JTCS Email Content Delivery
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Martens, S.
Right arrow Articles by Moritz, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Martens, S.
Right arrow Articles by Moritz, A.
Related Collections
Right arrow Extracorporeal circulation
Right arrow Myocardial protection

J Thorac Cardiovasc Surg 2002;124:387-391
© 2002 The American Association for Thoracic Surgery


Cardiopulmonary Support and Physiology (CSP)

Optimal carbon dioxide application for organ protection in cardiac surgery

S. Martens, MD, M. Dietrich, MD, M. Doss, MD, G. Wimmer-Greinecker, MD, PhD, A. Moritz, MD, PhD

From the Department for Thoracic and Cardiovascular Surgery, University Hospital J. W. Goethe, Frankfurt am Main, Germany.

Received for publication Nov 29, 2001. Revisions requested Dec 26, 2001; revisions received Jan 23, 2002. Accepted for publication Feb 3, 2002. Address for reprints: Sven Martens, MD, Klinikum der J. W. Goethe-Universität, Klinik für Thorax- Herz und thorakale Gefäßchirurgie, Theodor Stern Kai 7, D-60529 Frankfurt am Main, Germany (E-mail: martens.herz{at}gmx.de).


    Abstract
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
Background: Cardiac surgery is associated with an important risk of central or peripheral organ damage, attributed in part to air embolism from incompletely deaired cardiac chambers. Insufflation of carbon dioxide into the thoracic cavity is widely used for organ protection in cardiac surgery.
Methods: In patients operated on through a sternotomy, the gas was insufflated through a standard cardioplegia line (group I, n = 10) or a Jackson-Pratt drain (group II, n = 10), with flow rates of 2, 4, and 6 L/min. In patients undergoing mitral valve surgery through a right anterolateral minithoracotomy, application through a gas port (group III, n = 10) was compared with application through a Veress needle (group IV, n = 10). In groups I and IV measurements were repeated with a gauze sponge to divert the gas stream.
Results: At a flow of 2 L/min, carbon dioxide levels in the thoracic cavity reached 52% ± 30% in group I and increased to 81% ± 27% when a gauze sponge was used. In group II a level of 91% ± 5% was achieved. In minimally invasive procedures carbon dioxide levels reached 92% ± 6% in group III and 60% ± 25% in group IV without a gauze sponge and 97% ± 2% in group IV with a gauze sponge. Increasing flow rates from 2 to 6 L/min decreased carbon dioxide levels in the thoracic cavity. Arterial blood gas analysis did not reveal critical levels of partial pressure of carbon dioxide at any time.
Conclusions: For optimized carbon dioxide concentrations during cardiac procedures, jet effects in the thoracic cavity have to be avoided. The highest levels were achieved with infusion lines covered by a gauze sponge or a perforated drain for conventional operations and a sponge-covered Veress needle or a gas port for minimally invasive approaches.


    Introduction
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
Insufflation of carbon dioxide (CO2) to the operative field to prevent cerebral or myocardial damage by means of air embolism has been reported since 1957 in cardiac surgery.Go 1 With the advent of minimally invasive valve surgery, deairing of cardiac chambers has become more difficult, and replacing air in the operative field with a more soluble gas has become a widely used technique. In recent studies neurocognitive impairment has been observed in up to 88% of patients after cardiopulmonary bypass (CPB) procedures, and disturbances might be progressive and persistent.Go 2 Introduction of air into the CPB circuit, resulting in cerebral microembolization, might contribute to postoperative cognitive impairment.Go 3 Despite carefully performed deairing procedures, transcranial Doppler studies revealed a large amount of emboli during the first ejections of the beating heart.Go 4 Air emboli always remain in the arterial circulation until complete absorption, and they only occasionally dislodge and flow downstream.Go 5 CO2 fills the thoracic cavity by means of gravity and replaces air if adequately insufflated. Because solubility of CO2 is better than that of air, occlusion or flow disruption in arteries of the brain or the heart is thought to be diminished. In 1940, Moore and BraseltonGo 6 showed that the lethal dose of air was 12 times less than that of CO2 injected into the pulmonary vein. In our institution all patients undergoing cardiac surgery through limited incisions are operated on with CO2 insufflation as a matter of policy.

An indirect method to determine the effectiveness of CO2 application to the operative field by measurement of oxygen concentration was presented by Selman and colleaguesGo 1 in 1967. The aim of our study was to evaluate different application methods with regard to optimized gas flow by using direct measurements of CO2 in the thoracic cavity. Measurements were performed in standard procedures through a median sternotomy and in minimally invasive approaches through limited incisions. Because complications caused by systemic resorption of CO2 and an increase of arterial PCO2 with acidosis have been described, blood gas analyses were evaluated frequently in our patients.


    Patients and methods
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
In patients of groups I and II, operations were performed through a complete median sternotomy. Procedures in group I were aortic valve replacement (n = 4) combined with mitral valve replacement (n = 1) or with coronary artery bypass grafting (n = 2), mitral valve repair (n = 2), and atrial septal defect closure (n = 1). Procedures in group II were aortic valve replacement (n = 2) combined with coronary artery bypass grafting (n = 1), aortic valve replacement with a conduit (n = 2), aortic valve reconstruction (n = 1) combined with replacement of the ascending aorta (n = 1), and mitral valve reconstruction (n = 3). Preoperative and intraoperative data are summarized in Table 1.


View this table:
[in this window]
[in a new window]
 
Table 1. Preoperative and intraoperative data
 
After systemic anticoagulation (300 IU/kg Liquemin, Roche), cannulation of the ascending aorta was performed with a standard cannula (Jostra AG). For cannulation of the right atrium, we used a 2-stage cannula or bicaval cannulation for mitral valve procedures, respectively. CPB was instituted with a Quadrox (Jostra AG) capillary membrane oxygenator. The circuit was primed with 1500 mL of Ringer's lactate solution, 500 mL of 6% hydroxyethyl starch, 100 mL of 20% mannitol, and 150 IU/kg heparin. CPB was conducted with a flow of 2.4 L/min per square meter of body surface area, rectal temperature was lowered to 30°C ± 2°C, and alpha-stat management was applied. Antegrade and retrograde cold blood cardioplegia were administered in an intermittent fashion. The arresting dose was 1000 mL, and a maintenance dose of 400 mL was infused every 20 minutes. Anesthesia was maintained with sufentanyl, pancuronium, or propofol, and ventilation was performed with an O2/N2O mixture establishing an inspired oxygen fraction (FIO2) of 0.5 and aiming at a PCO2 of 35 to 40 mm Hg before and after CPB. We used intermittent cardiotomy suction and continuous venting through the apex of the left ventricle or the left atrium in mitral valve procedures, respectively.

CO2 was applied by using a perfusion line (2-mm inner diameter) sutured to the left side of the pericardium in group I. After consecutive measurements at flows of 2, 4, and 6 L/min, the perfusion line was covered with a gauze sponge to divert the gas stream, and measurements were repeated. Ten minutes after stable levels of CO2 in the operative field had been reached, arterial blood gas analyses were performed. Gas flow through the oxygenator and oxygen concentration (FIO2) were adjusted by the perfusionist, according to blood gas checks and expectation of PCO2 changes during gas application. In group II CO2 was applied with a commercially available perforated drain (Jackson-Pratt, Allegiance) placed in the pericardial sac left and posterior to the heart. Deairing of cardiac chambers was performed before release of the aortic crossclamp through the apex of the left ventricle, the ascending aorta, and the left atrium in mitral valve procedures. The procedure was repeated on the beating heart. Venting through the ascending aorta was continued until the heart ejected and extracorporeal circuit was reduced.

Patients in group III and IV were operated on through right anterolateral thoracotomies through a 7-cm skin incision. In group III 7 patients underwent mitral valve repair, and 3 patients underwent mitral valve replacement. In group IV 9 repairs and 1 replacement were performed on the mitral valve. For patients' preoperative and intraoperative data, see Table 1Go. CPB was instituted after cannulation of the femoral artery and vein (with cannulas fabricated by Jostra AG). Cold blood cardioplegic solution was applied through a needle in the ascending aorta. A transthoracic clamp was used for aortic crossclamping (Scanlan Inc). The management of CPB and anesthesia was not different from that in our patients operated on through a median sternotomy, except for selective single-lung intubation to allow deflation of the right lung. CO2 was insufflated beginning with the start of CPB until the atrium was closed. In group III a separate gas port with a 5-mm inner diameter (Endopath, Ethicon) was used. In group IV measurements were performed after introduction of a Veress needle with a 2-mm inner diameter. Afterward, the needle was covered with a gauze sponge, and measurements were repeated with gas flow of 2, 4 and 6 L/min. Blood gas analyses were performed as described above. For deairing of the cardiac chambers, a vent in the left atrium and another in the aortic root was used, and the ascending aorta was punctured with a needle.

We used an infrared CO2 analyzer (BUSE), allowing for measurements of CO2 concentration in the operative field, at the level of the left ventricular cavity. The analyzer aspirates gas samples through a plastic tube. Results are indicated with a time delay of a few seconds and a sensitivity of 0.5%. For blood gas analyses, ABL System 625 (Radiometer Medical) was used.

The study was approved by our institutional ethics committee. For statistical analysis, the SAS software package was used (SAS, Inc). The Duncan multiple range test was applied to test for significance between the different methods of application and results of blood gas analyses. Data are presented as mean and SD of the mean.


    Results
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
CO2 concentrations in the thoracic cavity are depicted for operations performed through a median sternotomy in Figure 1 and in Figure 2 for minimally invasive procedures. At 2-L/min flow, CO2 levels in the thoracic cavity reached 52% ± 30% in group I and increased to 81% ± 27% when a gauze sponge was used (P < .001). In group II a level of 91% ± 5% was achieved, which was significantly elevated compared with that in uncovered cardioplegia lines of group I (P < .001) and slightly higher than that with sponge-covered lines (not significant). Additional measurements with 10-L/min flow were performed according to the literature.Go 7 No increase of CO2 concentrations compared with lower flow rates could be shown (90% ± 7% vs 91% ± 5% with 2 L, not significant). In minimally invasive procedures CO2 levels reached 92% ± 6% in group III and 60% ± 25% in group IV without a gauze sponge and 97% ± 2% in group IV with a gauze sponge. The highest CO2 concentrations were achieved with sponge-covered Veress needles and gas port insufflation, respectively (P < .001 vs uncovered Veress needles).



View larger version (18K):
[in this window]
[in a new window]
 
Fig. 1. CO2 concentrations measured at the level of the left ventricular cavity, with operations performed through median sternotomy. Demonstrated are mean, SD (box plots), and range of values (T bars).

 


View larger version (15K):
[in this window]
[in a new window]
 
Fig. 2. CO2 concentrations measured at the level of the left ventricular cavity, with minimally invasive procedures. Demonstrated are mean, SD (box plots), and range of values (T bars).

 
Results of blood gas analyses are shown in Table 2. For groups I and IV, values are presented for the higher operative field concentrations achieved with gauze sponges. Because significant field levels of CO2 have already been achieved with a flow of 2 L/min, the measurements with 6 L/min were taken after approximately 30 minutes of exposure to a CO2 atmosphere. Our perfusionists increased gas flow through the oxygenator to avoid systemic accumulation of CO2. This was significant in minimally invasive procedures of group IV, in which gas flow was increased from 2.8 ± 0.5 to 3.6 ± 0.6 L/min (P = .0049). As a result of a slight overcorrection, PCO2 levels in serial blood gas analyses decreased from 42.3 ± 6.1 mm Hg with 2-L/min insufflation to 36.9 ± 6.7 mm Hg with 6-L/min CO2 insufflation (not significant). However, systemic PCO2 remained within normal ranges in all groups (Table 2Go). FIO2 was significantly reduced by the pump technicians in group IV, from 0.45 ± 0.07 to 0.39 ± 0.05 (P = .007), resulting in lower arterial PO2 values (251 ± 29 mm Hg with 2-L/min and 158 ± 45 mm Hg with 6-L/min gas insufflation, P = .016). All blood gas analyses remained within normal ranges, despite varieties of CO2 field concentration and adaptations of oxygenator gas flow and oxygen concentration.


View this table:
[in this window]
[in a new window]
 
Table 2. Blood gas analyses
 
There was no perioperative mortality in the study groups. One patient in group I was reoperated on for bleeding and recovered with prolonged confusion. One patient in group II presented with low cardiac output syndrome after double valve replacement and underwent respirator therapy for 6 days after the operation.


    Discussion
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 
CO2 is used to replace air in the operative field during cardiac procedures with the assumption that the more soluble CO2 will reduce the amount and sequelae of gaseous emboli. CO2 might be resorbed already in the cardiac chambers, and therefore the amount of gas bubbles in the ascending aorta can be reduced, as shown in an echocardiographic study.Go 7 In addition, embolic bubbles obstructing a peripheral vessel might be resorbed quicker when they consist of a high concentration of CO2. For the latter, however, we found no scientific proof in the literature. It is obvious that only a significant increase in the CO2 concentration in the cardiac chambers can influence resorption rates of residual gas and by this can influence clinical outcome.

There are several methods of CO2 application described in the literature. Many surgeons use standard perfusion lines with 2- to 3-mm inner diameter with a flow of 2 to 10 L/min CO2 to replace air in the thoracic cavity and cardiac chambers. This is done under the assumption that the heavy CO2 will fill the operative field by means of gravity. Shang and RosenGo 8 showed that a flow of greater than 5 L/min did not increase effectiveness of air replacement in the thoracic cavity. Webb and colleaguesGo 7 used a perforated drain (Jackson-Pratt) and a flow of 10 L/min CO2, and they demonstrated a reduction of gas bubbles in the cardiac chambers visualized by means of transesophageal echocardiography. Their group did not measure CO2 concentrations in the operative field directly but concluded from oxygen concentrations that they reached CO2 concentrations of 93% and greater. Our results indicate that concentrations of greater than 90% are achieved if only 2 L/min is applied. Selman and coworkersGo 1 showed that greater than 90% of air can be displaced from the operative field if a CO2 disperser is used. Increasing flow rates lead to reduced concentrations of CO2 in the thoracic cavity if 1/4-inch tubing was used for gas application. Even if a knitted blood vessel prosthesis was used as a disperser, CO2 levels decreased if more than 7 L/min CO2 were applied.

These results are confirmed by our direct measurements in the thoracic cavity. We showed that application is effective if a commercially available perforated drain (Jackson-Pratt) or a simple cardioplegia line covered with a gauze sponge is used in cardiac procedures. Our results in minimally invasive procedures indicate that insufflation through a gas port or dividing the gas stream of an endoscopic needle with a gauze sponge increases CO2 concentration in the operative field. Uncovered lines or endoscopic needles showed significantly lower CO2 concentrations. Most probably, high gas jets flushed the operative field, and the generated gas stream dragged air into the pericardial sac. Flooding of the operative field by means of gravity does not occur with these gas jets. Even with diverted gas jets, concentrations tended to decrease at higher flow rates.

Several authors described an increase of CO2 in blood with acidosis when CO2 was insufflated into the thoracic cavity.Go Go 9-12 This is in part dependent on the use of cardiotomy suction or venting. There were methods described to monitor CO2 levels in the cardiotomy reservoir and to prevent accumulation of CO2 by flushing excessive CO2 with oxygen in the oxygenator.Go 12 In our patients the PCO2 in arterial blood samples did not reach critical levels, although conventional cardiotomy suction and venting was used, and CO2 was applied from the start of extracorporeal circulation until release of the aortic crossclamp. Because of lack of experience and reliable data with pump management under these conditions, our perfusionists corrected the gas flow according to their expectations and results of blood gas analyses. However, this led to significant but clinically irrelevant overcorrection in group IV. With increasing gas flows, the oxygen concentration provided (FIO2) was reduced, and consecutively, PO2 levels in arterial blood samples declined to more physiologic levels.

Concentrations of greater than 90% of CO2 in the operative field are possible with our means of application to achieve effective reduction of gaseous cerebral and coronary artery emboli in cardiac surgery. A gas flow of 2 L/min is sufficient, as long as gas jets are diverted. Gas concentrations did not increase with flow, most probably because higher exit velocity causes jets, flushing the thoracic cavity with room air. Systemic accumulation of CO2 with severe acidosis was not observed. However, extracorporeal circulation management had to be adapted.


    References
 Top
 Abstract
 Introduction
 Patients and methods
 Results
 Discussion
 References
 

  1. Selman MW, McAlpine WA, Albregt H, Ratan R. An effective method of replacing air in the chest with CO2 during open heart surgery. J Thorac Cardiovasc Surg. 1967;53:618-22.[Medline]
  2. McKhann GM, Goldsborough LM, Borowicz MS, et al. Cognitive outcome after coronary artery bypass: a one-year prospective study. Ann Thorac Surg. 1997;63:510-5.[Abstract/Free Full Text]
  3. Borger MA, Peniston CM, Weisel RD, Vasiliou M, Green REA, Feindl CM. Neuropsychologic impairment after coronary bypass surgery: effect of gaseous microemboli during perfusionist interventions. J Thorac Cardiovasc Surg. 2001;121:743-9.[Abstract/Free Full Text]
  4. van der Linden J, Casimir-Ahn H. When do cerebral emboli appear during open heart operations? A transcranial Doppler study. Ann Thorac Surg. 1991;51:237-41.[Abstract]
  5. Branger AB, Eckmann DM. Theoretical and experimental intravascular gas embolism absorption dynamics. J Appl Physiol. 1999;87:1287-95.[Abstract/Free Full Text]
  6. Moore RM, Braselton CW. Injections of air and carbon dioxide into a pulmonary vein. Ann Surg. 1940;112:212.[Medline]
  7. Webb WR, Harrison LH, Helmcke FR, et al. Carbon dioxide field flooding minimizes residual intracardiac air after open heart operations. Ann Thorac Surg. 1997;64:1489-91.[Abstract/Free Full Text]
  8. Shang W, Rosen M. Carbon dioxide in the prevention of air embolism during open heart surgery. Thorax. 1968;23:194-6.[Abstract/Free Full Text]
  9. Burbank A, Ferguson TB, Burford TH. Carbon dioxide flooding of the chest in open heart surgery: a potential hazard. J Thorac Cardiovasc Surg. 1965;50:691-8.[Medline]
  10. Lippmann M. Complications of CO2 field flooding the surgical field in open heart surgery: an old technique revisited. Anesth Analg. 1998;87:978-9.[Free Full Text]
  11. O'Connor BR, Kussman BD, Kyung WP. Severe hypercarbia during cardiopulmonary bypass: a complication of CO2 flooding of the surgical field. Anaesth Analg. 1998;86:264-6.[Medline]
  12. Nadolny ED, Svensson LG. Carbon dioxide field flooding techniques for open heart surgery: monitoring and minimizing potential adverse effects. Perfusion. 2000;15:151-3.[Abstract/Free Full Text]



This article has been cited by other articles:


Home page
Ann. Thorac. Surg.Home page
S. Martens, K. Neumann, C. Sodemann, H. Deschka, G. Wimmer-Greinecker, and A. Moritz
Carbon Dioxide Field Flooding Reduces Neurologic Impairment After Open Heart Surgery
Ann. Thorac. Surg., February 1, 2008; 85(2): 543 - 547.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
K. Ak, T. Aybek, G. Wimmer-Greinecker, F. Ozaslan, F. Bakhtiary, A. Moritz, and S. Dogan
Evolution of surgical techniques for atrial septal defect repair in adults: A 10-year single-institution experience
J. Thorac. Cardiovasc. Surg., September 1, 2007; 134(3): 757 - 764.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
D. A. Murphy, J. S. Miller, D. A. Langford, and A. B. Snyder
Endoscopic robotic mitral valve surgery
J. Thorac. Cardiovasc. Surg., October 1, 2006; 132(4): 776 - 781.
[Abstract] [Full Text] [PDF]


Home page
Ann. Thorac. Surg.Home page
T. Aybek, S. Dogan, P. S. Risteski, A. Zierer, T. Wittlinger, G. Wimmer-Greinecker, and A. Moritz
Two hundred forty minimally invasive mitral operations through right minithoracotomy.
Ann. Thorac. Surg., May 1, 2006; 81(5): 1618 - 1624.
[Abstract] [Full Text] [PDF]


Home page
ICVTSHome page
J. Barnard and D. Speake
In open heart surgery is there a role for the use of carbon dioxide field flooding techniques to reduce the level of post-operative gaseous emboli?
Interactive CardioVascular and Thoracic Surgery, December 1, 2004; 3(4): 599 - 602.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
S. Martens, A. Theisen, J. O. Balzer, M. Dietrich, K. Graubitz, M. Scherer, C. Schmitz, M. Doss, and A. Moritz
Improved cerebral protection through replacement of residual intracavital air by carbon dioxide: A porcine model using diffusion-weighted magnetic resonance imaging
J. Thorac. Cardiovasc. Surg., January 1, 2004; 127(1): 51 - 56.
[Abstract] [Full Text] [PDF]


Home page
J. Thorac. Cardiovasc. Surg.Home page
J. van der Linden and M. Persson
A gauze sponge cannot act as a gas diffuser in cardiac surgery when it gets wet
J. Thorac. Cardiovasc. Surg., May 1, 2003; 125(5): 1178 - 1179.
[Full Text] [PDF]


This Article
Right arrow Abstract Freely available
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Add to Personal Folders
Right arrow Download to citation manager
Right arrow Permission Requests
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Martens, S.
Right arrow Articles by Moritz, A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Martens, S.
Right arrow Articles by Moritz, A.
Related Collections
Right arrow Extracorporeal circulation
Right arrow Myocardial protection


HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
ANN THORAC SURG ASIAN CARDIOVASC THORAC ANN EUR J CARDIOTHORAC SURG
J THORAC CARDIOVASC SURG ICVTS ALL CTSNet JOURNALS