J Thorac Cardiovasc Surg 2004;127:602-603
© 2004 The American Association for Thoracic Surgery
Brief communication |
a Division of Thoracic and Cardiovascular Surgery, Medical School, Hannover, Germany
Received for publication August 11, 2003; accepted for publication September 10, 2003.
* Address for reprints: Uwe Klima, MD, PhD, Department of Thoracic and Cardiovascular Surgery, Hannover Medical School, 30623 Hannover, Germany
klima{at}thg-mh.hannover.de
|
Several new automatic proximal and distal anastomotic devices are in practical use. Their clinical performance, feasibility, and angiographic follow-up have been described.1,2 Most authors report on a single automatic anastomosis using a saphenous vein graft. Total arterial revascularization using beating heart techniques is suggested by many authors to be the method of choice for revascularization.3 However, to our knowledge, such an operation by means of automatic anastomotic devices has not been reported.
We describe the clinical and angiographic follow-up of a patient who underwent total arterial revascularization on the beating heart in which all anastomoses were performed with an automatic anastomotic magnetic coupling system, the Ventrica Magnetic Vascular Positioner (MVP) Distal Anastomosis System (Ventrica, Inc, Fremont, Calif).
Clinical summary
A 69-year-old man with a 12-month history of angina (preoperative New York Heart Association III) underwent cardiac catheterization revealing 3 vessel disease with significant stenosis (>70%) of all 3 main coronary arteries. Total arterial revascularization was performed using the right and left internal thoracic arteries (ITAs) and the left radial artery (RA). The surgical approach was through a median sternotomy. All target arteries were exposed using the Octopus 4 Tissue Stabilizer and Starfish Heart Positioner (Medtronic, Inc, Minneapolis, Minn). The patient was given an antiplatelet regimen of 150 mg of clopidogrel (Plavix; Bristol-Myers Squibb, New York, NY) and 100 mg of aspirin the day before surgery and received 300 IE/kg of heparin to maintain an activated coagulation time level of more than 350 seconds during the procedure.
All 4 anastomoses were performed with the Ventrica MVP Distal Anastomosis System. The system consists of 6 magnetic clips, with 3 clips forming a set. One magnetic clip set is positioned into the target artery, and the other is positioned into the bypass graft using a preloaded delivery system. These ports then form an anastomosis by magnetic coupling (Figure 1). The first anastomosis created was the left ITA to LAD. The RA was then coupled distally to the circumflex artery, and after shortening the RA, a T-graft anastomosis was also performed to the left ITA with the magnetic coupler. The right coronary artery required an endarterectomy before the fourth automatic anastomosis with the right ITA was performed. The 4 connections were each created in less than 90 seconds.
|
|
The handsewn anastomosis is the "gold standard" for performing coronary artery bypass grafts. However, performing a handsewn anastomosis is more demanding and time-consuming when used in less-invasive approaches such as small access, totally endoscopic, or beating heart surgery.4 In conjunction with attempts to reduce the surgical trauma of coronary artery bypass grafting by using less-invasive approaches, alternative methods for constructing distal anastomoses should be explored and clinically evaluated. It may be hypothesized that by reducing the ischemic time per handsewn anastomosis from 8 to 10 minutes to less than 90 seconds with the automatic anastomosis, the risk of ischemia and its associated complications (eg, arrhythmia) might be decreased significantly.
The concept of magnetic coupling is vastly different from other contemplated micromechanical methods and has several advantages compared with handsewn anastomosis. The device is simple and easy to use, offering the benefit of a quick and repeatable anastomosis. The flexible applicator shaft also helps to facilitate less-invasive procedures. As shown in this case, the device seems to be compatible with ITAs and radial arteries, allowing total arterial revascularization without manipulation of the ascending aorta, as well as with saphenous vein grafts.1
One major advantage of the system seems to be the fact that the order of use (proximal, distal, or T-graft anastomoses first) is irrelevant because the design allows either technique. However, currently only end-to-side anastomoses are possible when the target arteries are not in 1 line (eg, obtuse marginal and diagonal branch). Sequential grafts are only possible when the coronary branches are "in line" (eg, diagonal branch and LAD).
One disadvantage of the device is the lack of long-term follow-up, which is common to all evolving technologies. As of today, predischarge angiographic and clinical follow-up are available1 for using magnetic anastomotic technology in coronary surgery. Currently 6-month postoperative angiographic and clinical follow-up data are also being collected as part of a clinical trial. Long-term follow-up is certainly necessary to make this new anastomotic technique comparable with the gold standard: the handsewn anastomosis. Whether intimal hyperplasia or local long-term reaction to the intravascular implant may occur also needs to be observed, because these factors may contribute to long-term patency rates.
Conclusion
Total arterial revascularization on a beating heart using exclusively automatic anastomotic coupling devices for all anastomoses has not been previously described. The Ventrica MVP System proved to be suitable for all target arteries and grafts including a T-graft anastomosis. The ischemic time during the anastomosis could be reduced significantly, thus decreasing the risk of potential intraoperative ischemia and its associated complications. This technology has the potential to facilitate beating heart procedures by creating a high-quality, reproducible, and rapid anastomosis. The long-term patency of these anastomoses needs to be investigated.
References
This article has been cited by other articles:
![]() |
U. Klima, J. Beilner, E. Bagaev, S. Fischer, T. Kofidis, and J. Lotz MRI-Based Safety Evaluation of the Ventrica Magnetic Coronary Anastomotic System Asian Cardiovasc Thorac Ann, February 1, 2007; 15(1): 24 - 29. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Klima, T. Kofidis, J. Beilner, M. Maringka, S. Kirschner, and A. Haverich Anastomotic Obstruction After Magnetic Vascular Coupling After Minimally Invasive Direct Coronary Artery Bypass Grafting Ann. Thorac. Surg., February 1, 2006; 81(2): 728 - 729. [Abstract] [Full Text] [PDF] |
||||
![]() |
U. Klima, H. MacVaugh III, E. Bagaev, M. Maringka, S. Kirschner, J. Beilner, and A. Haverich Magnetic Vascular Port in Minimally Invasive Direct Coronary Artery Bypass Grafting Circulation, September 14, 2004; 110(11_suppl_1): II-55 - II-60. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |