J Thorac Cardiovasc Surg 2005;130:1016-1021
© 2005 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
a Cardiac Surgery Department, University of Tor Vergata, Rome, Italy
b Helmholtz-Institute for Biomedical Engineering, RWTH Aachen, Aachen, Germany
Received for publication December 31, 2004; revisions received April 14, 2005; accepted for publication April 22, 2005. * Address for reprints: Ruggero De Paulis, MD, Cattedra di Cardiochirurgia, Università di Roma Tor Vergata, European Hospital, via Portuense 700, 00149 Roma, Italy (Email: depauli{at}tin.it).
| Abstract |
|---|
|
|
|---|
METHODS: With a computer-controlled pulse duplicator, the in vitro performance (pressure differences, closure and leakage volumes, and energy losses) of a 25-mm mechanical or biologic prosthesis implanted in a standard Dacron straight conduit or in the new Dacron graft with a sculpted sinus were analyzed and compared.
RESULTS: The mechanical and biologic prostheses at 7 L/min cardiac output showed pressure drops across the valve of 8.72 mm Hg and 13.45 mm Hg, respectively, when inserted in the new Valsalva-style graft and of 7.97 mm Hg and 12.94 mm Hg, respectively, when inserted in the standard graft. The closure and leakage volumes for mechanical valves were higher than those for biologic valves; however, the presence or absence of sinuses did not result in significant differences in closure and leakage volumes. The maximal total energy losses were 5.89% and 9.49% for mechanical and biologic valves, respectively. No differences were evident between the two different Dacron grafts for each prosthetic heart valve.
CONCLUSION: The normal opening and closing behavior of a prosthetic aortic valve was not altered or modified by a different root shape above the heart valve. The presence of vortices inside the pseudosinuses of Valsalva did not influence the hydrodynamic properties of the biologic and mechanical valves tested.
| Introduction |
|---|
|
|
|---|
We have recently introduced into clinical practice a new Dacron polyester fabric conduit (Gelweave Valsalva; Vascutek Terumo, Renfrewshire, United Kingdom) that incorporates sculpted sinuses of Valsalva and has been found particularly advantageous in both types of aortic valvesparing operation, where it contributes to a more anatomically and physiologically accurate reconstruction of the aortic root.
5,6
It has also been used in association with mechanical or biologic valves in a composite aortic root replacement type of aortic root reconstruction because of its improved adaptability to the patient's anatomy.
7,8
In fact, its peculiar shape is characterized by a lower section that stretches horizontally and tends to reproduce the bulged section of the aortic root corresponding to the sinuses. In this way it reduces the tension on the coronary anastomoses once the aorta is under pressure and the heart resumes beating. Unlike in the aortic valvesparing operation, in which the use of the new Dacron graft allows the reconstruction of three independent sinuses of Valsalva, when used as a composite aortic root replacement its final shape is that of a single, large, egg-shaped sinus to which the two coronary ostia are attached.
The aim of this study was to determine whether the presence of vortices inside this single sinus immediately above the prosthetic aortic valve could negatively influence the hydrodynamic in vitro performances of a prosthetic heart valve. We therefore analyzed and compared the in vitro performances of mechanical and biologic aortic valve prostheses inside a conventional straight Dacron conduit and inside the new Dacron graft that incorporates the sinuses of Valsalva.
| Materials and Methods |
|---|
|
|
|---|
|
For determination of pressure, volume, and energy losses, pressure and flow signals are measured at several positions within the pulse duplicator. Pressures are recorded within the atrium (Figure 1, 2), in the apex region of the ventricle (Figure 1, 8), and in the ascending aorta (Figure 1, 12) with strain gauge pressure transducers (Cobe Laboratories, Inc, Lakewood, Colo). Flow is recorded by electromagnetic flowmeters (Zedepa Instruments, Seattle, Wash). The mitral flowmeter (Figure 1, 5) is located upstream of the mitral valve (Figure 1, 6). For the aortic valve (Figure 1, 10), the flowmeter (Figure 1, 11) is located 4 diameters downstream. The sensor signals are processed by amplifiers and special data-acquisition software. Each signal was averaged over 64 consecutive beats, and the averaged signals were then transferred to a second computer for off-line analysis. Therefore all reported values are mean values for 64 cycles. Mean systolic pressure difference, regurgitation volume, and energy losses were determined according to the Food and Drug Administration Replacement Heart Valve Guidance.
11
Mean Systolic Pressure Difference
The mean systolic pressure difference is the time integral of ventricular pressure (PV) minus aortic pressure (PAo) during the systolic period divided by systolic duration time (TSYS):
|
|
Closure and Leakage Volume
The closure volume (VCL) and leakage volume (VL) equations represent the time integrals of aortic flow (
) during closure and leakage periods, respectively.
|
|
|
|
The closure volume is defined as that part of the regurgitant volume that flows back through the valve during the time interval of valve closing (TCL), in the time needed to move the occluder from the open to the closed position. The leakage volume is defined as that part of the regurgitant volume that passes during the rest of the cycle (TL) and the closed state of the valve through the gap between the occluder and the valve ring.
Energy Losses
Energy loss (
E) is derived as the time integral of the product of mean pressure difference and aortic flow during defined time intervals, such as the systolic period (systolic energy loss), closure interval (closure energy loss), and leakage interval (leakage energy loss):
|
|
All parameters were determined for four different cardiac outputs (COs): 2.0, 4.0, 5.0, and 7 L/min at a fixed beat rate of 70 beats/min, resulting in stroke volumes of 28, 57, 71, and 100 mL respectively. The systolic duration was 35% of the cycle period. The mean atrial and aortic pressures were kept constant at 10 mm Hg and 100 mm Hg, respectively, for all COs.
| Results |
|---|
|
|
|---|
|
|
|
Energy Loss
Figure 4
depicts the accumulated transvalvular energy losses for the systolic phase, the closing phase, and the leakage phase for the mechanical and biologic valves. The maximal total energy losses were 5.89% and 9.49% for mechanical and biologic valves, respectively. No differences were evident between the two different Dacron grafts for each prosthetic heart valve.
|
| Discussion |
|---|
|
|
|---|
The simultaneous replacement of the aortic valve and the entire ascending aorta with coronary reimplantation (composite aortic root replacement)
12
has become a standard operation for all cases of aortic valve disease associated with aneurysm or dissection of the aortic root and ascending aorta. Many modifications of the standard surgical technique have been described through the years. The popularity and the increased need for this type of operation have stimulated the development of composite valved conduits to facilitate and expedite the surgical procedure. The use of composite grafts has shown excellent long-term results with an extremely low rate of complications.
13,14
In recent years, we introduced into clinical practice an originally designed Dacron conduit with the aim of obtaining a more anatomically accurate reconstruction of the aortic root.
7
In particular, this newly designed conduit tends to reproduce the bulged section of the aortic root corresponding to the sinuses of Valsalva. The importance of the sinuses of Valsalva in regulating the aortic valve motion during the cardiac cycle has been extensively studied in recent years.
15
By using this Valsalva-type Dacron conduit, it was possible to obtain more physiologically accurate valve motion characteristics after either the remodeling or the reimplantation type of valve-sparing procedure.
5,6
Because of its peculiar shape, this modified Dacron graft has also been used in composite aortic root replacement, where it might facilitate the surgical technique and at the same time decrease the tension on coronary ostia anastomoses. As a consequence, its practical advantages would be more evident in those cases where the coronary ostia are more difficult to mobilize (eg, reoperative cases, calcification, and infection). In the case of composite aortic root replacement, this new vascular prosthesis does not reproduce three independent sinuses but rather a single, large, egg-shaped proximal portion. Although vortices are present within this single egg-shaped sinus, their effects on the mobile elements of mechanical or biologic valve prosthesis are less evident and more difficult to visualize.
For these reasons, we attempted to determine whether the presence of vortices generated inside this initial portion of the Dacron graft could somehow affect the hydrodynamic performance of a mechanical or a biologic valve. In fact, the presence of nonlaminar flow immediately above the prosthetic heart valve could potentially influence the opening angle of a mechanical valve, with a direct effect on the transvalvular gradient. A change in the opening angle might also have a direct effect on the closure volume, because a shorter time for valve closure leads to smaller closure volumes.
However, the results of these experiments showed that the differences in flow pattern immediately above the prosthetic heart valve did not modify or alter the normal opening and closing behavior of the valve and thus hydrodynamic performance. In particular, the presence of pseudosinuses of Valsalva in the new conduit did not appear to influence the hydrodynamic properties of prosthetic mechanical valves (pressure drop and energy loss across the valves, as well as the amount of back flow during and after the valve is closed). Svensson
16
recently noted that the gradient across composite valves appears to be greater than for equivalent "noncomposite" aortic valve sizes, suggesting the possibility that the presence of sinuses or the maintenance of a certain root elasticity could have a positive effect on postoperative gradients.
Similarly to the mechanical valves, the presence of vortices inside the newly reconstructed root did not appear to have any influence on the hydrodynamic behavior of biologic valve prostheses. In fact, the total energy loss, which can be considered a measure of the efficiency of the valve, was not different between a standard conduit and one with pseudosinuses of Valsalva. In particular, in both Dacron grafts the total energy loss for the biologic valve was due almost exclusively to the systolic energy losses, whereas closure and leakage energy losses were extremely small or absent.
In conclusion, this series of experiments clearly showed that the presence of vortices inside a single, large, egg-shaped portion of ascending aortic graft immediately above the aortic valve prosthesis did not modify the hydrodynamic performances of the biologic and mechanical valves tested. The advantages of the improved anatomic adaptability that is peculiar to this Valsalva-type graft can therefore be exploited without the risk of affecting prosthetic valve performance.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
L. de Kerchove, M. Boodhwani, D. Glineur, M. Vandyck, J.-L. Vanoverschelde, P. Noirhomme, and G. El Khoury Valve sparing-root replacement with the reimplantation technique to increase the durability of bicuspid aortic valve repair J. Thorac. Cardiovasc. Surg., December 1, 2011; 142(6): 1430 - 1438. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. D. de Tullio, G. Pascazio, L. Weltert, R. De Paulis, and R. Verzicco Evaluation of prosthetic-valved devices by means of numerical simulations Phil Trans R Soc A, June 28, 2011; 369(1945): 2502 - 2509. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. N. Azadani, N. Jaussaud, P. B. Matthews, L. Ge, T. S. Guy, T. A.M. Chuter, and E. E. Tseng Valve-in-Valve Implantation Using a Novel Supravalvular Transcatheter Aortic Valve: Proof of Concept Ann. Thorac. Surg., December 1, 2009; 88(6): 1864 - 1869. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Tabata, H. Takayama, M. E. Bowdish, C. R. Smith, and A. S. Stewart Modified Bentall Operation With Bioprosthetic Valved Conduit: Columbia University Experience Ann. Thorac. Surg., June 1, 2009; 87(6): 1969 - 1970. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||