J Thorac Cardiovasc Surg 2004;128:273-277
© 2004 The American Association for Thoracic Surgery
Surgery for acquired cardiovascular disease |
a Department of Cardiothoracic Surgery, The Edith Wolfson Medical Center, affiliated with Sackler School of Medicine, Tel Aviv University, Holon, Israel
Received for publication September 4, 2003; revisions received October 3, 2003; accepted for publication November 4, 2003.
* Address for reprints: Benjamin Medalion, MD, Department of Cardiothoracic Surgery, The Edith Wolfson Medical Center, PO Box 5, Holon, Israel 58100
medalion{at}wolfson.health.gov.il
| Abstract |
|---|
|
|
|---|
METHODS: Twelve patients with a diseased aorta who underwent off-pump myocardial revascularization and had their proximal anastomoses performed with the Heartstring device were studied for operative and postoperative outcomes and surgical technique.
RESULTS: The mean age of the patients was 76 ± 7 years. The estimated EuroSCORE perioperative mortality was 12% ± 8%. The median number of distal anastomoses was 3. In all patients, a saphenous vein graft was anastomosed to the aorta using the Heartstring device. In 5 patients, the radial artery was used as a conduit and connected proximally to a left internal thoracic artery as a Y graft, to a saphenous vein graft as a "horseshoe," or on the hood of a saphenous vein graft. The operative and postoperative courses were uneventful. Three seals developed cracks and were not used. One seal developed an unravel in its periphery but was used successfully.
CONCLUSIONS: The Heartstring proximal anastomotic system is a device that allows the surgeon to perform standard proximal clampless anastomoses. Elderly patients with a diseased aorta may benefit from this device.
This article summarizes our initial experience with the Heartstring system and emphasizes some tips and tricks for its use.
| Materials and methods |
|---|
|
|
|---|
Surgical technique
After mid-sternotomy, the left internal thoracic artery (LITA) simultaneously with the left radial artery or the greater saphenous vein graft (SVG) were harvested. A full heparin dose was administered. The ascending aorta was assessed by gentle palpation according to the technique described by Mills and Everson13 or with the aid of transesophageal or epiaortic ultrasound grading the aortic disease on a scale of I to V.14,15 In all patients, the aorta was found to be at least grade IV and judged to be diseased to such an extent that would endanger the likelihood of embolization if a partial clamp were to be used. The distal anastomoses were performed first, with the aid of a suction-assisted stabilizer and an apical suction device. The LITA was anastomosed to the left anterior descending artery in all cases and was performed first. After the distal anastomoses were completed, the proximal anastomoses were performed with the assistance of the Heartstring device.
Proximal anastomosis technique
For the placement of the proximal anastomosis, the aorta was assessed for an anterior circular segment of normal aorta with a diameter of approximately 1 cm. This assessment can be performed by gentle palpation, transesophageal echocardiography, epicardial echocardiography, or a combination of all of these.
The device is generally used according to the manufacturer instructions with some modifications. We found that immersing the seal in a bucket filled with normal saline at a temperature of approximately 50°C for 5 minutes before loading reduces the likelihood of cracks or unravels in the seal. The warm seal enhances and secures the manipulation and loading of the device. A 6-0 Prolene with a half-circle needle (W8815, Ethicon, Somerville, NJ) is our preferred suture. Because the distance between the edge of the seal and the outer part of the aortic hole is approximately 0.5 cm, the half-circle needle is easier to manipulate and reduces the likelihood of suturing the seal. A blower/mister is used to clear the field of oozing blood during the suturing.
Proximal anastomosis options
The technique suggested by the manufacturer is standard end-to-side anastomoses between a conduit and the aorta, using a new Heartstring device for each anastomosis. To perform proximal arterial anastomoses or to use only 1 hole on the aorta, we used 3 different modified techniques: (1) a standard end-to-side anastomosis between a conduit and the aorta, and then an end-to-side anastomosis between a second conduit to the first conduit (T graft); (2) a "horseshoe," side-to-side anastomosis, between a conduit and the aorta (Figure 1, A), and then a beveled end-to-end anastomosis between the proximal part of the anastomosed conduit to another conduit as long as the proximal part of the anastomosed conduit is valveless (Figure 1, B); and (3) a standard end-to-side anastomosis between a conduit and the aorta, but at the end of the anastomosis, the seal is left in place and the last suture is left untied (Figure 2, A). An incision is made on the hood of the anastomosed conduit, and a second conduit is anastomosed to the hood of the first one (Figure 2, B). After the completion of the second anastomosis, the seal is removed and the suture of the first proximal anastomosis is tied.
|
|
Data on the operative and postoperative courses were collected until the patient was discharged from the hospital.
| Results |
|---|
|
|
|---|
|
| Discussion |
|---|
|
|
|---|
The Heartstring proximal anastomotic device uses a different approach. It allows the surgeon to perform the anastomosis using standard techniques in a clampless environment. As a result, each bite can be placed appropriately whether the aorta is normal, thick, or irregular. In addition, the patency rate theoretically should not be different from standard anastomoses performed without the device. The surgeon can decide the proper sequence of anastomoses, proximal or distal first, according to his or her preference.
However, several limitations exist with the Heartstring system as well.
Bleeding
The seal does not give a complete bloodless field. A continuous oozing of blood was present in all cases and seemed to be blood-pressure dependent. Tozzi and colleagues18 used the Heartstring device for performing proximal anastomoses in 3 sheep that underwent OPCAB. They found the leakage to be 3 ± 0.5 mL/min with a blood pressure of 120/65 mm Hg and 9 ± 1.5 mL/min with a blood pressure of 200/95 mm Hg.18 We found it difficult to assess directly the amount of bleeding during the performance of the proximal anastomosis with the Heartstring device because of interference caused by an unknown amount of oozing from the sternum, mediastinal fat, and additional saline introduced to the surgical field by the blower/mister device. We attempted to indirectly estimate blood loss during the procedure by measuring blood-transfused, preoperative, and postoperative hemoglobin in patients who had their proximal anastomosis performed with and without the Heartstring device. Although the cause of intraoperative bleeding depends on many factors (eg, number of distal and proximal anastomoses), it is our impression that a significant amount of blood is lost during the use of the Heartstring device. The addition of a cell saver to the operation seems to be mandatory to reduce the amount of allogenic blood transfused. In addition, systemic blood pressure should be kept as low as possible during the performance of the proximal anastomoses to reduce blood loss and increase visibility. There may be rare cases of increased bleeding that obscures the operative field and that cannot be sufficiently cleared by the blower/mister or of device failure that prevents the anastomosis from being performed safely. In these cases, if the surgeon believes that partially clamping the aorta is too hazardous, the anastomosis can be closed with a pledgeted U stitch, the Heartstring can be removed, and an alternative site for placement of the proximal anastomosis can be considered. In our experience thus far, no such situation has occurred.
Technical failures
Because of the relative inflexibility of the seal in the first 2 cases, cracks appeared in the seal during the process of loading, which prevented us from using it. The process of seal loading seemed to be unpredictable. Subsequently, after immersing the seal in warm water, the loading procedure became easier and more predictable, and no cracks appeared thereafter. Nevertheless, in 1 case, despite warm water immersion, a minor unravel in the periphery of the seal was identified. The seal was used successfully, although the field was bloodier.
Anastomotic size
The aortic hole currently is relatively large (4.5 mm) and not suitable for most arterial grafts. An alternative approach has been suggested in this article that can overcome this limitation. Although theoretically, smaller Heartstring devices can be manufactured to allow the placement of arterial conduits directly on the aorta, the relatively inflexible seal does not seem to easily fit smaller delivery systems unless the seal is decreased significantly in size, a situation that will increase the difficulty of the anastomosis.
Cost
The Heartstring proximal anastomotic device is a relatively expensive device. Its cost is approximately equivalent to the cost of 1 hospitalization day. It certainly increases the cost of the procedure. However, if the use of this device in selected high-risk patients decreases postoperative neurologic complications, which definitely increase the length of hospitalization and complexity of treatment, it will probably prove to be markedly cost-effective.
| Conclusion |
|---|
|
|
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P. Lamm, S. Eifert, E. Kilian, B. Reichart, and G. Juchem Preventing Blood Loss During Application of the HEARTSTRING Proximal Seal System. Ann. Thorac. Surg., July 1, 2009; 88(1): 310 - 312. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Puskas, M. E. Halkos, H. Balkhy, M. Caskey, M. Connolly, J. Crouch, A. Diegeler, J. Gummert, W. Harringer, V. Subramanian, et al. Evaluation of the PAS-Port Proximal Anastomosis System in coronary artery bypass surgery (the EPIC trial) J. Thorac. Cardiovasc. Surg., July 1, 2009; 138(1): 125 - 132. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Shimokawa, S. Manabe, T. Sawada, S. Matsuyama, T. Fukui, and S. Takanashi Intermediate-term patency of saphenous vein graft with a clampless hand-sewn proximal anastomosis device after off-pump coronary bypass grafting. Ann. Thorac. Surg., May 1, 2009; 87(5): 1416 - 1420. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. G. Wolf, Y. Abu-Omar, B. P. Choudhary, D. Pigott, and D. P. Taggart Gaseous and solid cerebral microembolization during proximal aortic anastomoses in off-pump coronary surgery: The effect of an aortic side-biting clamp and two clampless devices J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 485 - 493. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. P. Martens, M. Argenziano, and M. C. Oz New Technology for Surgical Coronary Revascularization Circulation, August 8, 2006; 114(6): 606 - 614. [Full Text] [PDF] |
||||
![]() |