valve Replacement Crit. Care Nurse, February 1, 2009; 29(1): 22 - 36. [Full Text] [PDF] |
||||
![]() |
J. Ye, A. Cheung, S. V. Lichtenstein, L. A. Altwegg, D. R. Wong, R. G. Carere, C. R. Thompson, R. R. Moss, B. Munt, S. Pasupati, et al. Transapical transcatheter aortic valve implantation: 1-year outcome in 26 patients. J. Thorac. Cardiovasc. Surg., January 1, 2009; 137(1): 167 - 173. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Grube, L. Buellesfeld, R. Mueller, B. Sauren, B. Zickmann, D. Nair, H. Beucher, T. Felderhoff, S. Iversen, and U. Gerckens Progress and Current Status of Percutaneous Aortic Valve Replacement: Results of Three Device Generations of the CoreValve Revalving System Circ Cardiovasc Interv, December 1, 2008; 1(3): 167 - 175. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Quaden, T. Attmann, M. Schunke, D. Theisen-Kunde, J. Cremer, and G. Lutter Percutaneous aortic valve replacement: Endovascular resection of human aortic valves in situ. J. Thorac. Cardiovasc. Surg., May 1, 2008; 135(5): 1081 - 1086. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Davidson and D. S. Baim Percutaneous Aortic Valve Interventions Card. Surg. Adult, January 1, 2008; 3(2008): 963 - 971. [Full Text] |
||||
![]() |
W. R. Chitwood Jr. and E. Rodriguez Minimally Invasive and Robotic Mitral Valve Surgery Card. Surg. Adult, January 1, 2008; 3(2008): 1079 - 1100. [Full Text] |
||||
![]() |
J. G. Webb, S. Pasupati, K. Humphries, C. Thompson, L. Altwegg, R. Moss, A. Sinhal, R. G. Carere, B. Munt, D. Ricci, et al. Percutaneous Transarterial Aortic Valve Replacement in Selected High-Risk Patients With Aortic Stenosis Circulation, August 14, 2007; 116(7): 755 - 763. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. M. Flecher, J. W. Curry, T. M. Joudinaud, C. L. Kegel, P. A. Weber, and C. M.G. Duran Coronary flow obstruction in percutaneous aortic valve replacement. An in vitro study Eur. J. Cardiothorac. Surg., August 1, 2007; 32(2): 291 - 294. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. J. Antunes Off-pump aortic valve replacement with catheter-mounted valved stents.: Is the future already here? Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 1 - 3. [Full Text] [PDF] |
||||
![]() |
F. Beyersdorf Transapical transcatheter aortic valve implantation Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 7 - 8. [Full Text] [PDF] |
||||
![]() |
P. Tozzi, M. Pawelec-Wojtalic, D. Bukowska, V. Argitis, and L. K. von Segesser Endoscopic off-pump aortic valve replacement: does the pericardial cuff improve the sutureless closure of left ventricular access? Eur. J. Cardiothorac. Surg., January 1, 2007; 31(1): 22 - 25. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Handke, G. Heinrichs, U. Moser, F. Hirt, F. Margadant, F. Gattiker, C. Bode, and A. Geibel Transesophageal Real-Time Three-Dimensional Echocardiography: Methods and Initial In Vitro and Human In Vivo Studies J. Am. Coll. Cardiol., November 21, 2006; 48(10): 2070 - 2076. [Abstract] [Full Text] [PDF] |
||||
![]() |
U.A. Stock, I. Degenkolbe, T. Attmann, K. Schenke-Layland, S. Freitag, and G. Lutter Prevention of device-related tissue damage during percutaneous deployment of tissue-engineered heart valves J. Thorac. Cardiovasc. Surg., June 1, 2006; 131(6): 1323 - 1330. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Webb, M. Chandavimol, C. R. Thompson, D. R. Ricci, R. G. Carere, B. I. Munt, C. E. Buller, S. Pasupati, and S. Lichtenstein Percutaneous Aortic Valve Implantation Retrograde From the Femoral Artery Circulation, February 14, 2006; 113(6): 842 - 850. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Attmann, T. Jahnke, R. Quaden, A. Boening, S. Muller-Hulsbeck, J. Cremer, and G. Lutter Advances in Experimental Percutaneous Pulmonary Valve Replacement Ann. Thorac. Surg., September 1, 2005; 80(3): 969 - 975. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Ma, P. Tozzi, C. H. Huber, S. Taub, G. Gerelle, and L. K. von Segesser Double-crowned valved stents for off-pump mitral valve replacement Eur. J. Cardiothorac. Surg., August 1, 2005; 28(2): 194 - 198. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Lutter and J. Cremer Editorial comment: Off-pump mitral valve replacement: an attack on conventional heart surgery? Eur. J. Cardiothorac. Surg., August 1, 2005; 28(2): 198 - 199. [Full Text] [PDF] |
||||
![]() |
R. Quaden, T. Attmann, A. Boening, J. Cremer, and G. Lutter Percutaneous aortic valve replacement: resection before implantation Eur. J. Cardiothorac. Surg., May 1, 2005; 27(5): 836 - 840. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Lutter, R. Ardehali, J. Cremer, and P. Bonhoeffer Percutaneous Valve Replacement: Current State and Future Prospects Ann. Thorac. Surg., December 1, 2004; 78(6): 2199 - 2206. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. von Segesser, B. Marty, P. Tozzi, and P. Ruchat Impact of endoluminal stenting for aortic surgery Eur. J. Cardiothorac. Surg., December 1, 2004; 26(Suppl_1): S14 - S18. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. K. von Segesser Direct percutaneous valve replacement: the next step? Eur. J. Cardiothorac. Surg., November 1, 2004; 26(5): 873 - 874. [Full Text] [PDF] |
||||
![]() |
M Ferrari, H R Figulla, M Schlosser, I Tenner, I Frerichs, C Damm, V Guyenot, G S Werner, and G Hellige Transarterial aortic valve replacement with a self expanding stent in pigs Heart, November 1, 2004; 90(11): 1326 - 1331. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. H. Huber, P. Tozzi, A. F. Corno, B. Marty, P. Ruchat, P. Gersbach, M. Nasratulla, and L. K. von Segesser Do valved stents compromise coronary flow? Eur. J. Cardiothorac. Surg., May 1, 2004; 25(5): 754 - 759. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Vahanian and I. F. Palacios Percutaneous Approaches to Valvular Disease Circulation, April 6, 2004; 109(13): 1572 - 1579. [Full Text] [PDF] |
||||
![]() |
J. Q. Zhou, A. F. Corno, C. H. Huber, P. Tozzi, and L. K. von Segesser Self-expandable valved stent of large size: off-bypass implantation in pulmonary position Eur. J. Cardiothorac. Surg., August 1, 2003; 24(2): 212 - 216. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. F. Corno, J. Zhou, P. Tozzi, and L. K. von Segesser Off-bypass implantation of a self-expandable valved stent between inferior vena cava and right atrium Interactive CardioVascular and Thoracic Surgery, June 1, 2003; 2(2): 166 - 169. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Lutter and F. Beyersdorf Reply J. Thorac. Cardiovasc. Surg., March 1, 2003; 125(3): 742 - 743. [Full Text] [PDF] |
||||
![]() |
Y. Boudjemline and P. Bonhoeffer Percutaneous valve insertion: A new approach? J. Thorac. Cardiovasc. Surg., March 1, 2003; 125(3): 741 - 742. [Full Text] [PDF] |
||||
![]() |