J Thorac Cardiovasc Surg 2005;130:272-276
© 2005 The American Association for Thoracic Surgery
Surgery for Congenital Heart Disease |
a Department of Pediatric Cardiology and Congenital Heart Disease, German Heart Center, Technische Universität München, Munich, Germany
b Institute of Medical Statistics and Epidemiology, Technische Universität München, Klinikum r.d. Isar, Munich, Germany
Received for publication November 20, 2004; revisions received January 15, 2005; accepted for publication February 8, 2005. * Address for reprints: Ullrich Fakler, MD, German Heart Center, Department of Pediatric Cardiology and Congenital Heart Disease, Technische Universität München, Lazarettstraße 36, D-80636 Munich, Germany (Email: fakler{at}dhm.mhn.de).
| Abstract |
|---|
|
|
|---|
METHODS: The patient population consisted of 143 patients with a mean age of 11.3 years (age range, 2 days to 23.8 years) undergoing cardiac catheterization during general anesthesia and with mechanical ventilation. Oxygen consumption was measured with a standard commercial analyzing system (Deltatrac II; Datex, Engström, Helsinki, Finland). Assumed oxygen consumption values were calculated according to the formulas of Krovetz and Goldbloom and LaFarge and Miettinen. Comparisons between measurements and assumptions were performed by Bland-Altman plots. Two-sided paired t tests were used to assess a difference of the assumed and measured values.
RESULTS: The range of measured oxygen consumption values was between 55.2 and 249 mL·min1 ·m2. The Krovetz-Goldbloom formula led to systematically larger values compared with the measured values (P = .0001; mean difference of 53.3 mL·min1 ·m2; 95% confidence interval, 56.7 to 49.8 mL·min1 ·m2). The use of the LaFarge-Miettinen formula tends to overestimate oxygen consumption (P = .0037; mean difference of 15.9 mL·min1 ·m2; 95% confidence interval, 26.5 to 5.4 mL·min1 ·m2). A similarly poor agreement was found when analyzing a subgroup of 25 patients with Fontan-type circulation.
CONCLUSION: The use of assumed instead of measured oxygen consumption values introduces large errors in the determination of cardiac output.
| Introduction |
|---|
|
|
|---|
The Fick equation is as follows: CO = VO 2/(CaO 2 CvO 2), where CO is defined as cardiac output in liters per minute, VO2 is defined as oxygen consumption in liters per minute, CaO2 is defined as arterial oxygen content in milliliters per liter (1.36xHbg [g/L]xSaO 2) + (PaO 2[mmHg]x0.003), CvO2 is defined as mixed venous oxygen content in milliliters per liter (1.36xHbg [g/L]xSvO 2) + (PvO 2[mmHg]x0.003).
It is common practice to use an estimate of VO2 instead of measurements. Both Krovetz and Goldbloom,
2
carrying out multiple regression analysis, and LaFarge and Miettinen,
3
using multivariate analysis of covariance, derived empiric formulas to estimate VO2, which remain the most commonly used equations for calculating assumed VO2. Considerable errors introduced by using assumed VO2 have been reported and discussed previously.
47
General anesthesia is sometimes necessary to perform cardiac catheterization in pediatric patients under stable conditions, thus affecting oxygen consumption and other parameters substantially.
Therefore, the aim of this study was to assess the quantity of error that might be introduced by using assumed VO2 compared with measured VO2 in a pediatric population with congenital heart disease undergoing cardiac catheterization.
Subgroup analysis in patients with completed Fontan circulation was performed.
| Methods |
|---|
|
|
|---|
All patients received anesthesia, avoiding inhalant narcotics. Their lungs were mechanically ventilated in volume-controlled mode (IPPV ventilator, Siemens Servo Ventilator 900 D; Siemens, Erlangen, Germany), receiving an inhaled fraction of inspired oxygen of between 0.21 and 0.48. Twelve patients had to be excluded because of air leaks of the tube of more than 5% during the measurements. The LaFarge-Miettinen formula was derived in patients between 3 and 40 years of age. One hundred fifteen patients were of this age and were available for the comparison. A subset of 25 patients had a Fontan-type circulation.
Age, sex, height, weight, and heart rate (electrocardiography) were recorded, and body surface area was calculated according to the method of Dubois and Dubois.
8
VO2 was measured with the standard commercial analyzing system Deltatrac II (Datex-Engström, Helsinki, Finland). The Deltatrac Metabolic Monitor is an open-system, indirect calorimetry device equipped with a fast differential paramagnetic oxygen sensor to measure a differential signal between inspired and expired gases and a gas dilution system to measure flow.
9,10
The measurements were taken when the patient was in a stable state shortly after intubation over a period of 10 minutes, obtaining one measurement every minute. All sets of 10 single measurements showed an SD of less than 10%, so effects of a fluctuating FIO
2 on metabolic measurements could be excluded.
11
The mean of these single measurements was calculated and compared with the assumed VO2 values.
Assumed VO2 values were calculated according to the Krovetz-Goldbloom formula as follows: VO 2/BSA = (1.39xheight[cm] + 0.84xweight[kg] 35.6)/BSA(mL/min)/m 2. The formula of LaFarge-Miettinen was used as follows: VO 2/BSA = (138.117.04xln(age) + 0.378x HR)/BSA(mL/min)/m 2 for female subjects and VO 2/BSA = (138.1 11.49xln(age) + 0.378xHR_/BSA(mL/min)/m 2 for male subjects, where age is presented in years and HR is defined as heart rate (in minutes).
Comparisons between measurements and assumptions were performed by means of Bland-Altman plots
12
and comparative correlation plots. Two-sided paired t tests were used to assess a difference of the assumed and measured values. Because there were 2 such tests, the significance level was split to 2.5% for each test, according to the method of Bonferroni. Pearson correlation coefficients were calculated to assess the correlation of assumed and measured VO2. The percent error introduced by using assumed VO2 was calculated by dividing the difference of measured minus assumed VO2 by the corresponding measured VO2. Data are represented as means (SDs).
The subgroup of 25 patients with completed Fontan-type circulation was compared with the 118 patients without Fontan-type circulation by an unpaired t test.
Calculations were computed with SAS 6.12 software (Cary, NC).
| Results |
|---|
|
|
|---|
A Bland-Altman plot of the difference between VO2 measured minus VO2 Krovetz-Goldbloom and the mean of VO2 measured plus VO2 Krovetz-Goldbloom is shown in Figure 1. Figure 2 presents a comparative plot of VO2 measured versus VO2 Krovetz-Goldbloom and the corresponding correlation coefficient.
|
|
|
|
|
| Discussion |
|---|
|
|
|---|
In the patient in whom a Fontan-type repair is planned, the accurate determination of vascular resistance values is of even more substantial clinical importance. Looking at the results of the subgroup of patients with Fontan circulation, it can be demonstrated that the errors found in the determination of cardiac output will lead to wrong values of vascular resistance, especially pulmonary vascular resistance.
Calculations of vascular resistance are obtained by relating the mean pressure change across a circuit to the flow across the circuit. An error in the determination of VO2 as demonstrated (eg, in the Fontan subgroup; mean error, 59% with the Krovetz-Goldbloom formula and 22% with the LaFarge-Miettinen formula) affects the calculation of cardiac output (flow) and pulmonary vascular resistance. In patients with Fontan circulation, it might lead to an overestimation of cardiac output and an underestimation of vascular resistance. This quantity of error obviously cannot be accepted, considering the narrow limits of pulmonary vascular resistance values, planning a further management of a completed Fontan-type circulation.
16,17
A former study showed that in patients with bidirectional Glenn anastomoses, assumed VO2 led to underestimation of pulmonary vascular resistance to an extent that could significantly influence clinical decision making.
18
The poor agreement of measured and assumed VO2 from the formulas might be due to a difference of population and the fact that the patients in this study were undergoing general anesthesia and were mechanically ventilated. However, these formulas were used thus far in sedated patients, as well as in patients under the effects of anesthesia. LaFarge and Miettinen determined their formula in patients between 3 and 40 years of age. Therefore, the formula cannot be applied in younger children. If the Krovetz-Goldbloom formula is used in children less than 3 years of age (n = 28), we found a mean error of 42%. The use of this formula can be questioned, especially in neonates and infants.
Looking at recently published regression-based estimates, both bias and precision showed similar results.
19
A comparison of assumed VO2 with the Krovetz-Goldbloom or the LaFarge-Miettinen formula demonstrates poor agreement with measured VO2 in a pediatric population undergoing cardiac catheterization under general anesthesia and with mechanical ventilation. Routine use of assumed VO2 when calculating cardiac output with the Fick equation might frequently result in large errors in the determination of cardiac output and dependent parameters. Therefore, VO2 should be measured.
| References |
|---|
|
|
|---|
Related Article
This article has been cited by other articles:
![]() |
R.M.F. Berger and A. Bergstra Assumed oxygen consumption in the determination of cardiac output: assume carefully... J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 504 - 504. [Full Text] [PDF] |
||||
![]() |
U. Fakler and J. Hess Reply to the Editor J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 504 - 505. [Full Text] [PDF] |
||||
![]() |
H. Benallal A new modeling for each population J. Thorac. Cardiovasc. Surg., February 1, 2006; 131(2): 506 - 507. [Full Text] [PDF] |
||||
![]() |
J. Li, I. E. Konstantinov, G. S. Van Arsdell, and A. N. Redington Assumed oxygen consumption in the determination of cardiac output in children after cardiac surgery J. Thorac. Cardiovasc. Surg., January 1, 2006; 131(1): 251 - 251. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |