This assumption leads to nominal size being substituted for the measured left ventricular outflow tract diameter during the calculation of the continuity equation
7. Some studies suggest that this may be valid,
8 whereas others do not.
9 Similarly, echocardiography is often used for estimating the size of the substitute valve either before surgery or perioperatively,
10-13 although differences between nominal valve size and left ventricular outflow tract diameter are often large.
11 In fact, nominal size and patient tissue annulus diameter may not agree.
14 In terms of biologic equivalence, it may be more appropriate to compare valves of different nominal diameter for some valves made by different manufacturers. Therefore the aims of this study were to measure the orifices of a number of designs of bileaflet mechanical aortic valve and to compare nominal size with biologic equivalence, as assessed with an artificial patient annulus.
| Methods |
|---|
|
|
|---|
Photography
The leaflets were held fully open, and the inflow surface was photographed onto 35-mm transparency film (Kodak Ektachrome; Eastman Kodak Company, Rochester, NY) with an Olympus OM-4Ti camera with Olympus SZH Stereomicroscope (7x-40x magnification; Olympus Optical Co, Ltd, Tokyo, Japan). This was aligned with the central axis of the valve. The images were digitized with a Hewlett-Packard Scanjet 4C (Hewlett-Packard Company, Palo Alto, Calif) scanner (Figure 1) then analyzed manually with Adobe Photoshop (Adobe Systems Incorporated, San Jose, Calif). A program was then written in MATLAB (The MathWorks, Inc, Natick, Mass) to allow calibration and elimination of errors caused by variations in image magnification. For each image the maximum internal diameter in pixels was calibrated from the corresponding internal diameter measured directly with a digital caliper (resolution 0.01 mm; Mitutoyo Corporation, Kanagawa, Japan). Measurement errors resulting from manual edge detection and digitization were estimated from image analysis of a series of machined cylinders of known diameter between 19 and 31 mm. Computed areas were within ± 1.5% of the actual area calculated from the known diameter.
|
|
| Results |
|---|
|
|
|---|
|
|
|
|
| Discussion |
|---|
|
|
|---|
The measured orifice area ranged between 159 and 222 mm2 for the six size 19 valves and between 316 and 405 mm2 for the six size 25 valves (Table 1
). This is consistent with previous studies that have noted the disparity between nominal and actual size not only for bileaflet valves but also for tilting-disk mechanical and stented and stentless tissue valves.
14,15 Christakis and colleagues
14 therefore suggested that valves should be labeled by internal diameter. However, valves of the same internal diameter may have sewing cuffs that are very different in size and compressibility. This means that valves of a similar orifice area may not be capable of implantation in a given patient. We therefore compared valves with a measure of biologic equivalence provided by a series of machined polypropylene blocks.
The difference between nominal size and biologic equivalence varied widely. At one extreme, the biologic equivalence was as much as 3.5 mm smaller than nominal size for the Carbomedics supra-annular (Top Hat) valve; at the other, the biologic equivalence was 3.0 mm bigger for the Carbomedics standard valve. For the intra-annular valves, each nominal size of Carbomedics standard valve was approximately equivalent to the Carbomedics reduced cuff of the next nominal size larger. For the supra-annular valves, the Carbomedics supra-annular (Top Hat) was approximately equivalent to a St Jude EHP one size smaller and an On-X valve two sizes smaller. Discrepancies in comparing supra-annular and intra-annular valves were even greater. The Carbomedics standard size 19 valve was biologically equivalent to the Carbomedics supra-annular (Top Hat) size 25 valve, which fitted a similar sized polypropylene block.
It is clear that comparing hemodynamic function in bileaflet mechanical valves of the same nominal size is usually inappropriate. It would be tempting to compare valves with similar geometric orifice areas,
14 for example the 19 St Jude EHP with either the 21 Carbomedics standard, the 21 Carbomedics supra-annular (Top Hat), or the 21 St Jude Medical standard. This might be reasonable in the in vitro setting, but clinically it makes more sense to compare biologically equivalent valves that could be implanted into the same patient tissue annulus. Research studies investigating hemodynamic function should either use the tables of biologic equivalence established here (Tables 2
and 3
) or alternatively randomly assign the valve types being compared. An independent sizer could be used to measure the actual diameter of each individual patient's annulus, then the sizers for each of the investigational valve types should be used in turn before opening the randomization envelope. The different valve types should then be compared according to the patient tissue annulus, rather than the nominal size.
The difference between nominal size and biologic equivalence also means that extreme caution should be exercised when attempting to size a valve by echocardiography. Studies attempting this show mostly good agreement between left ventricular outflow tract diameter for stentless valves.
10 However, as much as 2 mm difference was shown in one study
13 of mechanical and stented biologic valves. In another,
11 the 95% range for the difference between left ventricular outflow tract diameter measured by transthoracic echocardiography and nominal valve size was between -8.5 and 5.1 mm. The difference between nominal size and diameter measurement depends on the valve design. The results in this study suggest that for the St Jude EHP valve there would be exact agreement, for the Carbomedics reduced cuff series there would be about 1 mm difference, but for the Carbomedics standard the left ventricular outflow tract diameter could be as much as 3 mm larger than the nominal size.
Similarly, it is usually inappropriate to substitute the nominal size of the replacement heart valve in place of direct measurement of the outflow tract when applying the continuity equation.
7,8 This has been advocated on the basis of better discrimination between valve sizes in St Jude Medical standard valves.
8 In fact, use of the nominal diameter caused a systematic overestimation of effective orifice area by 0.2 cm2 relative to direct measurement, and individual paired values were as much as 1.0 cm2 different.
8 That finding is consistent with this study, because we showed that cylindric fit, which is equivalent to left ventricular outflow tract diameter, was larger than nominal size by about 1 to 1.5 mm for St Jude Medical standard valves. A study of stented biologic prostheses
9 showed that the correlation with catheter-derived effective orifice area was poorer when the continuity equation was calculated with the nominal size than when the measured left ventricular outflow tract diameter was used. However, it is clear that the relationship between nominal size and outflow tract diameter depends on the type of valve and whether the valve is implanted in an intra- or supra-annular position.
Definitions of patient-prosthesis mismatch must also be tightened. It has been suggested
16 that a size19 valve should not be implanted in a patient with a body surface area larger than 1.7 m2. However the geometric orifice area of a size 19 valve depends on the site of implantation and the valve design and in this study varied by a factor of 1.4 between the largest and smallest bileaflet mechanical valves. Furthermore, the cylindric fit of a size 19 valve ranged from 17 mm to 21.5 mm, depending on position and valve type. Clearly, any guidance regarding prosthesis-patient mismatch must include geometric or preferably effective orifice area, rather than nominal size,
17 and we suggest also a measure of biologic equivalence as provided by cylindrical fit.
Toward a standardized nomenclature
It should be possible for the label on the case containing the valve to reflect the true size of that valve. We suggest that the most clinically useful measure would be the tissue annulus diameter, because this should reflect the diameter of the patient's annulus. This could be tested in vitro in terms of biologic equivalence as measured by cylindrical fit (Tables 2
and 3
). However, the relationship between the true valve size and the annulus depends on a number of factors, including suture technique and to what degree the surgeon compresses the cuff. There was a variation in biologic equivalence of 0.5 mm among the three observers in this study for 18 of the 29 valves, probably reflecting differences in assessing cuff compressibility. For valves intended for supra-annular implantation, the diameter of the aorta and position of the coronary arteries may also affect the size of valve suitable for implantation. An alternative approach is an alphanumeric code, as for the Starr-Edwards valve, with for example 1A for the smallest aortic valve of any given type. We suggest that whatever sizing convention is used, it would also be useful to have the geometric orifice area displayed on the label of the container, as suggested by Christakis and colleagues.
14
Limitations
The conditions in this study cannot completely reflect every clinical situation. The native aortic annulus is usually more flexible than the polypropylene cylinders used in this study. This may affect biologic equivalence, particularly for the On-X valve, which has a flared inlet requiring a degree of elasticity in the annulus. We assigned valves to either an intra-annular or supra-annular position, but the precise position may be intermediate to these depending on suture technique, and a valve may even on occasion be placed diagonally in the annulus.
| Conclusion |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
C. R. Bridges, S. M. O'Brien, J. C. Cleveland, E. B. Savage, J. S. Gammie, F. H. Edwards, E. D. Peterson, and F. L. Grover Association between indices of prosthesis internal orifice size and operative mortality after isolated aortic valve replacement J. Thorac. Cardiovasc. Surg., April 1, 2007; 133(4): 1012 - 1021. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Reply to the Editor J. Thorac. Cardiovasc. Surg., March 1, 2007; 133(3): 847 - 849. [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Small aortic annulus: The hydrodynamic performances of 5 commercially available tissue valves J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 1058 - 1064. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. P. Taggart Prosthesis patient mismatch in aortic valve replacement: possible but pertinent? Eur. Heart J., March 2, 2006; 27(6): 644 - 646. [Full Text] [PDF] |
||||
![]() |
J. Chambers, J. Roxburgh, C. Blauth, J. O'Riordan, F. Hodson, and H. Rimington A randomized comparison of the MCRI On-X and CarboMedics Top Hat bileaflet mechanical replacement aortic valves: Early postoperative hemodynamic function and clinical events J. Thorac. Cardiovasc. Surg., September 1, 2005; 130(3): 759 - 764. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-H. Sievers Prosthetic aortic valve replacement J. Thorac. Cardiovasc. Surg., May 1, 2005; 129(5): 961 - 965. [Full Text] [PDF] |
||||
![]() |
T. Bottio, L. Caprili, D. Casarotto, and G. Gerosa Small aortic annulus: The hydrodynamic performances of 5 commercially available bileaflet mechanical valves J. Thorac. Cardiovasc. Surg., September 1, 2004; 128(3): 457 - 462. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |