J Thorac Cardiovasc Surg 2003;126:2022-2031
© 2003 The American Association for Thoracic Surgery
Surgery for acquired cardiovascular disease |
a Division of Cardiothoracic Surgery, Medical University of South Carolina, Charleston, SC, USA
Read at the Eighty-third Annual Meeting of The American Association for Thoracic Surgery, Boston, Mass, May 4-7, 2003.
Received for publication April 23, 2003; revisions received June 27, 2003; accepted for publication July 7, 2003.
* Address for reprints: John S. Ikonomidis, MD, PhD, Assistant Professor, Cardiothoracic Surgery, Medical University of South Carolina, Suite 409 CSB, 96 Jonathan Lucas Street, Charleston, SC 29464, USA
ikonomij{at}musc.edu
| Abstract |
|---|
|
|
|---|
METHODS: We prospectively followed 837 valve recipients (aortic valve replacement; n = 478; mitral valve replacement; n = 359) from January 1979 to December 2000 at 12-month intervals.
RESULTS: Ages ranged from 19 to 84 years. Follow-up averaged (mean ± standard deviation) 7 ± 5 years (98% complete). Patients were in New York Heart Association class III or IV in 77% (aortic valve replacement) and 89% (mitral valve replacement) preoperatively. A 19-mm valve was implanted in 15.5% of aortic valve replacement patients. Coronary bypass was required in 31% of aortic valve replacements and 20% of mitral valve replacements. Operative mortality was 17/478 (3.6%) in aortic valve replacement and 19/359 (5.3%) in mitral valve replacement, and multivariable predictors were 19-mm valve size, 3 or more coronary bypass grafts, and New York Heart Association class IV for aortic valve replacement and New York Heart Association class IV and age for mitral valve replacement. Actuarial survivorship at 10 and 20 years was 57% ± 3% and 26% ± 5% for aortic valve replacement and 61% ± 3% and 39% ± 4% for mitral valve replacement. Multivariable predictors of late death were African-American ethnicity, New York Heart Association class III or IV, coronary bypass, and age for aortic valve replacement and New York Heart Association class III or IV, coronary bypass, and age for mitral valve replacement. For aortic valve replacement, effective orifice area was univariately (P = .002) but not multivariately (P = .378) predictive of late death. Structural valve deterioration was not observed. For aortic valve replacement, actuarial freedom (at 10 and 20 years) from reoperation was 93% ± 1% and 90% ± 2%; thromboembolism, 82% ± 3% and 68% ± 8%; bleeding events, 77% ± 3% and 66% ± 6%; prosthetic valve endocarditis, 94% ± 1% and 94% ± 1%; valve-related mortality, 94% ± 2% and 86% ± 4%; and valve-related mortality or morbidity, 58% ± 3% and 32% ± 8%. For mitral valve replacement, actuarial freedom (at 10 and 20 years) from reoperation was 96% ± 1% and 90% ± 3%; thromboembolism, 77% ± 3% and 59% ± 7%; bleeding events, 86% ± 2% and 65% ± 8%; prosthetic valve endocarditis, 98% ± 1% and 96% ± 2%; valve-related mortality, 89% ± 0.2% and 74% ± 8%; and valve-related mortality or morbidity, 63% ± 3% and 29% ± 7%.
CONCLUSIONS: After 2 decades of observation with close follow-up, the St Jude Medical mechanical valve continues to be a reliable prosthesis.
| Patients and methods |
|---|
|
|
|---|
|
In the AVR group, the standard St Jude Medical mechanical prosthesis was used in 378 patients. The Masters series valve was used in 78 patients and the Hemodynamic plus valve was used in 22 patients. In the MVR group, the standard St Jude Medical prosthesis was used in 266 patients and the Masters series valve was used in 93 patients. Thirty-three patients received AVR with valve sewing rings impregnated with Silzone, as did 37 patients undergoing MVR.
Heparin administration was started (5000 units subcutaneously every 6-12 hours depending on the surgeon preference) on the first postoperative day and continued until prothrombin time or more recently the International Normalized Ratio (INR) was regulated with the administration of warfarin sodium. Early in our experience, the INR was maintained at 2.5 to 3.5. More recently, we have recommended a target INR of 2.0 for patients who have undergone AVR and 2.0 to 2.5 for MVR. The warfarin dose was regulated by the surgeon during hospitalization and during early follow-up and thereafter by the referring physician.
Statistical analysis
Continuous variables are reported as the mean ± 1 standard deviation and the categorical variables are represented as percentages. Actuarial curves were constructed to describe mortality and the incidence of valve-related complications using the Kaplan-Meier technique and are reported with the standard error of the mean.12-14 Because actuarial analysis is known to overestimate the incidence of nonfatal end points, "actual" or cumulative incidence curves were constructed for both the AVR and MVR groups for the end points of reoperation, endocarditis, thromboembolic events, and anticoagulation-related hemorrhage.15 Linearized rates, used to describe multiple events, were reported as %/patient-year and those reported for the Silzone valve subset were compared to the remainder of the valve cohort using the exact binomial test. Independent factors associated with operative and late death were identified using logistic regression16 and the Cox proportional hazards regression model,17 respectively. Variables used for both the AVR and MVR groups are summarized in the Appendix. Effective orifice areas (EOAs) used for analysis were kindly provided by St Jude Medical, Inc.
| Results |
|---|
|
|
|---|
|
|
|
Kaplan-Meier analysis showed the 10- and 20-year freedom from thromboembolic events to be 82% ± 3% and 68% ± 8% in the AVR group and 77% ± 3% and 59% ± 7% in the MVR group. Cumulative incidence analysis showed 10- and 20-year freedom from thromboembolic events to be 86% ± 2% and 79% ± 3% in the AVR group and 80% ± 3% and 71% ± 4% in the MVR group. Actuarial and cumulative incidence freedom from thromboembolic events is shown in Figure 3.
|
|
Valve-related mortality
In the AVR cohort, 37 patients died of valve-related causes: embolism in 11 patients, valve thrombosis in 2 patients, bleeding events in 13 patients, prosthetic endocarditis in 6 patients, death related to reoperation for valve re-replacement in 4 patients, and death related to catheterization for perivalvular leak in 1 patient. In addition, 29 patients died of unknown causes. In the MVR group, 21 patients died of valve-related causes. These were due to embolism in 10 patients, bleeding events in 8 patients, prosthetic endocarditis in 2 patients, and death related to reoperation for valve re-replacement in 1 patient. Twelve patients died of unknown causes.
Kaplan-Meier analysis showed freedom from valve-related deaths to be 94% ± 2% and 86% ± 4% at 10 and 20 years in the AVR group and 89% ± 2% and 74% ± 8% at 10 and 20 years in the MVR group. Actuarial freedom from valve-related mortality is given in Figure 5.
|
Valve prostheses with Silzone-impregnated sewing rings
A total of 70 patients received valve prostheses with Silzone-impregnated sewing rings between May 1998 and December 1999 and were followed for a total of 111 patient-years. Analysis of the Silzone subset compared with the remaining cohort of 767 patients (5638 patient-years) showed a significantly higher incidence of prosthetic valve endocarditis (2.7%/patient-year vs 0.5%/patient-year, P = .02). All of these patients showed clinical and operative findings of endocarditis but were culture-negative. There were also nonsignificant trends toward increased reoperation (2.7%/patient-year vs 0.7%/patient-year, P = .053) and late death (9.0%/patient-year vs 5.0%/patient-year, P = 0.08) in the Silzone subset.
| Discussion |
|---|
|
|
|---|
Operative mortality in the current series is 3.6% in the AVR group and 5.3% in the MVR group. These results compare favorably with other long-term studies of this prosthesis.8-10 Our data continue to show that preoperative patient characteristics such as age, previous valve replacement, cardiac function, and coronary disease may have a greater influence on operative mortality than the type of valve prosthesis implanted. Similar operative mortality is seen in other series of tissue and other commonly used mechanical valves.20-23
As other studies have observed,8-10 our data show that late death is usually unrelated to the prosthetic valve. Multivariable analysis demonstrated that the major independent variables impacting the long-term outcome in this patient population were high NYHA class, concomitant CABG, and increasing age in the both the AVR and MVR groups. In addition, in the AVR group African-American ethnicity emerged as a strong predictor of outcome. The reason for this is unclear. Previous studies have shown that African-American ethnicity poses a significantly higher risk for hypertension, diabetes mellitus, and renal disease.24-26 Subgroup analysis within this cohort showed no significant differences between African-American and other ethnic groups with respect to these variables. Further analysis is required to identify factors that influence this predictor of outcome.
Thromboembolism remains a major complication after mechanical valve replacement. The linearized rate of thromboembolic complications in this series (AVR 2.0%/patient-year, MVR 3.4%/patient-year) is slightly lower than in our last report. This may reflect improved ability to reach and maintain acceptable anticoagulation status in our patients. The incidence of thromboembolism is comparable with other series of St Jude Medical mechanical valve replacement (0.8% to 3.5%/patient-year).3-6,8-10 These thromboembolic rates are also comparable with the composite rates for other mechanical valves and tissue valves.20-22
The incidence of anticoagulation-related bleeding complications has remained relatively low. The linearized rate was 2.5%/patient-year (AVR) and 1.8%/patient-year (MVR). This represents a considerable reduction in anticoagulant-related bleeding compared with our original 5-year study (AVR, 6.9%/patient-year; MVR, 2.4%/patient-year),1 a modest reduction compared with our 10-year report (AVR, 3.5%/patient-year; MVR, 2.2%/patient-year),2 and little difference from our 15-year report (AVR, 2.7%/patient-year; MVR, 1.6%/patient-year).7 Other investigators have also observed a higher incidence of anticoagulation-related bleeding after AVR compared with MVR.5 In general, we have been impressed with the low thromboembolic rate of the St Jude Medical mechanical prosthesis and have gradually lowered our target INR from 2.5 to 3.5 to 2.0 to 2.5 as have other groups,27,28 and this may account for the progressive decline in bleeding complications. Overall, however, thromboembolic and bleeding complications continue to be the Achilles' heel of mechanical valve replacement, with cumulative yearly rates of 4.5% for AVR and 5.3% for MVR, accounting for most of the valve-related morbidity seen in these patients.
The incidence of prosthetic valve endocarditis was quite low (AVR, 0.8%/patient-year; MVR, 0.3%/patient-year) and is similar to other series.8-10 It is interesting that after approximately 5 years, there is near complete freedom from endocarditis in both the aortic and mitral valve groups. It has been speculated that valve design, increased probability of prosthetic valve sterilization, and early incorporation of the cloth sewing ring by tissue in growth may offer some protection from late bacterial invasion.
Data from the 70 patients in this series who received Silzone-treated valves showed a higher linearized rate of prosthetic valve endocarditis. The association between Silzone treatment and valve-related complications is somewhat controversial but has been documented in other series.29-31 This valve was not used in our institution after December 1999 because of concerns regarding an increased rate of endocarditis and valve dehiscence.
Long-term follow-up of these patients has confirmed that undergoing prosthetic valve replacement does not confer a survival rate like that of the general population. To justify this within our data set, age-adjusted standard mortality rates32 (SMR) were calculated for mitral and aortic valves combined. The expected number of deaths was propagated using the age-specific national death rates for the year 200033 and the number of corresponding at risk patient-years for our valve population. Because the proportion of males to females in our study differed from that of the national population, a separate SMR was calculated for each gender. The results showed that our valve population had mortality rates significantly higher than those of the general population (men, SMR 4.3; women, SMR 5.4; P < .0001).
Although 20-year freedom from valve-related complications was only 32% (AVR) and 29% (MVR), data from this study and others6 suggest that overall long-term survival after prosthetic valve replacement may be more strongly influenced by preexisting comorbidities prior to surgery than by the presence of a mechanical valve. As increasing numbers of elderly patients with more comorbidities undergo valve replacement, the influence of valve-related factors on mortality rate is likely to continue to decline compared with patient-related factors. In addition, with the improved durability of later generation tissue valves, the long-term results of mechanical valve replacement are also likely to be affected by increased use of bioprosthetic valves in younger patients.
In summary, this series of patients undergoing AVR and MVR with the St Jude Medical mechanical prosthesis with follow-up to 20 years confirms the excellent performance of this valve that we have documented in our earlier reports. In addition, the complete absence of structural valve deterioration makes the St Jude Medical mechanical heart valve prosthesis an excellent choice for patients who require mechanical prosthetic valve replacement.
| Discussion |
|---|
|
|
|---|
Two years ago we published our 20-year comparison between St Jude Medical and tissue valves. There were 670 St Jude Medical valves and 720 tissue valves in the aortic position, about 500 St Jude Medical valves and 400 tissue valves in the mitral position. The mean age of patients with the St Jude Medical valve was almost 10 years greater. The 10- and 20-year actuarial survival for all the patients with mechanical and biological prostheses was similar to that presented here by you today.
Freedom from thromboembolic events was also similar for both St Jude Medical and tissue valves in both aortic and mitral position. Freedom from hemorrhage was also similar. The only difference is that in the aortic position, biological prostheses fared much better, with a freedom from hemorrhage of 94% and 92% at 10 and 20 years, respectively.
We have also learned from other retrospective studies, as well from the 2 randomized studies that are published, that in general, long-term survival is not influenced by the type of valve that we use, mechanical or biologic, but by other factors, as you mentioned, like age, coronary artery disease, New York Heart Association class, diabetes, and finally, and probably, valve to patient size mismatch.
In view of this background, justifiably or not, over the past 10 years, in the aortic position we have reduced the use of St Jude Medical valves from 55% in 1991 to less than 20% in 2002. In the mitral position the same things happened in favor of mitral valve repair.
Quality of life, freedom from warfarin, and increased mobility of the patient population, which makes follow-up more difficult, have steered patients and physicians away from warfarin. Has your experience been similar on the choice of valves? Could you also relate the thromboembolic and hemorrhagic complications to the patient's INR when they happen? Was the patient's INR out of range?
Now, the most important issue is the warfarin therapy. How low do you think you can push the INR? Do you think we can have a safe anticoagulation for the aortic valve with an INR between 1.5 and 2 for the aortic position and an INR around 2 for the mitral position? That would make warfarin therapy much more attractive.
Dr Ikonomidis (Charleston, SC). Thank you, Dr Trento, for those insightful comments and questions. With regard to comparing mechanical valves to tissue valves, what we have observed is what has generally been observed in most studies, namely that the durability of mechanical valves exceeds that of tissue valves, whereas tissue valves show superior freedom from thromboembolic and bleeding complications.
As far as relating thromboembolic and hemorrhagic complications to the patient's INR, we didn't specifically relate the incidence of the complication to the INR, but that would be something that worth looking at. As far as decreasing the targeted INR is concerned, despite the fact that we have decreased the INR in our institution, this has not resulted in an increase in thromboembolic rates; in fact, it has still dropped to a small extent. Therefore, at the present time we feel that our current target INRs are safe.
Dr Kit V. Arom (Minneapolis, Minn). I rise to congratulate the group from Charleston for their excellent long-term follow-ups, particularly with 98% completed. We, too, in Minneapolis, after Demetre Nicoloff implanted the first valve in 1977, have followed these people up to 20 years. At that time with 95.4% complete (21,342 patient years), we had 3937 patients available for the study. About 1300 of the patients were dead at that time or approximately one third of them.
The causes of death were 4.5% valve-related, 4% related to anticoagulation. The rest of them died from ongoing atherosclerosis and heart failure (48%). Linearized rate (percent per patient-year) included 1.24 for thromboembolism (TE), 1.00 for bleeding, 0.1 for perivalvular leak, 0.18 for prosthetic endocarditis, and valve thrombosis was 0.09.
If we translate this into actuarial analysis, we obtained results similar to yours, except in 2 areas. One is that at the end of 20 years our aortic valve had better survival rate than the mitral valve, being about 44%, comparing with about 33% for the mitral valve.
And the other area was the incidence free from bleeding and TE. There was about 82% incidence free from bleeding and 70% free from TE. If I remember correctly, this is quite different than yours, which it is about 60%.
We did lower the intensity of anticoagulation in 1985 by keeping the INR between 1.8 and 2.5 for AVR, 2.5 and 3.2 for MVR.
Dr Trento has suggested lowering intensity of anticoagulation, which I agree with. At the present time we still using a slightly lower intensity than what you just described, and I think this could be the reason why we see a slightly better TE and bleeding rate.
Dr Ikonomidis. How did you conduct the follow-up on your patients?
Dr Arom. Our follow-up is not quite as good as yours, which is being done yearly. We hope, however, that we can use the same approach that you do. In 1985, we also instructed the cardiologist and family physician to stop using prothrombin values and to have the INR strictly within the range that I have mentioned. I really believe that this approach has contributed to our recent outcomes.
Dr Ikonomidis. The less aggressive warfarin dosing that you described may help address your improved freedom from TE and bleeding compared with our series. In addition, conducting follow-up over longer time intervals may cause patients to forget some events, effectively increasing the documented freedom from TE and bleeding complications.
As for addressing the flip in terms of aortic and mitral valve survival, I think that this discrepancy may reflect inherent differences in the population as they relate to severity of comorbid factors such as hypertension, left ventricular hypertrophy, and perhaps even compliance with medications.
Dr W. R. Eric Jamieson (Vancouver, British Columbia, Canada). Congratulations on an excellent study. I just have 2 short questions. You shared with us your linearized occurrence rates for aortic and mitral valves. Could you share with us the linearized rates for major embolisms, including Reversible Ischemic Neurologic Deficits, and comment also on early and late events, because many patient mechanical populations can have major events in the first 30 days.
My second question is related: have you looked at actual versus actuarial freedoms when you assess the composites of complications, morbidity, mortality, and reoperation, which are probably better parameters when you are counseling patients to have a mechanical or any prosthesis?
Dr Ikonomidis. I am sorry, Dr Jamieson, can you repeat the second question?
Dr Jamieson. The concept of using actual freedoms compared with actuarial freedoms, which are better parameters to use when counseling patients as to which prosthesis they should have.
Dr Ikonomidis. All of the nondeath outcomes were analyzed by cumulative incidence analysis. I did not include them here in this presentation for purposes of simplicity but in all end points measured, the cumulative incidence curves were all shifted upward compared with the actuarial freedom curves, as one would expect. With regard to neurologic events, approximately one third were early and then the remaining two thirds occurred later on.
Dr Mohammed A. Quader (Cleveland, Ohio). Congratulations on your excellent presentation of this valuable data. I would like to ask you a question about the use of 19-mm valves. Fifteen percent of the patients had the 19-mm valve placed, and multivariate analysis has shown that there is an increased risk of death with that. Has there been a trend toward not using this valve more frequently in the recent past?
Also, have you looked for patient-prosthesis mismatch in that group of patients?
Dr Ikonomidis. We have trended toward not using 19-mm valves as much as possible. As far as patient-prosthesis mismatch is concerned, neither EOA nor EOA index were found to be multivariate predictors of death.
Appendix
Variables examined for association with operative mortality and late death
|
NYHA, New York Heart Association; CABG, coronary artery bypass grafting.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
R. C. Elkins, D. M. Thompson, M. M. Lane, C. C. Elkins, and M. D. Peyton Ross operation: 16-year experience J. Thorac. Cardiovasc. Surg., September 1, 2008; 136(3): 623 - 630. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Gudbjartsson, T. Absi, and S. Aranki Mitral Valve Replacement Card. Surg. Adult, January 1, 2008; 3(2008): 1031 - 1068. [Full Text] |
||||
![]() |
Y. S. Tjang, Y. van Hees, R. Korfer, D. E. Grobbee, and G. J.M.G. van der Heijden Predictors of mortality after aortic valve replacement Eur. J. Cardiothorac. Surg., September 1, 2007; 32(3): 469 - 474. [Abstract] [Full Text] [PDF] |
||||
![]() |
B.-K. Lam, V. Chan, P. Hendry, M. Ruel, R. Masters, P. Bedard, B. Goldstein, F. Rubens, and T. Mesana The impact of patient-prosthesis mismatch on late outcomes after mitral valve replacement J. Thorac. Cardiovasc. Surg., June 1, 2007; 133(6): 1464 - 1473. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Magne, P. Mathieu, J. G. Dumesnil, D. Tanne, F. Dagenais, D. Doyle, and P. Pibarot Impact of Prosthesis-Patient Mismatch on Survival After Mitral Valve Replacement Circulation, March 20, 2007; 115(11): 1417 - 1425. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Bryan, C. A. Rogers, K. Bayliss, J. Wild, and G. D. Angelini Prospective randomized comparison of CarboMedics and St. Jude Medical bileaflet mechanical heart valve prostheses: Ten-year follow-up J. Thorac. Cardiovasc. Surg., March 1, 2007; 133(3): 614 - 622. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Lau, W.R. E. Jamieson, C. Hughes, E. Germann, and F. Chan What Prosthesis Should Be Used at Valve Re-Replacement After Structural Valve Deterioration of a Bioprosthesis? Ann. Thorac. Surg., December 1, 2006; 82(6): 2123 - 2132. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. Lund and M. Bland Risk-corrected impact of mechanical versus bioprosthetic valves on long-term mortality after aortic valve replacement J. Thorac. Cardiovasc. Surg., July 1, 2006; 132(1): 20 - 26. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. R. Moon, M. K. Pasque, N. A. Munfakh, S. J. Melby, J. S. Lawton, N. Moazami, J. E. Codd, T. D. Crabtree, H. B. Barner, and R. J. Damiano Jr Prosthesis-Patient Mismatch After Aortic Valve Replacement: Impact of Age and Body Size on Late Survival Ann. Thorac. Surg., February 1, 2006; 81(2): 481 - 489. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Bernal, J. Lorca, D. Prieto-Salceda, I. Pulitani, A. Ponton, I. Garcia, and J. M. Revuelta Performance at 10 years of the CarboMedics 'Top-Hat' valve. Postclamping time is a predictor of mortality Eur. J. Cardiothorac. Surg., February 1, 2006; 29(2): 144 - 149. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Li, J. G. Dumesnil, P. Mathieu, and P. Pibarot Impact of valve prosthesis-patient mismatch on pulmonary arterial pressure after mitral valve replacement J. Am. Coll. Cardiol., April 5, 2005; 45(7): 1034 - 1040. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. A. Crawford Jr Residual pulmonary artery hypertension after mitral valve replacement: Size matters! J. Am. Coll. Cardiol., April 5, 2005; 45(7): 1041 - 1042. [Full Text] [PDF] |
||||
![]() |
R. W. Emery, C. C. Krogh, K. V. Arom, A. M. Emery, K. Benyo-Albrecht, L. D. Joyce, and D. M. Nicoloff The St. Jude Medical Cardiac Valve Prosthesis: A 25-Year Experience With Single Valve Replacement Ann. Thorac. Surg., March 1, 2005; 79(3): 776 - 782. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||