J Thorac Cardiovasc Surg 2005;129:1078-1083
© 2005 The American Association for Thoracic Surgery
Evolving Technology |
Heart Lung Center Utrecht, University Medical Center, Utrecht, The Netherlands.
Received for publication July 4, 2004; revisions received September 1, 2004; accepted for publication September 7, 2004. * Address for reprints: Paul F. Gründeman, MD, PhD, Associate Professor, Cardiothoracic and Vascular Research, Large Animals, University Medical Center Utrecht (Room G02.523), Heidelberglaan 100, 3584 CX Utrecht, The Netherlands. (E-mail: p.f.grundeman{at}hli.azu.nl).
| Abstract |
|---|
|
|
|---|
METHODS: In total, 120 internal thoracic arter-t-coronary artery anastomoses were constructed correctly (n = 60) or incorrectly (n = 60) with one technical error: suture crossover, purse-string or deep toe stitch (n = 20 each). Anastomoses were performed on ex vivo pressure-perfused porcine (96 anastomoses) and human hearts (24 anastomoses). Two blinded observers scanned and scored the anastomoses with epicardial ultrasonography. In 24 human and 24 porcine anastomoses, angiograms were made of 24 correct and 24 incorrect anastomoses and scored by two other blinded observers. Angioscopy and cast injection served as a reference.
RESULTS: Overall, 119 of 120 anastomoses were accurately scored as correct or incorrect within a median of 67 seconds (8381 seconds) by both observers (sensitivity 0.98, specificity 1.00,
1.00 (1.00, 1.00, and 1.00 in angiography subset, respectively). One deep toe stitch that induced outflow corner stenosis was spotted by both observers but regarded as insignificant and thus inaccurately scored as correct. In 5 anastomoses, unintended irregularities were detected. By angiography, anastomoses were accurately scored with a sensitivity of 0.75 and a specificity of 0.81 (P < .001 vs ultrasonography) and
of 0.54. Angioscopy and cast confirmed ultrasonographic findings and did not reveal irregularities other than detected by ultrasonography.
CONCLUSION: Ex vivo epicardial 13-MHz ultrasonography allowed rapid and accurate evaluation of coronary anastomoses and detected technical construction errors with higher sensitivity and specificity than angiography.
An intraoperative technique to assess anastomotic quality could improve patency rates in all approaches to CABG, because suboptimal anastomoses might be revised before chest closure.4 Currently, angiography is the gold standard, but it is rarely used intraoperatively.5 Epicardial ultrasonography for anastomosis visualization and assessment was described in the 1980s.6 Despite promising results,6 technical limitations prevented widespread use. Recently, we described a new high-frequency epicardial ultrasound minitransducer (15 x 9 x 6 mm) for assessment of the left anterior descending coronary artery (LAD) in patients7 and geometry assessment of anastomoses.8,9 Because of its small size and high frequency (up to 13 MHz in B-mode), this minitransducer has potential for routine intraoperative assessment of anastomoses at all locations on the heart.
The aim of this study was to investigate the ability of two blinded observers to detect and characterize three different standardized construction errors in coronary anastomoses on ex vivo hearts with the 13 MHz minitransducer and compare the results against the gold standard of angiography.
| Material and methods |
|---|
|
|
|---|
Anastomosis construction
All 120 anastomoses were randomly constructed by a single investigator (R.P.J.B.) deliberately either correctly (n = 60) or with a single, standardized construction error (n = 60), either suture crossover, purse-string, or deep toe stitch (n = 20 each). The distribution of anastomosis types is summarized in Table 1. The crossover anastomosis was constructed by interlocking two suture bites on opposite sides of the arteriotomy approximately a third of the anastomotic orifice length from the toe. The purse-string anastomosis was made by pulling heavily on the suture ends before fashioning the suture. In the deep toe suture bite anastomosis, the suture was passed through the posterior wall of the coronary artery at the toe.
|
Ultrasonographic equipment
As in our previous work,8,9 a commercially available, high-frequency (up to 13 MHz in B-mode), linear-array minitransducer (15 x 6 x 9 mm; Aloka Co, Ltd, Tokyo, Japan) was used. The image scan width was 10 mm. The minitransducer was placed in a handling tool that could be held like a pencil.7 Imaging was performed with an Aloka SSD 5000 Prosound ultrasound system (Aloka Co, Ltd). Ultrasound transmission gel (Parker Laboratories, Fairfield, NJ) was applied directly onto the anastomosis for proper contact.
Ultrasonographic scanning
Two observers (R.M. and T.C.D.) blinded to the anastomosis type scanned and scored all anastomoses as described. Both had extensive experience in echocardiographic scanning of coronary anastomoses. Scan time needed to obtain sufficient information for scoring was recorded.
Angiography
In a subgroup of 24 porcine anastomoses (12 correct, 4 of each type of error) and all 24 human heart anastomoses (12 correct, 4 of each type of error), angiography (C-arm, Pulsera; Philips Nederland BV Medical Systems, Eindhoven, The Netherlands) of the graft and anastomosis was performed in at least two different oblique projections by selectively injecting contrast media through the ITA (Table 1). Two independent observers (J.M.P.G.E. and C.B.) blinded to the anastomosis type scored the angiograms as described.
Anastomosis scoring by ultrasonography and angiography
Anastomoses were scored as either correct or a construction error. The location and appearance of detected technical errors and irregularities were indicated in a schematic drawing of the anastomosis. The location was categorized in one or more of the following categories: ITA, anastomotic orifice, inflow corner, outflow corner, and recipient coronary artery.
Angioscopy and cast
During perfusion, a 2.4-mm angioscope (11281A; Karl Storz GmbH & Co, Tüttlingen, Germany) was introduced through the free proximal end of the ITA to visualize the anastomosis.2,3 A polyvinylsiloxane impression material (Kerr Co, Romulus, Mich) was injected into the ITA to fill the coronary artery, the anastomosis, and the ITA itself. The resulting 3-dimensional cast of the anastomosis was removed from the vessels after hardening and visually inspected at 3.5x magnification. Angioscopic and cast findings combined served as a reference for ultrasonographic and angiographic findings.
Statistical analysis
The
value was calculated to rate the agreement between observers and interpreted according to Landis and Koch.10 Sensitivity and specificity were calculated for the scoring of both observers combined. Scores were compared with the
2 test. Scan times are presented as median with range and were compared with the Wilcoxon rank sum test.
| Results |
|---|
|
|
|---|
Ultrasonographic scanning and scoring
All 120 anastomoses were easily visualized within a median of 67 seconds (range 8381 seconds). Visualization of the porcine anastomoses (median 61 seconds, range 8381 seconds) was faster than that of the human anastomoses (median 92 seconds, range 14256 seconds, P < .001). Representative ultrasonographic images of a control anastomosis are presented in Figure 1.
|
1.00). In most anastomoses, the presence of a construction error was detected directly on first visualization of the anastomosis. In the single inaccurately scored anastomosis (deep toe stitch anastomosis), slight narrowing of the outflow corner was spotted but considered insignificant and thus scored as correct by both observers.
|
|
|
|
In several anastomoses, tissue was detected in the anastomosis only by the second observer. It is conceivable that sometimes tissue was introduced during the angioscopy procedure that was performed between scans by the two observers. This is a known complication of angioscopy.2 Those findings were not used in the analysis.
In all anastomoses scored as construction error, the error location was accurately scored by both observers (
1.0). The sort of error was correctly identified in 58 of 60 anastomoses by observer 1 and in 54 of 60 anastomoses by observer 2 (93% overall).
Presence of calcifications in the human hearts was not a confounding factor for image interpretation. Calcifications showed clear echocardiographic shadowing (Figure 5), making them easy to discriminate from construction errors.
|
0.54; Table 2). In these 48 anastomoses, the sensitivity and specificity of ultrasonography were both 1.0 (P < .001 vs angiography). The suture crossover error proved to be particularly difficult to detect by angiography (Figure 2). Of the inaccurately scored anastomoses, 5 were inaccurately scored by both observers. The error location in anastomoses accurately scored as construction error was accurately identified in 10 of 18 and 14 of 18 anastomoses (68% accurate overall) by observers 3 and 4, respectively. Three anastomoses were inaccurately scored because of the presence or presumed presence of plaque and calcification.
Angiographically, the specific kind of construction error could not be determined reliably. Only a distinction between outflow corner and orifice narrowing could be made.
Angioscopy and cast
Angioscopy and cast findings corresponded with ultrasonographic findings, and no irregularities other than those detected by ultrasonography were noted.
| Discussion |
|---|
|
|
|---|
Anastomosis quality control
With the advent of off-pump and minimally invasive CABG, there has been a renewed interest in techniques for intraoperative assessment of distal anastomosis quality.4,11,12 Currently, only graft flow measurement is used on a large scale.11 Graft flow measurement, however, can only detect severe stenosis (>75%) and provides no information about its location. Furthermore, there is no clear cutoff point for adequate graft flow, and thus it may underestimate the number of suboptimal anastomoses. For various reasons, the gold standard of angiography is not used frequently, most importantly because it is invasive, not readily available in the operating room, and time-consuming.5
The potential of epicardial ultrasonography for coronary anastomosis assessment was recognized almost 20 years ago.6 Technical transducer limitations, however, have prevented widespread clinical introduction. The current minitransducers allow easier access. To date, no detailed evaluation has been reported regarding the ability of epicardial ultrasonography to detect and characterize different technical errors in coronary anastomoses relative to the gold standard of angiography.
In the subgroup of 48 anastomoses evaluated with angiography, epicardial ultrasonography enabled detection of construction errors with significantly higher sensitivity (1.00) and specificity (1.00) than achieved with angiography (0.75 and 0.81, respectively). When calculated from all 120 anastomoses in the study, the sensitivity and specificity of epicardial ultrasonography were 0.98 and 1.00, respectively.
In addition, epicardial ultrasonography enabled discrimination between narrowing caused by suture errors or by calcifications, that present with a distinct echo shadowing (Figure 5). This was not possible with angiography.
Anastomosis evaluation with epicardial ultrasonography required only a median of 92 seconds in human anastomoses, which is clinically acceptable. In most cases, the detection of the error was instantaneous, and subsequent scanning time was spent characterizing the error and looking for additional irregularities. Clinically, revision would be performed directly after a serious error was spotted.
As we have shown before,7 the minitransducer can also be used to locate the coronary artery and select the optimal anastomotic site. Combined with its use for anastomotic quality assessment and possibly epiaortic scanning as well, the ultrasound minitransducer may prove a multipurpose diagnostic tool to improve the quality of CABG surgery.
A concern with the use of epicardial ultrasonography for anastomosis quality assessment is that image interpretation is subjective and operator dependent.4 In this study, however, the two observers scored all anastomoses identically with regard to the presence of a construction error (
1.0), indicating an almost perfect agreement,10 whereas the
for angiography was 0.54, indicating only moderate agreement. The observers did have extensive previous experience with the interpretation of ultrasonographic images of anastomoses. Preferably, the surgeon would be trained in a laboratory setting, including scanning of anastomoses on ex vivo hearts perfused with saline solution and image interpretation of off-line images on a computer. This might take a couple of hours. Alternatively, the surgeon might teach himself or herself by starting to acquire images during cardioplegic arrest and later studying the images off-line. This would permit peer review by a radiologist.
| Limitations |
|---|
|
|
|---|
| Conclusion |
|---|
|
|
|---|
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
P. K. Hol, K. Andersen, H. Skulstad, P. S. Halvorsen, P. S. Lingaas, R. Andersen, J. Bergsland, and E. Fosse Epicardial Ultrasonography: A Potential Method for Intraoperative Quality Assessment of Coronary Bypass Anastomoses? Ann. Thorac. Surg., September 1, 2007; 84(3): 801 - 807. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, P. F. A. Bakker, R. Meijer, C. Borst, and P. F. Grundeman Ultrasound Mini-Transducer with Malleable Handle for Coronary Artery Surgery Ann. Thorac. Surg., January 1, 2006; 81(1): 322 - 326. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. P.J. Budde, W. J.L. Suyker, P. T.W. Suyker, C. W.J. Verlaan, R. Meijer, C. Borst, and P. F. Grundeman Quality assessment of distal S2AS connector anastomosis by 13MHz epicardial ultrasound Eur. J. Cardiothorac. Surg., December 1, 2005; 28(6): 833 - 837. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||