J Thorac Cardiovasc Surg 2004;128:535-542
© 2004 The American Association for Thoracic Surgery
Surgery for acquired cardiovascular disease |
a Division of Cardiothoracic Surgery, Washington University School of Medicine, St Louis, Mo, USA
Read at the Twenty-ninth Annual Meeting of The Western Thoracic Surgical Association, Carlsbad, Calif, June 18-21, 2003.
Received for publication December 12, 2003; revisions received February 12, 2004; accepted for publication February 26, 2004.
* Address for reprints: Ralph J. Damiano, Jr, MD, Suite 3108 Queeny Tower, 1 Barnes-Jewish Hospital Plaza, St Louis, MO 63110, USA
damianor{at}msnotes.wustl.edu
| Abstract |
|---|
|
|
|---|
METHODS: Beginning January 2002, a total of 40 consecutive patients underwent a modified Cox maze III procedure with bipolar radiofrequency energy. Nineteen had a lone maze procedure and 21 had a maze procedure plus a concomitant operation. One month after the operation, the first 8 patients were investigated with high-resolution magnetic resonance imaging. Patients were followed up monthly with clinical examination and electrocardiography.
RESULTS: There was no operative deaths. The crossclamp times were 47 ± 26 minutes for the modified lone Cox maze III procedure and 92 ± 37 minutes for the Cox maze III procedure plus concomitant procedures. These were significantly shorter than our previous times for the traditional Cox maze III procedure (93 ± 34 minutes and 122 ± 37 minutes, respectively, P < .05). Follow-up magnetic resonance imaging showed no evidence of pulmonary vein stenosis, and atrial contractility was preserved in all patients. There were no late strokes. At 6-month follow-up, 91% of patients (21/23) were in sinus rhythm.
CONCLUSIONS: Bipolar radiofrequency ablation can be used to replace the surgical incisions of the Cox maze procedure. This energy source did not result in pulmonary vein stenosis. The modification of the Cox maze III procedure to use bipolar radiofrequency ablation simplified and shortened this procedure without sacrificing short-term efficacy.
|
Atrial fibrillation (AF), the most common sustained cardiac arrhythmia, affects as much as 6% of the population older than 65 years.1 It is estimated that 1.4 million outpatient hospital visits and 227,000 hospitalizations occur annually for the treatment of AF.2 Anticoagulation, rate control, antiarrhythmic drugs, cardioversion, and catheter ablation therapy all play a role in the treatment of this arrhythmia, but all of these strategies have had limited success.
The Cox maze procedure was introduced 15 years ago by James Cox for the surgical treatment of refractory AF. It was hypothesized that this operation would cure AF by abolishing the macroreentrant circuits responsible for its maintenance. This result was achieved by creating conduction block with transmural surgical incisions. Since its introduction, the operation has undergone two modifications. The Cox maze III procedure (CMP) is the most recent version and has had an impressive success rate of more than 95% at 10 years.3 The late stroke rate in our series has been less than 1%, with only 1 reported event among more than 200 consecutive patients. Normally, such success would have resulted in rapid adoption of a surgical procedure. However, implementation has been slowed by its perceived invasiveness and difficulty. The traditional "cut and sew" technique requires cardiopulmonary bypass and can be time-consuming. This has resulted in a hesitation by many surgeons to combine this procedure with valve repair or replacement. Furthermore, significant morbidity has been reported, including pacemaker requirement and left atrial dysfunction.4,5
Simplification of the CMP by eliminating the traditional cut and sew approach has been proposed as a way to increase the adoption of this curative operation. Surgical investigators have used a variety of different energy sources to create lines of ablation to replace several, and in some instances all, of the myriad incisions of the CMP. Radiofrequency (RF) energy has frequently been used in the ablation of cardiac tissue.6-8 It is capable of producing myocardial necrosis when applied to either the endocardium or epicardium. Unipolar RF energy, however, has been unable to consistently produce transmural lesions. Because the energy is unfocused, it also has the potential for unwanted collateral damage.9-12 To overcome these shortcomings, we have investigated a unique bipolar RF device. The RF energy is delivered between two closely approximated 1-mm electrodes embedded in the jaws of the clamp. By measuring the drop in conductance, transmurality of ablation can be measured on-line and used to control the time of energy delivery. Moreover, the lesions are thin and discrete, eliminating the possibility of collateral tissue injury.13-15 After extensive experimental evaluation, a prospective clinical trial was undertaken to examine the effectiveness of the bipolar RF device to replace some of the cut and sew lesions in the CMP. The objectives of this modification of the CMP were to simplify the procedure and reduce operating time while retaining the efficacy and safety of the procedure.
| Patients and methods |
|---|
|
|
|---|
|
|
Technique of the modified maze procedure
After induction of anesthesia, the patients were placed on cardiopulmonary bypass with bicaval cannulation. Initially, temperature was maintained at 36°C to encourage sinus rhythm and allow accurate determination of pacing thresholds. The left and right pulmonary veins were dissected and surrounded with umbilical tape. Patients in AF were electrically cardioverted. Pacing thresholds were recorded from the right and the left pulmonary veins. The bipolar RF device was then placed around first the right and then the left pulmonary vein (Figure 3).
The device was clamped twice to ensure complete encirclement of both the superior and the inferior veins. Occasionally, in patients with large pulmonary veins, it was necessary to clamp each vein separately. Electrical isolation was documented by pacing at a stimulus strength of 20 mA. Further ablations were performed if necessary until there was documented conduction block.
|
|
|
One month after discharge from the hospital, the first 8 patients underwent magnetic resonance imaging (MRI) to assess for pulmonary vein stenosis and atrial function. Follow-up electrocardiograms were obtained at 1, 3, and 6 months for all patients. All patients reporting palpitations or with documented atrial arrhythmias by electrocardiography underwent 24-hour Holter monitoring.
MRI imaging protocol
Subjects were placed supine in a clinical 1.5-T MRI scanner (NT Inter CV; Philips Medical Systems, Best, The Netherlands) The imaging protocol comprised multiple anatomic and functional cine images and multiple velocity-encoded images. The cine images, prospectively triggered to each R-wave peak, were based on a steady-state gradient echocardiographic technique, balance fast field echo. Vertical long-axis cine images were obtained by aligning the left ventricular apex with the center of the mitral valve. From this vertical long-axis image, the horizontal long-axis plane was aligned through the midpoint of the left ventricular apex and the mitral valve. A diastolic image at end-expiration provided the reference image from which a stack of contiguous short-axis slices was positioned. Slice thickness was 9 mm, and it took 8 to 10 slices to completely traverse the entire left atrium. The typical field of view was 288 x 360 mm with matrix size 154 x 256, giving a typical reconstructed pixel size of approximately 1.5 x 1.2 mm.
Quantitative flow study
Transmitral and transtricuspid blood flows were assessed with velocity-encoded MRI.16 Imaging was centered on the mitral valve plane, with the velocity-encoded imaging slice being 7 mm thick. Images were acquired at 15 to 20 phases during the cardiac cycle, depending on the heart rate. Retrospective gating was used to acquire data for 20% longer than the expected R-R interval.
Data analysis
Analysis of the acquired images was performed off-line with a commercial software package (EasyVision release 5, Philips Medical System). The volume of the left atrium was assessed by planimetry.
Analysis of quantitative flow study was performed by examining the velocity-encoded images at all time points. The peak velocity data from within these regions of interest were used to construct a velocity versus time curve, from which E and A waves could be identified. The data was then exported to a spreadsheet format for further data analysis. MRI images were used to assess anatomic and physiologic flow patterns in the pulmonary venous system. Basic cine images were used to define pulmonary venous anatomy in the short axis view. Velocity-encoded MRI images were used to quantify both direction and speed of pulmonary venous flow.
| Results |
|---|
|
|
|---|
Pulmonary vein isolation
Pacing thresholds for the right and left pulmonary veins were 0.9 ± 1.0 mA and 1.1 ± 1.6 mA, respectively. The left and right pulmonary veins were isolated in every instance, as documented by pacing at 20 mA. The right pulmonary veins underwent an average of 3.0 ± 1.2 RF applications, whereas the left pulmonary veins underwent 3.1 ± 1.3 RF applications. The mean total ablation times were 28 ± 13 seconds for the right pulmonary veins and 28 ± 12 seconds for the left pulmonary veins.
Perioperative results
The mean RF ablation time was 9.5 ± 3.8 seconds for each application. The mean crossclamp time was 54 ± 27 minutes for the lone modified CMP. In comparison, the crossclamp time for lone traditional CMP was 93 ± 34 minutes (P < .001). For the concomitant procedure, the average crossclamp time was 99 ± 30 minutes. In the traditional concomitant procedure group, the crossclamp time was 123 ± 36 minutes (P = .005).
The median intensive care unit stay was 2 days. The median hospital stay was 12 days.
Postoperative complications
There were no operative deaths. Five patients had reexploration for mediastinal bleeding. There were no strokes or sternal infections. Postoperative AF was present in 25 patients (63%), but this was usually transient. Thirty-four patients (85%) were discharged from the hospital in normal sinus rhythm. Six patients (15%) required postoperative permanent pacemaker placement for sick sinus syndrome. There were no complications attributable to the bipolar device in terms of collateral tissue injury.
MRI results
MRI studies were conducted on the first 8 patients 4 to 6 weeks after operation. There was no evidence of pulmonary vein stenosis or thrombosis in any patient (Figure 6).
Atrial contractility, as documented by the presence of an A wave on the transmitral flow trace, was preserved in all patients.
|
| Discussion |
|---|
|
|
|---|
Despite these excellent results, the CMP had shortcomings. Although mortality rates remained low (1.5% in our series), there was significant morbidity. The operation required a relatively long period of cardiopulmonary bypass and an arrested heart. At our institution, the average crossclamp time to perform a lone maze procedure was 93 ± 34 minutes. After CMP, 15% of the patients required postoperative pacemakers. The median stay was 9 days after a lone CMP and 12 days after a CMP concomitant with other surgery. Finally, left atrial dysfunction has been noted in many patients.4,20,21
These problems with the traditional CMP have led our group and others to begin evaluating strategies to simplify the surgical treatment of AF.22,23 A common strategy has been to replace the surgical incisions with linear lines of ablation made by mans of various energy sources, including RF, cryoablation, microwave, ultrasound, and laser energy.24,25 Elimination of the extensive cutting and sewing needed to make the atrial incisions of the CMP gives these new operations the potential to simplify the procedure and decrease the time of surgery. The hope is that this will reduce morbidity, encourage referrals, and speed the adoption of this operation by surgeons.
Many groups also have chosen to omit many of the incisions of the Cox maze procedure. This has made interpretation clinical results difficult, because failure could be due either to inadequate ablation technology or to an inadequate lesion set. It was our goal to perform all the basic lesions of the CMP but to replace many of the incisions with lines of ablation.
Our group has concentrated on a device that uses bipolar RF ablation. By measuring the fall in tissue conductance, this device was shown to have a definitive and verifiable end point of transmurality.13-15 Mean ablation times were always less than 10 seconds, even in thick tissue, and the lesions were only 1 to 3 mm in width. This was an improvement relative to the width of lesions and time required for ablation with unipolar devices. This report represents our first clinical series with this device and details our operative technique and early clinical results. The safety of this technology for pulmonary vein isolation was evaluated by MRI scans at 1 month.
Our modification of the CMP replaced more than half of the surgical incisions with bipolar ablation lines. This simplified procedure has a single long left atriotomy and only two right atrial incisions. Isolation of the pulmonary veins can be performed on the beating heart before crossclamping. The use of this device decreased mean crossclamp time in both lone CMP and concomitant CMP groups by more than 30 minutes. However, it should be noted that this comparison is with a historical cohort performed by different surgeons, making direct comparison problematic.
This procedure may have other advantages relative to the cut and sew procedure in addition to its technical simplicity. With separate isolation of the right and left pulmonary veins, a significant amount of functioning posterior atrial myocardium remains, theoretically improving atrial contractility. In the traditional CMP, the right and left pulmonary veins are isolated en bloc, removing virtually all function of the entire posterior left atrium. Although it is hoped that such advantages will translate into decreased morbidity, a much larger cohort of patients will be needed to prove this hypothesis. There were no operative deaths in this series, and at 6 months 91% of patients were free from AF, with only 10 patients requiring antiarrhythmic therapy.
This device was particularly effective in isolating the pulmonary veins, a key part of any procedure for AF. The device isolated the right and left pulmonary veins in every instance. Unfortunately, the device usually required several applications to isolate the atrial cuff around the pulmonary veins. We believe that this was due to the large amount of atrium clamped during this maneuver, which can result in poor clamp apposition. Experience with the device has supported not bunching the atria into the clamp to perform a single ablation, but rather coming across the atrial cuff from both above and below in cases where there appears to be excess tissue. Although several ablations were often required, total combined ablation time for both the right and left pulmonary veins was under 60 seconds. Isolation was documented in every instance by pacing. It is our belief that surgeons who use this technology should routinely use pacing to confirm isolation.
The question of late pulmonary vein stenosis remains an important issue. In the electrophysiologic laboratory, there has been a defined incidence of late pulmonary vein stenosis.26,27 Few surgical groups have rigorously evaluated patients for this problem after RF ablation. In this study, the first 8 patients were evaluated with high-resolution MRI scans at 1 month. At that interval there was no evidence of pulmonary vein stenosis, normal pulmonary vein flow patterns were maintained, and left atrial function was preserved.
Preliminary reports from a number of electrophysiologic laboratories have suggested that pulmonary vein isolation alone may be curative for selected patients with paroxysmal AF.28 Although it still remains unclear whether this simple procedure will be effective in surgical cases, particularly those of patients with valvular heart disease, this technology does allow pulmonary vein isolation to be performed quickly, without the need for any incisions or cardiopulmonary bypass.
The use of bipolar RF ablation represents a step toward a less invasive maze procedure. Results in our research laboratory have shown that this technology can be effectively used on the beating heart.14 If the excellent early results of the modified Cox maze procedure stand the test of time, the indications for this operation should be expanded. In our own practice, we believe that the operation is a viable alternative to a lifetime of anticoagulation for young patients with solitary AF. Moreover, this simplified operation should be offered to all patients undergoing valvular surgery with concomitant AF.
This study has limitations. It represents a relatively small number of patients with short follow-up. Pulmonary vein stenosis was evaluated at 1 month, which may be too short an interval to pick up this complication. Careful long-term observation will be needed to demonstrate the late efficacy of this new modification of the CMP. The strengths of this study are that it is a prospective study of a consecutive group of patients with 100% follow-up. It is the only study to date to examine postoperative pulmonary vein stenosis after bipolar RF ablation. In summary, this study suggests that bipolar RF ablation is safe and can effectively replace the surgical incisions of the CMP.
Discussion
Dr Roger C. Millar (Salt Lake City, Utah). Dr Damiano and I have a similar interest in the AF problem that dates back really to Dr Cox, who was a predecessor of both of us at Duke University.
Nine months ago we became interested in this same energy device, the bipolar RF, and have performed 35 cases, all concomitant procedures. Eighty-six percent had mitral valve surgery. In that group we had 3 triple valves and 16 double valves. You have averaged 3.2 and 3.3 applications on the pulmonary veins on the right or the left to prove transmurality by atrial pacing, or technically really pulmonary vein pacing. Why do you think this is, and why do you think you required three applications to obtain complete isolation of the pulmonary veins?
Dr Damiano. That's an absolutely excellent question. Part of it has been the learning curve with the technology. First of all, we basically are trying to replicate our animal work, where we would try in just one clamp to try to bunch all the pulmonary veins into what is only a 5-cm long clamp and take a big cuff of left atrium, because we were trying to make sure we didn't see any pulmonary vein stenosis. What we have learned and what I think is happening is that if you bunch what is actually quite a bit of atrial tissue into this clamp that if you get any type of folding over of the tissue and any air in there, because air is an incredibly great insulator, you will get a false high reading of tissue resistance, with the inverse of that being a low reading of the conductance. That will, I think in human subjects give you a false-positive indication. It is actually a real indication that conductance has fallen to a stable minimum, but that is probably due to having some air or bunched up tissue in the clamp. What we have gone to in the last group of patients is that if the actual width of those pulmonary veins is more than 2 thirds of the clamp width, in which the electrode length is 5 cm, we will then clamp the right superior and inferior and the left superior and inferior veins separately, or we will take clamps that go from both below and above. So by definition, you would have to have two clamps. By doing that we virtually always have conduction block after just two applications, but certainly in the beginning we had as many as five applications before we had complete conduction block. I would say that I certainly believe because of that that it is absolutely mandatory if you are using this technology to pace or record pulmonary vein electrograms to document that you have pulmonary vein isolation, because one would certainly consider that among the most important parts of any surgical treatment of AF.
Dr Millar. That seems to imply, then, that we need to be doing this on the beating heart, at least so we can test transmurality before proceeding with the other connecting lesions in the left atrium.
When you are doing your linear lesions to connect the pulmonary veins across the base of the left atrium, or even on the right side the superior to inferior vena cava longitudinal lesion that we construct with the bipolar device, do you do more than one burn in those areas because you cannot prove transmurality by pacing across those lines?
Dr Damiano. Usually not. I usually just do one burn. You can directly observe the ablation, and what we have learned in a couple of years of animal experimentation with this is that if you have uniformly thick tissue and encompass it without any bunching in this device, it is virtually 100% reliable in terms of creating a transmural lesion. If for any reason I am going over any type of tissue that appears to be a bit nonhomogeneous in terms of its thickness, to do a second ablation line doesn't take long. If I have any question in my mind, a lot of times I will put two ablation lines right there and still be done in 10 or 15 seconds. Generally for the other lines, we will perform simply one; especially along the right atrial free wall and in the posterior left atrium, the tissue is almost always very uniform in thickness, and usually just one ablation line is used in those areas.
Dr Millar. Related to this same issue, when I have placed a second burn, I have usually released the clamp and then reapplied it. It still takes almost the same amount of time. If this is full thickness to start with, why does it take several seconds again? Why, if it was transmural the first time, does it take time to signal transmurality?
Dr Damiano. The only explanation I can think of is if there is any moisture that would slip in after your ablation. Basically, the low conductance is due to a complete desiccation of the tissue. If you stop, perhaps there is some leeching in of interstitial fluid which you then have to boil off again. I probably would take that up with some of the engineers at the company. That is an interesting observation. I have never really tried that myself.
Dr Millar. Try it some time. I really can't explain it either, but that's my best explanation as well. Dr Gillinov presented a series of 190 patients undergoing left-sided procedures only and showed a 6-month 80% success rate. I noticed that in your series more than half of your patients had paroxysmal AF. Would pulmonary vein isolation and left atrial appendage removal also have given you an 80% success rate? What does that imply for the possible future application of such a device to do pulmonary vein isolation and left atrial appendage removal in less than 10 minutes? That could easily be added to any kind of left-sided operation you are doing. Would you speculate on the future?
Dr Damiano. I will speculate on the role of pulmonary vein isolation, both today and in the future. Certainly, there are provocative electrophysiologic data indicating that pulmonary vein isolation may be effective treatment for paroxysmal AF. I would caution all surgeons that the data from Dr Haissaguerre's laboratory in Bordeaux, from Dr Chen's laboratory in Taiwan, from the group in Michigan, and from Dr Morady that have justified that rapid conclusions are virtually all in patients without structural heart disease and very small left atria. So the role of pulmonary vein isolation and its possible success will probably be best determined by our surgical results and not necessarily by what has been reported in the electrophysiology literature. I think it is still a hypothesis and that we need to design good clinical studies to answer this. Clearly, with this device total ablation time to isolate both sides of the pulmonary vein was less than a minute. It can dissect, and the dissection of both the left and right sides can be performed in a couple of minutes each once you are used to doing it. We do perform this on the beating heart before crossclamping, as you have suggested, so that we can pace, and that would probably only add 5 minutes to the operation. I think that you have to amputate the left atrial appendage, although what our success rate will be in that group is unclear. Dr Gillinov reported a success rate clearly much lower than our success rate with the maze procedure. Their latest reported success is about 75%. They do have a drawback in that most of their patients really have not been able to be followed up, because they usually come from a long way off. Thus I don't think we know their success rate clearly, and I think we have to look at. Even in this study, I would tell you it is too soon. I am not sure what our success rate is going to be. It is promising that we are at 95% at 6 months, but I don't know what we will be long term. We have done 10 cases with just pulmonary vein isolation. This was a highly selected group. Our success rate at 6 months has only been 50%, although among patients with paroxysmal AF it was 66%, so I think there is some indication that there are some patients for whom it is going to work. I think we need to get a little bit smarter, and I think we are going to have to wait for our experience to determine who these patients are. I don't think I would jump to the conclusion right now that all patients with paroxysmal AF who are referred for surgical treatment of AF should only have pulmonary vein isolation. I think that is a question remaining to be answered, and I do feel strongly that for patients who are good candidates for a CMP or a modified procedure like the one I've shown today, that's what you should do. It historically has a 10-year efficacy rate greater than 90%. Now that it has been modified with bipolar ablation and some other technologies, it really does not add significantly to the procedure nor does it require a lot of lines. I do think that there certainly is going to be a group of patients who can be successfully treated with pulmonary vein isolation. I just heard a report from Dr Nitta's group in Tokyo, where they have done intraoperative mapping; they found the patients with pulmonary vein fociinterestingly not all patients with paroxysmal AF, but about 60% to 70% of those patientsunderwent pulmonary vein isolation with excellent results. So that may be where we are going in the future. Certainly I think that in those cases it will play a large role and probably can be done in a minimally invasive and rapid manner by surgeons.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
L. D. Sauren, M. la Meir, L. de Roy, L. Pison, F. H. van der Veen, W. H. Mess, H. J. Crijns, and J. G. Maessen Increased number of cerebral emboli during percutaneous endocardial pulmonary vein isolation versus a thoracoscopic epicardial approach Eur. J. Cardiothorac. Surg., November 1, 2009; 36(5): 833 - 837. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi, A. Galanti, S. Nascimbene, A. Fumero, E. Dorigo, V. Zerbi, and O. Alfieri Complete right atrial ablation with bipolar radiofrequency. Ann. Thorac. Surg., May 1, 2009; 87(5): 1573 - 1576. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Voeller, A. Zierer, S. C. Lall, S.-i. Sakamoto, N.-L. Chang, R. B. Schuessler, M. R. Moon, and R. J. Damiano Jr. The effects of the Cox maze procedure on atrial function. J. Thorac. Cardiovasc. Surg., November 1, 2008; 136(5): 1257 - 1264.e3. [Abstract] [Full Text] [PDF] |
||||
![]() |
S.-i. Sakamoto, R. K. Voeller, S. J. Melby, S. C. Lall, N.-l. Chang, R. B. Schuessler, and R. J. Damiano Jr. Surgical ablation for atrial fibrillation: The efficacy of a novel bipolar pen device in the cardioplegically arrested and beating heart. J. Thorac. Cardiovasc. Surg., November 1, 2008; 136(5): 1295 - 1301. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Fukunaga, H. Hori, T. Ueda, K. Takagi, E. Tayama, and S. Aoyagi Effect of Surgery for Atrial Fibrillation Associated With Mitral Valve Disease Ann. Thorac. Surg., October 1, 2008; 86(4): 1212 - 1217. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Groh, O. A. Binns, H. G. Burton III, G. L. Champsaur, S. W. Ely, and A. M. Johnson Epicardial Ultrasonic Ablation of Atrial Fibrillation During Concomitant Cardiac Surgery Is a Valid Option in Patients With Ischemic Heart Disease Circulation, September 30, 2008; 118(14_suppl_1): S78 - S82. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Voeller, M. S. Bailey, A. Zierer, S. C. Lall, S.-i. Sakamoto, K. Aubuchon, J. S. Lawton, N. Moazami, C. B. Huddleston, N. A. Munfakh, et al. Isolating the entire posterior left atrium improves surgical outcomes after the Cox maze procedure. J. Thorac. Cardiovasc. Surg., April 1, 2008; 135(4): 870 - 877. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. C. Lall, K. V. Foyil, S.-I. Sakamoto, R. K. Voeller, J. P. Boineau, R. J. Damiano Jr., and R. B. Schuessler Pulmonary vein isolation and the Cox maze procedure only partially denervate the atrium. J. Thorac. Cardiovasc. Surg., April 1, 2008; 135(4): 894 - 900. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. S.C. Geuzebroek, P. K.E.W. Ballaux, N. M. van Hemel, J. C. Kelder, and J. J.A.M.T. Defauw Medium-term outcome of different surgical methods to cure atrial fibrillation: is less worse? Interactive CardioVascular and Thoracic Surgery, April 1, 2008; 7(2): 201 - 206. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Voeller, R. B. Schuessler, and R. J. Damiano Jr. Surgical Treatment of Atrial Fibrillation Card. Surg. Adult, January 1, 2008; 3(2008): 1375 - 1394. [Full Text] |
||||
![]() |
M. A. Groh, O. A. Binns, H. G. Burton III, S. W. Ely, and A. M. Johnson Ultrasonic Cardiac Ablation for Atrial Fibrillation During Concomitant Cardiac Surgery: Long-Term Clinical Outcomes Ann. Thorac. Surg., December 1, 2007; 84(6): 1978 - 1983. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov and L. G. Svensson Ablation of Atrial Fibrillation With Minimally Invasive Mitral Surgery Ann. Thorac. Surg., September 1, 2007; 84(3): 1041 - 1042. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. J. Damiano Jr and M. Bailey The Cox-Maze IV procedure for lone atrial fibrillation MMCTS, July 23, 2007; 2007(0723): 2758. [Abstract] [Full Text] [PDF] |
||||
![]() |
A.M. Patwardhan Intraoperative ablation of atrial fibrillation using bipolar output of surgical radiofrequency generator (diathermy) and reusable bipolar forceps J. Thorac. Cardiovasc. Surg., June 1, 2007; 133(6): 1683 - 1683. [Full Text] [PDF] |
||||
![]() |
H. Calkins, J. Brugada, D. L. Packer, R. Cappato, S.-A. Chen, H. J.G. Crijns, R. J. Damiano Jr, D. W. Davies, D. E. Haines, M. Haissaguerre, et al. HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-Up: A report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation Developed in partnership with the European Heart Rhythm Association (EHRA) and the European Cardiac Arrhythmia Society (ECAS); in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Endorsed and Approved by the governing bodies of the American College of Cardiology, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, and the Heart Rhythm Society. Europace, June 1, 2007; 9(6): 335 - 379. [Full Text] [PDF] |
||||
![]() |
J. M. Stulak, J. A. Dearani, T. M. Sundt III, R. C. Daly, C. G.A. McGregor, K. J. Zehr, and H. V. Schaff Superiority of cut-and-sew technique for the Cox maze procedure: Comparison with radiofrequency ablation J. Thorac. Cardiovasc. Surg., April 1, 2007; 133(4): 1022 - 1027. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. C. Lall, S. J. Melby, R. K. Voeller, A. Zierer, M. S. Bailey, T. J. Guthrie, M. R. Moon, N. Moazami, J. S. Lawton, and R. J. Damiano Jr The effect of ablation technology on surgical outcomes after the Cox-maze procedure: A propensity analysis J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 389 - 396. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Martin-Suarez, B. Claysset, L. Botta, M. Ferlito, D. Pacini, C. Savini, G. Marinelli, and R. DiBartolomeo Surgery for atrial fibrillation with radiofrequency ablation: four years experience Interactive CardioVascular and Thoracic Surgery, February 1, 2007; 6(1): 71 - 76. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Melby, S. L. Gaynor, J. G. Lubahn, A. M. Lee, P. Rahgozar, S. D. Caruthers, T. A. Williams, R. B. Schuessler, and R. J. Damiano Jr Efficacy and safety of right and left atrial ablations on the beating heart with irrigated bipolar radiofrequency energy: A long-term animal study J. Thorac. Cardiovasc. Surg., October 1, 2006; 132(4): 853 - 860. [Abstract] [Full Text] [PDF] |
||||
![]() |
Writing Committee Members, V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, et al. ACC/AHA/ESC 2006 guidelines for the management of patients with atrial fibrillation: full text: A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation) Developed in collaboration with the European Heart Rhythm Association and the Heart Rhythm Society Europace, September 1, 2006; 8(9): 651 - 745. [Full Text] [PDF] |
||||
![]() |
V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, J. E. Lowe, et al. ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation) Developed in Collaboration With the European Heart Rhythm Association and the Heart Rhythm Society J. Am. Coll. Cardiol., August 15, 2006; 48(4): e149 - e246. [Full Text] [PDF] |
||||
![]() |
V. Fuster, L. E. Ryden, D. S. Cannom, H. J. Crijns, A. B. Curtis, K. A. Ellenbogen, J. L. Halperin, J.-Y. Le Heuzey, G. N. Kay, J. E. Lowe, et al. ACC/AHA/ESC 2006 Guidelines for the Management of Patients With Atrial Fibrillation: A Report of the American College of Cardiology/American Heart Association Task Force on Practice Guidelines and the European Society of Cardiology Committee for Practice Guidelines (Writing Committee to Revise the 2001 Guidelines for the Management of Patients With Atrial Fibrillation): Developed in Collaboration With the European Heart Rhythm Association and the Heart Rhythm Society Circulation, August 15, 2006; 114(7): e257 - e354. [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov, S. Bhavani, E. H. Blackstone, J. Rajeswaran, L. G. Svensson, J. L. Navia, B.G. Pettersson, J. F. Sabik III, N. G. Smedira, T. Mihaljevic, et al. Surgery for Permanent Atrial Fibrillation: Impact of Patient Factors and Lesion Set Ann. Thorac. Surg., August 1, 2006; 82(2): 502 - 514. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Jahangiri, G. Weir, K. Mandal, I. Savelieva, and J. Camm Current strategies in the management of atrial fibrillation. Ann. Thorac. Surg., July 1, 2006; 82(1): 357 - 364. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Akpinar, I. Sanisoglu, M. Guden, E. Sagbas, B. Caynak, and Z. Bayramoglu Combined Off-Pump Coronary Artery Bypass Grafting Surgery and Ablative Therapy for Atrial Fibrillation: Early and Mid-Term Results Ann. Thorac. Surg., April 1, 2006; 81(4): 1332 - 1337. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. di Virgilio, F. Onorati, P. Mastroroberto, and A. Renzulli Beating-heart ablation of the cavo-tricuspid isthmus Interactive CardioVascular and Thoracic Surgery, April 1, 2006; 5(2): 85 - 87. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi and O. Alfieri Concomitant ablation of atrial fibrillation during mitral surgery MMCTS, November 29, 2005; 2005(1129): 1081. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Sanchez-Quintana, J. A. Cabrera, V. Climent, J. Farre, M. C. de Mendonca, and S. Y. Ho Anatomic Relations Between the Esophagus and Left Atrium and Relevance for Ablation of Atrial Fibrillation Circulation, September 6, 2005; 112(10): 1400 - 1405. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Gammie, J. C. Laschinger, J. M. Brown, R. S. Poston, R. N. Pierson III, L. G. Romar, K. L. Schwartz, M. J. Santos, and B. P. Griffith A Multi-Institutional Experience With the CryoMaze Procedure Ann. Thorac. Surg., September 1, 2005; 80(3): 876 - 880. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Ninet, X. Roques, R. Seitelberger, C. Deville, J. L. Pomar, J. Robin, O. Jegaden, F. Wellens, E. Wolner, C. Vedrinne, et al. Surgical ablation of atrial fibrillation with off-pump, epicardial, high-intensity focused ultrasound: Results of a multicenter trial J. Thorac. Cardiovasc. Surg., September 1, 2005; 130(3): 803 - 803. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi, S. Nascimbene, G. Calori, P. Denti, Z. Ziskind, S. Kassem, G. La Canna, C. Pappone, and O. Alfieri Surgical ablation of atrial fibrillation with a novel bipolar radiofrequency device J. Thorac. Cardiovasc. Surg., August 1, 2005; 130(2): 491 - 497. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Benussi Editorial comment Eur. J. Cardiothorac. Surg., July 1, 2005; 28(1): 80 - 82. [Full Text] [PDF] |
||||
![]() |
P. Zimetbaum An Argument for Maintenance of Sinus Rhythm in Patients With Atrial Fibrillation Circulation, June 14, 2005; 111(23): 3150 - 3156. [Full Text] [PDF] |
||||
![]() |
R. H. Jones The Year in Cardiovascular Surgery J. Am. Coll. Cardiol., May 3, 2005; 45(9): 1517 - 1528. [Full Text] [PDF] |
||||
![]() |
E. Lim and S. Large What is the true (unbiased) percentage freedom from atrial fibrillation at 6 months after the modified Cox maze procedure using bipolar radiofrequency energy? J. Thorac. Cardiovasc. Surg., February 1, 2005; 129(2): 477 - 477. [Full Text] [PDF] |
||||
![]() |
S. L. Gaynor, R. B. Schuessler, M. S. Bailey, Y. Ishii, J. P. Boineau, M. J. Gleva, J. L. Cox, and R. J. Damiano Jr Surgical treatment of atrial fibrillation: Predictors of late recurrence J. Thorac. Cardiovasc. Surg., January 1, 2005; 129(1): 104 - 111. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||