J Thorac Cardiovasc Surg 2005;130:803
© 2005 The American Association for Thoracic Surgery
Evolving Technology |
a Hôpital Cardiologique Louis Pradel and Claude Bernard University, Lyon, France
b Hôpital Haut Lévêque and Victor Segalen University, Bordeaux, France
c AKH and Vienna University, Vienna, Austria
d Hospital Clinico and University of Barcelona, Barcelona, Spain
e Onze Lieve Vrouwziekenhuis, Aalst, Belgium
f Epicor Medical, Inc, Sunnyvale, Calif
g Division of Cardiothoracic Surgery, Washington University School of Medicine, St Louis, Mo
Received for publication February 6, 2005; revisions received May 7, 2005; accepted for publication May 18, 2005. * Address for correspondence: Gerard Champsaur, MD, 1430 Channing Ave, Palo Alto, CA 94301 (Email: jean.ninet{at}chu-lyon.fr; gchampsaur{at}pacbell.net).
* Address for reprints: Jean Ninet, MD, Cardiac Surgery, Hôpital Cardiologique Louis Pradel, 59 Blvd Pinel, 69003, Lyon, France (Email: jean.ninet{at}chu-lyon.fr; gchampsaur{at}pacbell.net).
| Abstract |
|---|
|
|
|---|
METHODS: From September 2002 through February 2004, 103 patients were prospectively enrolled in a multicenter study. Atrial fibrillation duration ranged from 6 to 240 months (mean, 44 months) and was permanent in 76 (74%) patients, paroxysmal in 22 (21%) patients, and persistent in 5 (5%) patients. All patients had concomitant operations, and ablation was performed epicardially on the beating heart before the concomitant procedure. The device automatically created a circumferential left atrial ablation around the pulmonary veins in an average of 10 minutes, and an additional mitral line was created epicardially in 35 (34%) patients with a handheld device by using the same technology.
RESULTS: No complications or deaths were device or procedure related. There were 4 (3.8%) early deaths and 2 late extracardiac deaths. The 6-month follow-up was complete in all survivors. At the 6-month visit, freedom from atrial fibrillation was 85% in the entire study group (80% in patients with permanent atrial fibrillation, 88% in the 35 patients who had the additional mitral line, and 100% in patients with paroxysmal atrial fibrillation). A pacemaker was implanted in 8 patients. Only the duration and type of atrial fibrillation significantly increased the risk of recurrence.
CONCLUSION: Epicardial, off-pump, beating-heart ablation with acoustic energy is safe and cures 80% of patients with permanent atrial fibrillation associated with long-standing structural heart disease.
| Introduction |
|---|
|
|
|---|
AF is associated with a mortality rate that is 1.5 to 2 times the expected rate in the general population after adjustment for other cardiovascular conditions,
1,2
and despite the benefits of anticoagulation, it is still responsible for thousands of strokes caused by systemic emboli with the subsequent burden on health care costs.
3,4
The medical management of AF is well established,
2
but little was offered for the cure of AF until the direct interventional therapy of AF was initiated by Cox and colleagues
5,6
in the form of the surgical maze procedure in 1987. The Cox maze procedure remains the gold standard for the treatment of AF, with 80% to 95% freedom from AF at 15 years of follow-up.
7
Haissaguerre and colleagues
8
subsequently demonstrated that most individual episodes of paroxysmal AF originate from premature atrial beats located in or near the orifices of the pulmonary veins. This concept set the stage for the development of catheter-based technologies aimed at isolating the orifices of the pulmonary veins from the remainder of the left atrial wall, thereby isolating the triggers that induce AF and curing the arrhythmia.
Because of the complexity of the Cox maze procedure, several modifications have been introduced in an effort to simplify and shorten the operation. Some authors abandoned the right-sided lesions completely yet still attained AF cure rates in the range of 75% to 85%.
9,10
In addition, alternatives to the cut-and-sew techniques using various energy sources (radiofrequency, microwave, laser, and cryotherapy) have been developed in an attempt to replicate the Cox maze lines of atrial conduction block more quickly.
11-16
However, these new energy sources are all faced with similar challenges. When the energy is applied from the endocardium (either through catheters or during surgical intervention), it is impossible to determine whether the lesion is transmural. Additional energy might ensure transmurality but might also result in damage of the surrounding structures, such as the esophagus.
13,17
If these various energy sources are applied from the epicardium, their problems might be even more complex because of their thermal gradient nature. These problems include overcoming the heat-sink effect from the cooler blood (or warmer in the case of cryotherapy) and difficulty in propagating through epicardial fat. Moreover, even when applied epicardially, they can still cause damage to vital nearby structures.
18
High-intensity focused ultrasound (HIFU) is an energy source that specifically addresses the problems associated with thermal gradient energy sources for the epicardial ablation of AF. This report details the results of the first clinical feasibility study with this new energy source in which HIFU is applied epicardially on the beating working heart without the use of cardiopulmonary bypass.
| Material and Methods |
|---|
|
|
|---|
Patients were candidates for an Epicor Medical HIFU ablation (Epicor Medical, Inc, is a St Jude Medical company) if they had any form of AF present for at least 6 months before the operation and were already scheduled to undergo concomitant cardiac surgery for the correction of valvular, ischemic, or congenital heart disease. Exclusion criteria included acute or active infection, severe heart failure, severe progressive extracardiac disease, a prior cardiac operation, and the presence of a previously implanted intracardiac device.
The primary end point for the study was feasibility determined by the time required for sizing, insertion, and epicardial ablation by using the Epicor Medical Ablation System on the beating heart. Secondary end points included (1) safety, defined as the incidence of device-related and nondevice-related complications and mortality during and 6 months after the operation, and (2) freedom from AF, atrial flutter, or both, as determined by means of surface electrocardiography (ECG) and Holter monitoring at the 6-month visit.
Device Description
The Epicor Medical Ablation System is a US Food and Drug Administrationapproved, CE-marked system designed to deliver HIFU energy to ablate cardiac tissue. The system consists of the Ablation Control System (ACS) generator, a family of disposable ablation devices and a set of accessories. The ACS is a microprocessor-based unit that provides acoustic power to the ultrasound transducers.
The UltraCinch device (Figure E1) is an array of multiple ultrasound transducers that is used to produce a transmural circumferential left atrial lesion around the pulmonary vein orifices. An additional tool, the UltraWand, is a handheld ablation device with 2 transducers only for the epicardial creation of additional linear lesions.
|
|
Patient Population
From September 2002 through February 2004, 103 patients (59 male and 44 female patients) were enrolled. AF was categorized as being permanent in 76 (74%) patients, paroxysmal in 22 (21%) patients, and persistent in 5 (5%) patients.
20
Fifty (49%) patients were 70 years of age or older, and 19 (18%) were older than 75 years (Table 1). Comorbidity was relatively common, with chronic obstructive pulmonary disease in 15 patients and renal insufficiency in 7 patients. A history of cerebral ischemia was documented in 5 patients, repeat transient ischemic attack was documented in 5 patients, and peripheral embolism was documented in 3 patients. Catheter ablation had been attempted previously in 3 patients.
|
|
Data were collected on standard case report forms approved by each institution's institutional review board. In addition to the demographic data collected at the preoperative visit, a standard 12-lead ECG, 24-hour Holter monitoring, and transthoracic echocardiography were performed at the time of hospital discharge and at the 3- and 6-month follow-up visits. Event-monitoring systems (Instromedix, San Diego, Calif) were required in patients with paroxysmal or persistent AF who became symptomatic between the 3- and 6-month visits. The 6-month follow-up period was completed for all 103 patients as of June 10, 2004, with no patient lost to follow-up during that period. The mean follow-up period was 177 days (median, 185 days), with a range of 2 to 232 days. Adverse events and their relatedness to the device or the procedure were collected and reviewed by each clinical investigator and the local clinical coordinator and further assessed by a clinical events committee of independent US thoracic and cardiovascular surgeons not involved in the study.
Statistical Analysis
Data were entered into a computerized database and analyzed with a statistical package (STATISTICA; StatSoft, Inc, Tulsa, Okla). The descriptive summary of data included mean ± standard deviation and 95% confidence intervals for continuous variables and proportions for categoric variables. Between-group differences were assessed with t tests and analysis of variance in more than 2 groups for continuous variables and
2 tests for nominal variables. All reported P values are 2 sided.
| Results |
|---|
|
|
|---|
Evaluation of Feasibility
Various aspects of the procedure were timed (Table 3). The combined time required to insert the UltraCinch device and perform the automated ablation averaged 11.9 minutes, all of which was off-pump time. During the actual ablation itself (mean time, 9.7 minutes), surgeons continued to prepare the heart for the subsequent concomitant procedure. Because this was a required part of the concomitant cardiac surgery, the AF ablation portion of the procedure added an average of only 2.2 off-pump minutes to the overall combined operative procedure. No extra time was required for epicardial fat dissection or removal.
|
Complications
Early complications (
30 days after surgical intervention) occurred in 20 patients. Postoperative bleeding necessitating surgical reexploration occurred in 6 (5.8%) patients and was unrelated to the AF ablation. A permanent pacemaker was required in 5 patients. Strokes occurred between postoperative days 1 and 4 in 3 patients, but each resolved rapidly with medical therapy, with no residual effects in any patient. Two patients experienced an episode of femoral artery embolism on postoperative days 1 and 10, respectively, both of which were resolved by means of embolectomy. One patient had a serious deep wound infection (sacral decubitus ulcer) that required repeated surgical debridement. Late complications (
30 days after surgical intervention) consisted of sinus node dysfunction requiring a permanent pacemaker in 3 patients, ventilation pneumopathy in 1 patient, delayed tamponade treated with pericardial drainage in 1 patient, and a transient ischemic attack in a patient in sinus rhythm but who was severely hypertensive.
Mortality
The operative mortality rate (30 days postoperatively) was 3.8% (n = 4), including 3 patients older than 75 years. Two patients died from left ventricular failure because of prolonged bypass and crossclamp times, and 2 other patients died from multiorgan failure after aspiration pneumonia and prolonged ventilation in one case and general sepsis caused by repeat excisions of a bed sore in the other case. Late mortality consisted of 2 patients who died from noncardiac causes: multiorgan failure in a 74-year-old man who died 2 months after a mitral valve replacement complicated by a prolonged period of assisted ventilation and bilateral pneumopathy and infection of an aortoiliac graft implanted 2 months after coronary artery bypass grafting in a severely diabetic patient.
Evaluation of Efficacy
Absence of AF or atrial flutter was evaluated in 94 patients available at the 6-month follow-up visit (excluding the 3 patients mentioned above who were not considered for the efficacy analysis and the early and late deaths). In addition to the American Heart Association/American College of Cardiology classes of AF, the surgical classes included intermittent (paroxysmal plus persistent) AF (n = 27) and continuous (permanent) AF (n = 76).
22,23
At the 6-month follow-up visit, freedom from AF was 85% in the entire study group (n = 94) and 88% in the 35 patients who received the additional mitral line. Given the late and noncontrolled introduction of the UltraWand, the study was incapable of discerning any potential added benefit of the mitral line lesion. Remarkably, in the group of 76 patients who had continuous AF with a mean duration of 52.9 months (95% confidence limit, 37.8-69 months), the freedom from AF at 6 months was 80%. Thus epicardial off-pump, beating-heart ablation performed by using an automated 10-minute algorithm with HIFU cured 80% of patients in permanent AF associated with long-standing structural heart disease requiring surgical treatment. Equally as gratifying was the observation that 100% of patients with intermittent AF were free from AF at 6 months. Two patients who were in atrial flutter at the 3-month visit underwent an external cardioversion and remained in stable sinus rhythm thereafter. There were no instances of left atrial reentrant tachycardia or left-sided flutter. Patients with intermittent AF who were free from AF at the 3-month visit were entered into an event-monitoring program. No episodes of AF or atrial flutter were documented in any of these patients between the 3- and 6-month visits.
Eight pacemakers were implanted during the follow-up period, 4 for complete heart block with normal regular atrial activity and 4 (double chamber) for sinus node dysfunction. The assessment of left atrial function was not part of the protocol, but transthoracic pulsed Doppler echocardiography was available in a small cohort of 23 patients. All patients in sinus rhythm were documented to have left atrial contraction by the presence of a transmitral a-wave 6 months after the operation. Diastolic left ventricular function was preserved, with a peak-to-peak atrial velocity ratio of 1.90 ± 0.56, with the usual restrictions caused by the high incidence of concomitant mitral valve surgery. By using univariate analysis of preoperative and intraoperative variables, predictors for residual AF 6 months after surgical intervention were preoperative AF duration (duration of 87.4 months in patients with residual AF vs 37.06 months in patients free from AF, P = .03) and type of AF (freedom from AF is 80% for continuous vs 100% for intermittent, P = .01).
| Discussion |
|---|
|
|
|---|
Over the past several years, the surgical treatment of AF has evolved rapidly, and a number of devices with various energy sources have been used to create linear lesions in the atria (endocardial, epicardial, or combined). These therapies have been used for both lone AF and AF associated with structural heart disease.
11-16
They have been used to replicate some or all of the lesions of the original Cox maze procedure and have evolved into something of a hybrid maze procedure that might or might not include the right-sided lesions of the original procedure (Table 4).
7,11-16,24-27
The goal of these techniques, however, is to create lines of intra-atrial conduction block that will (1) preclude the development of macroreentrant circuits in the atria, (2) isolate the trigger or triggers for AF in or near the pulmonary vein orifices, or (3) accomplish both goals 1 and 2, while at the same time allowing the atria to resume either a sinus or atrially paced rhythm. Recent clinical evidence has clearly demonstrated that the key mandatory step in the interventional treatment of AF is isolation of the pulmonary vein orifices.
10-18
Although this was a part of the original Cox maze procedure, that portion of the procedure was not designed specifically to isolate AF triggers because they were unknown at the time the Cox maze procedure was conceived and first applied.
5,6
On the contrary, since Haissaguerre and colleagues' seminal article was published in 1998,
8
virtually all subsequent lesion patterns have included isolation of the pulmonary veins.
|
|
The disadvantage of the device in its current embodiment is, during the first few experiences, the perception of a certain size and rigidity, which both turn out to be very positive in that this perception does not preclude its use through a minimally invasive approach. The relative rigidity allows for a precise setting of the device on the atrial wall behind both venae cavae, without the risk of sliding toward the pulmonary veins or twisting onto itself.
Recently, the use of additional right-sided lesions has been investigated by using monopolar
15,24
or bipolar
25
radiofrequency, either as an adjunct to the traditional surgical maze procedure to replace some of the surgical lines (hybrid maze) or as an isolated procedure (Table 4). Freedom from AF ranged from 73% in long-term survivors to 90% and 91% at 6 months with a complete lesion set, including bilateral appendage resection, right-sided lesions, and adjunctive cryotherapy.
25
In this latter report, AF was paroxysmal in 63% of the patients, and along with the extensive lesion set, this might have played a role in achieving the excellent results. However, the 6-month follow-up in 2 reports
24,25
was limited to a fraction (50%) of the population under study (Table 4), and the procedure was performed during cardiopulmonary bypass and an additional aortic crossclamp time of 20 to 40 minutes.
Perhaps the only way to compare the results of this preliminary study with those in the literature is to perform a meta-analysis of the reports in which thermal lesions were performed exclusively in the left side of the heart during concomitant cardiac surgery. Although the lack of standardization in the reporting of events and results makes accurate comparisons difficult, freedom from AF at 6 months in these reported series ranges from 65% to 92% (Table 4) using either various types of radiofrequency or microwave ablation. However, except for a study reporting on a limited number of patients treated during a fully beating-heart epicardial procedure and an incomplete follow-up,
29
current ablation technology still requires significant additional cardiopulmonary bypass and aortic crossclamp times, as well as the dissection of fat when some epicardial lines are applied, and occasional serious collateral damage can still occur.
18
Finally, the presence of thromboembolic complications raises the question of their prevention through exclusion of the left atrial appendage. The incidence of thromboembolic events after valvular surgery in this age group remains significant,
21
and the majority of patients in this report had mechanical valve replacements. At the time of the writing of the protocol, there were no studies justifying clearly the exclusion of the appendage over abstention or the advantage of an exclusion technique over another, so that the steering committee did not give a specific recommendation.
The limitations of this study are inherent in a report on the first clinical use of any technology, particularly the short follow-up period, although it was completed, with no patients lost to follow-up. Further studies will be required to assess the need for a full epicardial lesion set of the Cox maze procedure, the importance of the mitral line, the results in patients with lone AF, and the development of more minimally invasive approaches.
| Acknowledgments |
|---|
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S.-i. Sakamoto, R. K. Voeller, S. J. Melby, S. C. Lall, N.-l. Chang, R. B. Schuessler, and R. J. Damiano Jr. Surgical ablation for atrial fibrillation: The efficacy of a novel bipolar pen device in the cardioplegically arrested and beating heart. J. Thorac. Cardiovasc. Surg., November 1, 2008; 136(5): 1295 - 1301. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Groh, O. A. Binns, H. G. Burton III, G. L. Champsaur, S. W. Ely, and A. M. Johnson Epicardial Ultrasonic Ablation of Atrial Fibrillation During Concomitant Cardiac Surgery Is a Valid Option in Patients With Ischemic Heart Disease Circulation, September 30, 2008; 118(14_suppl_1): S78 - S82. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Johansson, B. Houltz, E. Berglin, G. Brandrup-Wognsen, T. Karlsson, and N. Edvardsson Short-term sinus rhythm predicts long-term sinus rhythm and clinical improvement after intraoperative ablation of atrial fibrillation Europace, May 1, 2008; 10(5): 610 - 617. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. K. Voeller, R. B. Schuessler, and R. J. Damiano Jr. Surgical Treatment of Atrial Fibrillation Card. Surg. Adult, January 1, 2008; 3(2008): 1375 - 1394. [Full Text] |
||||
![]() |
M. A. Groh, O. A. Binns, H. G. Burton III, S. W. Ely, and A. M. Johnson Ultrasonic Cardiac Ablation for Atrial Fibrillation During Concomitant Cardiac Surgery: Long-Term Clinical Outcomes Ann. Thorac. Surg., December 1, 2007; 84(6): 1978 - 1983. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lamotte, L. Annemans, B. Bridgewater, S. Kendall, and M. Siebert A health economic evaluation of concomitant surgical ablation for atrial fibrillation Eur. J. Cardiothorac. Surg., November 1, 2007; 32(5): 702 - 710. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Fleck, F. Wolf, T. Bader, R. Lehner, C. Aigner, G. Stix, E. Wolner, and W. Wisser Atrial Function After Ablation Procedure in Patients With Chronic Atrial Fibrillation Using Steady-State Free Precession Magnetic Resonance Imaging Ann. Thorac. Surg., November 1, 2007; 84(5): 1600 - 1604. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov Choice of Surgical Lesion Set: Answers From the Data Ann. Thorac. Surg., November 1, 2007; 84(5): 1786 - 1792. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. Wisser, G. Seebacher, T. Fleck, C. Aigner, C. Khazen, G. Stix, D. Hutschala, and E. Wolner Permanent Chronic Atrial Fibrillation: Is Pulmonary Vein Isolation Alone Enough? Ann. Thorac. Surg., October 1, 2007; 84(4): 1151 - 1157. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Calkins, J. Brugada, D. L. Packer, R. Cappato, S.-A. Chen, H. J.G. Crijns, R. J. Damiano Jr, D. W. Davies, D. E. Haines, M. Haissaguerre, et al. HRS/EHRA/ECAS Expert Consensus Statement on Catheter and Surgical Ablation of Atrial Fibrillation: Recommendations for Personnel, Policy, Procedures and Follow-Up: A report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation Developed in partnership with the European Heart Rhythm Association (EHRA) and the European Cardiac Arrhythmia Society (ECAS); in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), and the Society of Thoracic Surgeons (STS). Endorsed and Approved by the governing bodies of the American College of Cardiology, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, and the Heart Rhythm Society. Europace, June 1, 2007; 9(6): 335 - 379. [Full Text] [PDF] |
||||
![]() |
S. C. Lall, S. J. Melby, R. K. Voeller, A. Zierer, M. S. Bailey, T. J. Guthrie, M. R. Moon, N. Moazami, J. S. Lawton, and R. J. Damiano Jr The effect of ablation technology on surgical outcomes after the Cox-maze procedure: A propensity analysis J. Thorac. Cardiovasc. Surg., February 1, 2007; 133(2): 389 - 396. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov Advances in Surgical Treatment of Atrial Fibrillation Stroke, February 1, 2007; 38(2): 618 - 623. [Abstract] [Full Text] [PDF] |
||||
![]() |
I. Bakir, F. P. Casselman, P. Brugada, P. Geelen, F. Wellens, I. Degrieck, F. Van Praet, Y. Vermeulen, R. De Geest, and H. Vanermen Current Strategies in the Surgical Treatment of Atrial Fibrillation: Review of the Literature and Onze Lieve Vrouw Clinic's Strategy Ann. Thorac. Surg., January 1, 2007; 83(1): 331 - 340. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Jeanmart, F. Casselman, R. Beelen, F. Wellens, I. Bakir, F. Van Praet, G. Cammu, Y. Degriek, Y. Vermeulen, and H. Vanermen Modified Maze During Endoscopic Mitral Valve Surgery: The OLV Clinic Experience Ann. Thorac. Surg., November 1, 2006; 82(5): 1765 - 1769. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. M. Gillinov, S. Bhavani, E. H. Blackstone, J. Rajeswaran, L. G. Svensson, J. L. Navia, B.G. Pettersson, J. F. Sabik III, N. G. Smedira, T. Mihaljevic, et al. Surgery for Permanent Atrial Fibrillation: Impact of Patient Factors and Lesion Set Ann. Thorac. Surg., August 1, 2006; 82(2): 502 - 514. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. E.W. Hemels, Y. L. Gu, A. E. Tuinenburg, P. W. Boonstra, A. C.P. Wiesfeld, M. P. van den Berg, D. J. Van Veldhuisen, and I. C. Van Gelder Favorable long-term outcome of maze surgery in patients with lone atrial fibrillation. Ann. Thorac. Surg., May 1, 2006; 81(5): 1773 - 1779. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||