J Thorac Cardiovasc Surg 2004;127:234-241
© 2004 The American Association for Thoracic Surgery
Surgery for congenital heart disease |
a University of Nebraska/Creighton University, Joint Division of Pediatric Cardiology, Children's Hospital of Omaha, Omaha, Neb, USA
Read at the Eighty-third Annual Meeting of The American Association for Thoracic Surgery, Boston, Mass, May 4-7, 2003.
Received for publication May 2, 2003; revisions received August 14, 2003; accepted for publication August 25, 2003.
* Address for reprints: Zahid Amin, MD, Children's Hospital of Omaha, 8200 Dodge St, 4th Floor Health Care Pavilion, Omaha, NE, USA 68114
zamin{at}chsomaha.org
| Abstract |
|---|
|
|
|---|
METHODS: Recently, the Amplatzer membranous ventricular septal defect device (AGA Medical Corp, Golden Valley, Minn) was introduced. The device has a double-disc design with a short connecting waist. The left ventricular disc has an eccentric design to prevent encroachment on the aortic valve leaflets. Eight Yucatan miniature pigs with naturally occurring perimembranous ventricular septal defects underwent closure of the defect in the operating room by using the perventricular technique. After median sternotomy, a purse-string suture was placed on the free wall of the right ventricle. An angiocatheter was advanced in the right ventricle, and through the catheter, a wire was advanced from the right ventricle through the ventricular septal defect into the left ventricle. A delivery sheath and the dilator were advanced over the wire. The wire and catheter were removed, and an appropriately sized Amplatzer membranous device was advanced through the sheath. The device was deployed under echocardiographic guidance with the heart beating.
RESULTS: The procedure was successful in all animals. There was no incidence of device embolization, heart block, or aortic insufficiency. Angiograms at 3 and 6 months revealed no residual defects and no aortic insufficiency. Pathologically, the devices were completely endothelialized when examined grossly.
CONCLUSIONS: The perventricular technique appears to be excellent for closure of perimembranous ventricular septal defects in the operating room. The technique might be feasible in smaller babies, who are high-risk candidates for closure in the catheterization laboratory. Cardiopulmonary bypass and prolonged hospital stay are avoided.
| Methods |
|---|
|
|
|---|
Animal model
Yucatan miniature pigs with naturally occurring PMVSDs were used in this study. These pigs have naturally occurring PMVSDs that are similar to those of human beings. The defects are located under the anteroseptal commissure of the tricuspid valve and are subaortic in location when viewed from the left ventricular side.
Amplatzer PMVSD Occluder
The Amplatzer PMVSD Occluder device (AGA Medical Corp, Golden Valley, Minn) is made of 0.004-inch Nitinol wire mesh (Figure 1). The wires are shaped to form double discs, with a short connecting waist measuring 1.5 mm. Dacron is sewn inside the discs to decrease shunting through the wire mesh. The device is self-expandable and fully retrievable to the point of release. The size of the device corresponds to the waist of the device. The edge of the left ventricular disc, which is to face the aortic valve, is only 0.5 mm larger than the waist to prevent impingement of the aortic valve. The edge of the disc opposite the aortic valve is 5 mm larger than the waist. The right ventricular disc is 2 mm larger than the waist all around its circumference. A female screw is welded in the center of the right disc for attachment to the delivery cable. The device is designed to be deployed percutaneously from the right ventricular side with a slightly complex delivery system to ensure that the device orientation is maintained (the flat edge toward the aortic valve). For the perventricular approach, however, a simple delivery sheath is used because the device can be easily rotated at the time of deployment. The device is available in sizes ranging from 4 to 18 mm, with 2-mm increments. The delivery sheath size is dependent on the size of the device and ranges from 6F to 9F.
|
The perventricular technique for closure of muscular ventricular septal defects (VSDs) has been described in previous studies.7,8 The procedure remains the same for PMVSDs. The chest was opened through a median sternotomy. The pericardium was cradled. A purse-string suture of 4-0 or 5-0 Prolene sutures (Ethicon, Somerville, NJ) was placed on the free wall of the right ventricle. Epicardial echocardiography was performed to measure the exact size and location of the PMVSD. The integrity of the aortic valve was evaluated. An angiocatheter was passed into the right ventricle through the free wall of the right ventricle, and the needle was removed (Figure 2). A 0.035-inch floppy or glide wire was advanced through the angiocatheter and aimed toward the PMVSD. The PMVSD was crossed, and the wire was advanced through the aortic valve in the ascending aorta (Figure 3). The dilator was removed, and an appropriately sized delivery sheath was advanced over the wire into the ascending aorta (Figure 4). The wire and the dilator were removed, and the sheath was allowed to back bleed to ensure there was no air entrapment. The Amplatzer PMVSD device was immersed in saline and screwed on to the delivery cable. The device was loaded with the help of a loader. The loader was introduced into the delivery sheath, and the device was advanced to the tip of the sheath (Figure 5). The sheath was gently pulled back until the tip was in the left ventricle, just under the aortic valve (Figure 6). Under echocardiographic guidance, the left disc was deployed; the sheath and the device were pulled back until the left disc approximated the ventricular septum (Figure 6). The device was gently rotated only if the flat part of the disc was not directly under the aortic valve. The waist and the right disc were deployed while maintaining fair traction on the delivery cable (Figure 7, A). The position of the device was checked, the aortic valve was evaluated for aortic insufficiency, and the device was released by rotating the delivery cable anticlockwise (Figure 7, B). The sheath and the delivery cable were pulled out of the right ventricle, and the purse-string suture on the right ventricle free wall was tied. Epicardial echocardiography was used to evaluate tricuspid or aortic regurgitation. The pericardium was partially closed, and a chest tube was placed in the mediastinum. The sternum was approximated with the help of sternal wires. The chest wall was closed in routine fashion. The animal was extubated on the operating room table. The chest tube was removed before the animal was transported back to the recovery room.
|
|
|
|
|
|
|
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
The perventricular technique was introduced in 1997, and the first patient who underwent intraoperative device closure of a muscular VSD without CPB was reported by us in 1998.7 In 1999, we reported intraoperative device closure of a PMVSD through a left thoracotomy, but the results were not encouraging because a concentric device was used.8
Cumulatively, more than 19 patients have undergone closure of the muscular VSD with this technique, with good results.11,12 The success of this technique, when compared with that of previous attempts, lies in several factors. The first factor is that CPB is not used. The second factor is that the device placement and residual shunt can be evaluated by means of echocardiography because the heart is beating during the procedure. The third factor is that the delivery sheath is malleable and small in size when compared with that used in the previous experiments. The fourth factor is the design of the device. The Amplatzer device is a self-centering, self-expanding, and fully retrievable device with a circular design. It is devoid of hooks or arms. The length of the connecting waist is small, which helps the device to approximate to the ventricular septum.
The last several years have seen significant advancement in minimally invasive techniques and increased use of devices.12-16 The deleterious effects of CPB in adult patients range from microbubbles, total body inflammatory responses, and postoperative cerebral dysfunction.17 The deleterious effects of CPB in pediatric patients without complex lesions remain unknown; however, transient and permanent neuropsychiatric injury has been reported in as many as 25% of patients.6 There is also risk of silent brain injury, even in patients who underwent isolated secundum atrial septal defect closure, as reported by Visconti and colleagues.18 Therefore avoiding CPB might decrease postoperative morbidity and mortality.
PMVSDs are among the most common congenital cardiac lesions, and hemodynamically symptomatic patients require closure earlier in infancy. These patients are failing to thrive because of congestive heart failure. Catheterization laboratory closure of the PMVSD in these patients is difficult and carries a high risk of complications because of large sheath size relative to patient size. The catheters are difficult to manipulate, which increases cardiac catheterization time, radiation, and risk of arrhythmia. The perventricular approach simplifies PMVSD closure. The potential complications of cardiac catheterization and fluoroscopy are avoided because the procedure is performed under echocardiographic guidance. The procedure is simplified because there are no acute turns of the catheter inside the heart. Because the procedure is performed on the beating heart, the procedure time is significantly reduced; the closure rate and the integrity of the aortic valve are evaluated before the device is released. Postoperatively, these patients are expected to recover faster than patients who have undergone closure during CPB, as has been seen when patients underwent muscular VSD closure with this technique.
In summary, we believe that the perventricular technique is feasible and might open other avenues that can be used to close septal defects by using minimally invasive techniques, such as robotic surgery. Avoidance of CPB will decrease the immediate postoperative and long-term potential complications.
This study has some limitations. We did not evaluate tricuspid valve insufficiency by means of echocardiography at the time of death. The decision was based on poor transthoracic acoustics in pigs. The angiograms would have been unreliable because to assess tricuspid insufficiency, we had to cross the tricuspid valve with a catheter to perform an angiogram. The coaptation of the tricuspid valve leaflet was assuring after the heart was explanted.
Although there was no evidence of aortic valve insufficiency on the basis of angiography and no evidence of valve injury on the basis of gross or microscopic examination, there is potential for injuring the aortic valve with the delivery sheath or catheters. Similarly, we did not see any injury to the tricuspid valve, although the possibility exists. The wires and catheters used in the procedure are the usual catheters used during cardiac catheterizations in neonates and infants. Hence the chance of injuring the cardiac valves or other structures is low.
The risk of long-term problems with the aortic valve remains unknown. Long-term follow-up will be needed in human subjects. We do not believe it to be a problem because the device does not encroach on the aortic valve.
Discussion
Dr Emile A. Bacha (Chicago, Ill). Amin and colleagues performed PMVSD closure in 8 large animals, which were kept alive for 6 months. They were successful in all cases and have presented solid angiographic and histologic evidence that this technique is feasible and safe in the medium term. Dr Amin, I congratulate you for pioneering the field of intraoperative device closure. I also commend you on presenting your data to a surgical audience as opposed to an interventional meeting. After all, surgeons have been involved with VSD closure for more than 30 years.
On the basis of your data and the preliminary US Food and Drug Administration trial data, it appears that device closure of a membranous VSD is safe at medium term. However, the aortic valve is at risk for either immediate injury or long-term dysfunction. Long-term results are awaited in that respect. I disagree with your description of CPB as being deleterious. Most studies showing the deleterious effects of CPB were done either in adults with atherosclerosis or in neonates with complex malformations. Having said that, I believe that perventricular device closure is here to stay and that surgeons should firmly embrace and study this new technology. As you know, we have now performed perventricular muscular VSD closure in 6 patients, and all of those procedures were successful. I have 3 questions for you.
What is your opinion of the long-term effect of the device on aortic valve function given the proximity of the device? In neonates and infants, do you think that the disks protruding in the outflow tract might result in outflow obstruction? I noticed that the lowest weight of the pigs was 16 kg. Typically, the patients we operate on would be much smaller than that.
The third question is, because of the need for precise positioning of the device, with the flat side toward the aortic valve, and thus the need to rotate the entire delivery assembly, do you believe that this technique would be easier to apply through a perventricular approach in the operating room, as you have shown, as opposed to a percutaneous approach from the groin? Thank you very much.
Dr Amin. Thank you, Dr Bacha. As to your first question about the long-term follow-up, as you mentioned, the pigs weighed anywhere from 17 to about 53 kg. Excluding one pig that weighed 53 kg, all remaining pigs weighed less than 30 kg and were less than 3 months old. Therefore they really were small in size and young. During the follow-up period of 6 months, there was tremendous weight gain. One of the pigs that weighed only 16 kg at the time the device was placed weighed more than 100 kg at the time of death. I think that is a significantly long follow-up period in pig years. Therefore they really matured into adulthood, if you will.
But in human terms, I am not sure what the long-term follow-up will hold. If you do look at the devices that were placed in 1989 in muscular VSDsand sometimes those devices were also placed in the perimembranous areaalthough their immediate and acute results were not very good, long-term data are lacking. I totally agree with you that we will find out what the long-term results are in about 5 to 10 years. China has performed almost 300 perimembranous VSD closures with this device over the last 2
years. During follow-up, they had 6 patients who had trace to mild aortic insufficiency. Unfortunately, I cannot tell you exactly what their criteria for trace or mild insufficiency are unless and until I review the echocardiograms.
In regard to your second question about CPB, I simply believe that no matter how simple or complex the patient's condition is, no CPB is preferable.
As to your last question about positioning the device, how the flat edge has to face or should face the aortic valve, I agree with you that it is easier to do in the operating room because we can rotate the sheath very easily clockwise or anticlockwise as opposed to when we are performing the procedure in the cardiac catheterization laboratory because we lose the torque on the sheath. In the operating room it is relatively easier.
Dr Ross M. Ungerleider (Portland, Ore). Zahid, you are to be complimented on something very innovative, not just the use of a device but also the concept of truly collaborative work. The Institute of Medicine has been telling us for years that to improve patient safety, we need to become more team oriented and multidisciplinary in our approach to disease, and for you to get out of the silo of the catheterization laboratory and to begin to collaborate with surgeons in the operating room might open up the doors for several innovations for the treatment of congenital heart disease in babies as we combine our efforts. We have done that with intraoperative stenting and now with these devices. I have 2 questions for you.
Tell us a little bit about the kind of anatomy that these Yucatan pigs have with their perimembranous VSDs. It is my experience that human infants can sometimes have very complex relationships between the tricuspid valve chordal mechanisms and the VSD that might make it somewhat difficult to position the device without creating significant tricuspid insufficiency. Is the same kind of chordal relationship complexity true of the Yucatan pigs?
The second question is, have you considered using this device in an open fashion, although it does require the infants being placed on bypass, to close multiple muscular VSDs, or even to position it precisely in a perimembranous location, which might be easier than sewing a patch in place?
Again, I congratulate you on the true innovation, which is the collaboration with your surgical colleagues, because together I suspect we could innovate terrific things in the future.
Dr Amin. Thank you, Dr Ungerleider. The anatomy of the perimembranous VSDs in pigs is actually very similar to that seen in human subjects. The complexity of the chordal attachment of the septal leaflet of the tricuspid valve in pigs is very much similar to the chordal attachment of the leaflets in human subjects. Our pathology data revealed that some chordae were trapped underneath the right disc, but the tricuspid insufficiency was never more than mild, which I believe is acceptable and similar to the surgical results. The VSD is just under the anteroseptal commissure of the tricuspid valve, just as it is in human subjects.
The only difference that pops in my mind at the current time is that the pigs have a higher propensity for aneurysm formation and hence possibly a higher chance of spontaneous closure. The pig hearts have been examined by a few pathologists in the past, and they found that the pig hearts were similar to human hearts.
In regard to your second question, although we have not looked at closing the defects in an open fashion, I believe it will be very much feasible, especially in patients with muscular VSDs. The only disadvantage I can foresee is that we will not be able to determine whether the patient has a residual defect because the heart will be flaccid.
Dr Marian Zembala (Zabrze, Poland). I would compliment your work for 2 reasons on the basis of our experience, closure of 12 patients with post-VSD by nonoperating field by interventional cardiologists. We lost 12 patients; all of them were at higher risk. That is why we did not decide on surgery. One we lost in the sixth postoperative week because of multiorgan failure and septicemia caused by long-term radiation instead of short-term radiation. When you close an atrial septal defect, the VSD is much longer, and I think this teamwork helped to avoid this long radiation.
Anyway, I think it is a good approach, even for more difficult patients, such as those who have postinfarction VSD.
Dr Amin. Thank you.
Dr Carlos J. Troconis (Caracas, Venezuela). Congratulations for your innovative technique. I have 2 questions. The first one is, we have observed in those patients with a perimembranous VSD that some of them do not have an infundibular ventricular infold rim. And I noticed in your slides that these pigs have a rim of muscle (septum) between the tricuspid valve and the aortic valve. Therefore would the placement of your device sometimes be difficult if there is a lack of such morphology?
The second question is, does the 3-dimensional echocardiography enhance your technique?
Dr Amin. Thank you. I believe that you are asking me about the rim of tissue that we find under the aortic valve. If there is a complete absence of the rim, it is conceivable that the device might impinge on the aortic valve and cause insufficiency or the patient might have a residual defect. The perimembranous VSD device design has a flat edge, with the rim measuring only 0.5 mm more than the disk. In addition, in phase I trials in the United States, all patients with any evidence of aortic insufficiency will be excluded, and the patient's weight will have to be more than 8 kg. Therefore hopefully we can avoid this potential complication. I believe that this device is not going to be suitable for all the patients with perimembranous VSDs.
Concerning your second question about the 3-dimensional echocardiogram, I completely agree with you that it should enhance the visibility and precise placement, and we are working with one company at the current time.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
X. Quansheng, P. Silin, Z. Zhongyun, R. Youbao, L. Shengde, C. Qian, D. Shuhua, H. Kefeng, J. Zhixian, and W. Qin Minimally invasive perventricular device closure of an isolated perimembranous ventricular septal defect with a newly designed delivery system: preliminary experience. J. Thorac. Cardiovasc. Surg., March 1, 2009; 137(3): 556 - 559. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Gan, Q. An, K. Lin, H. Tang, R. C. Lui, K. Tao, W. Pan, and Y. Shi Perventricular Device Closure of Ventricular Septal Defects: Six Months Results in 30 Young Children Ann. Thorac. Surg., July 1, 2008; 86(1): 142 - 146. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. V. Vasilyev, I. Melnychenko, K. Kitahori, F. P. Freudenthal, A. Phillips, R. Kozlik-Feldmann, I. S. Salgo, P. J. del Nido, and E. A. Bacha Beating-heart patch closure of muscular ventricular septal defects under real-time three-dimensional echocardiographic guidance: A preclinical study J. Thorac. Cardiovasc. Surg., March 1, 2008; 135(3): 603 - 609. [Abstract] [Full Text] [PDF] |
||||
![]() |
X. J. Zeng, S. Q. Sun, X. F. Chen, X. J. Ma, Y. H. Luo, Y. P. Lim, and L. Tao Device Closure of Perimembranous Ventricular Septal Defects With a Minimally Invasive Technique in 12 Patients Ann. Thorac. Surg., January 1, 2008; 85(1): 192 - 194. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. S. Minette and D. J. Sahn Ventricular Septal Defects Circulation, November 14, 2006; 114(20): 2190 - 2197. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Amin Off-pump pulmonary valve placement. J. Thorac. Cardiovasc. Surg., November 1, 2006; 132(5): 1245 - 1246. [Full Text] [PDF] |
||||
![]() |
M A Walsh, J Bialkowski, M Szkutnik, M Pawelec-Wojtalik, W Bobkowski, and K P Walsh Atrioventricular block after transcatheter closure of perimembranous ventricular septal defects Heart, September 1, 2006; 92(9): 1295 - 1297. [Abstract] [Full Text] [PDF] |
||||
![]() |
Z. Amin, R. Woo, D. A. Danford, S. E. Froemming, V. M. Reddy, J. Lof, and D. Overman Robotically assisted perventricular closure of perimembranous ventricular septal defects: Preliminary results in Yucatan pigs J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 427 - 432. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||