J Thorac Cardiovasc Surg 2002;123:977-983
© 2002 The American Association for Thoracic Surgery
Evolving Technology (ET) |
From the Oxford Heart Centre,a Oxford, and The Heart Hospital, London, United Kingdom; Jarvik Heart Corp,b New York, NY; Texas Heart Institute,c Houston, Tex; and Department of Otolaryngology,d Radcliffe Infirmary, Oxford, United Kingdom.
Supported by charitable donations from the National Heart Research Fund and the Artificial Heart Fund, UK. Robert Jarvik, MD, is the chief executive of Jarvik Heart Inc. None of the other investigators has a financial interest in the company.
12/1/121045doi:10.1067/mtc.2002.121045 Address for reprints: Stephen Westaby, PhD, FRCS, MS, Oxford Heart Centre, John Radcliffe Hospital, Headington, Oxford OX3 9DU, United Kingdom (E-mail: swestaby{at}AHF.org.uk).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
|
|
| Patients and methods |
|---|
|
|
|---|
Device and power system
The Jarvik 2000 heart is implanted into the apex of the failing left ventricle, from which a 16-mm impervious Dacron graft conveys blood to the descending thoracic aorta (Figure 1
).
6 The loosely braided flexible power cable is composed of a bundle of 9 individually insulated pacemaker-type leads, which are extrusion insulated with polycarbonate polyurethane. The internal cable (Figure 2) terminates in a connector within the percutaneous titanium pedestal (Figure 3).
The external power cable, which can be replaced in the event of damage, connects from the skull-mounted pedestal to the belt-worn control system, which in turn is connected to the battery (Figure 4).
The cable uses a special connector that plugs directly into the skull-mounted pedestal and is keyed so that it cannot be connected improperly. The pump is controlled by the external all-analogue control system in which the motor pump speed is set manually and maintained at the set level by a pulse width, modulated, speed-control circuit. The control unit incorporates a watt meter, which displays power use and alarm functions for the pump under speed, low battery reserve, or circuit failure. The belt-worn lithium ion (Li-ion) battery provides more than 8 hours of support at nominal flows of 3 to 4 L/min under normotensive conditions. At higher flow or pressure, the belt-worn packs discharge faster. The controller and battery system weighs only 1 kg.
|
|
|
The pedestal is attached to the parietal bone about 5 cm behind and slightly above the ear (Figure 2
). A relatively flat area of skull is chosen, taking care to avoid the mastoid air cells, the mastoid emissary vein, or the lateral or sigmoid sinus. This part of the cranium behind the ear lies within the sphere of natural awareness and avoids trauma to the connector.
For the first patient, we used 3 separate neck incisions passing across the midline to zigzag the cable for tension relief (Figure 5). We now consider the zigzag to be unnecessary and use one incision over the trapezius muscle on the posterior aspect of the neck to provide a gentle curve between the thoracic exit and the skull pedestal.
|
The operation began with simultaneous exposure of the parietal bone and femoral vessels by different teams. In the event of profound hemodynamic deterioration, heparinization, cannulation, and cardiopulmonary bypass were then possible. The skull pedestal was implanted by an experienced cochlear implant surgeon (A.F.). A cork bore-type instrument the same diameter as the percutaneous part of the titanium pedestal was used to tunnel through the skin to the periosteum at the chosen site. An appropriately sized skin defect was thereby created. A C-shaped incision was then made around this on the basis of the occipital arterial supply, and a full-thickness flap was raised down to the periostium (Figure 2
, A). The periostium was elevated beneath the skin flap, and a template was used to define the position of the bone screws. Any skull irregularity was burred off to give a flat surface. A dental drill was used to penetrate the external table before inserting the self-tapping screws.
Meanwhile, the left thoracotomy was performed through the sixth intercostal space to expose the apex of the left ventricle and the descending thoracic aorta. The Jarvik 2000 heart and power cable were brought into the surgical field, and the incisions were made on the shoulder and neck to convey the 3-pin connector and power cable to the skull-pedestal site. This was achieved by inserting the 3-pin connector within the end of an intercostal drain. The drain was withdrawn out of the chest and through the neck to the scalp. The 3-pin connector was then inserted through the titanium pedestal (Figure 3
, A). This was then implanted firmly onto the external table by using 7- or 8-mm bone screws (Figure 3
, B). Bone dust from flattening of the external table or the screw hole was used to promote osseointegration. The postauricular skin flap was repositioned with the percutaneous pedestal through the punched-out defect. The scalp and skin incisions were then closed securely. The external power cable was attached to the skull pedestal, and with the Jarvik 2000 heart in a bowl of saline, the power was switched on to test the circuit.
A side clamp was then applied to the descending thoracic aorta, to which a 16-mm Dacron graft was anastomosed. Systemic heparinization was then undertaken, and the left femoral artery and vein were cannulated for cardiopulmonary bypass.
The pericardium was opened widely anterior to the left phrenic nerve. The apex of the globular left ventricle was identified and cored, and the Jarvik 2000 restraining cuff was sewn into position. The Jarvik 2000 heart was then inserted through the cuff and fixed in position with circumferential tapes. The Dacron graft attached to the pump was trimmed to length and anastomosed to the graft on the descending aorta. Meanwhile, the heart was defibrillated and filled to dispel air. When the whole system was thoroughly deaired, the pump was switched on, and the patient was weaned from cardiopulmonary bypass.
Postoperative care
The percutaneous pedestal required very little maintenance.
4 On the first postoperative day, the dressings were removed, and the site was cleaned and kept dry. The external cable was disconnected, and blood clot or debris was cleaned from the surfaces. The patients tolerated discontinuation of pump flow for up to 5 minutes, with a functional aortic regurgitation flow of 800 to 1000 mL/min. For the first 3 months during the process of osseointegration, every attempt was made not to apply weight or tension to the pedestal.
| Results |
|---|
|
|
|---|
In the second (112 kg) patient the implant seemed uneventful, and he initially recovered from anesthesia and moved all limbs to command. However, severe coagulopathy resulted in excessive blood transfusion with pulmonary hypertension and the need for prolonged sedation and positive-pressure ventilation. Ten days after the operation, he was anticoagulated with warfarin, but on discontinuing sedation, it was apparent that there was right-sided weakness. A CT scan revealed a subdural hemorrhage adjacent to the site of the pedestal (Figure 6). Craniotomy was performed, and the pedestal was detached and repositioned in a subcutaneous pocket. Warfarin was discontinued. The site healed well, but because of his size and neurologic impairment, it proved difficult to wean him from the ventilator. The combination of positive-pressure ventilation and pulmonary hypertension exacerbated tricuspid regurgitation and right heart failure. Although the pump continued to function satisfactorily without anticoagulation, the patient died 3 months after implantation.
|
The system is user friendly in the outpatient setting. The patients are able to shower and wash the hair normally. The pedestal site is checked and cleaned thoroughly by the LVAD nurse on outpatient visits. When the hair regrows, the external system is virtually invisible, passing from behind the ear and then under the shirt or jacket. The system is robust, and all external components can be replaced. Our first patient carries both controller and battery in a shoulder bag. He has carried his system overseas by air for a vacation. On one occasion, a thief snatched the bag outside a supermarket. The connecting cable was avulsed from the pedestal and activated the alarm. This caused the thief to drop the bag, after which the cable was reconnected. On another occasion, the controller and battery were dropped onto a hard surface, which damaged the cable connection. Again, he was able to exchange the cable and restarted the pump 20 minutes later without adverse effects.
| Discussion |
|---|
|
|
|---|
Bone-anchored implants that penetrate the epithelium have been successfully used for hearing aids and for the attachment of cosmetic prostheses in auricular malformations.
4,11 When the soft tissues adhere tightly to the titanium stem, the risk of cutaneous bacteria reaching the pedestal base and causing infection is small. The biologic acceptance of bone-anchored foreign-body implants is influenced by many factors, including bone density, fixing screw, host-tissue reaction, and biocompatibility of the implant material.
12 Osseointegration is the direct structural and functional connection between living bone and the surface of the load-carrying implant. In cochlear implants with a percutaneous button screwed to the skull, there is freedom from infection of greater than 90% at 10 years.
11 Many patients have had an implant without infection for longer than 20 years. Consequently, the chances of infection reaching the Jarvik 2000 heart through the cable are virtually nil.
Our early clinical experience is encouraging. The pedestal and connector are only a little more obtrusive than an old-fashioned hearing aid. The patient cannot hear or feel the Jarvik 2000 heart within the chest, and the power system allows normal life in the community. Our implant learning curve has been important and allowed others to begin using the system without complications. We soon abandoned the zigzag method, which was time consuming, difficult to keep sterile, and unnecessary. The gentle curve of the cable with only one incision between the thoracic exit and the pedestal is simpler and safe.
Penetration of the inner table of the skull in the second patient was an avoidable complication. Surgical exploration showed a single shallow drill hole that would not have resulted in significant bleeding in the absence of coagulopathy or anticoagulation. We responded to this complication by changing the type of drill used and carefully measuring the thickness of the skull by means of CT scanning and having a variety of different-length self-tapping titanium screws available. Alternative bone-anchoring points, such as the sternum, a rib, or the iliac crest site, may also be feasible, but the depth of subcutaneous fat is unpredictable in these areas and may undermine the benefit of the scalp site.
For permanent implants, the ability of the patient to change the external system is a distinct advantage. Discontinuation of the power supply for up to 30 minutes has been tolerated, and the degree of functional aortic regurgitation (retrogradely through the vascular graft) seems to be less than for other devices. These patients already had significant improvement in left ventricular function (increase in myocardial oxygen consumption from 5.7 mL · min-1 · kg-1 preoperatively to 12.0 mL · min-1 · kg-1 in patient 1). Significantly, there was no thromboembolism on restoration of pump flow.
In summary, the clinical experience to 1 year echos our freedom from infection in the laboratory. Postauricular power delivery is user friendly in the outpatient setting, and to date, we have had no sepsis. Permanent mechanical circulatory support is feasible only with a mechanically reliable pump and an acceptably low risk of infection. The Jarvik 2000 heart with postauricular power delivery is promising in both respects.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. J. Clegg, D. A. Scott, E. Loveman, J. L. Colquitt, P. Royle, and J. Bryant Clinical and cost-effectiveness of left ventricular assist devices as a bridge to heart transplantation for people with end-stage heart failure: a systematic review and economic evaluation Eur. Heart J., December 2, 2006; 27(24): 2929 - 2938. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Westaby, O.H. Frazier, and A. Banning Six years of continuous mechanical circulatory support. N. Engl. J. Med., July 20, 2006; 355(3): 325 - 327. [Full Text] [PDF] |
||||
![]() |
M. P. Siegenthaler, O.H. Frazier, F. Beyersdorf, J. Martin, H. Laks, J. Elefteriades, A. Khaghani, U. Kjellman, B. Koul, J. Pepper, et al. Mechanical reliability of the jarvik 2000 heart. Ann. Thorac. Surg., May 1, 2006; 81(5): 1752 - 1759. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. P. Siegenthaler, S. Westaby, O.H. Frazier, J. Martin, A. Banning, D. Robson, J. Pepper, P. Poole-Wilson, and F. Beyersdorf Advanced heart failure: feasibility study of long-term continuous axial flow pump support Eur. Heart J., May 2, 2005; 26(10): 1031 - 1038. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. J. Myers, K. Robertson, T. Pool, N. Shah, I. Gregoric, and O. H. Frazier Continuous flow pumps and total artificial hearts: management issues Ann. Thorac. Surg., June 1, 2003; 75(90060): S79 - 85. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||