J Thorac Cardiovasc Surg 2003;125:723-724
© 2003 The American Association for Thoracic Surgery
Brief Communications |
From the Department of Thoracic and Cardiovascular Surgery, Cleveland Clinic Foundation, Cleveland, Ohio.
Received for publication May 13, 2002. Accepted for publication June 13, 2002. Address for reprints: Inderjit S. Gill, FRCS(C), Department of Thoracic and Cardiovascular Surgery, Cleveland Clinic Foundation, 2500 MetroHealth Dr, Third Floor Hamann Building, Cleveland, OH 44109-1998 (E-mail: gillis{at}ccf.org).
|
Clinical summary
The patient was an 84-year-old woman who underwent cardiac catheterization after having abnormal results of a stress test before planned orthopedic surgery. Cardiac catheterization revealed triple-vessel coronary artery disease with 50% to 60% lesion of the left main trunk, mild and diffuse disease of the left anterior descending (LAD) coronary artery, normal left ventricular wall motion, and an ejection fraction of 79%. Treatment options were discussed, and the patient was referred for a hybrid procedure, an elective MIDCAB to be followed by percutaneous intervention to the right coronary and circumflex system.
The patient underwent MIDCAB through a left minithoracotomy. The operative technique has been previously discussed elsewhere.
1 The diameter of the left internal thoracic artery (LITA) was 1.5 mm. Stabilization of the target vessel was achieved with the Genzyme Immobilizer (Genzyme Corporation, Cambridge, Mass). Coronary arteriotomy was performed, and a 1.5-mm intracoronary shunt (Guidant Axius Coronary Shunt; Guidant, Cupertino, Calif) was placed with some difficulty (Figure 1). The anastomosis was carried out with 7-0 Prolene running suture (Ethicon, Inc, Somerville, NJ). After the last stitch was placed, the intracoronary shunt was carefully taken out and the suture was tied. Graft flow measurements were carried out with a Transonic Flow-QC meter (Transonic Systems, Inc, Ithaca, NY). Although the graft pulsation was good, the graft flow measurement showed a poor diastolic flow pattern and insufficient blood flow at 5 mL/min. The anastomosis was therefore taken down and examined. No reason was found for graft malfunction. A 1.5-mm vessel probe was passed through distally and proximally from the arteriotomy. The anastomosis was then carefully redone in the same fashion. However, a repeated graft flow measurement was still unsatisfactory with poor graft flow. The patient was taken for urgent cardiac catheterization, and the angiogram showed total occlusion of the LITA graft at the mid portion and total occlusion of the mid LAD.
|
|
With currently available stabilization systems and intracoronary shunts, MIDCAB can be performed safely with satisfactory graft patency even in elderly patients.
2,3 The use of an intracoronary shunt prevents deterioration in ventricular function and arrhythmias, presumably by maintaining distal myocardial perfusion.
4,5 However, there are few data available regarding complications caused by the use of such shunts.
Careful insertion and removal should be undertaken to prevent vessel injury. Selection of a properly sized shunt is also important in obtaining a blood-free surgical field and in preventing injury. The incidence of vessel injury caused by intracoronary shunt placement is extremely low with skilled surgical technique.
This report highlights the utility of intraoperative flow measurement to reveal acute graft failure in beating-heart surgery, allowing timely intervention. It also highlights the potential of traumatic intimal injury by intracoronary shunts. Extra caution should be taken in frail and elderly patients with diffusely atherosclerotic and fragile coronary arteries.
References
This article has been cited by other articles:
![]() |
H. Hangler, L. Mueller, E. Ruttmann, H. Antretter, and K. Pfaller Shunt or Snare: Coronary Endothelial Damage due to Hemostatic Devices for Beating Heart Coronary Surgery Ann. Thorac. Surg., December 1, 2008; 86(6): 1873 - 1877. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. G. Demaria and L. P. Perrault On Decreasing Distal Endothelial Damage After Intracoronary Shunt Insertion Ann. Thorac. Surg., May 1, 2005; 79(5): 1826 - 1826. [Full Text] [PDF] |
||||
![]() |
S. Verma, P. W.M. Fedak, R. D. Weisel, P. E. Szmitko, M. V. Badiwala, D. Bonneau, D. Latter, L. Errett, and Y. LeClerc Off-Pump Coronary Artery Bypass Surgery: Fundamentals for the Clinical Cardiologist Circulation, March 16, 2004; 109(10): 1206 - 1211. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |