J Thorac Cardiovasc Surg 2002;124:541-552
© 2002 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease (ACD) |
From the Department of Cardiovascular and Thoracic Surgerya and the Division of Cardiovascular Medicine,c Stanford University School of Medicine, Stanford, Calif, and the Laboratory of Cardiovascular Physiology and Biophysics,b Research Institute of the Palo Alto Medical Foundation, Palo Alto, Calif.
Supported by grants HL-29589 and HL-67025 from the National Heart, Lung, and Blood Institute and a contract from Medtronic Heart Valve Division, Minneapolis, Minn. Drs Timek, Tibayan, Dagum, and Lai are Carl and Leah McConnell Cardiovascular Surgical Research Fellows. Drs Timek and Dagum were supported by NHLBI Individual Research Service Awards HL-10452 and HL-10000. Dr Timek is also a recipient of the Thoracic Surgery Foundation Research Fellowship Award. Dr Lai was supported by a fellowship from the American Heart Association, Western States Affiliate.
Read at the Twenty-seventh Annual Meeting of The Western Thoracic Surgical Association, San Diego, Calif, June 20-23, 2001.
Received for publication June 28, 2001. Revisions requested Sept 19, 2001; revisions received Jan 4, 2002. Accepted for publication Feb 18, 2002. Address for reprints: D. Craig Miller, MD, Department of Cardiovascular Surgery, Falk Cardiovascular Research Center, Stanford University School of Medicine, Stanford, CA 94305-5247 (E-mail: dcm{at}stanford.edu).
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
The theoretic advantages of stentless mitral homografts and xenografts over traditional stented bioprostheses include better hemodynamic performance, lower thrombogenicity, resistance to infection, preservation of valvular-ventricular continuity
8 (which helps maintain left ventricular [LV] systolic pump function
10,11), and possibly enhanced structural durability. The availability of suitable mitral homograft valves is limited, however, and valve implantation techniques are demanding and require considerable experience. Although good results have been obtained in a few centers with mitral unstented homografts
12 and xenografts,
13 consistent clinical outcomes have not been universal.
14,15 The heterogeneity of papillary muscle (PM) anatomy in the human heart
16,17 also complicates homograft implantation techniques.
Porcine xenografts offer the advantage of plentiful availability and more consistent subvalvular anatomy, but their hemodynamic performance and optimal implantation technique are not known. We have previously reported on the feasibility of implanting a stentless porcine mitral xenograft (Medtronic physiologic mitral valve [PMV]) in a sheep model and its immediate postoperative hemodynamic performance (acute open-chest preparation).
18 In the current experiment, we placed radiopaque markers on this porcine mitral xenograft to evaluate valvular dynamics and gradients and leaflet 3-dimensional (3-D) motion at rest and during inotropic stimulation 10 days after implantation.
| Methods |
|---|
|
|
|---|
Porcine mitral xenograft
Porcine mitral valves were harvested from adult pig hearts and removed with the anulus, leaflets, chordae, and PMs intact by Medtronic Heart Valve Division (Minneapolis, Minn). The PMV design has been modified since our earlier preliminary experiment,
18 including UltiFix fixation, a pliable annular sewing cuff of UltiFix-preserved porcine pericardium, and Dacron velour sewing tubes attached to each PM head. The valve and subvalvular apparatus were fixed in UltiFix solution, a nonglutaraldehyde process that forms amide bonds with carbodiimide cross-linking in the presence of coupling agents. Tissue cross-linking is as good as glutaraldehyde on the basis of shrinkage temperature and enzymatic digestion analysis. It is biocompatible and resists calcification, as demonstrated in both the rat subdermal and juvenile sheep models.
19 The anatomy of the porcine mitral valve is different than that in sheep or human subjects: The orientation of the pig PMs is approximately 180° to each other, and the areas and heights of the porcine anterior and posterior leaflets are approximately equal. PMV size was selected on the basis of animal weight (PMV to native valve size correlations derived from previous acute, open-chest experiments), and 8 miniature radiopaque markers were sewn around the anulus of the xenograft: 1 on each PM head, 4 on the central portion of the anterior leaflet, and 2 on the central portion of the posterior leaflet (Figure 1, A).
|
Experimental protocol
After 10 ± 5 days' recovery, the sheep were sedated with propofol (150-300 mg/h) and diazepam (5-mg intravenous bolus as needed) and intubated. Intravenous esmolol infusion (20-50 µg · kg-1 · min-1) was used to minimize reflex sympathetic responses. A Swan-Ganz catheter was placed through the left internal jugular vein for measurement of cardiac output. Simultaneous biplane videofluoroscopy, hemodynamic data recordings, and transesophageal pulse-wave and color Doppler echocardiography were conducted under 2 experimental conditions in the PMV group: (1) baseline and (2) inotropic stimulation with dobutamine (10 µg · kg-1 · min-1). In the control animals only baseline data were obtained. The PMV animals were subsequently allowed to recover and then electively killed at 3 to 4 months after valve implantation. All animals received humane care in compliance with the "Principles of Laboratory Animal Care" formulated by the National Society for Medical Research and the "Guide for the Care and Use of Laboratory Animals" prepared by the Institute of Laboratory Animal Resources, National Research Council, and published by the National Academy Press, revised 1996. This study was approved by the Stanford Medical Center Laboratory Research Animal Review Committee and conducted according to Stanford University policy.
Data acquisition
Images were acquired with the animal in the right lateral decubitus position with a Philips Optimus 2000 biplane Lateral ARC 2/Poly DIAGNOST C2 system (Phillips Medical Systems, North America Company, Pleasanton, Calif) with the image intensifier in the 9-inch fluoroscopic mode. Data from 2 radiographic views were digitized and merged to yield 3-D coordinates for each of the radiopaque markers every 16.7 seconds by using custom-designed software.
21 Ascending aortic pressure, LV pressure, and electrocardiographic voltage signals were digitized and recorded simultaneously during data acquisition.
Data analysis
Two steady-state beats from each animal in the control and PMV groups were averaged and defined as control and PMV data, respectively. During each cardiac cycle, the time of end-systole was defined as the videofluoroscopic frame containing the point of peak rate of LV pressure fall (-dP/dt), whereas the time of end-diastole was defined as the videofluoroscopic frame containing the peak of the electrocardiographic R wave. Instantaneous LV volume was calculated from the epicardial LV markers with the use of a space-filling multiple tetrahedral volume method for each frame (ie, every 16.7 ms).
22 Although myocardial volume is included in this calculation of LV volume (thereby overestimating end-diastolic LV volume), this parameter accurately reflects relative changes in LV chamber size.
Mitral annular and leaflet dynamics
Mitral annular area (MAA) was calculated as the sum of multiple triangular areas, each from 2 adjacent annular markers and the 3-D centroid
20 of the annular markers (Nos. 1-8; Figure 1
, A). MAA contraction (
MAA) was determined as the percentage change throughout the cardiac cycle from maximum to minimum area. A plane was fitted to the posterior annular markers (Nos. 3-7; Figure 1
, A) by using least-squares estimation to quantify the amount of annular folding or tilting during the cardiac cycle. Septal (anterior) annular elevation was defined as the vertical distance of marker 1 (saddle-horn marker; Figure 1
, A) from that plane.
Anterior mitral leaflet (AML) and posterior mitral leaflet (PML) lengths were calculated as the sum of individual segments defined by the implanted markers on the central portion of each leaflet. Angular position of the anterior leaflet edge was calculated throughout the cardiac cycle as the angle (
AML) between the anterior leaflet edge marker (No. 12) and the midseptal (No. 1) and midlateral (No. 5) anulus. Posterior leaflet edge angular position (
PML) was calculated in a similar fashion. For 3-D reconstruction of annular and leaflet shape, an internal right-handed Cartesian coordinate system was defined with the origin located at the midseptal anulus marker (No. 1), the positive X-axis directed toward the midlateral anulus (marker No. 5), the positive Y-axis directed toward the LV apex, and the positive Z-axis directed toward the posterior commissure. The midseptal anulus was chosen as the origin because it is at the center of the fibrous anulus and represents an anatomic landmark that is relatively stable in laboratory coordinates.
Statistical analysis
All data are reported as means ± 1 SD, unless otherwise stated. Hemodynamic and marker-derived data from 2 consecutive steady-state beats were aligned at end-diastole (t = 0), and data from the 2 beats were averaged over all animals in each group. The data (time aligned at end-diastole) were analyzed from 20 frames before to 20 frames after end-diastole, thereby allowing evaluation of the studied variables over the entire cardiac cycle. Data were compared with the Student t test for paired observations (PMV baseline vs dobutamine) and independent observations (control vs PMV).
| Results |
|---|
|
|
|---|
Hemodynamics
Table 1 summarizes the hemodynamics of control and PMV animals during baseline conditions. There was no significant difference between the groups in heart rate, LV dP/dt, peak LV pressure, end-systolic and end-diastolic volumes, or stroke volume. These data indicate comparable loading and hemodynamic conditions between the 2 groups, thus allowing direct comparison of valvular performance and dynamics. Hemodynamic variables before and after dobutamine infusion in the PMV animals are shown in Table 2, and these indicated a markedly enhanced inotropic state. Cardiac output increased by 61% during dobutamine infusion, which allowed assessment of valvular hemodynamics and leaflet 3-D motion during high flow conditions.
|
|
Mitral annular dynamics
MAA and elevation of the septal (anterior) anulus above the lateral annular plane throughout the cardiac cycle for control and PMV animals are shown in Figure 2. After implantation, the motion of the stentless PMV mitral xenograft anulus was reduced, with only a 6% MAA change throughout the cardiac cycle (Table 3). This was most likely related to the significant reduction of maximum MAA by 35% (P = .002), resulting from suturing the PMV around the mitral anulus, similar to an annuloplasty ring effect. However, folding of the septal anulus, defined as the change in distance of the midseptal marker to the posterior annular plane during systole, was similar in the 2 groups, suggesting that annular flexibility or folding was maintained with the PMV valve. The 3-D geometric relationship between the mitral anulus and the PMs in the control and the PMV study animals are shown in Figure 3. Although annular shape was similar, the PMV PM sewing tubes were inserted closer to the septum because of the natural 180° orientation of the PMV (porcine) PMs compared with the angle between the native ovine PM positions.
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Surgical technique
Technically, satisfactory implantation of mitral homograft and xenograft valves presents a new surgical challenge, and considerable experience is required to achieve good results.
17 In our experiment most of the deaths early on resulted from pulmonary failure. Sheep lungs do not tolerate prolonged periods of CPB; pulmonary deaths were rare after we shortened the CPB times and started using aprotinin. On the other hand, 5 deaths were directly related to technical errors, mostly related to not inserting the PM sewing tubes apically enough. Because the porcine mitral valve has leaflets of almost equal height and longer chordae than the ovine valve, the PM heads, particularly the posterior PM head, must be implanted apically distant relative to the native ovine PM tips to avoid PMV leaflet prolapse. Because the ovine ventricle is conical in shape and thick walled, proper implantation of the PM sewing tubes is difficult because of limited visibility deep in the left ventricle. Furthermore, because of the 180° positioning of the porcine PM heads relative to the long axis of the ventricle, the PM sewing tubes of the PMV cannot be simply implanted on the recipient's native PM tips; in this preparation they ended up being inserted closer to the septal anulus (Figure 3
) to avoid twisting of the subvalvular apparatus. Size mismatch between the native and xenograft mitral anulus, on the other hand, proved to be less problematic and could be corrected by means of annular plication without obvious compromise of valve function. This resulted, however, in a substantial cinching effect on the mitral anulus, which dampened its 3-D motion, similar to what is seen after mitral ring annuloplasty in experimental animals.
28 Although there was a considerable learning curve for us, we found that with experience, the PMV valve can be implanted in a predictable and reproducible fashion with clinically acceptable CPB and aortic crossclamp times.
Previous experimental studies with stentless ovine allografts in growing sheep revealed that early valve failure was related to chordal rupture or technical error in PM implantation.
29 In patients with an implanted porcine stentless mitral xenograft, the high rate of valve failure and recurrent mitral insufficiency were due to valve tissue tears parallel to the mitral anulus, particularly on the posterior aspect of the anulus and near the commissures.
30 These findings were often accompanied by chordal rupture, suggesting excessive tension on the chordal apparatus, possibly related to the technical aspect of PM implantation. Chordal tears and PM disruption caused by improper PM implantation have also been reported in human mitral allografts,
14,15 which could also occur with PMV implantation. We did not observe any ruptured chordae, but 4 deaths were directly related to technical errors in PM implantation.
Valvular 3-D dynamics and leaflet motion
As reported in prior clinical
12,23,24 and experimental
18,29,31 studies of stentless mitral valves, the baseline mean and peak gradients across the PMV were low. There is a paucity of data on the hemodynamic performance of stentless mitral bioprostheses under high flow conditions; this experiment revealed acceptable mean gradients across the PMV xenograft in the face of a 67% increase in cardiac output. For perspective
32-35 (even though direct comparisons between human subjects and animals are difficult), the mean transvalvular gradient in the PMV animals during inotropic stimulation and high flow conditions was only 1 mm Hg higher than the average gradient of a quadrileaflet stentless mitral bioprosthesis in human subjects.
36
Although no clinical data exist at the present time, it is postulated that the superior hemodynamic performance of stentless mitral bioprostheses might, in large part, be related to maintained normal annular motion. Surprisingly, dynamic MAA change was damped after PMV implantation in the current study, which was different than that observed in our earlier acute, open-chest experiment.
18 This discordance might be due to the previous animals being relatively hyperdynamic and receiving inotropic drugs immediately postoperatively because of the differences in the experimental conditions (open-chest and anesthetized vs closed-chest and sedated conditions) or perhaps because of a greater reduction of MAA and more of an annular purse-string effect in the current PMV series. Similar degrees of MAA reduction (approximately 30%-35%) after flexible ring annuloplasty have also been shown to abolish mitral annular dynamics in sheep.
28 Folding of the septal anulus in the PMV xenograft was similar to that in the native ovine anulus; preservation of normal annular folding might be important for avoiding LV outflow tract obstruction after annular reduction,
37 particularly with the relatively large amount of tissue in the anterior leaflet of the PMV bioprosthesis.
This porcine xenograft functioned effectively as a monoleaflet valve in vivo, with a frozen posterior leaflet acting as a buttress against which the AML closed. The PMV anterior leaflet was concave to the left atrium during closure, with some relative billowing of the basal portion of the leaflet, which was different from the closing motion of the native ovine leaflet. Whether this altered shape reflects abnormal chordal forces, which could in turn affect leaflet durability and function, is unclear; however, displacement of the PMV anterior leaflet edge away from the annular plane did not differ from that of control hearts, indicating appropriate restoration of the valvular-subvalvular geometric relationship. Whether limited posterior leaflet motion is related to reduction of annular size, change in the annular-papillary distance, or both is speculative, but frozen posterior leaflet motion has been observed in ovine hearts after annular reduction with either a flexible or a semirigid annuloplasty ring
38 and also seen commonly in patients after a mitral ring annuloplasty. Perhaps both annular dynamics and posterior leaflet motion might be less abnormal if these valves had been oversized, as is done with mitral homografts clinically.
12,15
Although 3 months' follow-up is short, and there were only 4 animals, the preliminary pathologic findings were encouraging. Excluding the animal that had endocarditis, the PMV anulus was completely endothelialized and well incorporated into the native mitral anulus. The leaflets were pliable, with only minimal calcification, and the PM sewing tubes were covered with fibrous tissue and firmly attached to the LV wall.
Limitations
Leaflet dynamics of the porcine xenograft were compared with those of the native ovine valve, and the observed differences might not only be related to technical aspects of implantation but also to possible interspecies differences in porcine and ovine mitral valve closing mechanics in vivo. Radiopaque marker studies require suturing of small markers on leaflets, which could potentially affect normal leaflet motion, but both groups of animals had the same kind of leaflet markers placed in identical arrays. Therefore, any effects of the composite mass of the markers should be the same. Furthermore, it is unlikely that the markers interfered with mitral leaflet motion because they are very small (aggregate mass, 20 ± 6 mg). Even when we grossly overloaded the anterior leaflet with a very large number of excessively heavy markers (total mass, 184 mg) in another study, the peak anterior leaflet opening velocity determined by means of epicardial PW Doppler echocardiography was 0.47 ± 0.05 m/s compared with 0.45 ± 0.06 m/s for leaflets without any markers attached. We placed markers only on the central portion of each leaflet, and therefore no data on leaflet shape and dynamics in the commissural parts of the leaflets are available. Lack of preoperative sizing of the xenograft and having available a complete set of all sizes of PMV bioprostheses is another possible limitation; in certain cases we were forced to implant a valve that obviously was too long or too large for a particular sheep's ventricle. More refined PMV sizing methods (incorporating both mitral annular size and the length of the LV chamber) and availability of more sizes will provide better surgical results.
| Appendix: Discussion |
|---|
|
|
|---|
This is a very complex model, and I admire your surgical tenacity with the first 10 preparations dying. I think we can attribute that to the effects of putting a sheep on CPB and the lung damage that it will do, but I think we have to focus also on the 5 deaths that seem to be the result of technical problems. All the PMV animals were studied at 10 days, and I would like to focus on 3 points of the study: valve gradient, valve function, and valve motion.
I think points about durability can not be surmised from this study, although we did note there was some early calcification in some of these valves, as you pointed out in your article. As we look at valve gradient, we see minimal gradient, an average of about 3 mm during the control state, 6 peak gradients, and again on dobutamine infusion, it only rose to about twice that, or 6 mm. Now we look at valvular function, and this is going to lead to my first question: Of the valves studiedthese were 7 valveswe have had 1 valve with no leak, 1 valve with 1+, 3 valves with 2+, 1 valve with 3+, and 1 valve with 4+, and you mean those to a score of 1.2, and you said that on the basis of these data, you believed you had only mild regurgitation as a group. Could you comment on that? It would seem to me that if you average those out, we would probably rate most of those valves in the mild-to-moderate regurgitation group.
Dr Timek. The data in the article were on the first 5 valves, 1 with no regurgitation, 1 with trace regurgitation, and 3 with mild regurgitation. Of the 2 subsequent animals that were added, there was one with moderate and one with moderate-to-severe regurgitation. According to that scale, it was 1.2.
Dr Starnes. Another question I have, obviously, in doing this operation is that we had a lot of technical difficulties in the beginning, and then we ended up with 7 survivors. Could you tell us whether you learned during this preparation how to determine the length of the chordal attachments. It seemed to be the most difficult part of this operation.
Dr Timek. Thank you. That is a very insightful point. Indeed, that is the most difficult part of this operation, to decide how deep the papillary tips must be placed. Initially, we began by placing them as far apically as possible, and in sheep with normal ventricles, that did not seem to be a wrong thing to do. We also tried to measure them with a custom-designed tape, and that has also helped us. One of the problems that we have had is that because of the valve availability, we cannot always match the size of the valve with the size of the sheep, and therefore it is somewhat subjective where the papillary tip is placed. We are currently still working on how to refine that technique because that has really been the major portion of our technical failures, where the papillary muscle tips are placed.
Dr Starnes. With the sheep's heart being more conical than elliptical, would you anticipate the implantation in the human subject to be easier?
Dr Timek. We would like to think so; however, at this point, we can not comment on that, although it would seem reasonable that in a dilated ventricle, where you could excise the entire valve, you probably would have good exposure, and perhaps it would be easier to implant these valves. However, that is all wishful thinking.
Dr Starnes. I would like to turn now to the leaflet motion. Again, because of the excellent technical capabilities of the laboratory, you were able to follow the physiologic motion of the anterior and posterior leaflets. The results revealed a mobile anterior leaflet and a fixed posterior leaflet. This brings me to my second question. Do you believe the immobile posterior leaflet was due to chordal positioning, or do you believe it is due to the retained posterior leaflet and chordae apparatus that I noticed you left in these animals? In other words, you did not excise the posterior leaflet. Could that have been the effect of your immobile leaflet posteriorly?
Dr Timek. That is a very good question. The reason for leaving the posterior leaflet in is to provide extra tissue to sew the PMV valve because the ovine mitral anulus is rather fine. In terms of freezing the posterior leaflet, we have seen that with any type of annular reducing procedure, such as a ring annuloplasty, either a flexible or rigid anulus, when we reduce the annular size, we also freeze the PML. In this experiment you see that for weight-matched animals, the anulus is reduced by about 30% in the PMV valves, and that is approximately what we see with annuloplasty rings. We believe that one reason why the posterior leaflet is frozen is because of the annular reduction. Of course, the second possible and likely reason is the placement of the papillary muscle tips. How to differentiate those is difficult to say, but we suspect that both contribute.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. W.M. Frater Stentless mitral valves J. Thorac. Cardiovasc. Surg., April 1, 2007; 133(4): 861 - 864. [Full Text] [PDF] |
||||
![]() |