J Thorac Cardiovasc Surg 2002;124:1157-1164
© 2002 The American Association for Thoracic Surgery
Surgery for Congenital Heart Disease (CHD) |
From the Department of Surgery, Division of Cardiovascular Surgery, Toronto General Research Institute, Toronto General Hospital, and University of Toronto, Toronto, Ontario, Canada.
R.K.L. is supported by research grants funded by Heart & Stroke Foundation of Ontario (NA#4603 and NA#4829) and Canadian Institute of Health Research (MOP14795). R.K.L. is a Career Investigator of the Heart and Stroke Foundation of Canada.
Received for publication Feb 4, 2002. Revisions requested May 15, 2002; revisions received June 10, 2002. Accepted for publication June 14, 2002. Address for reprints: Ren-Ke Li, MD, PhD, Toronto General Hospital, CCRW 1-815, 200 Elizabeth St, Toronto, Ontario M5G 2C4, Canada (E-mail: RenKeLi{at}uhnres.utoronto.ca).
| Abstract |
|---|
|
|
|---|
-caprolactone and L-lactic acid reinforced with a poly-l-lactide knitted [KN-PCLA] or woven fabric [WV-PCLA]) with a nonbiodegradable polytetrafluoroethylene (PTFE) material. An animal heart model was tested that simulates the in vivo clinical condition to which a synthetic material would be used. | Introduction |
|---|
|
|
|---|
|
Since the biodegradable scaffold needs to be capable of resisting damage from the contracting myocardium and be bioresorbable at a rate compatible with the repair process, we studied the in vivo characteristics of acellular scaffolds used to repair a surgically created myocardial defect in the right ventricle of the adult rat. Nonbiodegradable PTFE and biodegradable gelatin sponge, biodegradable nonwoven polyglycolic acid (PGA) mesh, and biodegradable poly-L-lactide knitted or woven fabric with a 50%
-caprolactone and 50% L-lactide spongy copolymer (PCLA) were tested. At 8 weeks after surgical repair we found that material degradation, expansion, host cell migration, and tissue formation were material related. All the patches were endothelialized on the endocardial surface. Our results indicated that PCLA patches may be the most suitable of the biodegradable biomaterial that we studied for future myocardial cellular engineering.
| Methods |
|---|
|
|
|---|
Biomaterial
Five materials were used in the in vivo studies: sheets of expanded PTFE,* gelatin sponge (Gelfoam; Pharmacia Co, Kalamazoo, Mich), PGA (Medifit; JMS Inc, Hiroshima, Japan), and biodegradable poly-L-lactide knitted (KN) or woven (WV) fabric with PCLA (Gunze Ltd, Kyoto, Japan). Rectangular pieces (7 x 7 x 0.4 mm for PTFE, 7 x 7 x 3 mm for gelatin, 7 x 7 x 0.5 mm for PGA, 7 x 7 x 1 mm for KN-PCLA, 7 x 7 x 0.5 mm for WV-PCLA) were used for the right ventricular wall replacement.
Right ventricular outflow tract free wall resection and replacement with patches
The rats were anesthetized with an intramuscular injection of ketamine hydrochloride (22 mg/kg), followed by an intraperitoneal injection of sodium pentobarbital (30 mg/kg). The rats were intubated and ventilated at a rate of 60 cycles/min with a tidal volume of 3 mL under room air supplemented with oxygen (2 L/min) and 1% to 2.5% isoflurane.
The rat heart was exposed through a median sternotomy. A purse-string suture (5-6 mm in diameter) was placed in the free wall of the RVOT. Both ends of the suture were passed through a 22-gauge plastic vascular cannula (Angiocath; Becton Dickinson and Company, Franklin Lakes, NJ), which was used as a tourniquet. The tourniquet was tightened, and the bulging part of the RVOT wall inside the purse-string stitch was lopped off. The tourniquet was briefly loosened to assess whether massive bleeding occurred, indicating that a significant defect had been created in the RVOT.
5 A patch of nonbiodegradable PTFE or biodegradable gelatin, PGA, KN-PCLA, or WV-PCLA (n = 5 per group) was sutured along the margin of the purse-string suture with an over-and-over method with 7-0 polypropylene (Prolene; Ethicon, Inc, Somerville, NJ) to cover the defect in the RVOT. The tourniquet was released and the purse-string suture was removed. The chest incision was closed in layers with running sutures of 3-0 silk.
Penicillin G benzathine at 150,000 U/mL and penicillin G procaine at 150,000 U/mL (Penlong XL) was injected intramuscularly (0.4 mL per rat). Buprenorphine hydrochloride (0.01 mg/kg) was given subcutaneously every 8 hours for the first 48 hours after the operation. After the operation, rats were monitored in a warm environment until they had completely recovered from the anesthesia and were then returned to their cages.
Morphologic evaluations of the patches in vivo
Immediately after trimming of the patch to repair the defect in the RVOT, the preoperative patch area was measured with a scale and a digital video camera (DCR-PC110; Sony Corporation, Tokyo, Japan). Eight weeks later the animals were killed with an intraperitoneal injection of 0.2 mL of pentobarbital sodium (Euthanyl; MTC Pharmaceuticals, Cambridge, Ontario, Canada) after intramuscular injection of 200 units of heparin. Each heart was excised through a median sternotomy and fixed with 10% phosphate-buffered formalin solution for 2 days. The patch area of the RVOT was remeasured with a scale by using the digital video camera. The ratios of preoperative and postoperative patch areas were calculated and compared among all groups.
The right ventricular free wall was then cut in half through the center of the patch. The thickness of the center of the patch replacing the defect in the RVOT was measured. The ratios of the preoperative and postoperative patch thicknesses were calculated and compared among groups.
Histologic studies of the patches in vivo
The 8-week-old implanted patches were processed with JFC solution (Milestone Srl, Sorisole, Italy), embedded in paraffin, and sectioned to yield 3-µm thick specimens. The sections were stained with hematoxylin and eosin as described in the manufacture's specifications (Sigma Diagnostics, St Louis, Mo), with Masson trichrome to assess the extracellular matrix,
6,7 with a monoclonal antibody against vimentin (1:2000, monoclonal V9; Sigma) to assess fibroblast growth, and with an antibody against factor VIII (Polyclonal 1:2000; DAKO Corporation, Carpinteria, Calif) to identify endothelial cells.
Histologic analysis of the implanted patches
Five different 400-power microscopic fields (ECLIPSE-TE200; Nikon Corporation, Tokyo, Japan) of each section of the RVOT were randomly selected and photographed with a digital camera (Coolpix; Nikon). All the digital images of morphologic and histologic characteristics of the patch portion of the RVOT were analyzed with the National Institutes of Health Image program (version 1.62; National Institutes of Health, Springfield, Va). Briefly, after tracing of the outline of the target areas in the section and subtraction of the backgrounds, the color images were converted to corresponding gray images. A fixed threshold (90 in 255 grades) of gray color scale determined automatically the number of nuclei staining positively for hematoxylin.
Protein assay of scaffolds
The protein content in the remaining half of the patch was extracted. After being washed with buffer, the patch was homogenized with protein extraction solution (50-mmol/L tris(hydroxymethyl)aminomethane hydrochloride, 0.5-µg/mL aprotinin, 0.5-µg/mL pepstatin A, and 0.5-µg/mL leupeptin) and centrifuged at 4°C at 14,000 rpm for 15 minutes. The supernatant was collected. Serial bovine serum albumin solutions were prepared for the standard protein solution as follows: 2, 1, 0.5, 0.25, 0.125, and 0.0625 mg/mL with 50-mmol/L tris(hydroxymethyl)aminomethane hydrochloride (pH 7.4). The 495 µL of protein assay solution (BIO-RAD Dye and 0.9% sodium chloride at 5:1; Bio-Rad Laboratories Inc, Hercules, Calif) and 5 µL of each protein standard or sample solution were mixed and read with a spectrophotometer (model DU 640; Beckman Coulter, Inc, Fullerton, Calif) at a wavelength of 595 nm. A standard curve was constructed to measure protein concentrations. Gelatin patches were excluded in the assay, because gelatin is also measured by the protein assay. All other patch materials had no measurable protein content. The patch protein content per cubic millimeter was obtained by dividing the protein content by the patch cubic volume.
Data analysis
All data are expressed as mean ± SE. StatView statistical software (SAS Institute Inc, Cary, NC) was used to analyze statistical data. The changes between the preoperative and postoperative patch areas and thicknesses in each group were compared by a paired t test. Comparisons among the five groups were performed by 1-way analysis of variance. The Scheffé test was used to specify differences among groups when the F ratio of the analysis of variance was significant.
| Results |
|---|
|
|
|---|
Patch area
Eight weeks after implantation, the changes in the patch area (the ratios of postoperative to preoperative values) were different among the five materials (Figure 1, A). The nonbiodegradable PTFE patch did not change in area. However, the biodegradable gelatin and PGA patches expanded significantly (P < .01 for both) relative to the PTFE patch area. The KN-PCLA patch was larger (P < .05) than the PTFE patch but significantly smaller (P < .05) than the gelatin patch. The KN-PCLA and PGA patch areas did not differ. The WV-PCLA patch did not expend in area and was not statistically different in size from that of the KN-PCLA patch. The KN-PCLA patch size was smaller in area than the gelatin and PGA patches (P < .01 for both).
|
Cell migration and extracellular matrix formation in the biomaterials
Only the surfaces of the PTFE patch were coated with fibroblasts (Figure 2, A) and collagen (Figure 3, A). No cellular infiltration was evident within the PTFE patch. The PTFE patch remained unchanged. In contrast, fibroblast ingrowth (Figure 2
, B-E) and collagen formation (Figure 3
, B-E) were present throughout the gelatin, PGA, KN-PCLA, and WV-PCLA patches. Of the biodegradable materials, the highest cell number per unit area was measured in the KN-PCLA patch (Figure 4). No significant differences in cell numbers were found among the gelatin, PGA, and WV-PCLA patches. Two of five gelatin patches had foci of inflammatory cells. The gelatin, PGA, KN-PCLA, and WV-PCLA materials were partially absorbed and replaced by fibroblasts and collagen (Figure 2
, B- E, and Figure 3
, B-E).
|
|
|
Endothelialization
At 8 weeks after implantation all the patches were endothelialized on their endocardial surfaces (Figure 5). No thrombi were observed on the patches.
|
| Discussion |
|---|
|
|
|---|
A biodegradable scaffold should be preferable to a nonbiodegradable scaffold. A biodegradable scaffold can be replaced by the patient's own cells and will be treated as a foreign body for only a limited time. The PTFE patch has been accepted worldwide as one of the most reliable nonbiodegradable synthetic materials
10 in terms of durability, low thrombogenicity, and comfortable handling for surgery. Unfortunately, it does not grow as a child's heart or blood vessel grows, and it carries increased susceptibility to infection.
11 In this study PTFE was used as a control patch for comparison purposes with the biodegradable patches that we studied. The biodegradable gelatin sponge (Gelfoam) was tested because of our earlier works.
5,12,13 The use of fetal cardiomyocytes the pliable and fragile gelatin scaffold enabled us to create in vitro and in vivo beating patches. The biodegradable nonwoven PGA mesh was studied because it is one of the most common and attractive synthetic biodegradable polymers for tissue engineering field.
14 However, this PGA has recently been used in combination with another biodegradable polymer,
7,15,16 because nonwoven meshes of PGA fibers lack structural stability and often cannot maintain their original structure during tissue development.
17 Of special interest to us was the PCLA patch with a biodegradable spongy matrix, reinforced on the outside with a knitted or woven poly-L-lactide fabric that takes between 1 and 2 years to biodegrade (information provided by the manufacturer). We thought that the spongy matrix would favor in vitro and in vivo cell growth and might be suitable for in vitro autologous seeding of smooth muscle cells or stem cellderived cardiomyocytes. Also, the outer fabric of PCLA patch should provide structural stability and permit sufficient time for in vivo maturation of the seeded cells in the contracting ventricular wall. Kim and Mooney
17 demonstrated that the poly-L-lactidebonded PGA matrices had a high cellularity and maintained their predefined structure more than did the unbonded PGA matrices during the process of tissue development in vitro. Their report suggests that the combined scaffold structure, like the PCLA patch in our study, should allow the maintenance of the configuration and dimension of the original matrices and the development of a new tissue in a predefined 3-dimensional structure.
Because Carver,
18 Freed,
19 and their colleagues have shown that mechanical stimuli and pressure can enhance cellular formation of extracellular matrix and tissue regeneration, we used an animal model that simulates the in vivo conditions to which a tissue patch would be exposed clinically. In addition to the mechanical forces of myocardial contractility, the patch is influenced by the myocardial tissue milieu and is in direct contact with the blood in the right ventricular chamber. An 8-week period of in vivo testing was arbitrarily chosen as being sufficiently long to detect graft failure and to compare the various scaffolds.
The PTFE scaffold did not change in its dimensions or break down after 8 weeks as a patch in the RVOT. Compared with the biodegradable patches, PTFE permitted only surface cell growth that consisted predominantly of fibroblasts. Fibrous tissue was firmly attached onto both surfaces of the patch. No inflammatory infiltration was noted. The gelatin patch was unsuitable as a patch because of its marked expansion and thinning. The gelatin scaffold was almost completely degraded and infiltrated with fibroblasts. Additionally, some inflammatory foci were found. Because this phenomenon was found in one of our earlier studies,
5 we believe that the inflammatory reaction was caused by a xenogenic reaction to porcine skin proteins in the sponge. The nonwoven PGA patch increased in surface area and thinned after 8 weeks in the RVOT. The scaffold itself was almost completely degraded and infiltrated with fibroblasts. The expansion of the patch size during the 8 weeks indicated either that the scaffold biodegradation process was not in synchrony with the fibroblast colonization and collagen deposition or that fibroblast colonization and collagen deposition were insufficient to prevent the changes in patch surface area and thickness.
Similar results to the PGA patch were observed with the knitted PCLA patch except that fibroblast colonization per unit area was greater and more localized in the knitted matrix than in the PGA patch. Also, collagen formation appeared to be greatest in the KN-PCLA patch. The dimension of the WV-PCLA patch was almost unchanged during the 8 weeks, despite the spongy matrix scaffold degradation, fibroblast infiltration, and collagen deposition in the sponge. In contrast, the KN-PCLA patch permitted more host fibroblast infiltration than did the WV-PCLA patch. This finding is consistent with previous reports that studied woven and knitted Dacron polyester fabric grafts.
20,21
The endocardial surface was covered by endothelial cells in all the biomaterials tested. As Scott and colleagues
22 demonstrated, our findings indicate that host endocardial endothelial cells or endothelial progenitor cells were involved in endothelialization of the endocardium. The necessity of seeding endothelial cells in biomaterials remains controversial. To avoid thrombosis, some investigators have constructed conduits seeded with endothelial cells before implantation.
23 On the other hand, Shi and coworkers
24 demonstrated that the rapidity with which the specimens are fixed affects the identification of endothelium in explanted synthetic aortic grafts obtained from the patients. They also suggested that endothelialization of synthetic arterial grafts might occur more frequently in patients than previously recognized.
If a biodegradable scaffold is infiltrated primarily with fibroblasts that form a fibrous structure, localized dilatation of the patch may occur similarly to a ventricular aneurysm formed after a transmural myocardial infarction. However, Malm and colleagues
25 reported that unseeded absorbable, nonwoven patches made from polyhydroxybutyrate implanted as transannular patches into the RVOT in 13 weanling sheep could be used as a scaffold for tissue regeneration in low-pressure systems. Our histologic results demonstrated that the patch cells and extracellular matrix in the biodegradable scaffolds were predominantly fibroblasts and collagen. It has been unclear whether the nonmuscular regenerated tissue can tolerate right ventricular pressure without dilatation during the long term unless a strong and durable biodegradable material is used. Even after the RVOT reconstruction with a transannular patch or conduit, the implanted material may be exposed to residual stenosis because of remaining thick trabecular muscle through the RVOT to pulmonary artery or may be exposed to pulmonary hypertension. In these cases, the biomaterial would require more strength to prevent its dilatation because of increased right ventricular pressure overload.
To avoid host overcolonization of the scaffold with fibroblasts and eventual scar formation, in vitro autologous cell seeding may be essential in cardiac tissue engineering. In vitro seeding of a biodegradable material with cultured smooth muscle cells would permit formation of a muscular tissue within the scaffold and the formation of elastin extracellular matrix in addition to collagen formation
7,26 Shinoka and associates
23 have demonstrated the benefits of in vitro autologous cell seeding to form mature autografts from synthetic biodegradable tubular scaffolds. We have also found morphologic differences between the cell-seeded and nonseeded gelatin patches.
5 The elastic properties of elastin and strength of collagen formed by the implanted smooth muscle cells and the capacity of smooth muscle cells to proliferate and hypertrophy in response to increased pressure should prevent patch dilatation and thinning.
One limitation of this study is the small size of the patches in the right ventricular outflow tract. Our ultimate goal is to create a beating conduit that will provide pulsatile perfusion to the lungs. Our preliminary study
12 with seeding a gelatin patch with fetal cardiomyocytes suggests that a beating patch may be possible in the right ventricular outflow tract. However, allogeneic cells are eventually rejected, and a better biomaterial would be required before assessing cell-seeded grafts. Therefore this study reviewed alternative scaffolds that can be used for cell seeding. It should be possible to seed a biodegradable scaffold with beating cardiomyocytes derived from mesenchymal stem cells and to use the scaffold in vivo to induce pulsatile pulmonary perfusion. A challenge for all biodegradable patches is to seed the correct number and proportion of myocytes, endothelial cells and fibroblasts so that the graft can develop in vivo into a structurally strong tissue that does not dilate in response to intraventricular pressures. With a large graft, selective endothelial cell seeding of the endocardial surface may be necessary before implantation to ensure complete in vivo endothelial coverage of the graft's endocardial surface.
In summary, the unique structure of the PCLA patch with a spongy matrix favors in vivo cell colonization relative to the other patches tested. Its outer poly-L-lactide fabric layer also offers advantages relative to the other biodegradable materials we studied.
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C.-H. Chen, H.-J. Wei, W.-W. Lin, I. Chiu, S.-M. Hwang, C.-C. Wang, W.-Y. Lee, Y. Chang, and H.-W. Sung Porous tissue grafts sandwiched with multilayered mesenchymal stromal cell sheets induce tissue regeneration for cardiac repair Cardiovasc Res, October 1, 2008; 80(1): 88 - 95. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Mettler, V. L. Sales, C. L. Stucken, V. Anttila, K. Mendelson, J. Bischoff, and J. E. Mayer Jr Stem cell-derived, tissue-engineered pulmonary artery augmentation patches in vivo. Ann. Thorac. Surg., July 1, 2008; 86(1): 132 - 141. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Chang, P.-H. Lai, H.-J. Wei, W.-W. Lin, C.-H. Chen, S.-M. Hwang, S.-C. Chen, and H.-W. Sung Tissue regeneration observed in a basic fibroblast growth factor-loaded porous acellular bovine pericardium populated with mesenchymal stem cells J. Thorac. Cardiovasc. Surg., July 1, 2007; 134(1): 65 - 73. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Iwai, Y. Sawa, S. Taketani, K. Torikai, K. Hirakawa, and H. Matsuda Novel Tissue-Engineered Biodegradable Material for Reconstruction of Vascular Wall Ann. Thorac. Surg., November 1, 2005; 80(5): 1821 - 1827. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Chang, S.-C. Chen, H.-J. Wei, T.-J. Wu, H.-C. Liang, P.-H. Lai, H.-H. Yang, and H.-W. Sung Tissue regeneration observed in a porous acellular bovine pericardium used to repair a myocardial defect in the right ventricle of a rat model J. Thorac. Cardiovasc. Surg., September 1, 2005; 130(3): 705 - 705. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Kofidis, D. R. Lebl, E. C. Martinez, G. Hoyt, M. Tanaka, and R. C. Robbins Novel Injectable Bioartificial Tissue Facilitates Targeted, Less Invasive, Large-Scale Tissue Restoration on the Beating Heart After Myocardial Injury Circulation, August 30, 2005; 112(9_suppl): I-173 - I-177. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. H.L. Tang, S. Fazel, R. D. Weisel, G. S. Van Arsdell, and R.-K. Li Cardiovascular Tissue Engineering Therapy: So Near, So Far? Ann. Thorac. Surg., June 1, 2005; 79(6): 1831 - 1833. [Full Text] [PDF] |
||||
![]() |
T. Kofidis, J. L. de Bruin, G. Hoyt, D. R. Lebl, M. Tanaka, T. Yamane, C.-P. Chang, and R. C. Robbins Injectable bioartificial myocardial tissue for large-scale intramural cell transfer and functional recovery of injured heart muscle J. Thorac. Cardiovasc. Surg., October 1, 2004; 128(4): 571 - 578. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Iwai, Y. Sawa, H. Ichikawa, S. Taketani, E. Uchimura, G. Chen, M. Hara, J. Miyake, and H. Matsuda Biodegradable polymer with collagen microsponge serves as a new bioengineered cardiovascular prosthesis J. Thorac. Cardiovasc. Surg., September 1, 2004; 128(3): 472 - 479. [Abstract] [Full Text] [PDF] |
||||
![]() |