J Thorac Cardiovasc Surg 2005;130:1044-1049
© 2005 The American Association for Thoracic Surgery
Surgery for Acquired Cardiovascular Disease |
Department of Cardiac Surgery, University Clinic of Schleswig-Holstein, Campus Luebeck, Luebeck, Germany.
Received for publication April 19, 2005; revisions received June 3, 2005; accepted for publication June 7, 2005. * Address for reprints: Hans-H. Sievers, MD, Department of Cardiac Surgery, University Clinic of Schleswig-Holstein, Campus Luebeck, Ratzeburger Allee 160, 23538 Luebeck, Germany. (Email: herzchir{at}medinf.mu-luebeck.de).
| Abstract |
|---|
|
|
|---|
METHODS: Fresh porcine aortic roots were investigated in a pulsatile flow simulator, either native root or after different types of valve-sparing procedures (remodeling, sinus prosthesis, and reimplantation). Functional parameters, such as transvalvular pressure gradient, closing volume, cusp-bending deformation, and distensibility at different levels of the root were analyzed.
RESULTS: The mean pressure gradient was highest in reimplantation techniques (8.4 ± 1.8 mm Hg) compared with sinus prostheses (7.2 ± 0.9 mm Hg, P = .01) and remodeling techniques (6.8 ± 1.0 mm Hg, P = .002), mirror imaging the closing volume (reimplantation, 1.5 ± 0.4 mL; sinus prostheses, 2.3 ± 0.7 mL [P < .001]; remodeling, 3.4 ± 1.1 mL [P < .001]). Bending deformation indices increased significantly from remodeling (0.45 ± 0.05) and sinus prostheses (0.58 ± 0.06) to reimplantation techniques (0.73 ± 0.09). Dynamic changes in area of all techniques were decreased at the sinotubular junction and the commissural and sinus levels when compared with those seen in native roots but increased at the annular level for techniques with unfixed annulus (remodeling and modified sinus prosthesis).
CONCLUSIONS: In vitro the various aortic valvesparing operations differed characteristically in their ability to spare valve function, none of them completely meeting native valve behavior. The remodeling techniques exhibited valve dynamics closest to those of the native aortic root. The more the aortic valve is fixed with noncompliant prosthetic material, the more the native root dynamics are impaired.
| Introduction |
|---|
|
|
|---|
| Materials and Methods |
|---|
|
|
|---|
|
Surgical Techniques
For each technique, the size of the Dacron tube (Boston Scientific Corp, Wayne, NJ) was calculated by multiplying the measured outer diameter of the unloaded sinotubular junction by 1.2 to roughly achieve the diameter at systemic pressure.
6
The sizes of the prostheses as thus calculated were 30 to 32 mm in all procedures.
Figure 1 illustrates the principles of the different surgical procedures. In all aortic roots, the 3 sinuses of Valsalva were resected, leaving a rim of sinus tissue of 2 mm to the fibrous ring of the cusps. For the remodeling technique, the Dacron tube was trimmed in a 3-tongue fashion for replacement of the sinuses (Figure 1, a). The graft was sewn on the remaining aortic root by using a continuous 4-0 Prolene suture (Ethicon, Norderstedt, Germany). The remodeling technique was modified by sewing the 3 tongues separately into the resected sinuses, enabling more precise matching and easier surgical handling of the sinus reconstruction (Figure 1, b). Finally, the remaining Dacron tube was sewn onto the reconstructed root at the height of the sinotubular junction. The sinus prosthesis was self-manufactured. A Dacron tube was trimmed by inserting sinus of Valsalvashaped pieces of the same tube into the graft for creation of the sinuses. The 3 neosinuses were inserted by turning the prosthetic material at a 90° angle to use the distensibility of the Dacron material in circumferential direction (Figure 1, c). The aortic valve was implanted into this prosthesis by using interrupted, U-shaped 4-0 Prolene sutures at the annulus in a planar fashion and a continuous 4-0 Prolene suture for attachment of the valve from inside the prosthesis. The sinus prosthesis was further modified by inserting 3 square pieces of the same graft into the tube after turning them by 90°, with no rim of the Dacron tube left at the base of the sinus (Figure 1, d). This prosthesis enables the surgeon to adapt the height of the graft to the varying heights of the commissures of the pig aortic root. The aortic valve was then inserted into the prosthesis, as described for the sinus prosthesis.
For the original reimplantation technique, the aortic valve was implanted into an untrimmed Dacron tube (Figure 1, e), as described above. For the modified reimplantation technique (Figure 1, f), a 6-mm-larger graft was downsized to the calculated diameter during the proximal anastomosis by using mattress Teflon felt-pledgeted sutures placed in a horizontal plane at the annulus. The valve was then implanted from inside the Dacron tube with 4-0 Prolene continuous sutures. At the sinotubular junction, another graft tailored to the calculated diameter was sutured to the larger proximal prosthesis.
Test Circuit
The aortic roots were tested in a pulsatile flow simulator, details of which have been described previously.
7
In brief, the flow simulator consists of an open atrial reservoir, a cam-platedriven piston pump with low mass inflow valves, and a compliance simulating ventricular impedance. The aortic root section has a ventricle-shaped inflow configuration and an aortic impedance including a height-variable fluid column, an adjustable aortic air compliance chamber, and a special elastoresistive element for peripheral resistance.
The roots were mounted in a fluid reservoir, fixed between 2 spigots in the vertical position, and then tested at a heart rate of 64 cycles per minute, with a stroke volume of 54 mL for a systemic pressure of 125/80 mm Hg. For each valve-sparing technique, the dynamic function and cusp configuration of at least 6 aortic roots were investigated and compared with those of native aortic roots. Data were collected from 5 consecutive cycles each, and measurements were repeated 10 times for each root and test condition.
Data Acquisition and Analysis
Pressure measurements were performed by using Envec Ceracore M pressure transducers (Endress+Hauser, Maulburg, Germany) at the inflow and outflow spigots. The flow through the valves was measured with a HT207 ultrasonic flow-meter (Transonic Systems Inc, Ithaca, NY) positioned upstream of the valve. Data were collected digitally at a rate of 500 samples per second. Mean pressure gradient and closing volume from 5 cycles were determined as required in the ISO 5840
8
and calculated from the data by using a commercial spreadsheet analysis program.
Movements of the cusps were recorded with a Motionscope HR-1000 high-speed camera (Redlake Imaging Corp, Morgan Hill, Calif) positioned straight above the valve at a rate of 500 frames per second. Video data were digitized and analyzed with a custom-made motion evaluation software. Open cusp-bending deformations were determined from the still image in midsystole, where the strongest deformation was observed, and quantified by taking 3 points along the edges of the largest fold (Figure 2). The ratio of the maximal fold depth and the shorter distance BC was used as a cusp-bending deformation index.
9
|
|
| Results |
|---|
|
|
|---|
|
Bending Deformation Index
Concerning bending deformation, none of the surgical procedures was comparable with the native aortic root, and all techniques differed significantly from each other (Figure 5). Bending deformation indices for all valve-sparing techniques were more than twice those of the native aortic root and increased in relation to the degree the root was fixed with synthetic noncompliant material.
|
|
| Discussion |
|---|
|
|
|---|
The transvalvular pressure gradient was highest with the reimplantation techniques. Because the diameter of the implanted prosthesis was the same for all techniques investigated, other parameters, like the increased cusp-bending deformation and decreased root distensibility at the sinotubular junction and commissural level, could have contributed to these results. It is also possible that reimplanting the valve into the prosthesis slightly narrowed the annulus, leading to somewhat higher pressure gradients. Some increased pressure gradients could also be found in the clinical situation measured at rest in patients who had undergone the reimplantation technique. However, these findings do not seem to have clinical relevance in the midterm follow-up.
4
Nevertheless, long-term follow-up and exercise studies are of interest to further clarify this item.
The closing volume roughly mirrored the transvalvular pressure gradient, which differed significantly from that of the native root only for the reimplantation and modified reimplantation techniques and should have no clinical effect because the absolute volume is small compared with the stroke volume. Whether a causative relation exists between pressure gradient and closing volume remains to be established.
The bending deformation index was significantly increased with the reimplantation techniques and sinus prostheses compared with that seen in the remodeling techniques and native root, which is probably related to the restricted distensibility at the sinotubular and commissural level of these implantation techniques more than the configuration of the sinus. This is supported by the finding that bending deformation was also increased in the modified sinus prosthesis and the modified reimplantation technique, which provide some kind of neosinus. However, there are some other modifications of the reimplantation technique with more prominent neosinuses not investigated in this study. Whether these nonphysiologic bending stresses on the cusps during systole accelerate the cusp-degeneration process
10
and thus affect the durability of the valve remains to be investigated and seem to have at least no clinical significance in the mid-term follow up. In this context Sripathi and associates
11
found an unsymmetric opening of wrinkled aortic cusps in a stiff root with increased compressive stresses that might promote the degeneration process. Furukawa and coworkers
12
also found experimentally and Leyh and colleagues
13
found clinically that the cusp motion characteristics might be better preserved with the remodeling compared with the reimplantation technique, supporting the hypothesis of better-preserved valve function with the remodeling technique. Further investigations, also including opening and closing velocities, will be needed.
Regarding distensibility at the same area in the root, namely at the sinus and commissural levels, we found a reduced change of area that was pronounced in the reimplantation and modified reimplantation techniques both for the sinus and commissural levels but somewhat preserved for the commissural level with the remodeling techniques and sinus prostheses. Whether this was related to the more distensible native intervalvular tissue in the remodeling techniques, the stretchable prosthetic material in the sinus prostheses, or the cyclic change of cross-sectional configuration from the sinus shape to the circular shape remains unknown. Interestingly, the distensibility at the annular level was increased compared with that seen in the native root with the remodeling techniques and with the modified sinus prosthesis. Because the preparation techniques at the annular level were identical for all procedures, this finding remains difficult to understand. Whether this finding is related to a somewhat nonphysiologic increase in wall tension at the annular level as a response to restricted distensibility downstream remains to be determined. But only these techniques preserve the potential for unrestricted dilatation of the annulus, which the fixed annulus by prosthetic tube in the other techniques does not. Nevertheless, in 2 reoperations after the remodeling technique, we found the newly developed aortic insufficiency to be caused by an outward movement of the base of the sinuses, rendering the cusps noncoapting. Further clinical and experimental studies on this matter are necessary.
It needs to be stated that on the basis of our results, the techniques that imitate the hemodynamics of the native aortic root most adequately are the remodeling and the modified remodeling techniques. In certain pathologies, however, as in acute type A aortic dissection with enlarged roots and more definitely in Marfan syndrome, surgical techniques that stabilize the aortic root at the annular level and prevent it from further dilatation should be considered. However, only prospective randomized studies will show which procedure offers the best long-term results for which group of patients.
There are some limitations to this study. First, in vivo results could well be different from those of this in vitro study with nonliving subannular material. This potential drawback, however, could also be an advantage because it could unmask effects that are not quite so clear when living subannular tissue, like septal muscle, partially compensates for increased procedural wall tension at the annular level. Second, the retained tissue of the porcine root has functional characteristics that differ from those of the diseased tissue of a patient; however, because most of the tissue is replaced with prosthetic material, this possible effect is minimized. Third, we did not use blood but rather saline water, which has different viscosity features that might have influenced valve-bending deformation. Finally, alterations in distensibility caused by, for example, pseudoneointima proliferation could not be evaluated in this study.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
T. Hanke, E. I. Charitos, U. Stierle, D. Robinson, A. Gorski, H.-H. Sievers, and M. Misfeld Factors associated with the development of aortic valve regurgitation over time after two different techniques of valve-sparing aortic root surgery. J. Thorac. Cardiovasc. Surg., February 1, 2009; 137(2): 314 - 319. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Katayama, N. Umetani, S. Sugiura, and T. Hisada The sinus of Valsalva relieves abnormal stress on aortic valve leaflets by facilitating smooth closure. J. Thorac. Cardiovasc. Surg., December 1, 2008; 136(6): 1528 - 1535. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David Aortic Valve Repair and Aortic Valve Sparing Operations Card. Surg. Adult, January 1, 2008; 3(2008): 935 - 948. [Full Text] |
||||
![]() |
R.-U. Kuehnel, U. A. Stock, M. O. Wendt, I. Degenkolbe, U. Jainski, M. Hartrumpf, M. Pohl, and J. M. Albes Physiological function of stentless aortic valves is altered by trimming and removal of aortic wall components Interactive CardioVascular and Thoracic Surgery, April 1, 2007; 6(2): 182 - 187. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Scharfschwerdt, H.-H. Sievers, J. Greggersen, T. Hanke, and M. Misfeld Prosthetic Replacement of the Ascending Aorta Increases Wall Tension in the Residual Aorta Ann. Thorac. Surg., March 1, 2007; 83(3): 954 - 957. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David, C. M. Feindel, G. D. Webb, J. M. Colman, S. Armstrong, and M. Maganti Aortic Valve Preservation in Patients With Aortic Root Aneurysm: Results of the Reimplantation Technique Ann. Thorac. Surg., February 1, 2007; 83(2): S732 - S735. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. C. Miller Valve-Sparing Aortic Root Replacement: Current State of the Art and Where Are We Headed? Ann. Thorac. Surg., February 1, 2007; 83(2): S736 - S739. [Full Text] [PDF] |
||||
![]() |
A. W. Erasmi, H.-H. Sievers, J.F. M. Bechtel, T. Hanke, U. Stierle, and M. Misfeld Remodeling or Reimplantation for Valve-Sparing Aortic Root Surgery? Ann. Thorac. Surg., February 1, 2007; 83(2): S752 - S756. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. E. David, C. M. Feindel, G. D. Webb, J. M. Colman, S. Armstrong, and M. Maganti Long-term results of aortic valve-sparing operations for aortic root aneurysm. J. Thorac. Cardiovasc. Surg., August 1, 2006; 132(2): 347 - 354. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||