J Thorac Cardiovasc Surg 2005;130:351-354
© 2005 The American Association for Thoracic Surgery
Cardiopulmonary Support and Physiology |
a Department of Cardiothoracic Surgery, Stanford University, Stanford, Calif
b Department of Research and Development, Thoratec Inc, Pleasanton, Calif
Presented in part at the 50th annual ASAIO Conference, Washington, DC, June 2004.
Received for publication September 10, 2004; revisions received December 20, 2004; accepted for publication December 29, 2004. * Address for reprints: Olaf Reinhartz, MD, Department of Cardiothoracic Surgery, Stanford University, Stanford, CA 94305-5407 (Email: orx{at}stanford.edu).
| Abstract |
|---|
|
|
|---|
METHODS: Eight sheep (4248 kg) were studied. After pilot studies in 3 animals, 5 underwent both pump-supported and nonsupported total cavopulmonary connection in alternating sequence for up to 2 hours. This was achieved with a 12-mm polytetrafluoroethylene graft from the (distally ligated) superior vena cava to the main pulmonary artery and a cannula placed in the inferior vena cava with an attached 16-mm Dacron graft to the main pulmonary artery. Pressures (arterial, inferior vena cava, left atrium, and pulmonary artery) and flows (ascending aorta and inferior vena cava) were recorded over 1 hour both with unsupported total cavopulmonary connection and after placing an axial flow pump (Thoratec HeartMate II) between the inferior vena caval inflow cannula and the main pulmonary artery.
RESULTS: Under nonsupported total cavopulmonary connection circulation, inferior vena caval and aortic blood flow decreased by nearly 50%. Inferior vena caval pressure nearly doubled, whereas arterial pressure decreased by one third. Pulmonary artery pressure became nonpulsatile; however, mean pulmonary artery pressure and left atrial pressure did not change significantly. With pump-supported Fontan circulation, cardiac output, inferior vena caval flow, and arterial pressure returned to baseline. Inferior vena caval pressure decreased to below baseline levels. Mean pulmonary artery pressure and left atrial pressure again remained unchanged.
CONCLUSIONS: Axial flow pump support from the inferior vena cava to the pulmonary artery can prevent the substantial decrease of aortic flow and pressure associated with total cavopulmonary connection and can reverse its poor hemodynamics. This is a simple model that can be used to further evaluate the potential of mechanical support as a treatment option in failing Fontan circulation.
| Introduction |
|---|
|
|
|---|
Up to 40% of patients undergoing TCPC experience heart failure.
2
Late abnormalities of liver function are frequent.
3
Protein-losing enteropathy is diagnosed in up to 15% of patients and has a 5-year mortality rate of 50%.
4
Medical treatment options are limited. In rare cases of Fontan pathway obstruction, revision is indicated. Cardiac transplantation is the only remaining option for patients refractory to all other treatments. This option is, however, associated with increased risk because of the poor condition of patients with failing Fontan circulation.
5
Mechanical circulatory support for left ventricular failure with pulsatile and, more recently, axial flow pumps is now state of the art clinically. However there is little experience with axial flow pumps in low-pressure circuits, such as support of the right ventricle or even a TCPC circulation. Pump support could conceivably reverse poor hemodynamics and secondary organ failure in patients with failing Fontan circulation, could make these patients more successful transplant candidates, or could be used as destination therapy. In this study we compared the acute physiologic response to passive TCPC with that of mechanically supported TCPC by using an axial flow blood pump in a sheep model. Our objective was to define the initial adaptive response of the circulation to passive TCPC and determine the feasibility of mechanical support to restore hemodynamics.
| Methods |
|---|
|
|
|---|
Surgical Preparation
We performed a midline sternotomy; inserted pressure catheters into the left atrium, right atrium (RA), inferior vena cava (IVC), and main pulmonary artery (PA); and placed perivascular ultrasonic flow probes (Transonics Systems, Inc, Ithaca, NY) around the IVC distal to the anastomosis and on the ascending aorta. All pressures and flows were continuously monitored and recorded.
We mobilized the IVC, superior vena cava (SVC), and main PA and ligated the azygous vein. After heparinization with 300 U/kg, we fashioned a bidirectional Glenn connection by anastomosing a 12-mm polytetrafluoroethylene (PTFE; Gore-Tex vascular graft, registered trademark of W. L. Gore & Associates, Inc, Newark, Del) vascular graft from the SVC to the main PA in side-to-side fashion and ligating the SVC-RA junction. For nonsupported TCPC, we connected the IVC and main PA with a 16-mm PTFE graft and ligated the IVC-RA connection (Figure 1). To test pump-supported TCPC, we placed a right-angled Thoratec inflow cannula connected to an axial flow pump (Thoratec HeartMate II; Thoratec Corporation, Pleasanton, Calif) into the IVC. Its outflow end was connected to a 16-mm Dacron graft anastomosed to the proximal main PA. We then again ligated the IVC-RA connection (Figure 2). These graft diameters were chosen to match the sheeps SVC and IVC diameters. The chest was approximated with clamps. All animals were studied with and without mechanical support in random order. The PTFE graft was clamped before insertion of the axial flow pump, or vice versa, the pump was removed before opening the PTFE graft. Therefore, all animals served as their own controls. Hemodynamic variables were recorded over a period of 1 to 2 hours each before intervention and with unsupported and supported TCPC, respectively.
|
|
Metabolic Analyses
Hourly arterial blood gas measurements were obtained to monitor ventilation and base excess. In addition, oxygen saturations in PA and IVC blood, serum lactate levels, and aspartate aminotransferase and alanine aminotransferase concentrations were measured before and during supported and nonsupported TCPC after 30-minute equilibration periods.
Statistical Analysis
Hemodynamic values were analyzed by means of analysis of variance with paired comparison between baseline, unassisted, and ventricular assist device-assisted states. The analysis was conducted with the JMP software package (SAS Institute, Inc, Cary, NC). A significance level of .05 was the minimum criterion for acceptance of differences between groups. All data values are expressed as means ± standard error unless otherwise stated.
| Results |
|---|
|
|
|---|
|
Venous Collapse
At a pump speed of 6000 rpm, the IVC was subject to intermittent collapse and oscillation in all animals. However, this did not appear to affect overall IVC flows. In preliminary studies we verified that further increasing pump speed induced more prolonged IVC collapse, leading to brief and intermittent decreases in flow.
Metabolic Changes
During the relatively brief course of the experiment, we did not observe significant metabolic or lactic acidosis or increases of aspartate aminotransferase or alanine aminotransferase concentrations. Mixed venous saturations were measured in 5 animals: the median was 87% at baseline, 61% at the passive Fontan state, and 82% during Fontan circulation with pump support. Because of the small number of observations, we did not perform any statistical analyses for mixed venous saturations.
| Discussion |
|---|
|
|
|---|
A limitation of this model is the incomplete bypass of the right heart in that coronary sinus flow is still being pumped by the right ventricle. However, this did not lead to any appreciable pulsatility of the PA pressure tracing in our study. It appears that coronary sinus flow is not substantial enough to affect the hemodynamics of this experiment significantly.
Hemodynamics in TCPC
In the previously reported studies, cardiac index was substantially reduced with TCPC; however, output could be maintained by substantial volume loading.
6,7
In our sheep model we have seen an even greater decrease of cardiac output and blood pressure with TCPC without volume loading. As expected, IVC pressure was significantly increased in TCPC. Additionally, we measured IVC flow, which was significantly reduced along with cardiac output.
As in the previous studies in dogs, all animals survived in the acute setting. Chronic studies (already underway) will have to be performed to determine whether sheep can survive TCPC long term.
Pump Support of TCPC
Axial flow pump support from the IVC to the PA reversed the negative hemodynamics of passive TCPC. Cardiac output, IVC flow, and arterial pressure were similar to biventricular circulation without volume administration, illustrating the potential feasibility of pump support of Fontan circulation. IVC pressure in pump-supported TCPC is low, potentially allowing for recovery from liver failure or protein-losing enteropathy.
Rodefeld and coworkers
8
were the first to apply pump support of TCPC in an animal model, using 2 Hemopumps in the SVC and IVC. They were able to show feasibility of mechanical support; however, they failed to provide a control group for comparison of hemodynamics. This study uses each animal as its own control, clearly proving the improvement of hemodynamics of pump-supported versus passive Fontan circulation.
IVC Collapse
The Thoratec HeartMate II is a powerful axial flow pump designed for use in the systemic circulation of adults. We would have considered using a smaller pump in this experiment if it had been available to us. However, despite negative IVC pressures, we did not encounter as many problems with IVC collapse and associated reduction of pump inflow as we anticipated. Even though we consistently experienced oscillations of the IVC wall and moderate IVC collapse, pump flow remained stable throughout at the recommended minimum of 6000 rpm. Even with increased revolutions per minute, there was only intermittent and self-limiting inflow reduction.
Thrombosis
During the short duration of this study in heparinized animals, we did not encounter any thromboembolic problems. This, however, remains a concern, particularly when using a pump designed for higher flows. The need for anticoagulation will have to be evaluated in subsequent chronic studies.
Other Limitations
Anesthetic drugs are known to adversely affect the hemodynamics of TCPC. Therefore the concentration of inhaled anesthesia was kept constant during the periods of passive and pump-supported TCPC, minimizing its influence on the data.
Positive pressure ventilation could also affect hemodynamics unfavorably. However, in our experience chest approximation did not change pressures and flows significantly. The data in these series of experiments did not suggest major compression of the grafts by means of chest approximation, although we have seen graft compression in chronic preparations when the animals position was changed.
The relatively large size of the Thoratec HeartMate II pump could obviously be of concern in smaller animals and, eventually, in small patients. It was not a limitation in this study with large sheep.
| Conclusions |
|---|
|
|
|---|
| Acknowledgments |
|---|
| Footnotes |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
F. G. Lacour-Gayet, C. J. Lanning, S. Stoica, R. Wang, B. A. Rech, S. Goldberg, and R. Shandas An artificial right ventricle for failing fontan: in vitro and computational study. Ann. Thorac. Surg., July 1, 2009; 88(1): 170 - 176. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| ANN THORAC SURG | ASIAN CARDIOVASC THORAC ANN | EUR J CARDIOTHORAC SURG |
| J THORAC CARDIOVASC SURG | ICVTS | ALL CTSNet JOURNALS |