J Thorac Cardiovasc Surg 2003;126:1568-1574
© 2003 The American Association for Thoracic Surgery
Evolving technology |
a Division of Thoracic and Cardiovascular Surgery, Medical School, Hannover, Germany
b Department of Cardiac Surgery, Heart Center University, Leipzig, Germany
c Department of Thoracic and Cardiovascular Surgery, JW-Goethe University, Frankfurt, Germany
Received for publication May 22, 2002; revisions received August 15, 2002; revisions received March 12, 2003; accepted for publication April 10, 2003.
* Address for reprints: Uwe Klima, MD, PhD, Department of Thoracic and Cardiovascular Surgery, Hannover Medical School, 30623 Hannover, Germany
klima{at}thg-mh.hannover.de
| Abstract |
|---|
|
|
|---|
METHODS: In a multicenter trial, 32 patients (mean age: 65 ± 9 years; 85% men) requiring multivessel coronary artery bypass surgery had 1 of the anastomoses performed using a novel anastomotic technology. The Magnetic Vascular Positioner System was used in 1 of the bypass grafts and the other bypasses were completed by conventional hand-sewn technique. The Magnetic Vascular Positioner System consists of 4 magnetic, gold-plated implants and 2 delivery devices that facilitate the creation of a functional end-to-side anastomosis. A predischarge angiogram was performed to evaluate graft patency.
RESULTS: There were no device-related major adverse events. The application of the Magnetic Vascular Positioner device was successful in 32 of 41 cases (78%). Nine patients were intended for treatment but did not receive the Magnetic Vascular Positioner System. In 5 of the cases the coronary artery was too small; 1 case had a posterior wall plaque in the target artery; and 3 patients had a nonhemostatic anastomosis after coupling of the port and were subsequently converted to hand-sewn anastomoses. The median total Magnetic Vascular Positioner anastomotic time was 137 seconds with a range from 65 to 370 seconds. Overall patency rate of the Magnetic Vascular Positioner anastomosis was 93.5% versus 91.7% (P = not significant) in hand-sewn grafts. One patient (3.1%) died due to low cardiac output but had patent grafts at autopsy. One myocardial infarction (3.1%) occurred the day after a percutaneous transluminal coronary angioplasty of a hand-sewn graft. One prolonged mechanical ventilation (3.1%) was required because of pneumonia and adult respiratory distress syndrome.
CONCLUSIONS: Magnetic vascular coupling in coronary surgery is safe and effective and has acceptable early patency rates. This new technique may facilitate beating heart and minimally invasive coronary artery bypass grafts.
One goal in coronary surgery is to achieve complete revascularization through a small incision preferably on the beating heart. Currently, few centers are performing these technically demanding and still developmental procedures.9,10 Widespread adoption is unlikely without significant advances in technology such as stabilizers and easily deployable and reliable anastomotic devices.
This multicenter clinical trial investigated the safety and efficacy of a novel anastomotic device, the Magnetic Vascular Positioner (MVP) System (Ventrica, Inc, Fremont, Calif) to facilitate the construction of a distal coronary anastomosis.
| Methods |
|---|
|
|
|---|
80%, and free of plaque or calcification at the anastomotic site. Subjects excluded from the study had a left ventricular ejection fraction <30% or a contraindication to either aspirin or clopidogrel and those patients undergoing emergent CABG, redo CABG, or requiring additional noncoronary procedures. One anastomosis in each patient was performed by means of a magnetic vascular coupling device as described below. All other anastomoses were hand-sewn. Left internal thoracic arteries (LITAs) and saphenous vein grafts (SVGs) were used with the MVP device.
|
Operative procedure
Standard techniques for induction and maintenance of anesthesia during the procedure were used in all 3 centers. Twenty-six patients were operated on using standard cardiopulmonary bypass techniques and cardiac arrest. Heparin (350 U/kg) was administered initially; activated clotting time levels were thereafter targeted to remain above 400 seconds. Six patients had an off-pump beating heart procedure with mechanical stabilization using the Axius stabilization system (Guidant, Santa Clara, Calif). Aprotinin (Trasylol; Bayer, Leverkusen, Germany) or epsilon amino capron acid was not routinely used.
Target vessel and intention to treat
The target vessel and site of anastomosis was chosen intraoperatively based on the surgeons' personal subjective decision. The anastomotic site was required to be free of plaque and to have a minimum intraluminal diameter of 2.0 mm. Intention to treat was defined as intraoperative decision to attempt to perform an MVP anastomosis.
Description of device
The MVP System consists of 4 magnetic implants and 2 delivery systems for completion of an anastomosis. A pair of magnetic, elliptically shaped, gold-plated implants with an open area of 8.1 mm2 is placed into a 5.0 mm longitudinal incision near the end of a ligated graft vessel to create an anastomotic port (Figure 1).
One magnetic implant lies just within the graft lumen at the incisional site (intravascular magnetic implant). The other is held in place on the external surface of the vessel by mutual magnetic attraction, precisely above the first implant (extravascular magnetic implant). The implants hold the elliptically shaped anastomotic port open. The second pair of magnetic implants forms an identical port in the coronary artery and is introduced in the same manner as was done at the graft port. The 2 ports formed by the magnetic implant pairs are then brought together, allowing the magnetic attraction of the 2 pairs to maintain apposition and form a reliable side-to-side anastomosis (functional end-to-side). The distal end of the bypass was either clipped or tied off. In addition, 2 to 4 tacking sutures were placed according to the protocol to secure the position of the MVP graft.
|
Statistical analysis
Data are expressed as mean ± standard deviation. In addition, 95% confidence intervals were calculated to facilitate comparison of patency rates between groups. After testing for normal distribution the paired Student t test was used for intragroup comparison of the Canadian Cardiovascular Society (CCS) angina class; P < .05 was considered significant.
| Results |
|---|
|
|
|---|
|
|
Major adverse events
There was 1 death (3.1%) in this series, unrelated to the device. The patient had a hand-sewn LITA graft to the left anterior descending artery (LAD), a hand-sewn SVG to the second obtuse marginal branch (OM), and an MVP SVG to the intermediate branch. After an initially uneventful operation the patient's electrocardiogram (ECG) showed ischemia of the anterior wall and septum. Cardiac enzymes were within normal limits. The patient was reexplored and no graft occlusions were found. The cause of death has been classified as low cardiac output syndrome. At autopsy all grafts were confirmed as patent. No histological studies of the MVP device were performed, as the device was not explanted at the time of autopsy.
One patient (3.1%) suffered a myocardial infarction with positive changes in ECG leads II, III, and aVF. This patient had a hand-sewn LITA graft to the LAD, a hand-sewn SVG to the first OM, and an MVP SVG to the posterior descending artery. At the predischarge angiogram a stenosis was detected in the midsection of the SVG to the OM1 hand-sewn graft. Though the patient was asymptomatic, a percutaneous transluminal coronary angioplasty was performed in the stenosed graft. The following day, the patient sustained a myocardial infarction. The patient was hemodynamically stable and did not undergo any further reintervention. This event was not device related.
One patient (3.1%) required prolonged mechanical ventilation due to pneumonia and adult respiratory distress syndrome. The patient was discharged to rehabilitation for ventilator weaning.
Other adverse events
There were 3 (9.3%) reexplorations for bleeding. One patient sustained a cardiac tamponade 2 hours after operation and was returned to the operating room. No source of bleeding could be identified; there were no anastomotic leaks. One patient was reexplored for bleeding on the ninth postoperative day. No source of bleeding could be identified. The third patient experienced bleeding from a side branch of a radial artery graft leading to a hand-sewn anastomosis. These 3 patients reached the 30-day clinical end point and continued to be followed. There were 2 (6.1%) transluminal coronary interventions in this series: 1 patient with an occluded MVP anastomosis and 1 patient with a stenosed vein graft. Both patients were asymptomatic and were diagnosed during the routine predischarge angiogram. The interventions were performed at the time of the diagnostic procedure. The patient with the MVP occlusion was treated by percutaneous stenting of the native coronary artery (SVG to OM1). The patient with the stenosed hand-sewn graft had a percutaneous angioplasty procedure to the graft (SVG to OM1).
There was 1 (3.1%) transient ischemic attack prior to hospital discharge. The patient's cerebral computed tomography showed no morphologic changes. Recovery was complete.
Intention to treat
A successful anastomosis was performed in 78% of cases (32/41). In addition to the 32 patients who received an MVP anastomosis, 9 patients were classified as intention to treat patients. Four of the 9 patients initially received the MVP implant but were converted to a hand-sewn anastomosis for the following reasons: posterior plaque formation in 1 patient and 3 patients had inadequate hemostasis at the anastomotic site after the coupling due to insufficient tissue capture within the magnetic implants. This is an error in deployment technique and was noted upon immediate test flushing of the anastomosis. In the other 5 patients the magnets were not inserted into the coronary artery due to improper judgment of the target vessel diameter. The target vessels appeared to be larger from the outside; however, after incising the vessels, the internal diameter was unacceptable for inserting the magnets due to sclerotic narrowing.
Angiographic follow-up
A total of 32 patients (32 MVP and 74 hand-sewn grafts) received the MVP device. Of these, 31 patients (1 death) had a postoperative coronary angiogram prior to hospital discharge (mean of 6.3 ± 5.4 days). Thirty-one MVP and 72 hand-sewn anastomoses were evaluated. The overall patency rate was 93.5% (29 of 31) in the MVP group versus 91.7% (66 of 72) in the hand-sewn group (P = not significant). In addition to the occluded grafts, 3 hand-sewn grafts were found to have stenoses >50%. There was a 100% (9 of 9) patency rate in left internal thoracic to LAD MVP anastomoses.
Thirty-day clinical follow-up
There was significant clinical improvement at the 30-day evaluation with a mean CCS status of 0.37 ± 0.5 at 30 days postoperatively versus 2.3 ± 0.8 preoperatively (P < .0001). Angina modification, as assessed by CCS classification, measured a secondary efficacy end point. On the preprocedure evaluation, 81.2% of assessments indicated a class II or class III angina status, with the remainder falling into class I or class IV. At the 30-day evaluation 63.3% of patients were evaluated as class 0 and 36.7% as class I.
No new major adverse events occurred within the 30-day follow-up period. Minor adverse events included new onset of atrial fibrillation in 2 patients (6.5%), pneumonia in 1 patient (3.2%), and leg wound infection in 1 patient (3.2%). The overall minor adverse event rate was 13.1%. Of the 31 surviving patients, ECGs demonstrated no new myocardial infarction or Q-wave changes during a mean follow-up of 32 days.
| Discussion |
|---|
|
|
|---|
There was not an appreciable learning curve associated with handling of the device. Only a few deployments in the animal model were needed to become familiar with the different delivery aspects of the MVP system. An important part of the MVP procedure is the arteriotomy. A targeted arteriotomy length of 4.5 to 5.0 mm was desired. There is approximately 1.0 mm of variability that can be accommodated by the device. No cases had to be aborted due to improper incision length.
As described in Table 2 the MVP device could be applied to all 3 coronary arteries and their side branches. Both LITAs and SVGs were used as bypass conduits for the MVP anastomosis. The radial artery was not used with the MVP as the preclinical testing had been done on internal thoracic arteries and SVGs only. However, the device appears to be compatible with radial arteries based on our evaluation of remnant pieces of radial artery during the clinical investigation. Consequently, the MVP system seems to be applicable to all target vessels and graft conduits.
Limitations of study and device
This iteration of the device was targeted for coronary vessels with an intraluminal diameter
2.0 mm. Of the vessels treated with the MVP, 30 were within this range and 2 were below. At the moment only 1 size magnet is available, limiting the applicability to target vessels with smaller diameters. A second generation of downsized magnets is currently being evaluated with early results in animals demonstrating similar patency rates in target vessels as small as 1.5 mm. A device for vessels smaller than 1.0 mm will not likely be needed due to questionable perfusion benefit as well as poor patency rates in these coronary targets.12 The current version of the device with a straight delivery system limits its flexibility on the posterolateral wall. Even though an anastomosis could be performed in the circumflex artery area, an angled delivery system will certainly facilitate MVP placement in posterolateral vessels. For purposes of this study, anastomotic target sites with diffuse disease, calcification, and visible plaques were avoided. However, we were routinely able to find an acceptable target site. In cases where an ideal deployment was not achieved (eg, too much tissue in the lumen), the device could be easily removed with no apparent trauma to the tissue. Following removal of the MVP, the device could be redeployed or a hand-sewn anastomosis could be performed if desired. This bailout feature helped to provide an extra level of comfort when performing an anastomosis with a new, mechanical device.
Angiographic results
The predischarge angiographic patency of 93.5% for the MVP system was very encouraging and compares favorably with other CABG trials with angiographic end points at this time point.13,14 Of particular note was the 100% patency rate of the MVP in the LITA-LAD position. In the group of hand-sewn anastomosis 3 grafts had a stenosis >50%. These anastomoses were classified as patent in our analysis. The reason for doing so was that no quantitative coronary flow analysis was performed and the results could not be objectively assessed in either group. Due to the radiopacity of the MVP system, a stenosis in the MVP anastomosis could not be assessed angiographically, and hence an anastomosis was classified as patent when good runoff could be demonstrated.
Anticoagulation regimen
Preoperative antiplatelet therapy was instituted because the MVP system has exposed foreign material in the blood flow path. Furthermore, it is yet unknown in humans if the application of a magnet into the coronary arteries may cause similar problems as seen after stent application with intimal hyperplasia formation and restenosis. The combination of aspirin and clopidogrel was chosen for several reasons. Clopidogrel, a platelet adenosine diphosphate receptor antagonist, has been shown to inhibit platelet aggregation and to prevent the recurrence of ischemic events more effectively than aspirin. This is due to a short-lasting metabolite produced by activation of hepatic cytochrome P450-1A.15 Recent experimental data demonstrated that the antiaggregating, antithrombotic, and antiproliferative activity of clopidogrel was potentiated by aspirin.16 Patients received preoperative loading dose of clopidogrel and aspirin to obtain the protective antithrombotic effect immediately at the end of the operation. Prior to commencing clinical trials with the MVP system, extensive preclinical testing was performed. Acute and chronic animal studies (swine and ovine) were performed to demonstrate acceptable performance of the device. In yet unpublished experimental data a patency rate of 97% after a 1-month follow-up period was demonstrated using this antiplatelet regimen. Additionally, animals killed at 3 and 6 months were found to have similar patency rates and demonstrated a fully healed and stable tissue layer that lined the lumen and covered the magnets. Histological analysis showed a smooth, uniform endothelial covering, lining the entire lumen.17 Hence, we decided to administer a combination of clopidogrel and aspirin for 12 weeks, and thereafter a single antiplatelet therapy with aspirin only.
Prior to surgery, many patients are currently taking antiplatelet drugs for various reasons. These include routine antiplatelet administration following prophylactic loading in the catheterization laboratory, stent implantation, and clopidogrel therapy for multivessel disease. To minimize potential thrombotic problems we thought it prudent and reasonable to use full antiplatelet therapy. Since commencement of this trial, recent studies have been published regarding reexploration rates in CABG patients on antiplatelet therapy. Hongo and colleagues18 reported a 6.8% reoperation rate and Yende and colleagues19 reported a rate of 9.8%. We now have additional animal data to support using reduced antiplatelet dosage in subsequent trials.
| Conclusion |
|---|
|
|
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
W. J.L. Suyker and C. Borst Coronary Connector Devices: Analysis of 1,469 Anastomoses in 1,216 Patients Ann. Thorac. Surg., May 1, 2008; 85(5): 1828 - 1836. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Klima, J. Beilner, E. Bagaev, S. Fischer, T. Kofidis, and J. Lotz MRI-Based Safety Evaluation of the Ventrica Magnetic Coronary Anastomotic System Asian Cardiovasc Thorac Ann, February 1, 2007; 15(1): 24 - 29. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Vicol, S. Eifert, M. Oberhoffer, P. Boekstegers, and B. Reichart Mid-term patency after magnetic coupling for distal bypass anastomosis in coronary surgery. Ann. Thorac. Surg., October 1, 2006; 82(4): 1452 - 1456. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. J.L. Suyker, J. P. Matonick, P. T.W. Suyker, A. Brutel de la Riviere, M. P. Buijsrogge, R. P.J. Budde, C. W.J. Verlaan, G. Pasterkamp, P. F. Grundeman, and C. Borst S2 Connector Versus Suture: Distal Coronary Anastomosis Remodeling, Patency, and Function in the Pig Circulation, July 4, 2006; 114(1_suppl): I-390 - I-395. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Klima, T. Kofidis, J. Beilner, M. Maringka, S. Kirschner, and A. Haverich Anastomotic Obstruction After Magnetic Vascular Coupling After Minimally Invasive Direct Coronary Artery Bypass Grafting Ann. Thorac. Surg., February 1, 2006; 81(2): 728 - 729. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, W. J.L. Suyker, P. T.W. Suyker, C. W.J. Verlaan, R. Meijer, C. Borst, and P. F. Grundeman Quality assessment of distal S2AS connector anastomosis by 13MHz epicardial ultrasound Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 833 - 837. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Klima, T. Kofidis, A. Boening, and A. Haverich Long term angiographic monitoring reveals safety and efficacy of the Converge end-to-side anastomotic coupler Interactive CardioVascular and Thoracic Surgery, December 1, 2005; 4(6): 505 - 508. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Boening, F. Schoeneich, A. Lichtenberg, E. Bagaev, J. T. Cremer, and U. Klima First clinical results with a 30{degrees} end-to-side coronary anastomosis coupler Eur. J. Cardiothorac. Surg., May 1, 2005; 27(5): 876 - 881. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Vicol, S. Eifert, M. Oberhoffer, P. Boekstegers, A. Knez, F. Christ, and B. Reichart Early Clinical Results With a Magnetic Connector for Distal Coronary Artery Anastomoses Ann. Thorac. Surg., May 1, 2005; 79(5): 1738 - 1742. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. P. Kypson and W. R. Chitwood Jr INVITED COMMENTARY Ann. Thorac. Surg., May 1, 2005; 79(5): 1742 - 1743. [Full Text] [PDF] |
||||
![]() |
U. Klima, H. MacVaugh III, E. Bagaev, M. Maringka, S. Kirschner, J. Beilner, and A. Haverich Magnetic Vascular Port in Minimally Invasive Direct Coronary Artery Bypass Grafting Circulation, September 14, 2004; 110(11_suppl_1): II-55 - II-60. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Carrel, L. Englberger, D. Keller, S. Windecker, B. Meier, and F. Eckstein Clinical and angiographic results after mechanical connection for distal anastomosis in coronary surgery J. Thorac. Cardiovasc. Surg., June 1, 2004; 127(6): 1632 - 1640. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||