J Thorac Cardiovasc Surg 2003;125:419-420
© 2003 The American Association for Thoracic Surgery
Brief Communications |
From the Department of Cardiovascular Surgery, Heart Institute of Japan, Tokyo Women's Medical University, Tokyo, Japan.
Received for publication Feb 22, 2002. Accepted for publication July 23, 2002. Address for reprints: Yuji Naito, MD, Department of Cardiovascular Surgery, Heart Institute of Japan, Tokyo Women's Medical University, 8-1, Kawada-cho, Shinjyuku-ku, Tokyo, 162-8666, Japan (E-mail: ujinaito{at}aol.com).
Currently satisfactory clinical results are observed with the extracardiac Fontan operation (ECFO); however, some issues associated with graft materials must be addressed, such as infection, thrombogenicity, and the lack of growth potential.
1 We report here the case of a patient with occlusion of a graft implanted at initial operation and successfully underwent reoperative ECFO with a tissue-engineered graft.
Clinical summary
A 12-year-old boy had the diagnoses of polysplenia, complete atrioventricular septal defect, small left ventricle, common atrium, bilateral superior vena cava, and hemiazygous connection. A modified Blalock-Taussig shunt and ECFO with a Hemashield graft (Boston Scientific Corporation, Meadox Division, Boston, Mass) were performed at the ages of 9 and 11 years, respectively. The patient had general fatigue 7 months after ECFO. Catheterization demonstrated occlusion of the extracardiac graft. The clinical application of tissue engineering in this patient was approved by the ethical committee at Tokyo Women's Medical University, and the patient's parents were thoroughly informed and consented to the procedure. Graft replacement with a tissue-engineered graft was planned, and graft preparation was conducted.
An approximately 2-cm segment of peripheral vein was harvested, and cells from its walls were isolated. The cells were cultured and expanded as previously described elsewhere.
2 The tubular scaffold for these cells was composed of a polycaprolactone-polylactic acid copolymer (weight ratio 1:1) reinforced with woven polyglycolic acid. The synthetic biodegradable polymer conduit (20 mm in diameter and 1 mm in thickness) was designed so that it would be degraded within 8 weeks. Ten days after seeding, the replacement of the occluded conduit was performed by transplantation of the tissue-engineered graft.
Intraoperative findings showed that the conduit was completely collapsed and flattened, with white thrombus contained within. Pump perfusion was established with ascending aortic and single left superior vena caval cannulation. Finally, the conduit was completely removed, but hepatic veins were completely occluded and no lumen was found. Therefore the diaphragm was cut open in the midline, and the incision was extended to the hepatic veins. Tissue-engineered conduit 20 mm in diameter was anastomosed with 4-0 Prolene suture (Ethicon, Inc, Somerville, NJ) to the hepatic venous stump (Figure 1). The main pulmonary artery end was opened along its previous anastomotic orifice into the right pulmonary artery, and the other end of the conduit was anastomosed with the orifice. Intraoperative transesophageal echocardiography showed patent graft flow, and the patient was discharged from the hospital in good clinical condition. Postoperative computed tomography done 4 months after the operation revealed a patent graft (Figure 2).
|
|
Significant developments have been made in the field of tissue engineering, and many clinical applications have been found in recent years. Previously we have reported the successful results of surgical implantation of tissue-engineered pulmonary valve leaflet, pulmonary artery, and inferior vena cava in large animal models.
2,3 On the basis of such reliable evidence, the first clinical application was successfully done in May 2000 with a tissue-engineered patch for pulmonary arterial angioplasty.
4 The creation of tissue-engineered structures from autologous cells offers many potential advantages: elimination of the problems of rejection and donor scarcity, greater durability and longevity because of the tissue's natural mechanism for repair and remodeling, lack of foreign body response or need for long-term anticoagulation, and potential for growth, which last is important in the pediatric age group. The major drawbacks and main points of criticism of the ECFO are lack of growth potential and thrombogenicity of the conduit. Hemiazygous continuation of the inferior vena cava in which conduit carries only hepatic venous flow may especially have a higher risk of thrombosis.
5 In response to concerns related to the Dacron polyester fabric conduit in the venous system that we used in the initial operation, we believed that there could be less chance of obstruction because the conduit interposed between the inferior vena cava and the pulmonary artery relies on a better hydrodynamic pattern with uniform caliber and less turbulence, regardless of graft material used.
6
We believe that tissue-engineered prostheses could be useful in such complicated cases, and their indications would be increased for other kinds of cases. There remain limitations of this method in the clinical setting. First, a tissue-engineered graft cannot be prepared for an emergency operation because of the prolonged in vitro times necessary for development of tissues. Second, sufficient cell proliferation cannot be accomplished in all patients, and in such cases we have to prepare cells from other part of the body. Third, tissue-engineered grafts cannot be used in systemic circulation in which proper biodegradable scaffold is now developing. Finally, small diameter vascular grafts may occlude during early follow-up. To find solutions for these issues, we are continuing experimental studies.
References
This article has been cited by other articles:
![]() |
G. N. Nelson, T. Mirensky, M. P. Brennan, J. D. Roh, T. Yi, Y. Wang, and C. K. Breuer Functional Small-Diameter Human Tissue-Engineered Arterial Grafts in an Immunodeficient Mouse Model: Preliminary Findings Arch Surg, May 1, 2008; 143(5): 488 - 494. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Lee, C.-H. Lee, S. W. Hwang, H. G. Lim, S.-J. Kim, J. Y. Lee, W.-S. Shim, and W.-H. Kim Midterm follow-up of the status of Gore-Tex graft after extracardiac conduit Fontan procedure Eur. J. Cardiothorac. Surg., June 1, 2007; 31(6): 1008 - 1012. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Ikada Challenges in tissue engineering J R Soc Interface, October 22, 2006; 3(10): 589 - 601. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Ishii, M. Shin, T. Sueda, and J. P. Vacanti In vitro tissue engineering of a cardiac graft using a degradable scaffold with an extracellular matrix-like topography J. Thorac. Cardiovasc. Surg., November 1, 2005; 130(5): 1358 - 1363. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Isomatsu, T. Shin'oka, G. Matsumura, N. Hibino, T. Konuma, M. Nagatsu, and H. Kurosawa Extracardiac total cavopulmonary connection using a tissue-engineered graft J. Thorac. Cardiovasc. Surg., December 1, 2003; 126(6): 1958 - 1962. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||