|
|
||||||||
J Thorac Cardiovasc Surg 2005;129:1056-1063
© 2005 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
German Heart Center Munich, Department of Cardiovascular Surgery, Munich, Germany.
Received for publication May 21, 2004; revisions received July 30, 2004; accepted for publication August 18, 2004. * Address for reprints: Walter B. Eichinger, MD, Department of Cardiovascular Surgery, German Heart Center Munich, Lazarettstr 36, 80636 Munich, Germany. (E-mail: eichinger{at}dhm.mhn.de).
| Abstract |
|---|
|
|
|---|
METHODS: Between August of 2000 and December of 2002, 136 patients underwent aortic valve replacement with the porcine Medtronic Mosaic (n = 66) or the bovine Carpentier-Edwards Perimount (n = 70) bioprosthesis. Transthoracic echocardiography was performed to assess hemodynamic and dimensional data preoperatively and 10 months postoperatively; the latter follow-up included stress echocardiography with treadmill exercise.
RESULTS: At rest and during exercise (25 and 50 W), there was a significant difference in mean pressure gradient between the bovine and the porcine valves with labeled sizes 21 and 23, with superiority of the Perimount prosthesis. There was no difference in effective orifice area and incidence of patient-prosthesis mismatch among all sizes. The left ventricular mass index decreased significantly within 10 months postoperatively in the size 23 bovine group and the size 25 porcine group.
CONCLUSIONS: Our data show a significant superiority of pressure gradients for the bovine bioprosthesis, especially with small valve sizes, when compared with the porcine device, which is more distinctive during exercise.
| See related editorial on page 961.
|
The major concern when implanting a biologic heart valve prosthesis are primary tissue failure after implantation and long-term hemodynamic performance. Most bioprostheses are made of either bovine pericardium or porcine heart valve tissue. Both types have been shown to warrant satisfactory hemodynamic results and tissue durability over 10 to 15 years. However, up to now, no prospective randomized comparison of both valve types under stress conditions has been available. Hence the aim of this study was to compare porcine and bovine bioprostheses in a prospective randomized investigational series with regard to differences in hemodynamic performance at rest and during exercise and postoperative left ventricular (LV) mass regression. The porcine valves are represented by the Medtronic Mosaic bioprosthesis (Medtronic, Inc, Minneapolis, Minn), and the bovine valves are represented by the Carpentier-Edwards Perimount bioprosthesis (Baxter Healthcare Corp, Edwards Division, Santa Ana, Calif).
| Methods |
|---|
|
|
|---|
|
|
Echocardiographic measurement
Echocardiographic measurements performed at rest included the transvalvular flow velocity with continuous-wave Doppler scanning and left ventricular outflow tract (LVOT) flow velocity with pulsed-wave Doppler scanning. LVOT diameter was assessed from a parasternal long-axis view. The same measurements were performed during exercise, except for LVOT diameter, which was assumed to remain constant.1 The velocity recording was first performed in the transvalvular jet and then in the LVOT. From these measurements, we calculated the LV stroke volume as the product of LVOT velocity time integral and cross-sectional area and the cardiac output as the product of stroke volume and heart rate. The transvalvular pressure gradient was calculated by using the modified Bernoulli equation, with inclusion of subvalvular velocity and the effective orifice area (EOA) by using the standard continuity equation. If the difference in heart rate between the time of transvalvular velocity recording and that of LVOT velocity recording was greater than 5%, the data were rejected to avoid potential errors in EOA and mean gradient because of changing hemodynamics. The EOA was indexed for body surface area.
LV end-systolic and end-diastolic dimensions and the thickness of the LV posterior wall and interventricular septum were assessed in the short axis of the parasternal view by means of multiple M-mode measurements with calculation of shortening fraction. LV mass was calculated by using the corrected formula of the American Society of Echocardiography and was indexed by body surface area.2
Implanted porcine and bovine bioprostheses
The Medtronic Mosaic is a stented heart valve fixed with glutaraldehyde at zero pressure and treated with
-amino-oleic acid to reduce tissue calcification. It has been in clinical use since 1994 (Europe) and 2000 (United States), respectively, and has shown satisfactory hemodynamic performance and freedom rates from adverse events.36 The Carpentier-Edwards Perimount is also a stented heart valve fixed with glutaraldehyde at low pressure, and the cusps are treated with surfactant to restrict calcification. The Perimount received US Food and Drug Administration approval in 1991. Long-term studies have constantly shown excellent hemodynamic and clinical results.710
Surgical procedure
Aortic valve replacement was undertaken by using standard cardiopulmonary bypass at mild hypothermia with cold crystalloid cardioplegia. Intraoperatively, aortic annulus diameter was measured by the surgeon by inserting a Hegar dilator into the annulus. Mean aortic annulus diameter was equal in the Mosaic and the Perimount groups. Prosthetic valve size was determined by using the original sizer provided by each manufacturer. The Mosaic valve is designed to allow for a complete supra-annular implantation technique made possible by a low-profile stent design and a construction involving no stent material reaching into the aortic annulus and disturbing the physiologic flow pattern. The Perimount bioprosthesis is sized for and implanted in the intra-supra-annular position. There was no difference in the operative technique between both valve types. The bioprostheses were implanted with pledget-supported, interrupted, noneverting mattress sutures.
Statistical analysis
Data are expressed as means ± 1 SD. Comparisons between both groups were made by the t test for independent samples in case of normal data distribution and the Mann-Whitney U test in case of nonnormal data distribution. Comparisons within one group were performed by using the t test for dependent samples or the Wilcoxon matched pairs test, respectively. Correlations were tested with the Pearson product-moment method. Because there were few patients in the size 19 groups, no statistical comparison was performed for this group.
| Results |
|---|
|
|
|---|
|
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
Hemodynamic data at rest and at stress
No statistical comparison was performed in the size 19 group because of the small number of patients enrolled. Concerning the hemodynamic results at rest, there is a significant superiority in mean pressure gradient of the bovine bioprosthesis in general (P = .001) with valve sizes 21 (P = .022) and 23 (P = .031) compared with the porcine valve sizes 21 and 23. No significant difference was found in mean pressure gradient comparing valve size 25 (P = .139), but an obvious trend toward lower pressure gradients was seen in the bovine group. However, Vitale and colleagues14 and Bortolotti and associates15 reported on higher pressure gradients with the size 19 and 21 Perimount valves.
During exercise, the size 21 and 23 bovine prostheses again showed significantly lower mean pressure gradients than the porcine valves, with even more distinct differences (size 21, P = .0006 in 25 W and P = .0005 in 50 W; size 23, P = .002 in 25 W and P = .0007 in 50 W). In the size 19 and 25 groups, there is a trend towards better hemodynamics in the bovine group but without significant results in t tests for independent samples for the size 25 group (P = .132 in 25 W and P = .061 in 50 W).
Mean pressure gradient was calculated according to the modified Bernoulli formula, which is based on the blood flow velocity measured by means of continuous Doppler scanning at the prosthesis and by means of pulsed Doppler scanning in the LVOT. To exclude differences in mean pressure gradient between both investigated bioprostheses caused by different levels of blood flow at the same level of exercise, we added the mean stroke volume for each valve size at each exercise level to the table of results. As depicted in Table 4, there was no significant difference in the mean stroke volume of the size 21 and 23 prostheses, which differed significantly in mean pressure gradients. We preferred to refer the hemodynamic results at stress to the stroke volume to omit bias caused by interindividual performance levels during exercise. However, often cardiac output (stroke volume x heart rate in milliliters per minute) is used as a reference. We observed that old patients, in particular, often increase their cardiac output, mainly by an increase in heart rate and not by enhancing the stroke volume. Finally, cardiac output increases during exercise, with steady stroke volume. We think that the stroke volume is the decisive parameter, which mainly influences the pressure gradients, and therefore, the stroke volume is the most suitable parameter as a reference value for comparisons of hemodynamic data at stress.
It should be mentioned that the comparison of 2 bioprostheses according to their labeled valve size is difficult because the labeled valve size does not represent the geometric dimensions of the valve. In general, the inner diameter of the Mosaic prosthesis is smaller across all sizes than the inner diameter of the Perimount prosthesis, whereas the sewing ring diameter of the Mosaic valve is smaller in sizes 19, 21, and 23; equal in size 25; and larger in sizes 27 and 29 in comparison with that of the Perimount valve. Moreover, the Mosaic is intended for complete supra-annular position and the Perimount for intra-supra-annular position. For example, in the same aortic root either a size 21 Perimount or a size 23 Mosaic prosthesis, allowing for lower pressure gradients, might be implanted. However, this upsizing is not considered in a hemodynamic comparison simply on the basis of labeled valve size. Such difficulties must be considered when assessing the hemodynamic performance of these prostheses and additionally must be remembered in all comparisons between different heart valve prostheses.16
For this reason, the hemodynamic results are additionally depicted according to the patient annulus diameter, which was measured intraoperatively by the cardiac surgeon (Tables 4, B, and 5, B). In this series the mean aortic annulus diameter was identical for the Mosaic and the Perimount groups (22.6 mm). The mean labeled valve size did not differ significantly (Mosaic, 22.6 mm; Perimount, 22.2 mm), with a tendency toward larger valve sizes in patients receiving Mosaic prostheses. These results do not confirm the expectation that in general a Mosaic prosthesis larger in labeled valve size can be implanted than a Perimount prosthesis. The fact that upsizing of the Mosaic valve is not possible in all cases can be explained by looking at the geometric dimensions of both prostheses. The sewing ring diameter of the size 23 Perimount prostheses is 31 mm, the sewing ring diameter of a size 23 Mosaic prosthesis is 30 mm, and the sewing ring diameter of a size 25 Mosaic prosthesis is 33 mm. Therefore, an objective comparison of the hemodynamic results between different prostheses should be based on the patients aortic annulus diameter measured intraoperatively with a neutral instrument, such as a Hegar dilator. Regarding all patients, there is a significant superiority in mean pressure gradient of the Perimount prosthesis at rest (P = .001), which becomes even more distinct during exercise (P < .001 at 25 W and P < .0001 at 50 W).
Patient-prosthesis mismatch
Regarding EOA and EOAI, there is no significant difference between both valve types. EOAI is an important variable because it includes EOA, representing the prosthetic hemodynamic performance and body surface area as an indicator of the patients estimated need for performance. EOAI is the decisive parameter to describe the incidence of patient-prosthesis mismatch. We rated the extent of patient-prosthesis mismatch as not present in patients with an EOAI of greater than 0.85 cm2/m2, moderate in patients with an EOAI of 0.85 cm2/m2 or less, and severe in patients with an EOAI of 0.65 cm2/m2 or less.11 This graduation corresponds to the general concept that moderate aortic stenosis of a native valve is present in patients with an EOAI of less than 0.90 cm2/m2.17 Although the aim of aortic valve replacement in patients with aortic stenosis is to remove the stenosis and to realize nearly physiologic transvalvular gradients, a residual stenosis represented by low EOAI is frequent, especially with small valve sizes. Patient-prosthesis mismatch is present in up to 52% of the patients with a stented aortic bioprosthesis.18 This corresponds to the incidence of patient-prosthesis mismatch seen in this series with valve sizes 23 and 25. Because patient-prosthesis mismatch might be associated with higher pressure gradients, less LV mass regression, and increased morbidity and mortality,1921 various strategies to reduce the incidence of mismatch were introduced: aortic root enlargement, stentless bioprostheses, and full root replacement. The application of aortic root enlargement procedures must be recommended, especially for patients with an aortic annulus that would only allow the implantation of a size 19 valve. In this series there is a mismatch in 100% of the patients with a size 19 valve. Thus, neither a bovine nor porcine stented bioprosthesis of size 19 can be recommended for implantation, apart from patients with increased risk for aortic root enlargement or with such a small body surface area that mismatch does not occur.
Left ventricular mass regression
All patients in our series showed a regression in LV mass and mass index, irrespective of prosthesis size, although a significance in t test results was only seen with the size 23 bovine and the size 25 porcine valves. There was no significant difference between both valve types regarding the absolute amount of LV mass regression. Although there was a clear reduction in LV mass, the average postoperative LV mass index for the entire series remained greater than normal in 34.9% of the patients, with LV hypertrophy defined as an LV mass index of greater than 130 g/m2 in male subjects and greater than 100 g/m2 in female subjects.22 Similar observations have been made in other studies of various heart valve devices.2325 The reasons for incomplete regression of hypertrophy include a residual aortic gradient caused by patient-prosthesis mismatch, persistent high blood pressure, and nonhemodynamic factors, such as genotype.25 Preoperative LV mass in patients with aortic stenosis is markedly increased. The magnitude of absolute LV mass index regression is closely related to preoperative LV mass index.26 In our series the correlation between preoperative LV mass index and postoperative LV mass reduction was an r value of 0.74 (r2 = 0.55, P < .01).
| Conclusions |
|---|
|
|
|---|
We recommend that the pressure gradients obtained during stress echocardiography be referred to the stroke volume instead of cardiac output because cardiac output is strongly influenced by increasing heart rate in elderly patients during exercise.
Patient-prosthesis mismatch is a common complication in aortic valve replacement with small valve sizes. This experience was confirmed in this series without significant differences between both valve types. Mismatch was present in 100% of the patients with valve size 19, and thus aortic root enlargement procedures and implantation of a larger valve should be recommended instead of implantation of a valve labeled size 19.
|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P.-Y. Litzler, A. Cribier, A. Zajarias, D. Comte, H. Eltchaninoff, C. Tron, C. Haas-Hubscher, and J.-P. Bessou Surgical aortic valve replacement after percutaneous aortic valve implantation: What have we learned? J. Thorac. Cardiovasc. Surg., September 1, 2008; 136(3): 697 - 701. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Bottio and G. Gerosa Conflicting Echocardiographic Evaluations of Supra-Annular Tissue Valves: A Mystery Not Yet Resolved Ann. Thorac. Surg., August 1, 2007; 84(2): 713 - 714. [Full Text] [PDF] |
||||
![]() |
M. J. Dalmau, J. Maria Gonzalez-Santos, J. Lopez-Rodriguez, M. Bueno, A. Arribas, and F. Nieto One year hemodynamic performance of the Perimount Magna pericardial xenograft and the Medtronic Mosaic bioprosthesis in the aortic position: a prospective randomized study Interactive CardioVascular and Thoracic Surgery, June 1, 2007; 6(3): 345 - 349. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Reply to the Editor J. Thorac. Cardiovasc. Surg., March 1, 2007; 133(3): 847 - 849. [Full Text] [PDF] |
||||
![]() |
W. B. Eichinger, I. M. Hettich, and R. Lange Small aortic annulus: The hydrodynamic performances of 5 commercially available tissue valves J. Thorac. Cardiovasc. Surg., December 1, 2006; 132(6): 1499 - 1500. [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Reply to the Editor J. Thorac. Cardiovasc. Surg., December 1, 2006; 132(6): 1500 - 1501. [Full Text] [PDF] |
||||
![]() |
M. J. Dalmau, J. MariaGonzalez-Santos, J. Lopez-Rodriguez, M. Bueno, and A. Arribas The Carpentier-Edwards Perimount Magna aortic xenograft: a new design with an improved hemodynamic performance Interactive CardioVascular and Thoracic Surgery, June 1, 2006; 5(3): 263 - 267. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Gerosa, V. Tarzia, G. Rizzoli, and T. Bottio Small aortic annulus: The hydrodynamic performances of 5 commercially available tissue valves J. Thorac. Cardiovasc. Surg., May 1, 2006; 131(5): 1058 - 1064. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-H. Sievers Prosthetic aortic valve replacement J. Thorac. Cardiovasc. Surg., May 1, 2005; 129(5): 961 - 965. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |