J Thorac Cardiovasc Surg 2005;130:830-836
© 2005 The American Association for Thoracic Surgery
Surgery for Congenital Heart Disease |
a Department of Surgery, University of California, San Francisco, Calif.
b Department of Anesthesia, University of California, San Francisco, Calif.
c Department of Neurology, University of California, San Francisco, Calif.
d Department of Pediatrics, University of California, San Francisco, Calif.
Presented at Scientific Sessions of the American Heart Association, New Orleans, La, 2004.
Received for publication February 23, 2005; revisions received April 1, 2005; accepted for publication April 12, 2005. * Address for reprints: Patrick McQuillen, MD, 505 Parnassus Ave, Room M680, San Francisco, CA 94143-0106. (Email: psmcq{at}itsa.ucsf.edu).
| Abstract |
|---|
|
|
|---|
METHODS: In 18 children having repair of aortic coarctation, we measured the maximum change and integral for hemoglobin D (difference of oxyhemoglobin and deoxyhemoglobin), total oxygenation index, and the redox state of cytochrome aa3. Thirteen subjects had recordings from the left hemisphere to test the hypothesis that aortic occlusion impairs left hemispheric oxygen balance. Five subjects had recordings from the right hemisphere for comparison.
RESULTS: After aortic clamping, a significant decrease in hemoglobin D was observed in recordings from the left hemisphere compared with those from the right hemisphere (P = .03, maximum change in hemoglobin D). Total oxygenation index and cytochrome aa3 were generally preserved. There was an inverse linear relationship for the change in hemoglobin D during clamp application and after removal (Spearman rho = 0.74), with increased hemoglobin D after clamp removal in those subjects with the greatest decrease of hemoglobin D during arch occlusion. Linear regression analysis identified nitroprusside administration as significantly associated with a decrease in hemoglobin D (P < .001).
CONCLUSIONS: Significant impairment in left hemispheric cerebral oxygen balance was identified during arch clamping. The neurodevelopmental significance of impaired cerebral oxygen balance detected by near-infrared spectroscopy during aortic coarctation repair remains to be elucidated.
| Introduction |
|---|
|
|
|---|
Near-infrared spectroscopy (NIRS) is a noninvasive optical method that measures changes in tissue chromophore concentration
9,10
that correlate with cerebral blood flow,
11
oxygenation,
10
and cellular energetics,
12
which allow an estimation of the balance of cerebral oxygen delivery and use. The objective of this study was to measure cerebral oxygenation with NIRS during aortic clamping to determine whether cerebral oxygen balance is impaired.
| Methods |
|---|
|
|
|---|
NIRS
NIRS measurements were performed with a NIRO-300 (Hamamatsu Photonics, Hamamatsu City, Japan) equipped with a single monitoring unit. This limitation precluded simultaneous recording from the right and left hemispheres. Therefore, each subject was recorded from either the left hemisphere or the right hemisphere to determine the effects both distal (left hemisphere) and proximal (right hemisphere) to the aortic clamp. Recordings were obtained from the left hemisphere in the first 13 subjects and from the right hemisphere in the subsequent 5 subjects as a comparison group. Each recording was started before aortic clamp application and continued after release of the clamp, allowing comparison of NIRS variables both during clamping and after release within and between left and right hemisphere groups.
Immediately after induction of anesthesia, the NIRS probe was applied to the skin of the forehead below the hairline. The probe consists of a transmitting optode emitting 4 infrared wavelengths and an array of 3 photodiode detectors separated by a fixed distance of 4 cm in a probe holder. The NIRO-300 monitoring unit measures relative change in the concentration of oxygenated and deoxygenated hemoglobin and cytochrome aa3 (CytOx). The array of photodiode detectors allows calculation of an absolute value for total oxygenation index (TOI) by the method of spatially resolved spectroscopy.
13
Continuous measurement of NIRS values with a sample rate of 1 second were digitally recorded, along with patient physiologic variables (heart rate, blood pressure, electrocardiographic, and pulse oximetry), onto a portable computer (Dell Inc, Round Rock, Tex) running BioBench software (National Instruments, Temecula, Calif). Patient data, including hematocrit value, core body temperature, and vasoactive medications, were recorded from the anesthesia record. From NIRS measurements, hemoglobin difference (HbD; ), a measure of cerebral blood flow,
11
was derived. Absolute values of the primary NIRS variables are reported as micromoles per liter times the differential pathlength factor, a value that might vary between NIRS measurements. Therefore, a baseline was calculated as the average NIRS value for 5 minutes before application and removal of the aortic clamp, and only measures of the relative change in NIRS variables were analyzed. Specifically, during aortic clamping and after removal of the clamp, the maximum change and integral from baseline throughout the period (Figure 1) were calculated for HbD, TOI, and CytOx by BioBench software.
|
Data Analysis
Processed NIRS and physiologic data and patient information recorded manually were inserted into a common database for statistical analysis with Stata software (Stata Corp, College Station, Tex). The maximum change and integral (Figure 1) for HgD, TOI, and CytOx of recordings from the right and left forehead during and after aortic clamping were compared by using the Mann-Whitney U test for continuous or ordinal data. Nominal data were compared with the Fisher exact test. Linear regression (univariate) was used to assess the relationships between the changes in each NIRS variable with patient (age, weight, sex, and coarctation type), pharmacologic (use of antihypertensive or inotropic medications), or operative (aortic crossclamp time and surgical procedure) characteristics.
| Results |
|---|
|
|
|---|
Aortic Clamping Decreases Hemoglobin D Levels
Figure 1 presents typical raw data tracings for blood pressure (upper panels) and HgD (lower panels) in subjects monitored over the right hemisphere (Figure 1, A) and those monitored over the left hemisphere (Figure 1, B). Occlusion of the aorta (first arrow; Figure 1, A) increases blood pressure proximal to the clamp as the entire cardiac output is delivered to the proximal aorta. After release of aortic clamps (second arrow; Figure 1, A), there is a transient decrease in blood pressure as the cardiac output is redistributed to the descending aorta before regulation back to baseline. In all 5 subjects monitored over the right hemisphere, HbD increased during aortic clamping (Figures 1, A, and 2, A and B) and transiently decreased after removal of the aortic clamp (Figure 1, A, and 2, A and B). In contrast, 10 of 13 subjects monitored over the left hemisphere had a significant decrease in HbD as measured by maximum change (Figure 1 and 2, A; P = .03, Mann-Whitney test) or the integral during aortic clamping (Figures 1 and 2, B; P = .03, Mann-Whitney test). Subjects monitored over the left hemisphere showed a more variable response after release of the clamp, with individual subjects showing an increase in HgD with clamp removal (n = 5) and others having a decrease in HgD (n = 8). HgD was not significantly different after clamp removal among subjects measured over the left and right hemispheres (Figure 2, A; P = .15, Mann-Whitney test). An inverse linear relationship is observed in the change in HgD with the aortic clamp and removal (Figure 3), with those subjects with the largest decrease in HgD after aortic clamping having an increase in HgD after clamp removal (Spearman rho = 0.74).
|
|
Similar changes to those in TOI were observed from the subset of subjects monitored over the left hemisphere (n = 8) and subjects monitored over the right hemisphere (n = 4) who had CytOx recorded during their procedure. Only one subject monitored over the right side had a decrease in the maximal change or integral CytOx. However, the majority of subjects with left-sided monitoring (n = 6) had a small decrease (median, 0.2 µmol/L*dpf; range, 2.8 to 0.6) in the maximum change of CytOx during aortic clamping. After clamp removal, all subjects monitored over the right hemisphere had a decrease in maximum change or integral of CytOx. Subjects monitored over the left hemisphere had a variable response of maximum change or integral CytOx after clamp removal. The patient monitored over the left hemisphere with the largest decrease in maximum change for CytOx during clamping (2.8 µmol/L*dpf) continued to experience a decrease after clamp removal, resulting in the largest observed change in CytOx during the study (6.6 µmol/L*dpf; Figure 2, E).
Nitroprusside Treatment During Aortic Clamping Decreases Cerebral Oxygen Balance
On univariate analysis, the 2 patients treated with nitroprusside during aortic clamping had a significant decrease in HbD (P = .001, maximum change; P < .001 integral; Table 1) and in TOI (P = .04), but not in CytOx. This effect was specific for nitroprusside administered during aortic clamping because nitroprusside administered before or after clamping was not associated with a decrease in HbD (all P > .2, Table 1). In light of this relationship of nitroprusside to cerebral oxygen balance during aortic clamping, we examined the effects of other vasoactive infusions. The effect is unique to nitroprusside because esmolol or inotropes (dopamine and dobutamine) administered at any time did not affect HbD (all P > .1).
|
Clinical Outcomes and Follow-up
There were no hospital deaths, and all children are alive at a median follow-up of 15 months. No patients have required balloon dilation for recoarctation.
| Discussion |
|---|
|
|
|---|
Repair of coarctation of the aorta with or without arch hypoplasia requires extensive mobilization of arch vessels, the descending aorta, and intercostal branches. All ductal tissue is excised, and a large tension-free anastomosis to the undersurface of the aorta can be accomplished to prevent recoarctation, especially in neonates with transverse arch hypoplasia. Such an operative strategy usually requires control of the arch proximal to the left carotid artery, just distal to the innominate artery, or both. This assumes that collateral blood flow and completeness of the circle of Willis allows for a favorable and even distribution of cerebral blood flow.
Here we show that in a significant subset of patients undergoing coarctation repair, proximal occlusion of the aortic arch results in transient but significant impairment in contralateral cerebral oxygen balance. Monitoring of NIRS variables proximal to the aortic clamp (right side) consistently showed that on aortic arch occlusion, HbD, an indirect marker of cerebral blood flow, increased. On removal of the clamp, right-sided HbD decreased to baseline levels. However, NIRS monitoring distal to the aortic clamp (left side) showed that on application of the arch clamp, a progressive decrease in left-sided HbD occurred, despite increases in right radial arterial blood pressure. Nitroprusside use exacerbated the decrease in HbD in 2 patients. Use of nitroprusside to treat hypertension resulted in a 10-fold greater reduction in left-sided cerebral blood flow, as measured on the basis of HbD. Andropoulos and colleagues
18,19
have reported decreased saturation in the left cerebral hemisphere in most patients during regional cerebral perfusion for neonatal aortic arch repair. This group uses phenoxybenzamine for afterload reduction as part of their standard bypass strategy for arch repair. This might explain the high frequency of impaired left cerebral oxygen balance during regional cerebral perfusion if phenoxybenzamine impairs cerebral autoregulation in the same manner that we have observed for nitroprusside.
On completion of the coarctation repair and removal of clamps, HbD decreased because of relative systemic hypotension, except in those patients with the greatest magnitude of decrease during clamping. In these patients HbD increased even in the setting of systemic redistribution of flow after clamp removal, suggesting a left-sided cerebral vasodilatation to compensate for the decreased flow during arch occlusion. Despite decreases in left-sided HbD during proximal arch occlusion, indicators of cerebral oxygenation and cerebral energetics appeared to be preserved.
Rodriguez and associates
20
reported their experience with the cerebral effects of aortic occlusion during coarctation repair in children. Using transcranial Doppler scanning to monitor middle cerebral artery blood velocity, the authors found transient central nervous system hypoperfusion as a consequence of flow redistribution during aortic declamping in young infants. Older children usually showed faster autoregulatory compensation to the hemodynamic changes that occurred after unclamping the aorta. The authors inferred age-related physiologic differences, suggesting that young infants might require higher perfusion pressures during declamping to prevent decreases in cerebral blood flow. In the current study changes in left-sided cerebral oxygen balance during clamp application and early after its removal were not age dependent and were uniformly observed in neonates, infants, and older children. It should be noted, however, that after clamping, the decrease in HbD noted on left-sided NIRS monitoring was not absolute. There was a minority of patients in whom left-sided cerebral oxygen balance did not change during proximal arch occlusion, implying that cerebral autoregulatory mechanisms were active and also suggesting a completeness of the circle of Willis.
This study has notable limitations. The NIRS device used in this study was only capable of monitoring a single site, preventing simultaneous recording proximal and distal to the aortic clamp. HbD and CytOx are only indirect estimates of cerebral blood flow and cellular energetics, respectively. Current NIRS technology only allows for relative measurements of the primary NIRS variables. NIRS monitors use device-specific strategies (eg, spatially resolved spectroscopy) to derive an absolute measure of oxygen saturation.
13
Continuous transcranial Doppler (TCD) insonation of the middle cerebral artery and jugular venous oximetry are alternative techniques that have been used during cardiac surgery to estimate cerebral blood flow. However, TCD measures cerebral blood flow velocity and not flow,
21
and TCD does not monitor oxygen delivery or extraction. Continuous jugular venous oximetry, in contrast, can estimate cerebral oxygen extraction but is invasive and insensitive to regional cerebral ischemia.
We did not observe any clinical neurologic sequelae of the NIRS findings of impaired left-sided cerebral oxygen balance during a proximal arch occlusion, which is consistent with the fact that only 2 patients had a maximum decrease in TOI of greater than 10 µmol/L*dpf, a magnitude of decrease
22
that has been associated with postoperative neurologic dysfunction.
23
In particular, there was no evidence of postoperative neurologic deficit, stroke, or seizures, although this cohort was not studied with postoperative magnetic resonance imaging. Therefore, the question remains how to determine the neurodevelopmental effects of transient and asymmetric but significant impairment in cerebral oxygen balance during a coarctation repair. Significant cerebral injury can occur in neonates and can be undetectable by means of clinical examination.
17
Preoperative and postoperative magnetic resonance imaging will now be applied to determine whether brain injury occurs after transient decreases in left-sided cerebral blood flow measured by means of HbD. To avoid impairments in cerebral blood flow, it can be argued that a transsternal approach with relative hypothermia and selective cerebral perfusion might be optimal for children who have coarctation of the aorta, especially with arch hypoplasia. The current data also support the avoidance of afterload reduction to treat right radial arterial hypertension during aortic clamping because normalization of right radial arterial blood pressure can result in significant impairment of cerebral oxygen balance, possibly because of disruption of cerebral autoregulatory mechanisms. Furthermore, as previously reported, flow redistribution and hypotension occurs after declamping of the aorta, and persistent mild descending aortic compression might be indicated after clamp removal to avoid decreases in blood pressure and thus reduction in cerebral blood flow.
In conclusion, we have identified a significant decrease of HbD, a surrogate marker of cerebral blood flow, in the left hemisphere during aortic clamping in selected patients. Ongoing studies with intraoperative NIRS monitoring and postoperative magnetic resonance imaging and neurodevelopmental follow-up are needed.
| Footnotes |
|---|
* These authors contributed equally to this work. ![]()
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. C. Hirsch, J. R. Charpie, R. G. Ohye, and J. G. Gurney Near-infrared spectroscopy: What we know and what we need to know--A systematic review of the congenital heart disease literature J. Thorac. Cardiovasc. Surg., January 1, 2009; 137(1): 154 - 159. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. Farouk, M. Karimi, M. Henderson, J. Ostrowsky, E. Siwik, and H. Hennein Cerebral regional oxygenation during aortic coarctation repair in pediatric population. Eur. J. Cardiothorac. Surg., July 1, 2008; 34(1): 26 - 31. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Gargiulo, G. Oppido, E. Angeli, and C. P. Napoleone Neonatal aortic arch surgery MMCTS, July 23, 2007; 2007(0723): 2345. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. S. McQuillen, A. J. Barkovich, S. E.G. Hamrick, M. Perez, P. Ward, D. V. Glidden, A. Azakie, T. Karl, and S. P. Miller Temporal and Anatomic Risk Profile of Brain Injury With Neonatal Repair of Congenital Heart Defects Stroke, February 1, 2007; 38(2): 736 - 741. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||