J Thorac Cardiovasc Surg 2005;130:41-47
© 2005 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
a Academic Department of Cardiothoracic Surgery, National Heart and Lung Institute, Imperial College, University of London, London, United Kingdom
b Royal Society/Wolfson Foundation Medical Image Computing Laboratory, London, United Kingdom
Received for publication December 5, 2004; revisions received January 28, 2005; accepted for publication February 10, 2005. * Address for reprints: Professor Magdi Yacoub, The Magdi Yacoub Institute, Harefield Research Foundation, Hill End Rd, Harefield, Middlesex UB9 6JH, United Kingdom (Email: m.yacoub{at}imperial.ac.uk).
| Abstract |
|---|
|
|
|---|
METHODS: The technique relied on the use of hierarchic 3-dimensional free-form volume registration with fast global optimization between normally acquired and contrast-enhanced electron beam tomographic volume. A total of 66 contrast-enhanced electron beam tomographic scans of the aortic root were performed in 27 patients, 10 with native aortic valve disease (group A) and 17 from a prospective randomized trial of aortic root replacement (group B, 9 Freestyle grafts [Medtronic, Inc, Minneapolis, Minn] and 8 homografts). To validate the in vivo electron beam tomographic measurements, 5 patients from group A underwent electron beam tomographic scans before the operation and then had their own valves, explanted at the time of surgery, analyzed for calcium quantification by ex vivo electron beam tomography.
RESULTS: In group A, the mean (± SE) calcification score was 6560 ± 2388, which correlated with peak gradients measured at echocardiography (r = 0.93, P = .02). In group B, the mean (± SE) calcification score was 168 ± 27, showing a tendency toward a lower calcification for Freestyle valves than for homografts at 2 years after implantation (P = .052). A mean variability of 6% was found between in vivo electron beam tomographic scores of calcification and those measured on valve specimens after explantation.
CONCLUSION: We describe a novel method to characterize the degree and location of calcification in both native valves and postoperative valve implants. The technique may be useful in the management of patients with aortic valve disease and has potential as a screening tool for high-risk patients to diagnose early valve calcification and possibly institute corrective measures.
| Introduction |
|---|
|
|
|---|
| Methods |
|---|
|
|
|---|
|
For the in vivo EBT, a set of 20 to 25 transverse tomograms of 3-mm thickness was obtained through the aortic root with the subject breath holding. Acquisitions were made at 80% of the RR interval. The scan was then repeated during intravenous injection of a contrast medium. Omnipaque 240 (Nycomed Imaging AS, Oslo, Norway) was injected into a vein in the antecubital fossa at a rate of 3 mL/s, and scans were acquired at 30 seconds after the beginning of the injection. The explanted native aortic valves were analyzed in accordance with our previous studies.
15,16
Echocardiographic Measurements
Transthoracic echocardiograms were performed at same time intervals as the EBT investigations by means of a Hewlett Packard Sonos (Hewlett-Packard Ltd, Bracknell, UK) machine. Echocardiographic Doppler imaging was used to analyze transvalvular and subvalvular flows. Peak transvalvular and subvalvular flow velocities were recorded, and gradients were derived with the modified Bernoulli equation:
P = 4v2. Mean gradients were calculated from the flow profile.
EBT 3D Serial Image Registration
The examinations were based on a sequential back-to-back scan. To circumvent the geometric inconsistencies introduced by the respiratory movements of the base of the heart during the cardiac cycle, we adapted a previously developed rapid intensitybased multiresolution image registration scheme.
17
As extended from the previous study,
17
a free-form image registration based on 3D B-spline transformation model was adopted to correct the respiration-induced cardiac deformation. In this study, intensity based cross-correlation coefficient was used as a measure of the degree to which the two images were aligned. The optimal B-spline parameters to maximize this similarity measure were determined with a multihierarchical quasi-Newton optimization strategy.
17
The definition of the cross-correlation coefficient and a detailed optimization strategy for image registration are described in Appendix E1.
Interactive Volume Visualization and Calcium Scoring
For interactive volume visualization of the general trends of valve calcification with time, a software-accelerated 3D interactive rendering technique called the "bubble model" with shear-warp factorization was used. The normally scanned volume was segmented with calcium-level thresholding (130 Hounsfield units) with isosurface rendering.
For quantitative analysis, interactive calcium quantification was conducted with CMRTools (Imperial College, London, UK) according to the Agatston scoring scheme.
6,9
Two uninformed, independent observers (G.M and P.H.) were asked to interactively identify the location of the calcified leaflets and define regions of interest on the enhanced images showing the exact location of the valve leaflets. These regions of interest were then automatically overlaid on the corresponding unenhanced ones for calculation of the calcium score. A calcium score was then calculated for each valve by multiplying the calcified lesion area in square millimeters by a density score determined from the peak CT number. For each observer, the process was repeated twice. Both interobserver and intraobserver variabilities were calculated. In addition, on the basis of the free-form registration results, each aortic valve was normalized as a circular disk with sectors labeled as right, left, and noncoronary leaflets, respectively. For each subject, calcified lesions were spatially mapped onto the disk according to their actual locations on the leaflets. The use of this normalizing scheme not only facilitates intrasubject follow-up assessment but also simplifies localization comparisons. Calcium within aortic sinuses, aortic wall, or both was carefully excluded from the analysis.
Statistical Analysis
Continuous variables were analyzed by means of the Student t test for unpaired data as appropriate. A 2-way analysis of variance was used to assess serial changes in continuous variables. Correlations between the calcification scores and peak transvalvular gradients were determined by means of Pearson product-moment correlation analysis.
| Results |
|---|
|
|
|---|
|
|
The quantitative results for the 17 patients in group B are provided in Figure 1. There was a trend for calcification to increase with time in the homografts but not in the Freestyle valves. For the total 122 volume pairs analyzed, the interobserver variability was 9.2%, and intraobserver variabilities were 18% and 13.5%.
|
Calcium Localization
Figure 2
shows the calcium content for each cusp (right, left, and noncoronary) with time in all the patients from group B. Interestingly, the left coronary leaflet showed a higher degree of calcification than did the right and noncoronary leaflets at each time point.
|
|
|
Relation to Hemodynamic Data
In group A (n = 10), the mean peak transvalvular gradient measured at echocardiography was 54 mm Hg, and it correlated well with the scores of calcification (r = 0.94, P = .02), with the highest calcification score to cause significant stenosis being 16,332. No patient in group B (n = 17) showed pathologic gradients across the aortic valve. The mean gradients were always less than 3 mm Hg, and the echocardiographic investigations demonstrated a good leaflet excursion during the cardiac cycle at each interval. The highest calcification score in group B was 821.
| Discussion |
|---|
|
|
|---|
The aortic valve performs sophisticated functions, which are dependent on the dynamic behaviors of its components.
18
Calcific stenosis is a common failure mode for both native human aortic valves and the biologic aortic bioprostheses, but in neither case are the etiology and mechanism of progression well understood. In native valves, calcification is an active process resembling atherosclerosis.
3,19
The calcification process has some similarities in glutaraldehyde-fixed bioprosthetic valves, but it does not involve living interstitial cells because glutaraldehyde is cytotoxic. In addition, free aldehyde within the tissue matrix has also been thought to be an indicator for calcification.
20
The biomechanical environment within which the bioprosthetic valve is placed determines the underlying trend that the structure is altered both morphologically and functionally. Existing research in the area underscores the complexity of the bioprosthetic valve calcification process and highlights the need for combined biochemical, genetic, biomechanical, histologic, and accurate in vivo imaging studies. The method described in this article could help in that regard.
EBT has proved to be highly sensitive for calcium measurements in the coronary arteries,
21
and it has the advantage of being noninvasive, allowing a rapid acquisition of the images. By using EBT, we demonstrated the pattern of calcification of the aortic root wall after aortic root replacement with homografts or Medtronic Freestyle porcine xenografts.
9
The technique was then improved for the visualization of the aortic valve leaflets.
14
In two different in vitro experiments, we then validated our EBT method for detection and quantification of aortic valve leaflets.
15,16
We found high correlations between the EBT scores and of biochemically measured calcium content with experimental valves
15
or failed biologic prostheses explanted from patients.
16
The validation of the technique has been extended in this study, in which there was only a 6% mean variability between the EBT measures in vivo and the quantification of calcium on the aortic valve samples after explantation from the patients during the operation for aortic valve replacement. This variability could have been due, at least in part, to inability to excise the valve intact with all the calcium. The quantification of calcium was further improved in this investigation. We have used a fast, multiresolution 3D registration scheme based on the piecewise linear joint- histogram relationship between normally acquired and contrast-enhanced volumes. This simplifies the treatment of the optimization process, which involves a large number of free variables, and permits the use of Broyden-Fletcher- Goldfarb-Shanno (BFGS) algorithm, which is an extension to the Newton method and has a quadratic convergence property (Appendix E1). It is worth noting that the chosen similarity function is justified by the use of blood pools as registration landmarks. The local joint histogram of intensities within the blood pool has been shown to have nearly linear correlation between the standard and contrast-enhanced CT volumes. The abundance of the blood signal at the imaging volume therefore has a strong influence toward the correct alignment of both volumes. This patient study indicates the practical value of the developed method.
Some recent studies have suggested the use of EBT to detect and quantify calcification of the native aortic valve.
3,22
These studies, however, tended to use scans performed primarily to assess coronary artery calcification
3
or were done in patients undergoing catheterization before aortic valve replacement or who were found to have at least mild aortic stenosis at echocardiography
22
and therefore did not take into account the movement of the aortic root, the position of the aortic cusps, and the different positioning of the patients on the scanner. These difficulties can be circumvented by localizing the aortic cusps with contrast enhancement on one occasion.
14
Shavelle and associates
11
showed in a retrospective study that patients with EBT aortic valve calcification scores of at least 150 should be referred for echocardiography, because these are more likely to have significant aortic stenosis.
11
In our series of bioprosthetic valves (both homografts and Freestyle valves), 8 patients had a calcification score of at least 150 at different time points, but in none were peak gradients higher than 7 mm Hg. This rule of thumb, however could be true for the native valves, where the lowest score was 235 and the agreement with the echocardiographic gradients across the aortic valve was high. We think this demonstrates the importance of the accurate determination of the aortic valve leaflets and the quantification of calcium by one method.
Calcification is thought to be a major determinant of hemodynamic function and durability of biologic devices, which leads to stenosis or regurgitation after implantation. Localized calcification can alter the physical properties of the valve and encourage tearing in bioprostheses.
In an attempt to enhance the performance of bioprostheses, different anticalcification treatments are being used.
23,24
Although these techniques have met with success in animal models, the efficacy in human beings has not been investigated. A quantitative evaluation of the impact of such treatments is therefore needed, and the technique described in this article provides a powerful tool to achieve this goal. In addition, the technique allows evaluation of the rate of progression of calcification in both the native and bioprosthetic aortic valves and screening of people at risk such as middle-aged individuals, atherosclerotic patients, patients undergoing treatment of renal failure, postmenopausal women, and patients affected by familial hypercholesterolemia, which could be of prognostic importance. In addition, the technique described in this article provides for the first time data relating to the in vivo 3D localization of the calcium deposits in the valve. This provides new additional data relating to the pattern of calcification in different types of valves and could help in the understanding of the pathogenesis of calcification and possibly of its prognosis.
Until recently, the most accurate methods of quantifying bioprosthetic valve calcification consisted of in vitro analyses such as histologic assessments, radiographic and atomic absorption spectroscopy, and dual energy x-ray absorptiometry,
25
which did not allow sequential examinations in the same patient. This study, which was conducted in patients, confirmed that calcification of the bioprosthetic aortic leaflets occurs as early as 6 months after implantation for both homografts and Freestyle prostheses. There was a trend for the calcification to progressively increase in the homograft group up to 2 years after the operation. The trend was different in the Freestyle group, where calcification tended to remain constantly low throughout all the investigation period. The amounts of calcium documented in the bioprosthetic valves were much lower than those observed in the diseased native valves. This could be explained by the fact that the homograft and Freestyle valves studied were well functioning and had no echocardiographic evidence of stenosis. However, the progressive accumulation of calcium in the homografts is a cause for concern and warrants further investigation. One of the limitations of this study is that it does not provide information about the dynamics of the valve leaflets because of the intrinsic constraint of the imaging modality itself. Recent developments of object-tracked magnetic resonance imaging
26
and 3D ultrasonography
27
are likely to play an important role in future investigations in linking mechanical properties to the progression of valve calcification. Another limitation is that the use of a contrast medium may be a potential risk for patients with renal failure and is also a contraindication for allergic patients.
In conclusion, the technique of in vivo 3D characterization of aortic valve calcification described in this article appears to be a sensitive, reliable method of measuring the amount and localization of calcification in both native and bioprosthetic valves. Sequential use of this technique could be valuable in understanding the pathogenesis of valve calcification and in the management of patients with aortic valve disease or bioprosthetic valve replacement.
| Appendix E1 |
|---|
|
|
|---|
Extended from our previous study of fast 2-dimensional image registration, the registration procedure was carried out by using the cross-correlation coefficients. It was chosen as the similarity measure because its mathematical formulation can be exploited to speed up the optimization process, as discussed by Veeser and colleagues.
17
That is:
|
| (1) |
is the overlapping voxels in the intersection region of the reference and the transformed volume, E {·} is an expecting function of the voxels intensity, and I1, I2 is the referent and target images to be registered.
A linear tensor product B-spline defined by equation 2 was subsequently used as the transformation matrix because it enables adaptive subdivision of the control grid for a multiresolution implementation; that is:
|
| (2) |
p(u,v,w) c
ijk
is the coordinate of control points surrounding the point
p(u,v,w), and T(
p) is the transformation that map point
p(u,v,w) in target image to referent image.
A quasi-Newton optimization algorithm, the BFGS algorithm, was then used to find the optimum transformation parameter vectors. In contrast to the Newton method, the algorithm does not need explicit second order derivatives. Instead, it approximates these values by using information from previous iterations; that is:
|
| (3) |
f(x
k
) is the vector of partial derivative of the function with respect to each parameter, and H
k
is the estimated Hessian matrix.
As indicated by equation 3, the BFGS algorithm, however, does require the evaluation of the first order derivatives of the objective function, which can be computationally expensive. Fortunately, with the structure of the similarity matrix, the partial derivatives of the similarity measure with respect to the control points can be expressed analytically; that is:
|
| (4A) |
|
| (4B) |
|
| (4C) |
|
| (4D) |
|
| (4E) |
| Footnotes |
|---|
G.M. and P.H. have contributed equally to this article.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
I. El-Hamamsy, M. Zaki, L. M. Stevens, L. A. Clark, M. Rubens, G. Melina, and M. H. Yacoub Rate of Progression and Functional Significance of Aortic Root Calcification After Homograft Versus Freestyle Aortic Root Replacement Circulation, September 15, 2009; 120(11_suppl_1): S269 - S275. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David, S. Armstrong, M. Maganti, J. Butany, C. M. Feindel, and J. Bos Postimplantation morphologic changes of glutaraldehyde-fixed porcine aortic roots and risk of aneurysm and rupture. J. Thorac. Cardiovasc. Surg., January 1, 2009; 137(1): 94 - 100. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. E. Dapunt, J. Easo, P. P.F. Holzl, P. Murin, M. Sudkamp, M. Horst, and E. Natour Stentless full root bioprosthesis in surgery for complex aortic valve-ascending aortic disease: a single center experience of over 300 patients Eur. J. Cardiothorac. Surg., April 1, 2008; 33(4): 554 - 559. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |