J Thorac Cardiovasc Surg 2004;127:1602-1607
© 2004 The American Association for Thoracic Surgery
Surgery for acquired cardiovascular disease |
a Department of Clinical Physiology, Linköping University Hospital, Linköping, Sweden
b Department of Cardiothoracic Surgery, Linköping Heart Center, University Hospital Linköping, Linköping, Sweden
c Department of Medicine, University of California San Francisco, San Francisco, Calif, USA
This work was presented in part at the 52nd Annual Scientific Session of the American College of Cardiology, Chicago, Ill, March 30April 2, 2003.
Received for publication July 16, 2003; revisions received September 29, 2003; accepted for publication October 6, 2003.
* Address for reprints: John-Peder Escobar Kvitting, MD, Department of Clinical Physiology, Linköping University Hospital, SE-581 85 Linköping, Sweden
johkv{at}imv.liu.se
| Abstract |
|---|
|
|
|---|
METHODS: By using a time-resolved, 3-dimensional, phase-contrast magnetic resonance imaging technique, aortic root and aortic blood velocity data were acquired from 2 patients with Marfan syndrome 6 months after aortic valvesparing surgery with straight Dacron grafts and contrasted with data from 6 normal volunteers.
RESULTS: In normal aortas vortical blood flow became apparent in the individual sinuses after peak systole. The vortices filled the available space behind the valve leaflets and persisted until diastole, expanding and moving inward during aortic valve closure. In contrast, no vortices were observed in the postoperative patients with Marfan syndrome with negligible sinuses.
CONCLUSIONS: Changes in supravalvular flow accompany loss of sinus architecture. Whether the presence, size, and velocity of supravalvular vortices affects the function or durability of the preserved aortic valve remains to be studied.
An alternative to the composite valve conduit approach is to preserve the native aortic valve. Sarsam and Yacoub6 have presented a method in which they conserve the native aortic valve and its sinuses by connecting the annulus to a tailored graft. Another technique, advocated by David and Feindel,7 consists of a reconstruction in which the valve is mounted within a straight Dacron graft. The main difference between the 2 techniques is how the annulus is treated and whether the sinuses are preserved.
The importance of the sinuses of Valsalva in the function of the aortic valve has been debated since the time of Leonardo da Vinci.8 Modern investigators have theorized that the sinuses and their vortices adapt the valve for closure, thereby facilitating early closure and improving coronary blood flow.9,10 The dynamic and intricate function of the sinuses and of the blood flow within them has not been well studied in human subjects.
The aim of the present study was to investigate the in vivo flow patterns in the normal aortic root and sinuses of Valsalva by using magnetic resonance imaging (MRI) and to contrast them with flow patterns associated with valve-sparing aortic surgery with straight Dacron grafts.
| Material and methods |
|---|
|
|
|---|
We used 3-dimensional (3D) phase-contrast MRI data to visualize the flow pattern in the aortic root and ascending aorta. Velocity vector information was obtained by using a retrospectively gated 3D phase-contrast pulse sequence.11,12 We used a 1.5 T scanner (Signa LX EchoSpeed; GE Medical Systems, Milwaukee, Wis) and the following acquisition parameters: TR of 18 ms; TE of 7 ms; velocity encoding range (VENC) of 100 cm/s for healthy subjects; VENC of 180 cm/s for patients; and field of view (FOV) of 300 (SI) x 300 (AP) x 128 (RL) mm with a spatial resolution of 1 x 4 x 4 mm. Thirty-two time frames were reconstructed. The phase contributions from concomitant gradient (Maxwell) terms and eddy current effects were subtracted.13
The acquired velocity data were transferred to the Ensight visualization program (CEI Inc, Research Triangle Park, NC). Streamlines, on the basis of the velocity information from a single time frame, were generated to visualize the instantaneous 3D flow field in systole. Streamlines were calculated by using a fourth-order Runge-Kutta numeric integration technique.12
Timing during the cardiac cycle was obtained from the flow velocity for each time step at a reference point within the aortic lumen near the aortic valve. Stroke volume was estimated from velocity vectors perpendicular to a circular grid positioned within the aortic lumen a few centimeters above the aortic valve. Integration of this volume for each time step over the entire systolic time interval was taken as the stroke volume. Figure 1 shows a schematic drawing of the blood flow curve in the ascending aorta. S1 is the time from upstroke of the aortic flow curve to peak flow, and S2 is the time from peak flow to the end of ejection.
|
| Results |
|---|
|
|
|---|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Of interest, the apparent size of the vortical flow differed for the 3 sinuses, with relatively larger vortices associated with the right and noncoronary sinuses. This might reflect asymmetry of the aortic root, the swirling quality of the ejected blood, or both.16 Whether the volume or velocity of the supravalvular vortical blood flow affects the specific closing behavior of the individual leaflets has not been studied.
Leyh and colleagues17 demonstrated that composite aortic grafts that incorporate a sinus portion result in a more normal pattern of aortic valve opening and closure. In contrast, they found that the normal gradual inward motion of the aortic leaflets during systole was impaired after aortic root replacement with a straight graft without sinuses. Others have also concluded that incorporation of sinuses in aortic root grafts promotes more normal function of the valve.18,19
We did not observe supravalvular vortices in our 2 patients who did not have sinuses because of their aortic root replacement with a straight graft. Possible explanations for the absence of vortical flow in our patients include a relative lack of space, lack of appropriately coved architecture to promote circular flow, and changes in aortic compliance. These discrete changes in supravalvular flow behavior might explain earlier observations of delayed valve closure.17
It has been proposed that smooth and rapid aortic valve closure might minimize the stress on the valve leaflets.20,21 Perturbation of the normal valve closure mechanisms might contribute to long-term leaflet degeneration.22 In addition, it has been suggested that the normal aortic root, consisting of the aortic valve, corresponding sinus, and the sinotubular junction, provides a "pull and release" function. Loss of this function after root replacement might lead to stress overload on the aortic leaflets and eventual cusp fibrosis and calcification.23 Medium-term results from the Toronto and Hannover experiences with straight aortic root replacement have not demonstrated a trend toward valve dysfunction, however.24,25
Options for aortic root replacement with compliant sinuses are available. Homografts preserve normal geometry and compliance early on. They have been shown to have progressive calcification over time, however, and this is often associated with progressive aortic valve dysfunction.26 A second option is a conduit that incorporates compliant sinuses.19,27 Whether this type of conduit will retain its elasticity in the long term is uncertain.
The temporal resolution of these data is limited to 72 ms. This interferes with the assessment of rapid acceleration and very short-lived events. In the case of our measurements, the exact timing of peak systole and the measurement of the true peak velocity might have been affected by this. Another limitation relates to the optimization of the range of measured velocities. The VENC of the MRI scanner has to be matched to the individual's particular hemodynamics to capture the highest velocities in the aorta. The patients with aortic root replacement were therefore imaged with a higher VENC than the normal subjects. A higher VENC limits low-velocity sensitivity.28 A numerical simulation of the 2 VENC settings was performed to assess whether the higher VENC might interfere with recognition of vortices. In addition, a normal subject was imaged with both the higher and lower VENC settings. In both simulated and in vivo conditions, the higher VENC setting did not impair the recognition of vortical flow.
In conclusion, our findings provide new insight into the function of the aortic sinuses and of the flow within them. There is theoretic support for the concept that prompt and smooth closure of the valve might rely on the presence of normal aortic root flow. In our patients without sinuses undergoing root replacement, vortices were not demonstrable above the aortic valve. This phenomenon might explain some of the changes in aortic valve motion that have been observed in patients undergoing valve-sparing surgery. The effect of the presence, size, and velocity of supravalvular vortices on the function or durability of the aortic valve will require future investigation with methods that can describe their dynamic behavior in 3 dimensions.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
E. Weigang, F. A. Kari, F. Beyersdorf, M. Luehr, C. D. Etz, A. Frydrychowicz, A. Harloff, and M. Markl Flow-sensitive four-dimensional magnetic resonance imaging: flow patterns in ascending aortic aneurysms. Eur. J. Cardiothorac. Surg., July 1, 2008; 34(1): 11 - 16. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David Aortic Valve Repair and Aortic Valve Sparing Operations Card. Surg. Adult, January 1, 2008; 3(2008): 935 - 948. [Full Text] |
||||
![]() |
M. Markl, C. Benk, D. Klausmann, A. F. Stalder, A. Frydrychowicz, J. Hennig, and F. Beyersdorf Three-dimensional magnetic resonance flow analysis in a ventricular assist device. J. Thorac. Cardiovasc. Surg., December 1, 2007; 134(6): 1471 - 1476. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Cheng, P. Dagum, and D. C. Miller Aortic root dynamics and surgery: from craft to science Phil Trans R Soc B, August 29, 2007; 362(1484): 1407 - 1419. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David, C. M. Feindel, G. D. Webb, J. M. Colman, S. Armstrong, and M. Maganti Aortic Valve Preservation in Patients With Aortic Root Aneurysm: Results of the Reimplantation Technique Ann. Thorac. Surg., February 1, 2007; 83(2): S732 - S735. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Miller Valve-Sparing Aortic Root Replacement: Current State of the Art and Where Are We Headed? Ann. Thorac. Surg., February 1, 2007; 83(2): S736 - S739. [Full Text] [PDF] |
||||
![]() |
F. Settepani, W. Y. Szeto, D. Pacini, R. De Paulis, L. Chiariello, R. Di Bartolomeo, R. Gallotti, and J. E. Bavaria Reimplantation Valve-Sparing Aortic Root Replacement in Marfan Syndrome Using the Valsalva Conduit: An Intercontinental Multicenter Study Ann. Thorac. Surg., February 1, 2007; 83(2): S769 - S773. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David, C. M. Feindel, G. D. Webb, J. M. Colman, S. Armstrong, and M. Maganti Long-term results of aortic valve-sparing operations for aortic root aneurysm. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 347 - 354. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Ranga, O. Bouchot, R. Mongrain, P. Ugolini, and R. Cartier Computational simulations of the aortic valve validated by imaging data: evaluation of valve-sparing techniques Interactive CardioVascular and Thoracic Surgery, August 1, 2006; 5(4): 373 - 378. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Frydrychowicz, E. Weigang, M. Langer, and M. Markl Flow-sensitive 3D magnetic resonance imaging reveals complex blood flow alterations in aortic Dacron graft repair Interactive CardioVascular and Thoracic Surgery, August 1, 2006; 5(4): 340 - 342. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. G. Gleason Aortic annuloplasty and valve-sparing root replacement: Details of the primary suture line J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 502 - 503. [Full Text] [PDF] |
||||
![]() |
M. Markl, M. T. Draney, D. C. Miller, J. M. Levin, E. E. Williamson, N. J. Pelc, D. H. Liang, and R. J. Herfkens Time-resolved three-dimensional magnetic resonance velocity mapping of aortic flow in healthy volunteers and patients after valve-sparing aortic root replacement J. Thorac. Cardiovasc. Surg., August 1, 2005; 130(2): 456 - 463. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Maselli, A. Montalto, G. Santise, G. Minardi, C. Manzara, and F. Musumeci A normogram to anticipate dimension of neo-sinuses of valsalva in valve-sparing aortic operations Eur. J. Cardiothorac. Surg., May 1, 2005; 27(5): 831 - 835. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||